Strategy for Correction of Matrix Effect on the Determination of Pesticides in Water Bodies Using SPME-GC-FID

Abstract

This paper investigates a strategy as a quality control parameter, using standard surrogate and determination of a relation factor for determination of parathion-methyl, chlorpyriphos and cypermethrin pesticides in environmental aqueous matrices with distinct characteristics (river water, estuarine, seawater and weir water), using the technique of solid phase microextraction gas chromatography with flame ionization detector (SPME-GC-FID). Pesticides were very susceptible to matrix effects promoted by environmental samples. The salinity and the organic matter seem to have been the main sources of interference in the method. For chlorpyriphos, in middle and high levels, the values of relation factor (Rf) for estuarines, seawater and weir matrices were statistically similar. For cypermethrin, the statistical equality occurred in estuarine matrices in medium and high levels of concentration. That indicates proportional behavior between the pesticide and the surrogate recovery, suggesting that a single value of Rf can be used as correction factor recovery for any of these matrices.

Keywords:
correction of matrix effect; pesticides in estuarine; pesticide analysis by SPME-GC-FID; methyl-parathion; relation fator


Introduction

The abusive use of pesticides in agricultural production may be responsible for the contamination of food, soil and water. The occurrence of herbicide residues in natural waters may present a risk to water resources with a negative impact on aquatic life.11 Lushchak, V.; Kubrak, I.; Storey, M.; Storey, B.; Lushchak, I.; Chemosphere 2009, 76, 932.

2 Hiller, E.; Tatarková, V.; Šimonovičová, A.; Bartal', M.; Chemosphere 2012, 87, 437.
-33 Maloschik, E.; Ernst, A.; Hegedus, G.; Darvas, B.; Székács, A.; Microchem. J. 2007, 85, 88. The pesticides may trigger ecotoxicological problems acting as endocrine disruptor.44 Carlos, A.; Neves, A.; Reis, C.; Queiroz, R.; Quim. Nova 2011, 34, 272. Therefore, the concern and monitoring of pesticide residues in aqueous environmental matrices is necessary.

The technological advances of the analytical instrumentation have provided a great simplification to implementing of the methods of analysis. For example, eliminating laborious steps in the sample preparation prior to analysis.55 Caldas, S.; Gonçalves, F.; Primel, G.; Prestes, D.; Martins, L.; Zanella, R.; Quim. Nova 2011, 34, 1604.,66 Rizzetti, M.; Kemmerich, M.; Martins, L.; Prestes, D.; Adaime, B.; Zanella, R.; Food Chem. 2016, 196, 25. However, the procedures of sample preparation still represent an obstacle to development of an efficient analytical method.77 Souza-Silva, É. A.; Jiang, R.; Rodríguez-Lafuente, A.; Gionfriddo, E.; Pawliszyn, J.; TrAC, Trends Anal. Chem. 2015, 71, 224. The constituents or coextractives which are present in the sample, be food or environmental,88 Arruda, J.; Guedes, C.; Oliveira, R.; Gomes, C.; Aparecida, M.; Milhome, L.; Ferreira, R. ; Food Chem. 2016, 199, 380.

9 Oliveira, R.; Carius, R.; Castro, D.; Aparecida, M.; Milhome, L.; Ferreira, R.; LWT - Food Sci. Technol. 2014, 59, 21.
-1010 Pinheiro, S.; da Rocha, O.; de Andrade, B.; Microchem. J. 2011, 99, 303. may be responsible for promote the imprecision and/or inaccuracy in quantification of the species of interest, such as pesticides. This type of interference is known as matrix effect.

The matrix effect is a type of interference generated by combination of all other sample components which decrease or increase the analytical signal detected, and directly affects the analyte quantification of interest.1111 Paschoal, R.; Rath, S.; Airoldi, S.; Reyes, R.; Quim. Nova 2008, 31, 1190.

12 Shakouri, A.; Yazdanpanah, H.; Shojaee, H.; Kobarfard, F.; Iran. J. Pharm. Res. 2012, 13, 927.

13 Picó, Y.; Blasco, C.; Font, G.; Mass Spectrom. Rev. 2004, 23, 45.

14 IUPAC; Gold Book; Iupac: Genebra, 2014, p. 1670. Available at http://goldbook.iupac.org/M03759.html, accessed in September 2016.
http://goldbook.iupac.org/M03759.html...

15 Kruve, A.; Künnapas, A.; Herodes, K.; Leito, I.; J. Chromatogr. A 2008, 1187, 58.
-1616 Kebarle, P.; Tang, L.; Anal. Chem. 1993, 65, 972A. This can affect the precision and accuracy of the method.

Strategies have been proposed to minimize or eliminate the problems of matrix effect.1717 Kruve, A.; Leito, I.; Herodes, K.; Anal. Chim. Acta 2009, 651, 75.

18 Ferrer, C.; Lozano, A.; Agüera, A.; Girón, J.; Fernández-Alba, R.; J. Chromatogr. A 2011, 1218, 7634.

19 Van Eeckhaut, A.; Lanckmans, K.; Sarre, S.; Smolders, I.; Michotte, Y.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2009, 877, 2198.
-2020 Cavalcante, M.; Lima, M.; Fernandes, M.; Duaví, C.; Talanta 2012, 93, 212. Furthermore, also should be considered the problems involving injection techniques, coated inlet liners, labeled internal standards, extensive sample cleanup, GC priming, compensation factors and matrix matching and surrogate standard.2121 Rimayi, C.; Odusanya, D.; Mtunzi, F.; Tsoka, S.; Chemosphere 2015, 118, 35.,2222 Ribani, M.; Bottoli, G.; Collins, H.; Jardim, F.; Melo, C.; Quim. Nova 2004, 27, 771.

