Pre-concentration of lithium prior to its determination in honey by flame optical emission spectrometry

Abstracts

A procedure based on solid phase extraction with a strongly acidic cation-exchanger Dowex 50W×8-200 was proposed to pre-concentrate and determine ultra-traces of Li in different floral honeys. When loading 10% (m/v) honey solutions onto columns packed with the cation-exchanger, Li+ ions were retained on the resin along with K+ and Na+, while simple sugars were passed through columns and were entirely found in effluents. In turn, columns were washed with water and eluted with a 1.0 mol L-1 HCl solution to strip Li+ that was subsequently determined in respective eluates by flame optical emission spectrometry (FOES). The proposed method was characterized by a limit of detection (LOD) of 0.05 ng g-1 of Li and later used for analysis of several raw and commercial mono- and multi-flower honeys. It enabled to determine from 0.4 to 5.3 ng g-1 of Li with precision being within 8-11%. Accuracy of this sample pre-treatment procedure and analysis method was assessed applying spike recovery tests. Recoveries of added Li were found to be within 98 to 106%, demonstrating trustworthiness of results achieved.

lithium; honey; pre-concentration; determination; flame optical emission spectrometry


Foi proposto um procedimento para a pré-concentração e determinação de traços de lítio em diferentes méis florais, empregando para a extração contínua em fase sólida uma resina de troca catiônica fortemente ácida, a Dowex 50W×8-200. Ao se eluir as soluções de méis a 10% em massa, na coluna empacotada com a resina de troca catiônica, os íons Li+, K+ e Na+ foram retidos na resina enquanto que os açúcares simples não foram retidos e se encontraram integralmente nos eluatos. Os cátions Li+, Na+ e K+ foram eluídos com uma solução de HCl 1,0 mol L-1 após a lavagem das colunas com água; os analitos presentes nesse eluato foram determinados por espectrometria de emissão ótica com chama (FOES). O limite de detecção instrumental (LOD) apresentado pelo método foi de 0,05 ng g-1 de lítio e o método foi empregado para a quantificação de Li em diversos méis naturais e comerciais. A faixa de concentração ótima para a determinação de Li foi de 0,4 a 5,3 ng g-1 com precisão de 8-11%. A exatidão do método proposto foi avaliada por estudos de recuperação do analito e está no intervalo de 98 a 106%, demonstrando a confiabilidade dos resultados obtidos.


ARTICLE

Pre-concentration of lithium prior to its determination in honey by flame optical emission spectrometry

Helena Stecka; Pawel Pohl* * e-mail: pawel.pohl@pwr.wroc.pl

Department of Analytical Chemistry, Faculty of Chemistry, Wroclaw University of Technology, Wybrzeze Stanislawa Wyspianskiego 27, 50-370 Wroclaw, Poland

ABSTRACT

A procedure based on solid phase extraction with a strongly acidic cation-exchanger Dowex 50W×8-200 was proposed to pre-concentrate and determine ultra-traces of Li in different floral honeys. When loading 10% (m/v) honey solutions onto columns packed with the cation-exchanger, Li+ ions were retained on the resin along with K+ and Na+, while simple sugars were passed through columns and were entirely found in effluents. In turn, columns were washed with water and eluted with a 1.0 mol L-1 HCl solution to strip Li+ that was subsequently determined in respective eluates by flame optical emission spectrometry (FOES). The proposed method was characterized by a limit of detection (LOD) of 0.05 ng g-1 of Li and later used for analysis of several raw and commercial mono- and multi-flower honeys. It enabled to determine from 0.4 to 5.3 ng g-1 of Li with precision being within 8-11%. Accuracy of this sample pre-treatment procedure and analysis method was assessed applying spike recovery tests. Recoveries of added Li were found to be within 98 to 106%, demonstrating trustworthiness of results achieved.

Keywords: lithium, honey, pre-concentration, determination, flame optical emission spectrometry

RESUMO

Foi proposto um procedimento para a pré-concentração e determinação de traços de lítio em diferentes méis florais, empregando para a extração contínua em fase sólida uma resina de troca catiônica fortemente ácida, a Dowex 50W×8-200. Ao se eluir as soluções de méis a 10% em massa, na coluna empacotada com a resina de troca catiônica, os íons Li+, K+ e Na+ foram retidos na resina enquanto que os açúcares simples não foram retidos e se encontraram integralmente nos eluatos. Os cátions Li+, Na+ e K+ foram eluídos com uma solução de HCl 1,0 mol L-1 após a lavagem das colunas com água; os analitos presentes nesse eluato foram determinados por espectrometria de emissão ótica com chama (FOES). O limite de detecção instrumental (LOD) apresentado pelo método foi de 0,05 ng g-1 de lítio e o método foi empregado para a quantificação de Li em diversos méis naturais e comerciais. A faixa de concentração ótima para a determinação de Li foi de 0,4 a 5,3 ng g-1 com precisão de 8-11%. A exatidão do método proposto foi avaliada por estudos de recuperação do analito e está no intervalo de 98 a 106%, demonstrando a confiabilidade dos resultados obtidos.

