Non-Enzymatic Amperometric Determination of Glucose by CuO Nanobelt Graphene Composite Modified Glassy Carbon Electrode

Abstract

We have successfully developed for the first time an electrochemical sensor for glucose by using a simple and effective design of growing CuO nanobelts graphene composites (CuO@G). The as prepared composites were characterized by X-ray diffraction studies (XRD), field emission scanning electron microscopy (FESEM) and transmission electron microscopy (TEM). The prepared composite material exhibits higher electrocatalytic activity for the oxidation of glucose compared to chemically reduced graphene (C rGO) and bare CuO nanobelts. The optimized protocol has linearity with current response for glucose concentration in the range 0.5-6.5 μmol L−1 and detection limits were found to be 0.05 μmol L−1 (3σ). The method exhibits good reproducibility and selectivity towards glucose in the presence of other electro active compounds that are normally present in blood serum like dopamine, ascorbic acid, uric acid and sodium chloride. The protocol has been successfully applied for glucose determination in clinic human blood serum samples.

Keywords
graphene; CuO nanobelts; non-enzymatic sensor; glucose


Introduction

The development of reliable, fast and cheap sensing devices for monitoring of glucose is of considerable importance. This is because the blood glucose level is the basis of diagnosis and treatment of diabetes mellitus, a global health care problem that seriously affects normal life of hundreds of millions of people.1Hellers, A.; Feldman, B.; Chem. Rev. 2008, 108, 2482. Besides the clinical diagnosis, the glucose detection finds applications in the food industries, fermentation analysis, textile industries and bionanotechnology.1Hellers, A.; Feldman, B.; Chem. Rev. 2008, 108, 2482. Commercial pocket sized glucose sensors based on glucose oxidase electrodes have been used by diabetic patients for monitoring their blood glucose levels at any time conveniently.1Hellers, A.; Feldman, B.; Chem. Rev. 2008, 108, 2482. But enzyme based electrodes have several drawbacks like chemical (sensitive to oxygen concentration in the environment) and thermal instability, high cost, complicated immobilization procedures and critical operating situations.2Wilson, R.; Turner, A. P. F.; Biosens. Bioelectron. 1992, 7, 165. In recent years, the development of enzyme less electrochemical sensors for glucose monitoring utilizing various nano particles of metals (Au, Pt, Pd, Cu and Ni), metal oxides (CuO and NiO) and metal alloys (Pt-Pb, Pt-Ir, Pt-Au) as electrode materials are becoming popular due to their practical applicability compared to the enzyme based electrodes.3Si, P.; Huang, Y.; Wanga, T.; Ma, J.; RSC Adv. 2013, 3, 3487. Among these, copper oxide (CuO) a p-type semiconductor with a narrow band gap of 1.2 eV has been extensively studied due to its applications in various fields like catalysis, field-effect transistors and biosensors.4Umar, A.; Rahman, M. M.; Hajry, A. A.; Hahn, Y. B.; Electrochem. Commun. 2009, 11, 278. Many reports are based on various nano structures (nanowires, nanoribbons, nanocubes, nanofibers, nanospindles, nanospheres, nanoflowers and nanourchins) of CuO for amperometric glucose determination.5Wang, X. J.; Wang, G. F.; Liu, X. W.; Wu, J. J.; Li, M.; Gu, J.; Liu, H.; Fang, B.; J. Phys. Chem. A 2008, 112, 16845.

Zhang, L.; Li, H.; Ni, Y. H.; Li, J.; Liao, M.; Zhao, G. C.; Electrochem. Commun. 2009, 11, 812.

Wang, W.; Zhang, L. L.; Tong, S. F.; Li, X.; Song, W. B.; Biosens. Bioelectron. 2005, 25, 708.

Wang, X.; Hu, C. G.; Liu, H.; Du, G. J.; Xi, X. S.; Sens. Actuators, B 2010, 144, 220.