The standard surrogate (SS) is a pure compound with similar physical and chemical characteristics of the target analyte (e.g., same substance, but isotopically modified) spiking to the sample to be analyzed. The surrogate amount recovered is used in the calculation of the recovery factor (%Rss), and this factor is applied to correct the target analyte concentration.2323 Cuadros-Rodríguez, L.; Gámiz-Gracia, L.; Almansa-López, M.; Bosque-Sendra, M.; TrAC, Trends Anal. Chem. 2001, 20, 620. Thus, the correction of the analyte concentration is made through the recovery of the surrogate standard. This method is very useful when it does not have a certified reference material (CRM).2222 Ribani, M.; Bottoli, G.; Collins, H.; Jardim, F.; Melo, C.; Quim. Nova 2004, 27, 771.

As mentioned, it is important that the standard surrogate has chemical behavior similar to the target analyte. However, may occur the unavailability of such compound.

This work investigated a strategy as a quality control parameter (QC), using standard surrogate with physical and chemical characteristics different of the analytes of interest, to minimize the effect matrix in the determination of pesticides, by solid phase microextraction gas chromatography with flame ionization detector (SPME-GC-FID), in environmental aqueous matrices with different characteristics: river water, estuarine, seawater and weir water.

Experimental

Chemical and reagents

The pesticides parathion-methyl (95%), chlorpyriphos (98%) and cypermethrin (96%) were ceded by Nufarm (Maracanaú, Ceará), and used to prepare five mix stock solutions in varied concentrations.

Each stock solution was used to prepare, respectively, a point of the calibration curve by fortification of pure water and environmental samples in vials. It was used 35 mL of water and 35 µL sample mix solution. Thus, it remained constant volume of solvent added to five levels of concentration of the calibration curve.

The concentration of the standard solutions varied of 100 to 500 µg L-1 for parathion-methyl, 5 to 300 µg L-1 for chlorpyriphos and 5 to 100 µg L-1 for cypermethrin, which corresponded to a linear range study. The azobenzene (99%) (Sigma-Aldrich, St. Louis, USA) was used as a standard surrogate (SS) in extractions being employed in all solutions at 100 µg mL-1. The standard mix and of the recovery solutions were prepared in acetone (Vetec, grade HPLC-UV) due to efficiency of solubilization. Nonpolar solvents were avoided because they cause damage to the fiber. Ultra pure water, obtained by Milli-Q® system (São Paulo, Brazil) was used in other procedures.

Sampling

Water samples used in this study were collected from the Ceará river, in Caucaia and weir Santo Anastácio, in Fortaleza, both in Ceará State, Brazil. The samples from the Ceará river were collected at four different points reaching the river zone to estuary. The weir samples were collected at a point near its margin. Thus, water samples with distinct characteristics were investigated. The values of pH, salinity and total organic carbon (TOC) dissolved are described in Table 1.

Table 1
Physicochemical parameters of samples

The samples were stored in glass bottles and refrigerated at 4 ºC for further filtering in PTFE membrane with porosity of 0.45 µm and recovery studies using SPME.

Method of extraction

In SPME experiments were used fibers with the PDMS (polydimethylsiloxane) coating, 30 µm (Supelco). This polymer, with nonpolar characteristics, is widely used in pesticide analysis in aqueous matrices and demonstrated satisfactory results in many extractions of organochlorine and organophosphate, as well as pyrethroid.2424 Abdulra'uf, B.; Tan, H.; Food Chem. 2015, 177, 267.

25 Alpendurada, F.; J. Chromatogr. A 2002, 968, 177.

26 Filho, M.; dos Santos, N.; Pereira, P.; Microchem. J. 2010, 96, 139.
-2727 Tankiewicz, M.; Morrison, C.; Biziuk, M.; Talanta 2013, 107, 1. The extraction was performed by direct mode, using 35 mL of solution, 25 minutes for extraction time, stirring at 900 rpm and temperature 55 ºC. Preliminary extraction tests were performed at room temperature, but the best results were obtained at 55 ºC.

The Figure 1 summarizes the steps of the process of extraction and desorption in the injector.

Figure 1
Scheme of process of extraction and desorption on analysis.

The samples of pure water used in verification of method performance and environmental samples were fortified with 35 µL of the standard solution that varied of 100 to 500 µg L-1 for parathion-methyl, 5 to 300 µg L-1 for chlorpyriphos and 5 to 100 µg L-1 for cypermethrin. The surrogate standard (azobenzene) was added for all solutions at 100 µg mL-1.

A blank solution (pure water) was prepared to investigate possible losses to the glass walls of the container. Also was not observed the formation of precipitates at pH studied.

Analytical conditions

Chromatographic analysis were carried out using a Shimadzu gas chromatograph (GC17) with a flame ionization detector (FID). The system was equipped with a capillary column DB-5 (Agilent J&W, 30 m, 0.25 mm i.d., 0.25 µm film thickness). The GC oven was operated with the following temperature program: initially at 60 ºC held for 2 min, followed by a ramp of 15 ºC min-1 to 300 ºC and then held for 6 min. The injection port was at 260 ºC in splitless mode, and the detection was carried out at 300 ºC. Hydrogen 4.5 FID (White Martins) was used as carrier gas at constant flow rate of 1.2 mL min-1.