Introduction

Lithium in honey is present in trace concentrations and can primarily reach it from soil through nectar and pollen of blossom plants or secretions of living parts of plants. The content of this metal is highly variable and greatly depends on honey provenance as demonstrates data found in literature and related to concentration ranges of Li in honeys of different floral origin and from different countries, e.g., from not detected (ND) to 0.24 μg g-1 in French honeys,1 from 0.30 to 1.5 μg g-1 in Turkish honeys2 or from ND to 110 μg g-1 in Spanish honeys.3-9 Because Li, like other alkaline or alkaline earth metals present in honey, is associated with soil characteristics and specificity, and agricultural practices, it certainly well reflects geographical origin and different locality conditions.4-6 Accordingly, Li was recognized to be a useful feature accounting for successful discrimination between honeys from different regions, i.e., coastal and mountain territories, rural and urban areas. For the same reason, Li was also found to be a good descriptor for classification between honeys of a stated origin and those originated from an industrial processing.7,8

It appears however that Li is occasionally determined in honeys what proves a limited number of works reported so far and devoted to this metal. Inductively coupled plasma optical emission spectrometry (ICP OES) is commonly used for measuring concentrations of Li in honeys, typically along with other metals.1-6 Flame optical emission spectrometry (FOES), although more inexpensive than ICP OES is curiously rarely used.7-10 In case of both mentioned techniques, samples of honey are usually calcined, followed by dissolving resulted ash in solutions of mineral acids3-6 or digested in concentrated oxidizing reagents1,2 before measuring Li concentrations.

Flame OES with C2H2-air flame seems to be quite expedient technique for measurements of Li with low limit of detection (LOD).11 In contrast to K and Na, Li is hardly ionized in flame due to relatively high ionization potential (5.39 eV), hence, it could be expected that high concentrations of primarily K and Na as well as Ca and Mg would have a reduced effect on Li determinations. By contrast, high organic matter contents could deteriorate the analytical response for Li and result in high background emission levels.12

For that reason, a solid phase extraction (SPE) procedure was developed in the present work to overcome interferences coming from simple sugars and pre-concentrate traces of Li before its determination in honeys by means of FOES. In addition, it was expected that this procedure would be a good alternative to commonly used acid decomposition or calcinations of honey.1-6 In advance, the effect of various concentrations of K, Na, glucose and fructose, being the most important constituents of honeys, on the Li response was studied. The sorption behavior of Li and mentioned constituents toward different sorption materials, including an adsorbing resin Amberlite XAD-7HP, a strongly basic anion-exchange resin Dowex 1×4-200, a strongly acidic cation-exchange resin Dowex 50W×8-200 and a mixed bed ion-exchange resin Dowex Retardion 11A8, was examined as well. For selected strong cation-exchanger, retention and recovery conditions for Li were optimized. Validity of a Dowex 50W×8-200 column SPE-based procedure was verified using spike recovery tests. Finally, this procedure was applied to determine Li in several raw and commercially marked mono- and multi-flower honeys.

Experimental

Reagents and materials

All reagents used in this work were of analytical grade. Chemicals were supplied by POCh (Gliwice, Poland). Ultra-pure water, obtained from a WIGO PRO-11G reverse osmosis water purification system (Wroclaw, Poland) was used throughout. Single element stock standard solutions of 1000 mg L-1 of K, Li and Na were obtained from Merck KGaA (Darmstadt, Germany) and used to prepare mixed standard solutions of these metals for calibration. 100 mL working standard solutions, containing K (10.0 mg L-1), Li (0.1 mg L-1) and Na (1.0 mg L-1) in addition to glucose and fructose, both at 10 g L-1, were prepared to study sorption properties of polymeric resins used. These solutions were acidified using a 0.010 mol L-1 HCl solution to pH 3.5, 4.0, 4.5 and 5.0 which are close to the acidity of honey solutions.