Li, C. I.; Su, Y.; Zhang, S. W.; Lv, X. Y.; Xia, H. L.; Wang, Y. Z.; Biosens. Bioelectron. 2010, 26, 903.
-1010 Sun, S.; Zhang, X.; Sun, Y.; Yang, S.; Song, X.; Yang, Z.; ACS Appl. Mater. Interfaces 2013, 5, 4429. In recent years, metal oxide‑carbon nanotube (single walled/multi walled carbon nanotubes) composites are becoming promising materials in various fields like electrocatalysis, electronics and fuel cells. This may be due to enhanced properties of the composite material due to the high surface area and excellent electron transfer properties of the base material.1111 He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767. Some of the composite materials prepared using noble metal (Pd, Pt and Au) and carbon nanotube are reported for glucose sensing.1212 Meng, L.; Jin, J.; Yang, G. X.; Lu, T. H.; Zhangand, H.; Cai, C. X.; Anal. Chem. 2009, 81, 7271.,1313 Xiao, F.; Zhao, F. Q.; Mei, D. P.; Mo, Z. R.; Zeng, B. Z.; Biosens. Bioelectron. 2009, 24, 3481. These noble metals are highly expensive, hence non-noble metalic and more earth abundant metals like Cu in the form of copper oxide and carbon nanotube composites has been reported for glucose sensing.1414 Yang, Z.; Feng, J.; Qiao, J.; Yan, Y.; Yu, Q.; Sun, K.; Anal. Methods 2012, 4, 1924.

15 Myung, Y.; Jang, D. M.; Cho, Y. J.; Kim, H. S.; Park, J.; J. Phys. Chem. C 2009, 113, 1251.

16 Zhang, X.; Wang, G.; Zhang, W.; Wei, Y.; Fang, B.; Biosens. Bioelectron. 2009, 24, 3395.

17 Wu, H. X.; Cao, W. M.; Li, Y.; Liu, G.; Wen, Y.; Yang, H. F.; Yang, S. P.; Electrochim. Acta 2010, 55, 3734.

18 Yang, J.; Zhang, W. D.; Gunashekaran, S.; Biosens. Bioelectron. 2010, 26, 279.

19 Kang, X. H.; Mai, Z. B.; Zou, X. Y.; Cai, P. X.; Mo, J. Y.; Anal. Biochem. 2007, 363, 143.

20 Ziang, L. C.; Zhang, W. D.; Biosens. Bioelectron. 2010, 25, 1402.
-2121 Jiang, F.; Wang, S.; Lin, J. J.; Jin, H. L.; Zhang, L. J.; Huang, S. M.; Wang, J. C.; Electrochem. Commun. 2011, 13, 363.

Graphene, a 2D unique planar structure and extensively used in the field of electrochemistry due to its exceptional electronic, thermal and mechanical properties. Further, the electron transfer rate can be accelerated with the addition of different metal nanoparticles (MNPs). In recent years, designing of novel composites materials for bio sensing application, using metal nanoparticles modified graphene has become a hot topic. Some of the attempts have been made to grow CuO nanoparticle graphene composites for glucose sensing.2222 He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767.