Validation parameters

The performance of the extraction method was evaluated for linearity, sensibility, repeatability, limits of detection and quantification, recovery and matrix effect. The quantitation of the analytes was performed by external standard. A procedure for adjustment of the calibration curve was also used to allow a better prediction of the concentration of analytes in real samples.2828 Caulcutt, R.; Richard, B.; Statistics for Analytical Chemists; Chapman & Hall: London, 1994. Thus, sensitivity and limits of detection and quantification were recalculated based on the corrected curve.

The limits of detection (LOD) and quantification (LOQ) were calculated using the equation: LOD = 3sblank / m and LOQ = 10sblank / m, respectively, where sblank is the standard deviation of the three blank concentration values and m is the slope of the calibration curve.

To assess the recovery and matrix effect, the aqueous samples were spiked with standard mix of pesticide, in three concentrations levels, and surrogate standard azobenzene in concentration constant. Always been added 35 µL for both.

The recovery percentages (R(%)) for pesticides and azobenzene (SS) were determined according to the equation 1.

(1) R % = C spiked sample C non spiked sample C spiked sample water pure × 100

From the R(%) value obtained, the correction of the analyte concentration can be made on a sample by equation 2:

(2) C c = C e × 100 R %

However, with the use of surrogate standard for correction of pesticide concentration, R(%) value is now called Rss(%). Since the standard surrogate investigated (azobenzene) has physicochemical properties different of the pesticides investigated, it is necessary to calculate a relation between these compounds. This relation is called relation factor (Rf). It represents a correction or adjustment due to different chemical behavior between these compounds in the matrix relative to extraction by the analytical method. The Rf is calculated by the ratio of recoveries of the pesticide (R) and the recovery of surrogate standard (Rss).2020 Cavalcante, M.; Lima, M.; Fernandes, M.; Duaví, C.; Talanta 2012, 93, 212.

Thus, the correction of the analyte concentration using a surrogate standard of different chemical class and its respective Rf value can be obtained by equation 3:

(3) C c = C e × 100 R ss % × 1 R f

To evaluate the influence of the aqueous matrix by SPME analysis, was used the F-test (Fisher-Snedecor) (95% confidence level), to assess the homogeneity of variance between peak areas of pesticides in pure water (degrees of freedom, υ = 2) and environmental water (degrees of freedom, υ = 2) in three recovery levels. Thus, can be checked if the environmental aqueous matrix affect the accuracy of the method.2929 Brasil, MAPA/ACS; Manual de Garantia da Qualidade Analítica, 1ª ed.; Ministério da Agricultura, Pecuária e Abastecimento, Secretaria de Defesa Agropecuária: Brasília, 2011.

In the case where the constituents of the matrix does not affect the accuracy of the results obtained (Fcalculated < Fcritical), the standard deviations (s) of two analysis groups were grouped for obtained standard deviation combined (scomb) (equation 4), for application of the Student's t-test (equation 5). N is the number of points of the curve.

(4) s 2 comb = N i , 1 1 s i , 1 2 + N i , 2 1 s i , 2 2 N i , 1 + N i , 2 2

(5) t calc = x i , 1 x i , 2 s comb 2 11 N i , 1 + N i , 2

The t-test allowed the evaluation of statistical equality of mean results obtained in pure water and in water samples for each level of concentration i.

In cases where, for a given level of fortification, the variances of the two sample groups (pure water and environmental samples) were statistically different (Fcalculated > Fcritical), the matrix effect was evaluated by Student's t-test using equation 6, and the number of degrees of freedom (ν) was obtained by equation 7:

(6) t calc = x i , 1 x i , 2 s i , 1 2 N i , 1 + s i , 2 2 N i , 2

(7) v = s i , 1 2 N i , 1 + s i , 2 2 N i , 2 2 s i , 1 2 N i , 1 2 N i , 1 + 1 + s i , 2 2 N i , 2 N i , 2 + 1 2

Results and Discussion

SPME-GC-FID method

SPME is an extraction technique that stands out for its simplicity in terms of sample treatment because there is no need for the sample to pass through a rigorous clean-up and avoids the use of organic solvents.3030 Arthur, L.; Pawliszyn, J.; Anal. Chem. 1990, 62, 2145.,3131 Ouyang, G.; Pawliszyn, J.; Anal. Chim. Acta 2008, 627, 184. Its use is widespread in many analytical applications, such as environmental analysis,3232 Holopainen, S.; Luukkonen, V.; Nousiainen, M.; Sillanpää, M.; Talanta 2013, 114, 176.

33 Huang, S.; He, S.; Xu, H.; Wu, P.; Jiang, R.; Zhu, F.; Luan, T.; Ouyang, G.; Environ. Pollut. 2015, 200, 149.
-3434 Noventa, S.; Barbaro, J.; Formalewicz, M.; Gion, C.; Rampazzo, F.; Brusà, B.; Gabellini, M.; Berto, D.; Anal. Chim. Acta 2015, 858, 66. food,3535 Menezes, C.; Paulo, P.; Paiva, N.; de Barcelos, R.; Macedo, D.; Cardeal, L.; Microchem. J. 2015, 118, 272.