Polymer-based resins were purchased from Sigma-Aldrich (Saint Louis, MO, USA) and included a weakly polar macro-reticular adsorbent Amberlite XAD-7HP (polyacrylic ester matrix, 20-60 mesh size, 450 m2 g-1 surface area), a strongly basic anion-exchanger Dowex 1×4-200 (styrene-divinylbenzene co-polymer matrix, 100-200 mesh size, 1.0 mequiv. g-1 total exchange capacity for -N(CH3)3+ functional group), a strongly acidic cation-exchanger Dowex 50W×8-200 (styrene-divinylbenzene co-polymer matrix, 100-200 mesh size, 1.7 meq mL-1 total exchange capacity for -SO3- functional group) and an amphoteric exchanger Dowex Retardion 11A8 (styrene-divinylbenzene co-polymer matrix, 50-100 mesh size, 1.2 and 1.1 mequiv. g-1, respectively, for -COO- and -N(CH3)3+ functional groups). Resins were packed into Sigma-Aldrich glass columns (10 mm i.d.) with glass coarse frits and polytetrafluoroethylene (PTFE) stopcocks. A Cole-Parmer 4-channel MasterFlex L/S peristaltic pump (Vernon Hill, IL, USA) was employed to control flow rates of solutions passed through SPE columns.

Instrumentation and measurements

Measurements of K, Li and Na concentrations in effluents and eluates collected, and honey sample solutions were made with a Perkin Elmer (Norwalk, CT, USA) flame atomic absorption spectrometer, model 1100B. It was equipped with a 10 cm single slot burner head for a C2H2-air combustion flame. A sampling unit integrated with the burner comprised a plastic coated mixing chamber with a flow spoiler and an end cap with a drain interlock assembly. Solutions were continuously aspirated through a mounted stainless steel nebulizer. Working conditions for instrument operation (Table 1) were selected according to recommendations of the spectrometer manufacturer given for K, Li and Na measurements in an oxidizing C2H2-air flame in emission mode. Time-average integration (a hold mode) with an integration time of 1 s and 3 read cycles was applied for signal processing. Ten replicate measurements at maximum absorption were carried out at 0.1 s intervals and averaged in each read cycle. Sample solutions were directly analyzed against aqueous standard solutions. Three replicate measurements were made and averaged.

Concentrations of glucose and fructose in column effluents collected were measured using the Somogyi-Nelson spectrophotometric method. In a two-step procedure, a sample portion containing glucose and fructose with added an alkaline solution of Cu(II) ions was heated in a boiling water bath to reduce Cu(II) to Cu(I) ions. In turn, an arsenomolybdate reagent, prepared by reacting ammonium molybdate [(NH4)6Mo7O24 ] with sodium arsenate (Na2HAsO4) in H2SO4, was added to produce a polymolybdate complex of intensive blue color due to a redox reaction with resulting Cu(I) ions.13 The absorbance of molybdenum blue produced was measured at 520 nm using a Thermo Scientific Spectronic 20D+ digital spectrophotometer (Bremen, Germany). The concentration of simple sugars was determined against suitable standard solutions containing glucose at concentrations ranging from 1 to 50 mg L-1. Alkaline Cu(II) and arsenomolybdate reagent solutions used were prepared according to reference 13.

Resin conditioning and column operation

Amberlite XAD-7HP resin was initially dried in an oven at 110 °C for 2 h. Portions containing 1.5 g of this resin were soaked with methanol and then with water, and finally poured into SPE columns as water slurries. Resin beds were conditioned by washing with 10 mL of methanol, followed by 20 mL of water. After that, they were rinsed with 10 mL of a 4.0 mol L-1 HCl solution, and again with 40 mL of water to remove excessive HCl. In case of ion-exchangers, 1.5 g portions were soaked with water and poured into SPE columns. Resin beds were washed with 10 mL of a 4.0 mol L-1 HCl solution, followed by 40 mL of water.

Thus, counter ions of functional groups of ion-exchangers were H+ and/or Cl- ions. Water and conditioning solutions were passed through SPE columns at 2.0 mL min-1.

