Acessibilidade / Reportar erro

Validation of the analytical method for analysis of the amitraz in honey using GC-TSD

This paper describes the validation of method for analysis of the acaricide amitraz in honey using n-hexane/acetone extraction, cleanup in cartridge of solid phase extraction (SPE) and analysis by gas chromatography with nitrogen-phosphorus detector (GC -TSD). The method can be used to monitor the presence of the pesticide amitraz in honey because of its good analytical parameters, precision (CV <11%) and accuracy (> 80%) in the range of 100-400 ng g-1. The limit set in the European Union, by Committee for Veterinary Medicinal Products, for amitraz in honey (200 ng g-1) is within this range. The response of the analytical system was linear between 100 and 800 ng mL-1 with correlation coefficient of 0.9996, limit of detection (LD -7 ng g-1) and quantification (LQ -100 ng g-1). The sample was fortified at three concentrations 100 ng g-1, 200 ng g-1 and 400 ng g-1, giving an average recovery of 80, 88 and 92% respectively. The validated method was used to assess the quality of honey sold in the city of Araraquara/SP, in which there were no residues of amitraz.

Amitraz; validation method; honey; GC-TSD


Fundação Editora da Universidade Estadual Paulista Júlio de Mesquita Filho - UNESP Instituto de Química / UNESP, 14801-970 Araraquara SP Brazil, Tel.: +55 16 3301-9636/3301-9631 - São Paulo - SP - Brazil
E-mail: ecletica@iq.unesp.br