36 Adiani, V.; Gupta, S.; Padole, R.; Variyar, S.; Sharma, A.; Postharvest Biol. Technol. 2014, 98, 34.

37 Cecchi, T.; Alfei, B.; Food Chem. 2013, 141, 2025.

38 Lin, J.; Zhang, P.; Pan, Z.; Xu, H.; Luo, Y.; Wang, X.; Food Chem. 2013, 141, 259.

39 Moreira, A.; André, C.; Cardeal, L.; Food Chem. 2015, 178, 195.
-4040 Da Silva, C.; Da Silva, S.; Da Silva, N.; Godoy, O.; Nogueira, C.; Quitério, L.; Raices, L.; Food Chem. 2015, 167, 71. aromatic,4141 Dawidowicz, L.; Szewczyk, J.; Dybowski, P.; Talanta 2016, 146, 195.

42 Miró, P.; Buxaderas, S.; Food Res. Int. 2014, 57, 196.
-4343 Soto, C.; Maldonado, B.; Jofré, P.; Galmarini, R.; Silva, F.; Microchem. J. 2015, 122, 110. forensic4444 Aleksa, K.; Walasek, P.; Fulga, N.; Kapur, B.; Gareri, J.; Koren, G.; Forensic Sci. Int. 2012, 218, 31.

45 Brown, D.; Rhodes, J.; Pritchard, J.; Forensic Sci. Int. 2007, 171, 142.
-4646 Vazquez-Roig, P.; Blasco, C.; Picó, Y.; TrAC, Trends Anal. Chem. 2013, 50, 65. and pharmaceutical.4747 Aresta, A.; Bianchi, D.; Calvano, D.; Zambonin, G.; J. Pharm. Biomed. Anal. 2010, 53, 440.

48 Olszowy, P.; Szultka, M.; Fuchs, P.; Kegler, R.; Mundkowski, R.; Miekisch, W.; Schubert, J.; Buszewski, B.; J. Pharm. Biomed. Anal. 2010, 53, 1022.
-4949 Olszowy, P.; Szultka, M.; Ligor, T.; Nowaczyk, J.; Buszewski, B.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2010, 878, 2226. Also, low detection limits can be easily reached with this technique due to their ability in preconcentration of analytes,5050 Zhao, S.; Liu, L.; Chen, F.; Yuan, P.; Bai, Y.; Zhou, B.; Anal. Chim. Acta 2013, 769, 65. since it uses a polymeric coating which favors extraction.

The parathion-methyl, chlorpyriphos and cypermethrin pesticides were identified by GC-FID through the retention time, using individual standard solutions. In Figure 2 can be seen chromatograms obtained from environmental water samples (river, estuarine, seawater and weir) as well as pure water spiked. In Figure 2f, the peaks obtained for chlorpyriphos and cypermethrin are more intense, indicating good detection efficiency by the method.

Figure 2
Chromatograms obtained by SPME-GC-FID of the environmental water samples in blank: (a) river; (b) estuary P1; (c) estuary P2; (d) marine; (e) the weir and (f) pure water spiking in the middle level of concentration.

In fact, from Table 2 it can be seen good sensitivity of the method for chlorpyriphos and cypermethrin pesticides, with LOD of 0.4 and 0.3 µg L-1 and LOQ of 1.4 and 1.0 µg L-1, respectively.

Table 2
Validation parameters for method in pure water

In Brazil, CONAMA Resolution 357/20055151 www.mma.gov.br/port/conama/res/res05/res35705.pdf, accessed in August 2016.
www.mma.gov.br/port/conama/res/res05/res...
does the regulation of surface water and CONAMA Resolution 396/20085252 http://www.mma.gov.br/port/conama/legislacao/CONAMA_RES_CONS_2008_396.pdf, accessed in August 2016.
http://www.mma.gov.br/port/conama/legisl...
for groundwater. The MS Ordinance No. 2914/2011 sets potability standards for human consumption and establishes maximum value allowed 30 µg L-1 for chlorpyriphos. This ordinance does not report values for parathion-methyl and cypermethrin.

Physical and chemical properties of pesticides are shown in Table 3. Other studies indicate that there is a direct and positive relationship between log Kow (octanol/water partition coefficient) and log Kfw (fiber/water partition coefficient) for non-polar polymers.5353 Chao, P.; Lu, T.; Yang, W.; Int. J. Mol. Sci. 2014, 15, 2585.,5454 Difilippo, L.; Eganhouse, P.; Environ. Sci. Technol. 2010, 44, 6917.

Table 3
Physicochemical properties of pesticides

Thus, knowing that Kow for parathion-methyl can be considered relatively low,5555 FAO; http://www.fao.org/docrep/w5715e/w5715e03.htm, accessed in September 2016.
http://www.fao.org/docrep/w5715e/w5715e0...
the partition equilibrium for this pesticide between the PDMS coating of the fiber and the water, tends to be displaced into the aqueous phase.

The solubility value of this pesticide is a fact that corroborates their tendency to remain in the aqueous phase, due to non-polar nature of the fiber coating. This may explain the low sensitivity for the extraction of methyl-parathion. A better extraction for this pesticide may be possible using fibers with intermediate polarity phase, as polyacrylate (PA) and polydimethylsiloxane/divinylbenzene (PDMS/DVB).5757 Li, X.; Gan, P.; Peng, R.; Huang, C.; Yu, H.; J. Chromatogr. Sci. 2010, 48, 183.,5858 Lambropoulou, D.; Sakellarides, T.; Albanis, T.; Fresenius J. Anal. Chem. 2000, 368, 616.