To investigate sorption behavior of resins toward Li ions and simple sugars under pH changed from 3.5 to 5.0, working standard solutions described above were passed through SPE columns at 1.0 mL min-1. After passing about 80% of solutions, column effluents (10 mL) were collected and analyzed for K, Li, Na and glucose with fructose concentrations (Ceffluent). Recovery efficiencies of K, Li, Na and simple sugars (%) were evaluated relating concentrations of these constituents determined in effluents (Ceffluent) to their original concentrations present in loaded solutions (C) according to equation 100×Ceffluent/C. The effect of HCl concentrations (0.5, 1.0, and 2.0 mol L-1) in eluting solutions on recoveries of K, Li and Na ions from Dowex 50W×8-200 cation-exchanger was examined as well. Accordingly, 100 mL working solutions (pH 4.0) were passed through SPE columns at 6.0 mL min-1. In turn, metal ions retained were recovered using 5 or 10 mL portions of HCl solutions that were passed through SPE columns at 1.0 mL min-1. Volumes of 5 and 10 mL eluate portions were collected and analyzed for the contents of K, Li and Na (Celuate) against adequate matrix matching standard solutions. Recovery efficiencies (%) were assessed considering concentrations of K, Li and Na found in eluates (Celuate) and those in loaded solutions (C) according to equations 100 × Celuate / 20 × C or 100 × Celuate / 10 × C. Retention and recovery efficiencies were average values of three independent replicates. Respective column blanks were run and considered in calculations.

Honey samples and pre-concentration procedure

Four raw honeys, i.e., acacia flower, linden flower, rape flower and multi-flower, directly taken from holding tanks just a few days after their extraction from combs, were from an apiary located in Kostrzyn by the Oder (western Poland, Lubusz Voivodeship, Gorzow County). Additionally, four commercially available (Apis Apiculture Cooperative, Lublin, Poland) honeys of the same type were analyzed. Honeys were stored in plastic containers (raw honeys) or original glass jars in which they were dispatched (commercial honeys).

Before sampling, honeys were homogenized by vigorously stirring with glass rods. Prior to analysis, different amounts of honeys (1.0, 1.5 and 2.5 g) were dissolved in water and diluted to 100 mL. A volume of 100 mL 10% (m/v) honey solutions were loaded onto SPE columns and driven through 0.7 g Dowex 50W×8-200 resin beds at 6.0 mL min-1 to retain Li. Then, SPE columns were washed with water (10 mL) and submitted to elution with 5.0 mL of a 1.0 mol L-1 HCl solution at 1.0 mL min-1. Respective 5.0 mL eluate portions, after discarding column dead volumes, were collected and analyzed. Finally, concentrations of Li in eluates were determined using FOES versus matrix matching (containing 1.0 mol L-1 HCl) standard solutions. Procedural blanks were considered in the final calculations.

Results and Discussion

Effect of K, Na, glucose and fructose on Li response

Potassium and Na have the highest share in different Polish floral honeys,14i.e., to over 3200 μg g-1 (K) and up to 100 μg g-1 (Na), these metals are also easily ionized in C2H2-air flames. For that reason, the effect of K and Na concentrations on the analytical performance of FOES for Li was studied in detail. Concentrations of K and Na in standard working solutions were changed from 0.50 to 200 mg L-1 and from 0.05 to 10.0 mg L-1, respectively. This corresponded to 5.0% (m/v) solutions of honey in which K is present at 10 to 4000 μg g-1 while Na at 1 to 200 μg g-1. The concentration of Li in all these solutions was 0.10 mg L-1.

Determining Li concentrations using calibration with aqueous standard solutions it was found that K and Na present at studied concentration intervals practically had no effect on intensity signals obtained for the Li emission line. The average Li concentration measured in these conditions was 0.10 mg L-1 with the relative standard deviation (RSD) of 0.5% for K and 1.0% for Na.

In the same way, the separate effect of glucose and fructose as the most abundant constituents of the organic honey matrix was examined. Concentrations of glucose or fructose in standard working solutions were in the range of 3 to 120 g L-1 corresponding to 0.4 and 15.0% (m/v) solutions of honey, respectively, in which these simple sugars account for 0.80 g g-1.15 Both monosaccharides were found to have a detrimental effect on the analytical performance of FOES for Li when using calibration with aqueous standard solutions (Figure 1). Physical and chemical interferences related to increasing contents of glucose or fructose in solutions likely affected Li intensity signals since they were gradually suppressed. As a result, concentration of Li in solutions containing glucose or fructose at 30 g L-1 were found to be lower by about 6-7% than its initial concentration of 0.10 mg L-1 in these solutions. In case of solutions containing 120 g L-1 of one of simple sugars, concentrations of Li determined under these conditions were lower by 19-22% as compared to its initial concentration.