23 Qian, Y.; Ye, F.; Xu, J.; Le, Z. G.; Int. J. Electrochem. Sci. 2012, 7, 10063.

24 Luo, J.; Zhang, H. Y.; Jiang, S. S.; Jiangand, J. Q.; Liu, X. Y.; Microchim. Acta 2012, 177, 485.

25 Ritz, E.; Jia, W. Z.; Gentile, M.; Wang, Y.; Lei, Y.; Electroanalysis 2008, 20, 2482.

26 Luo, J.; Jiang, S.; Zhang, H.; Jiang, J.; Liu, X.; Anal. Chim. Acta 2012, 709, 53.

27 Hsua, Y. W.; Hsua, T. K.; Suna, C. L.; Nienc, Y. T.; Pud, N. W.; Gere, M. D.; Electrochim. Acta 2012, 82, 152.

28 Chen, Q.; Zhang, L.; Chen, G.; Anal. Chem. 2012, 84, 171.
-2929 Wang, X.; Liu, E.; Zhang, X.; Electrochim. Acta 2014, 130, 253. Zhenget al.2222 He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767. proposed sonication assisted electro deposition of copper nano flower graphene composite in the presence of ethaline on to glassy carbon (GC) and used for glucose sensing. Qian et al.2323 Qian, Y.; Ye, F.; Xu, J.; Le, Z. G.; Int. J. Electrochem. Sci. 2012, 7, 10063. reported synthesis of copper oxide nanoparticle graphene composite by in situ reduction and its application to glucose sensing using sodium citrate as reducing agent. Luo et al.2424 Luo, J.; Zhang, H. Y.; Jiang, S. S.; Jiangand, J. Q.; Liu, X. Y.; Microchim. Acta 2012, 177, 485. reported in situ electro chemical reduction and synthesis of copper nanoparticle graphene modified electrode using graphene oxide (GO) and copper ions. Ritz et al.2525 Ritz, E.; Jia, W. Z.; Gentile, M.; Wang, Y.; Lei, Y.; Electroanalysis 2008, 20, 2482.reported CuO nano cubes graphene nano composite modified electrode in which graphene was casted on GC and then electrochemical oxidation was performed for deposition of CuO nano cubes. Luoet al.2626 Luo, J.; Jiang, S.; Zhang, H.; Jiang, J.; Liu, X.; Anal. Chim. Acta 2012, 709, 53. reported Cu nanoparticle graphene sheet modified glassy carbon in which chemically reduced graphene (C rGO) was prepared and drop casted on the GC and then Cu nanoparticles were electrochemically coated on to the GC. Hsua et al.2727 Hsua, Y. W.; Hsua, T. K.; Suna, C. L.; Nienc, Y. T.; Pud, N. W.; Gere, M. D.; Electrochim. Acta 2012, 82, 152. reported CuO/graphene nanocomposites as glucose sensor. This protocol involves the chemical reduction of GO to graphene followed by the preparation of CuO/graphene nanocomposites. Chenet al.2828 Chen, Q.; Zhang, L.; Chen, G.; Anal. Chem. 2012, 84, 171. reported graphene-copper nanoparticle composite by in situ chemical reduction. In this protocol, GO and CuII ions were treated with potassium borohydride in order to get graphene copper nanoparticle composites. Wanget al.2929 Wang, X.; Liu, E.; Zhang, X.; Electrochim. Acta 2014, 130, 253.reported copper oxide-reduced graphene oxide nanocomposites synthesized from water-isopropanaol medium in which GO and CuII were heated in the water-isopropanaol mixture. Some of the methods mentioned above involve two steps for preparation of CuO or Cu nanoparticle graphene composites.2222 He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767.

23 Qian, Y.; Ye, F.; Xu, J.; Le, Z. G.; Int. J. Electrochem. Sci. 2012, 7, 10063.

24 Luo, J.; Zhang, H. Y.; Jiang, S. S.; Jiangand, J. Q.; Liu, X. Y.; Microchim. Acta 2012, 177, 485.

25 Ritz, E.; Jia, W. Z.; Gentile, M.; Wang, Y.; Lei, Y.; Electroanalysis 2008, 20, 2482.

26 Luo, J.; Jiang, S.; Zhang, H.; Jiang, J.; Liu, X.; Anal. Chim. Acta 2012, 709, 53.
-2727 Hsua, Y. W.; Hsua, T. K.; Suna, C. L.; Nienc, Y. T.; Pud, N. W.; Gere, M. D.; Electrochim. Acta 2012, 82, 152. Even though in situ reduction and formation of CuO or Cu nanoparticle graphene sheets are also reported, but these procedures involve use of organic solvents like isopropanaol or harsh reducing agents like potassium borohydride.2828 Chen, Q.; Zhang, L.; Chen, G.; Anal. Chem. 2012, 84, 171.,2929 Wang, X.; Liu, E.; Zhang, X.; Electrochim. Acta 2014, 130, 253.Qianet al.2323 Qian, Y.; Ye, F.; Xu, J.; Le, Z. G.; Int. J. Electrochem. Sci. 2012, 7, 10063. protocol is interesting, it involves single step and simple salt like sodium citrate is used as reducing agent in aqueous media but protocol requires 24 h for complete conversion of GO to reduced GO (rGO) and to get CuO graphene composites.

In order to overcome the above said problems, simple sonochemical protocol involving in situ reduction and growth of CuO nanobelts reduced graphene oxide in aqueous media has been reported. Recently, the chemically reduction of GO to rGO using NaOH has been proposed.3030 Zhou, H.; Wang, X.; Yu, P.; Chen, X.; Mao, L.; Analyst 2012, 137, 305. This method is found to be easier and greener compared to earlier chemical methods. Based on this, we are proposing a new method where NaOH and surfactant (cetyltrimethyammonium bromide) medium has been used for growing CuO nanobelts as well as for conversion of GO to graphene. NaOH and surfactant (cetyltrimethyammonium bromide, CTAB) medium assists in growing CuO nanobelts sonochemically.3131 Chen, Y. J.; Meng, F.; Yu, H.; Zhu, C.; Wang, T.; Gao, P.; Ouyang, Q.; Sens. Actuators, B 2013, 176, 15. The same medium is used for conversion of GO to graphene by conventional heating in which NaOH assist in reduction of oxygen moieties of GO.3030 Zhou, H.; Wang, X.; Yu, P.; Chen, X.; Mao, L.; Analyst 2012, 137, 305.

Herein, a facile and straight forward strategy is designed based on sonochemical method in which in situ reduction of GO and growing of CuO nanobelts on reduced graphene sheets in the presence of CTAB and sodium hydroxide. As prepared CuO nanobelts@G showed higher sensitivity and excellent selectivity than C rGO or CuO nanobelts in alkaline media for electrochemical glucose sensing. To the best of our knowledge, only a little research on the fabrication of CuO graphene nano hybrids in a sodium hydroxide system has been reported so far.