The application of appropriate statistical tests can contribute to a more reliable measure and provide a better prediction of concentration in samples. For this, the confidence interval for the intercept and the slope may be calculated. Also, in some cases it is possible to carry out a fitting of calibration curve. This adjustment is made when the intercept of the calibration curve does not cross the origin.

It can be seen that the intercept (a) of the calibration curves, in Table 2, does not pass through the origin. However, one can consider that the intercept pass through the origin if the confidence interval for the intercept contains zero. Thus, the curve would have the form y = bx.2828 Caulcutt, R.; Richard, B.; Statistics for Analytical Chemists; Chapman & Hall: London, 1994. This procedure would be considered as an adjustment of the calibration curve.

As an example to estimate the range of the intercept at 95% confidence, we used the curve for parathion-methyl in pure water.

Initially, it must be determined the residual standard deviation (RSD), according to equation 8:

(8) RSD = residual sum of squares residual degrees of freedom

The residual sum of squares (RSS) can be obtained from the Table 4. The degree of freedom is equal to (N - 2) where n is the number of points of the curve (N = 5). It loses two degrees of freedom because it is being estimated both the slope and intercept.

Table 4
Calculation of , Σxy, Σx2, Σy2, RSS, Sxy, Sxx and Syy

Then it was estimated the confidence interval for the true intercept (a), equation 9:

(9) a ± t RSD 1 N + x ¯ 2 S xx

where: a is intercept of calibration curve, t is tcritical, RSD is residual standard deviation, -x22 Hiller, E.; Tatarková, V.; Šimonovičová, A.; Bartal', M.; Chemosphere 2012, 87, 437. is square average of concentrations values, N is the number of points of the curve and S xx is sum of squares of x. The value of tcritical (95%) used in the equation above was 3.18, which corresponds to 3 degrees of freedom (υ), because υ = N - 2. Solving equation 9, there is obtained -76.6 ± 1604.5.

That is, the confidence interval is -1681.1 to 1527.9. The wide range may reflect the lower value of measurements (based on five points).2828 Caulcutt, R.; Richard, B.; Statistics for Analytical Chemists; Chapman & Hall: London, 1994. As noted, the interval contains zero. Thus, can be considered to fit the line through the origin and take it to the form y = bx.

The confidence interval of the slope for equation non adjusted can be calculated according to equation 10, obtaining 23.77 ± 4.84:

(10) b ± t RSD S xx

Proceeding with the fitted equation, which now has the form y = bx is necessary to calculate the new slope (b) and its confidence interval, from equations 11 and 12, respectively.

(11) b = Σ xy Σ x 2

(12) b ± t RSD Σ x 2

Therefore, the equation of the curve became y = 23.6x with slope b = 23.6 ± 1.6.

The true concentration (Y / b) can be given by equation 13:

(13) Y b ± t RSD b 1 + Y b 2 / Σ x 2

Substituting values of Y in the above equation, the intervals for concentration were obtained (Table 5). Perceive that the value of RSD is calculated as above, however, the number of degrees of freedom is now equal to (N - 1).

Table 5
Confidence interval for the true concentration of parathion methyl in pure water

Likewise, preceded to pesticides chlorpyriphos and cypermethrin in pure water and the results of the adjusted calibration curve with the new detection and quantification limits can be seen in Table 6. The objective of the adjusted curve is to provide a better prediction of concentration in unknown samples. The graph of calibration curves can be seen in Figure 3.

Table 6
Adjusted calibration curve by the origin for the determination of pesticides in pure water and news limits of detection and quantification

Figure 3
Calibration curve for (a) parathion-methyl; (b) chlorpyriphos and (c) cypermethrin in pure water (◆) with a confidence interval of 95% (- - - -).

Recovery and matrix effect

The matrix effect is dependent on the physicochemical properties of pesticides and characteristics of endogenous matrix components.5959 Pinho, P.; Silvério, O.; Neves, A.; Queiroz, R.; Starling, M.; Quim. Nova 2010, 33, 909. Thus, in aqueous samples, the matrix effect can become more pronounced with increasing complexity. Thus, rainwater, groundwater and drinking water tend to have less matrix effect than estuarine waters and weir.

In Figure 4, it can be seen a trend in decreased recovery of pesticide while the aqueous samples tend to become more complex. At low level concentration the parathion-methyl was not detected in weir, probably due to the competitive effects of dissolved organic matter by PDMS. The weir from which water has been collected presented evidence of eutrophication, which suggests some amount of dissolved organic matter.

Figure 4
Recovery of pesticides in three fortification levels in river, estuary, marine and weir waters.

Although the poor distribution of this compound to the non-polar PDMS fiber, it can be expected that the increase in concentration favors the analyte extraction, resulting in increased recovery.

Although the chlorpyriphos and cypermethrin have low water solubilities and higher values of Kow, the extraction efficiency of the method was unsatisfactory for these pesticides, considering the non-polar nature of the fiber. Albanis and Hela6060 Albanis, A.; Hela, G.; J. Chromatogr. A 1995, 707, 283. also observed that salinity and the total organic compound content increase the interferences in environmental analysis of water samples using solid-phase extraction discs.