However, when allowable 1.0, 1.5 or 2.5% (m/v) solutions of honey were analyzed, it was impossible to determine Li because its concentration was lower than the respective instrumental LOD assessed for FOES (Table 1). Quite opposite, concentrations of K and Na in analyzed honeys, evaluated to be within 190-470 μg g-1 (K) and 5-170 μg g-1 (Na), were so high that prepared honey solutions required to be subsequently diluted before FOES measurements.

For that reason, it was decided to prepare concentrated honey solutions and separate Li from simple sugars matrix and/or pre-concentrate it before measurements by FOES using SPE.

Sorption properties of different polymeric resins toward K, Li, Na, glucose and fructose

The sorption behavior of four different sorbents, i.e., non-ionic Amberlite XAD-7HP, and ionic Dowex 1×4-200, Dowex 50W×8-200 and Dowex Retardion 11A8, toward K, Li and Na ions, and glucose and fructose was tested as describe above.

On the basis of these measurements (Table 2) it was possible to conclude that only strongly acidic cation-exchanger Dowex 50W×8-200 led to a quantitative separation of glucose and fructose from Li ions at pH 3.5-5.0. In these conditions, Li ions were completely retained by the resin along with K and Na ions. Considering that much more concentrated solutions of honey would have to be passed through SPE columns to pre-concentrate traces of Li, the effect of glucose and fructose concentrations on retention efficiency of Li by Dowex 50W×8-200 cation-exchanger was studied. Working standard solutions (pH 4.0) containing glucose and fructose at concentrations of 40, 60, 80, 100 and 120 g L-1 were used for that purpose.

After analysis of respective column effluents against calibration with matrix matching standard solutions containing the same concentrations of simple sugars, it was found that glucose and fructose did not affect the sorption of alkaline metals by Dowex 50W×8-200. Accordingly, K, Li and Na were not detected in collected effluents, while glucose and fructose were determined in these effluents at the same levels as in original solutions loaded onto SPE columns. In the similar way, it was ascertained that retention of Li ions as well as of K and Na ions was quantitative when working standard solutions (pH 4.0) were driven through SPE columns with Dowex 50W×8-200 at much higher flow rates, i.e., 2.0, 4.0, 6.0, 8.0 and 10 mL min-1. Additionally, it was assessed that such ions as Ca (1.0 mg L-1), Fe (0.2 mg L-1), Mg (2.0 mg L-1), Mn (0.1 mg L-1) and Zn (0.1 mg L-1), added to working standard solutions of K, Li and Na (pH 4.0), were also retained by Dowex 50W×8-200 with efficiencies ranged from 92 to 103% at different flow rates studied, i.e., 1, 2, 4, 6, 8 and 10 mL min-1.

A corresponding behavior of Dowex 50W×8-200 toward K, Li and Na ions, glucose and fructose at different flow rates was also established in case of 0.7 g resin beds. This could result from certain advantages of this gel-type resin, including a high loading capacity and a fast reactivity due to lack of discrete pores.16 Consequently, solute ions easily diffuse through particles to interact with ion exchange sites.

Finally, 0.5, 1.0 and 2.0 mol L-1 HCl solutions were tested for recoveries of Li from 0.7 and 1.5 g resin beds of Dowex 50W×8-200 applying 5.0 and 10 mL portions of these solutions. In case of 1.5 g resin beds it was found that 10 mL of a 2.0 mol L-1 HCl solution produce total desorption of Li ions from SPE columns (Figure 2). Potassium and Na ions were also quantitatively recovered from the resin under these conditions. For 0.7 g resin beds, already 5.0 mL portions of 1.0 or 2.0 mol L-1 HCl solutions resulted in total recovery of Li.


Analytical application, pre-concentration and determination of Li in honeys

To apply the devised Li separation/pre-concentration SPE procedure, 10% (m/v) solutions of analyzed honeys were driven at 6.0 mL min-1 through 0.7 g resin beds of Dowex 50W×8-200. After loading of honey solutions, SPE columns were washed with 10 mL of water and then, they were eluted using 5.0 mL of a 1.0 mol L-1 HCl solution. Respective eluates were collected for determination of the concentration of Li by FOES.