Experimental

Chemicals and reagents

Graphite flakes were purchased from Sigma Aldrich. D-(+)-glucose, dopamine (DA), ascorbic acid (AA), uric acid (UA), NaCl, NaOH and Cu(NO3)2, potassium permanganate (99%) were purchased from SD Fine chemicals. Cetyltrimethylammonium bromide (CTAB), K2HPO4, KH2PO4, sodium nitrate (≥ 98%), hydrogen peroxide (≥ 30%), sulphuric acid (95-98%) and ethanol (synthesis grade) were purchased from Merck. All the other reagents were of analytical grade and used without further purification. Double distilled ultrapure water with an electrical resistance > 18.3 MΩ was used for all reagent preparations. Phosphate buffer solutions were prepared from the 0.1 mol L−1 K2HPO4 and KH2PO4 solutions.

Instrumentation

Electrochemical measurements were performed at a standard three-electrode electrochemical cell with a CHI 800 electrochemical work station (CH Instrument, USA) with a CuO nanobelts graphene composite modified glassy carbon electrode as the working electrode, a platinum wire as the counter electrode and Ag/AgCl as the reference electrode. A field emission scanning electron microscopy (FESEM) image was prepared using a JEOL MICROSCOPE (JSM-7000F). A transmission electron microscopy (TEM) was done with a JEOL JEM-2010 F. X-ray diffraction analysis (XRD) was recorded on a Smart APEXCCD, Bruker, Germany. All pH measurements were recorded using digital pH meter MK VI of Systronics make.

Preparation of GO

Graphite oxide was synthesized from oxidation of graphite powder by using modified Hummers and Offmann method.3232 Hummers, W. S.; Offeman, R. E.; J. Am. Chem. Soc. 1958, 80, 1339. In a typical procedure, 3 g of graphite powder, 1.5 g of NaNO3 and 69 mL of concentrated H2SO4 were mixed by constant stirring and the mixture was cooled to below 0 °C in an ice bath. Then, 9 g of KMnO4 was gradually added under vigorous stirring and the temperature was maintained below 10 °C. The reaction mixture was stirred for 60 min at room temperature followed by a slow addition of 138 mL of deionised water with stirring. The temperature of the mixture was increased to 98 °C and maintained at this temperature for 15 min. Finally, the reaction was terminated by adding 3 mL of H2O2followed by continuous stirring for another 30 min at room temperature. The resultant precipitate was washed with 5% HCl followed by distilled water until the supernatant of the solution become neutral. The obtained graphite oxide was sonicated in deionised water (1 mg mL−1) for 30 min to get GO, then centrifuged and dried at room temperature.

Preparation of (CuO nanobelts@G) composite

In a typical procedure, 50 mg GO, 25 mg Cu (NO3)2, 100 mL distilled water and 100 mL EtOH were mixed in a 1000 mL beaker. Then, 8 mL NaOH aqueous solution (0.5 mol L−1) and 0.1 g of CTAB were added slowly with vigorous stirring. Then the mixture was irradiated with an ultrasonication for 1 h at room temperature. It was then removed from the ultrasonic bath and refluxed for 4 h at 80 °C. Finally, the obtained precipitates were collected by centrifugation and washed several times with absolute ethanol and distilled water. The precipitates were then dried at 60 °C for 12 h before morphology characterization and fabrication of sensor. For comparison, C rGO was also synthesized by chemical reduction. Similarly, the CuO nanobelts were synthesized by reported procedure.3030 Zhou, H.; Wang, X.; Yu, P.; Chen, X.; Mao, L.; Analyst 2012, 137, 305.

Fabrication of (CuO nanobelts @G) composite modified glassy carbon electrode

Glassy carbon electrode (3 mm diameter) was polished on a micro cloth with 1.0, 0.3 and 0.05 micron alumina slurry for 5 min, followed by sonication in double distilled water and ethanol for 5 min respectively. 1 mg of CuO nanobelts@G was uniformly dispersed in the solution (0.5 mL water + 0.5 mL EtOH + 10 µL of 5% nafion) by sonicating for 30 min. 20 µL of above solution was casted on to the pre-treated glassy carbon and dried at room temperature. Similar procedure was applied to get chemically reduced graphene glassy carbon electrode as well as CuO nanobelt modified glassy carbon electrode. Before the experiment, the electrolytic solution was purged with nitrogen gas in order to remove dissolved oxygen. Working electrode was cycled 20 times in 0.1 mol L−1 NaOH solution between 0 and 0.9 V through cyclic voltammograms (CV) until a reproducible background voltammograms were obtained. Amperometric experiments for glucose measurement were carried out in 0.1 mol L−1 NaOH at a desired potential (+0.55 V vs. Ag/AgCl).