Chiou et al.6161 Chiou, T.; Malcolm, L.; Brinton, I.; Kile, E.; Environ. Sci. Technol. 1986, 20 (5), 502. report evidence that the solubility and stability of many hydrophobic organic compounds may be significantly improved by the presence of low concentrations of dissolved and/or suspended particulate-bound natural organic matter in water. These factors could have contributed to the low recovery values for pesticides by SPME.6262 Lambropoulou, A.; Sakkas, A.; Albanis, A.; Anal. Bioanal. Chem. 2002, 374, 932. In fact, the salinity and dissolved organic matter content are the main features between aqueous matrices assessed. Also, minerals with high adsorption capacity are present in estuarine matrices. They can interact with the organic contaminants and interfere with the extraction process.6363 Schroeder, A.; Wiesner, G.; Liu, Z.; Earth Planet. Sci. Lett. 2015, 430, 30.,6464 Weber, J.; LeBoeuf, J.; Young, M.; Huang, W.; Water Res. 2001, 35, 853.

In Table 7, it can be seen the significance of the matrix effect in the aqueous matrices based on the statistical tests F and t. The marine water and weir water promoted matrix effects in pesticide at all concentration levels. The chlorpyriphos and cypermethrin were the most affected by the matrix interferences. The parathion-methyl was the pesticide that has suffered minor occurrence of matrix effects.

Table 7
Significance matrix effect in aqueous samples for the determination of pesticides at three concentration levels, according to the statistical tests F and t

Relation factor (Rf)

Strategies using standard substitution (SS) can be adopted to correct the concentration of analytes due to the matriz effect. For SS compound to be considered adequate, it must have similar behavior to the analyte in the evaluated matrix, because it will be subject to the same conditions of interaction with the endogenous matrix components and extraction conditions.

In general, a deuterated compound is used. The SS is added to the sample (with known concentration) at the beginning of the analytical procedure, following every step of the process.

As low cost alternative, when there is no available deuterated or similar compound to the analyte of interest, it can be used another generic substance. Thus, a relation between these compounds (analyte and surrogate) is called relation factor (Rf). The Rf can be calculated by the recovery ratio between the SS and analyte. The Rf represents a correction or adjustment due to different chemical behavior between these compounds in the matrix to the analytical method of extraction.

As reported by Cavalcante et al.,2020 Cavalcante, M.; Lima, M.; Fernandes, M.; Duaví, C.; Talanta 2012, 93, 212. the use of compounds as standard surrogate has shown satisfactory results for correction of concentration of the pesticides in estuarine matrix. In Table 8 are shown the values of the Rf obtained by the ratio between the recovery of pesticides and recovery standard. The standard deviation (SD) value was obtained from error propagation formula:

(14) s y = y s a a 2 + s b b 2

Table 8
Relation factor (Rf) of pesticides with the standard surrogate in three fortification levels (low, medium and high levels)

where y is the result of the ratio of recoveries, a and b are the recovery values and sa and sb are their respective standard deviations.

In order to understand the matrix effect in the pesticides in aquatic environments from different origins, such as marine water and weir water the relation factors at three recovery levels of pesticides were studied. These aquatic environments usually are characterized by high salt content and high organic matter, respectively. As seen in Table 7 the chlorpyriphos and cypermethrin, which are hydrophobic compounds, were the most affected by the matrix interferences. On the other hand, although the parathion-methyl has unfavorable distribution on the PDMS fiber, it was verified a less pronounced matrix effect.

High recoveries for parathion-methyl were observed for higher concentration levels (middle and high), as shown in Figure 4, indicating that the matrix effects promoted by aquatic environments did not affect significantly the parathion-methyl extraction as noticeable for the chlorpyriphos and cypermethrin.

The matrix effect on the recoveries from pesticides in marine water and weir water with high organic matter were verified for a proportional increase of the relation factor (Rf) values for methyl-parathion among the aqueous environments from the middle levels of concentrations to high levels as shown in Table 8. It can be noted that lower Rf values may be associated with the increase on the occurrence of matrix effect.

The influence of organic matter of aquatic environments from various origins on the pesticide recoveries has implications and correlations with the total organic carbon (TOC).6161 Chiou, T.; Malcolm, L.; Brinton, I.; Kile, E.; Environ. Sci. Technol. 1986, 20 (5), 502. The difference in magnitude for Koc (mL g-1) shown in Table 3 indicates that a more pronounced matrix effect due to organic matter is expected for chlorpyriphos (8151 mL g-1) and cypermethrin (57889 mL g-1) than for the parathion-methyl (240 mL g-1).

In this case, it can be expected that an increase in parathion-methyl concentration favors its extraction, resulting in increase on recovery, because of the increased availability of the compound. Thus, in high recovery levels (300 and 500 µg L-1, respectively) the matrix effect has become less pronounced for parathion-methyl.

For parathion-methyl, the Rf value was similar in estuarine and marine water, in low levels. For chlorpyriphos, the Rf values in middle and high levels, for estuarines, seawater and weir waters were statistically similar, indicating proportional behavior between the pesticide and the surrogate recovery. For cypermethrin, the statistical equality occurred in estuarine matrices in medium and high levels of concentration. In addition, the data suggest that a single value of Rf can be used in calculations of recovery of pesticides in cited matrices.

The use of compounds as surrogate belonging to the same chemical class or group of analyte of interest can produce best results between the analyte and surrogate standard due to the similarity between both. Above all, more important is that the analyst knows the properties of the compounds being investigated so that it can make a prediction of their behavior in the matrix investigated. Thus, more efficient strategies can be designed in order to develop a methodology that meets the study requirements.

Conclusion

A real estimate of analyte concentration levels in samples should be based on an appropriate quality control. Thus, the application of adequate statistical tests and strategies that make it possible to offset the effect matrix in complex samples can contribute to a more reliable analysis.