Results of Li determinations are given in Table 3 and they demonstrate that precision (as RSD) of analysis with this method is satisfactory; with typical variations from 8 to 12%, while in case of very low concentrations of Li it is within 20-25%. To evaluate the accuracy of this method of Li pre-concentration and determination in honey, 10% (m/v) solutions of honeys were spiked with Li so as to double its original content. Next, these sample solutions were treated with Dowex 50W×8-200 SPE columns. It was found that recoveries of added Li were within 98.0-106%. Additionally, eluates achieved were analyzed by flame atomic absorption spectrometry (FAAS) using the same instrument and against calibration with matrix matching solutions. It was found that comparable concentrations of Li were determined; differences in results were lower than 20%. This proves the reliability of the proposed sample preparation SPE-based procedure and FOES analysis method. The established method LOD for Li (3×SD of average intensity signal for a SPE procedural blank resulted from passing 100 mL of water (pH 4.0) through a Dowex 50W×8 column and eluting the resin bed with 5 mL of a 1.0 mol L-1 HCl solution, n = 3) was 0.05 ng g-1. This LOD also considered the honey concentration in an analyzed 10% (m/v) solution and the enrichment factor. Considering analysis of a 2.5% (m/v) honey solution, the enrichment factor was 80, i.e., (100 mL / 5 mL) × (10.0 g / 2.5 g).

Conclusions

Providing a fast SPE procedure based on use of a gel-type strongly acidic cation-exchanger Dowex 50W×8-200, it was possible to precisely and accurately determine Li in honeys at a ng g-1 level by using FOES after its pre-concentration and separation from simple sugars. Considering the retention behavior of the exchanger used, the method proposed could be even simplified when applying syringe cartridges or implementing the procedure into a flow-injection analysis mode.

Submitted: May 18, 2010

Published online: December 14, 2010

  • 1. Devillers, J.; Dore, J. C.; Marenco, M.; Poirier Duchene, F.; Galand, N.; Viel, C.; J. Agric. Food Chem. 2002, 50, 5998.
  • 2. Bagci, Y.; Arslan, D.; Ozcan, M. M.; Dursun, N.; Int. J. Food Sci. Nutr. 2007, 58, 567.
  • 3. Terrab, A.; Recamales, A. F.; Gonzalez Miret, M. L.; Heredia, F. J.; Food Chem. 2005, 92, 305.
  • 4. Terrab, A.; Hernanz, D.; Heredia, F. J.; J. Agric. Food Chem. 2004, 52, 3441.
  • 5. Gonzalez Miret, M. L.; Terrab, A.; Hernanz, D.; Fernandez Recamales, M. A.; Heredia, F. J.; J. Agric. Food Chem. 2005, 53, 2574.
  • 6. Rodriguez Garcia, J. C.; Rodriguez Iglesias, R.; Pena Crecente, R. M.; Garcia, J. B.; Garcia Martin, S.; Herrero Latorre, C.; J. Agric. Food Chem. 2006, 54, 7206.
  • 7. Latorre, M. J.; Pena, R.; Pita, C.; Botana, A.; Garcia, S.; Herrero, C.; Food Chem. 1999, 66, 263.
  • 8. Latorre, M. J.; Pena, R.; Garcia, S.; Herrero, C.; Analyst 2000, 125, 307.
  • 9. Hernnandez, O. M.; Fraga, J. M. G.; Jimenez, A. I.; Jimenez, F.; Arias, J. J.; Food Chem. 2005, 93, 449.
  • 10. Nasiruddin Khan, M.; Qaiser, M.; Mubashir Raza, S.; Rehman, M.; Int. J. Food Sci. Technol. 2006, 41, 775.
  • 11. Welz B.; Sperling, M.; Atomic Absorption Spectrometry, 3rd ed., Wiley-VCH Verlag GmbH.: Weinheim, 1999.
  • 12. Price, W. J.; Spectrochemical Analysis by Atomic Absorption, Heyden & Son Ltd.: London, 1979.
  • 13. BeMiller, J. N.; Low, N. H. In Food Analysis; Nielsen, S. S., ed.; Aspen Publishers: Gaithersburg, 1998, pp. 167-188.
  • 14. Chudzinska M.; Baralkiewicz, D.; Food Chem. Toxicol. 2010, 48, 284.
  • 15. Fournier, E. In Handbook of Food Analytical Chemistry; Wrolstad, R. E.; Acree, T. E.; Decker, E. A.; Penner, M. H.; Reid, D. S.; Schwartz, S. J.; Shoemaker, C. F.; Smith, D.; Sporns P., eds.; John Wiley & Sons: New York, 2005, pp. 651-660.
  • 16. Zagorodni, A. A.; Ion Exchange-Materials, Properties and Applications, Elsevier: Amsterdam, 2007.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      14 Apr 2011
    • Date of issue
      2011

    History

    • Reviewed
      14 Dec 2010
    • Received
      18 May 2010