Pesticides were very susceptible to matrix effects promoted by environmental samples. This showed the need for the use of a surrogate standard for the correction of their concentrations. Thus, the use of a relation factor seemed viable as recovery correction parameter for pesticides, although belonging to another chemical group.

Furthermore, it has been found the possibility of the same Rf can be used in matrices of the same nature but with different characteristics.

Acknowledgments

The authors thank CNPq for the financial support. Also, we thank the Laboratório de Análise de Contaminantes Orgânicos (LABOMAR-UFC) and Laboratório de Tecnologia de Biomassa - Embrapa Agroindústria Tropical.

References

  • 1
    Lushchak, V.; Kubrak, I.; Storey, M.; Storey, B.; Lushchak, I.; Chemosphere 2009, 76, 932.
  • 2
    Hiller, E.; Tatarková, V.; Šimonovičová, A.; Bartal', M.; Chemosphere 2012, 87, 437.
  • 3
    Maloschik, E.; Ernst, A.; Hegedus, G.; Darvas, B.; Székács, A.; Microchem. J. 2007, 85, 88.
  • 4
    Carlos, A.; Neves, A.; Reis, C.; Queiroz, R.; Quim. Nova 2011, 34, 272.
  • 5
    Caldas, S.; Gonçalves, F.; Primel, G.; Prestes, D.; Martins, L.; Zanella, R.; Quim. Nova 2011, 34, 1604.
  • 6
    Rizzetti, M.; Kemmerich, M.; Martins, L.; Prestes, D.; Adaime, B.; Zanella, R.; Food Chem. 2016, 196, 25.
  • 7
    Souza-Silva, É. A.; Jiang, R.; Rodríguez-Lafuente, A.; Gionfriddo, E.; Pawliszyn, J.; TrAC, Trends Anal. Chem. 2015, 71, 224.
  • 8
    Arruda, J.; Guedes, C.; Oliveira, R.; Gomes, C.; Aparecida, M.; Milhome, L.; Ferreira, R. ; Food Chem. 2016, 199, 380.
  • 9
    Oliveira, R.; Carius, R.; Castro, D.; Aparecida, M.; Milhome, L.; Ferreira, R.; LWT - Food Sci. Technol. 2014, 59, 21.
  • 10
    Pinheiro, S.; da Rocha, O.; de Andrade, B.; Microchem. J. 2011, 99, 303.
  • 11
    Paschoal, R.; Rath, S.; Airoldi, S.; Reyes, R.; Quim. Nova 2008, 31, 1190.
  • 12
    Shakouri, A.; Yazdanpanah, H.; Shojaee, H.; Kobarfard, F.; Iran. J. Pharm. Res. 2012, 13, 927.
  • 13
    Picó, Y.; Blasco, C.; Font, G.; Mass Spectrom. Rev. 2004, 23, 45.
  • 14
    IUPAC; Gold Book; Iupac: Genebra, 2014, p. 1670. Available at http://goldbook.iupac.org/M03759.html, accessed in September 2016.
    » http://goldbook.iupac.org/M03759.html
  • 15
    Kruve, A.; Künnapas, A.; Herodes, K.; Leito, I.; J. Chromatogr. A 2008, 1187, 58.
  • 16
    Kebarle, P.; Tang, L.; Anal. Chem. 1993, 65, 972A.
  • 17
    Kruve, A.; Leito, I.; Herodes, K.; Anal. Chim. Acta 2009, 651, 75.
  • 18
    Ferrer, C.; Lozano, A.; Agüera, A.; Girón, J.; Fernández-Alba, R.; J. Chromatogr. A 2011, 1218, 7634.
  • 19
    Van Eeckhaut, A.; Lanckmans, K.; Sarre, S.; Smolders, I.; Michotte, Y.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2009, 877, 2198.
  • 20
    Cavalcante, M.; Lima, M.; Fernandes, M.; Duaví, C.; Talanta 2012, 93, 212.
  • 21
    Rimayi, C.; Odusanya, D.; Mtunzi, F.; Tsoka, S.; Chemosphere 2015, 118, 35.
  • 22
    Ribani, M.; Bottoli, G.; Collins, H.; Jardim, F.; Melo, C.; Quim. Nova 2004, 27, 771.
  • 23
    Cuadros-Rodríguez, L.; Gámiz-Gracia, L.; Almansa-López, M.; Bosque-Sendra, M.; TrAC, Trends Anal. Chem. 2001, 20, 620.
  • 24
    Abdulra'uf, B.; Tan, H.; Food Chem. 2015, 177, 267.
  • 25
    Alpendurada, F.; J. Chromatogr. A 2002, 968, 177.
  • 26
    Filho, M.; dos Santos, N.; Pereira, P.; Microchem. J. 2010, 96, 139.
  • 27
    Tankiewicz, M.; Morrison, C.; Biziuk, M.; Talanta 2013, 107, 1.
  • 28
    Caulcutt, R.; Richard, B.; Statistics for Analytical Chemists; Chapman & Hall: London, 1994.
  • 29
    Brasil, MAPA/ACS; Manual de Garantia da Qualidade Analítica, 1ª ed.; Ministério da Agricultura, Pecuária e Abastecimento, Secretaria de Defesa Agropecuária: Brasília, 2011.
  • 30
    Arthur, L.; Pawliszyn, J.; Anal. Chem. 1990, 62, 2145.
  • 31
    Ouyang, G.; Pawliszyn, J.; Anal. Chim. Acta 2008, 627, 184.
  • 32
    Holopainen, S.; Luukkonen, V.; Nousiainen, M.; Sillanpää, M.; Talanta 2013, 114, 176.
  • 33
    Huang, S.; He, S.; Xu, H.; Wu, P.; Jiang, R.; Zhu, F.; Luan, T.; Ouyang, G.; Environ. Pollut. 2015, 200, 149.
  • 34
    Noventa, S.; Barbaro, J.; Formalewicz, M.; Gion, C.; Rampazzo, F.; Brusà, B.; Gabellini, M.; Berto, D.; Anal. Chim. Acta 2015, 858, 66.
  • 35
    Menezes, C.; Paulo, P.; Paiva, N.; de Barcelos, R.; Macedo, D.; Cardeal, L.; Microchem. J. 2015, 118, 272.
  • 36
    Adiani, V.; Gupta, S.; Padole, R.; Variyar, S.; Sharma, A.; Postharvest Biol. Technol. 2014, 98, 34.
  • 37
    Cecchi, T.; Alfei, B.; Food Chem. 2013, 141, 2025.
  • 38
    Lin, J.; Zhang, P.; Pan, Z.; Xu, H.; Luo, Y.; Wang, X.; Food Chem. 2013, 141, 259.
  • 39
    Moreira, A.; André, C.; Cardeal, L.; Food Chem. 2015, 178, 195.
  • 40
    Da Silva, C.; Da Silva, S.; Da Silva, N.; Godoy, O.; Nogueira, C.; Quitério, L.; Raices, L.; Food Chem. 2015, 167, 71.
  • 41
    Dawidowicz, L.; Szewczyk, J.; Dybowski, P.; Talanta 2016, 146, 195.
  • 42
    Miró, P.; Buxaderas, S.; Food Res. Int. 2014, 57, 196.
  • 43
    Soto, C.; Maldonado, B.; Jofré, P.; Galmarini, R.; Silva, F.; Microchem. J. 2015, 122, 110.
  • 44
    Aleksa, K.; Walasek, P.; Fulga, N.; Kapur, B.; Gareri, J.; Koren, G.; Forensic Sci. Int. 2012, 218, 31.
  • 45
    Brown, D.; Rhodes, J.; Pritchard, J.; Forensic Sci. Int. 2007, 171, 142.
  • 46
    Vazquez-Roig, P.; Blasco, C.; Picó, Y.; TrAC, Trends Anal. Chem. 2013, 50, 65.
  • 47
    Aresta, A.; Bianchi, D.; Calvano, D.; Zambonin, G.; J. Pharm. Biomed. Anal. 2010, 53, 440.
  • 48
    Olszowy, P.; Szultka, M.; Fuchs, P.; Kegler, R.; Mundkowski, R.; Miekisch, W.; Schubert, J.; Buszewski, B.; J. Pharm. Biomed. Anal. 2010, 53, 1022.
  • 49
    Olszowy, P.; Szultka, M.; Ligor, T.; Nowaczyk, J.; Buszewski, B.; J. Chromatogr. B: Anal. Technol. Biomed. Life Sci. 2010, 878, 2226.
  • 50
    Zhao, S.; Liu, L.; Chen, F.; Yuan, P.; Bai, Y.; Zhou, B.; Anal. Chim. Acta 2013, 769, 65.
  • 51
    www.mma.gov.br/port/conama/res/res05/res35705.pdf, accessed in August 2016.
    » www.mma.gov.br/port/conama/res/res05/res35705.pdf
  • 52
    http://www.mma.gov.br/port/conama/legislacao/CONAMA_RES_CONS_2008_396.pdf, accessed in August 2016.
    » http://www.mma.gov.br/port/conama/legislacao/CONAMA_RES_CONS_2008_396.pdf
  • 53
    Chao, P.; Lu, T.; Yang, W.; Int. J. Mol. Sci. 2014, 15, 2585.
  • 54
    Difilippo, L.; Eganhouse, P.; Environ. Sci. Technol. 2010, 44, 6917.
  • 55
    FAO; http://www.fao.org/docrep/w5715e/w5715e03.htm, accessed in September 2016.
    » http://www.fao.org/docrep/w5715e/w5715e03.htm
  • 56
    http://sitem.herts.ac.uk/aeru/iupac/, acessed in August 2016.
    » http://sitem.herts.ac.uk/aeru/iupac/
  • 57
    Li, X.; Gan, P.; Peng, R.; Huang, C.; Yu, H.; J. Chromatogr. Sci. 2010, 48, 183.
  • 58
    Lambropoulou, D.; Sakellarides, T.; Albanis, T.; Fresenius J. Anal. Chem. 2000, 368, 616.
  • 59
    Pinho, P.; Silvério, O.; Neves, A.; Queiroz, R.; Starling, M.; Quim. Nova 2010, 33, 909.
  • 60
    Albanis, A.; Hela, G.; J. Chromatogr. A 1995, 707, 283.
  • 61
    Chiou, T.; Malcolm, L.; Brinton, I.; Kile, E.; Environ. Sci. Technol. 1986, 20 (5), 502.
  • 62
    Lambropoulou, A.; Sakkas, A.; Albanis, A.; Anal. Bioanal. Chem. 2002, 374, 932.
  • 63
    Schroeder, A.; Wiesner, G.; Liu, Z.; Earth Planet. Sci. Lett. 2015, 430, 30.
  • 64