Results and Discussion

The formation of CuO nanobelts by epitaxial growth process from Cu(OH)2nanowires to Cu(OH)2/CuO 1D core-shell nanostructure followed by CuO nanobelts has been reported.3131 Chen, Y. J.; Meng, F.; Yu, H.; Zhu, C.; Wang, T.; Gao, P.; Ouyang, Q.; Sens. Actuators, B 2013, 176, 15. Similarly, the reduction of GO to rGO in the presence of NaOH is reported.3030 Zhou, H.; Wang, X.; Yu, P.; Chen, X.; Mao, L.; Analyst 2012, 137, 305. Based on these two reports, we have developed new protocol for preparation of rGO-CuO nano belt composites by in situ reduction of GO and CuO nano belt formation using sonication method. In the presence of ethanol, water system and CTAB help the growth of Cu(OH)2 nano wires morphology on the GO surface. Then, epitaxial growth process of these materials resulted in the formation of Cu(OH)2/CuO 1D core-shell nanostructure. The Oswald ripening of these structures results in the formation of CuO nanobelts. Finally, refluxing the above reaction mixture at 80 °C for 4 h leads to the formation of CuO nanobelts graphene composite.

Characterization of CuO nanobelts graphene composite

XRD patterns of synthesized GO and CuO nanobelt graphene composite are shown in Figure 1a and in Figure 1b. The reflection peak (002) at 2θ = 11.88° corresponding to the interlayer basal spacing of 7.3 Å due to presence of oxygen functionalities present on the surface of GO and the reflection peak (100) at 2θ = 42.57° suggesting that GO exhibits turbostratic disorder.3333 Muralikrishna, S.; Sureshkumar, K.; Varley, T. S.; Nagaraju, D. H.; Ramakrishnappa, T.; Anal. Methods 2014, 6, 8698.,3434 Nethravathi, C.; Viswanath, B.; Michael, J.; Rajamath, M.; Carbon 2012, 50, 4839. In the case of CuO nanobelt graphene composite the reflection peak (002) shifted to 2θ = 20.18° corresponding to basal spacing of 4.32 Å. The decrease in basal spacing suggests that GO is converted into reduced graphene oxide. The peaks at 2θ values of 31.91°, 35.69°, 38.7°, 47.87°, 53.9°, 58.8°, 65.5° and 67.33° corresponding to (110), (002), (111), (202), (020), (202), (113) and (220) crystal planes of pure CuO, which is in good agreement with the reported values for the bulk CuO (JCPDS, 41-0254) indicating high degree purity and crystallinity quality.3030 Zhou, H.; Wang, X.; Yu, P.; Chen, X.; Mao, L.; Analyst 2012, 137, 305. The small intense peak at 2θ = 13° corresponds to incomplete conversion of GO to graphene and 42.6° indicates retainment of turbostratic disorder.

Figure 1
(a) XRD pattern of synthesized GO and (b) XRD pattern of synthesized CuO nanobelt graphene composite.

The morphologies and microstructures of the products were characterized by FESEM and TEM. Figures 2aand2b represents FESEM images of CuO nanobelt graphene composites. The one-dimensional (1D) nano structures, with several µm in length spread over the surface of graphene layer (Figure 2b). The EDX spectrum (Figure 2c) confirmed that the product was composed of 85.5% C, 13.13% O and 1.31% Cu (atomic %). The TEM images in Figure 3 illustrate CuO nanobelts with the length and width in the range of 38‑120 nm and 7-15 nm, respectively. The image in Figure 3b indicates the nanobelt to be of single crystal, which can be indexed to monoclinic CuO phase, which is in accordance with the XRD result.

Figure 2
(a) and (b) FESEM images and (c) EDX spectrum of CuO nano belt graphene composite.
Figure 3
(a) and (b) are the TEM images of CuO nano belts graphene composite.

Electrochemical characterization of CuO nanobelts graphene composites

Before applying the CuO nanobelts graphene composite modified glassy carbon electrode to electrocatalytic oxidation of glucose, we carried out cyclic voltammetric studies. From Figure 4, it is clear that the electrode kinetics is faster for CuO nanobelts graphene composite modified glassy carbon electrode than CuO nanobelts modified glassy carbon electrode. This indicates that graphene assists in faster electron transport in addition to providing large surface area compared to CuO nanobelts alone. These results were similar to that of several reported results regarding metal oxide graphene composites.2222 He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767.

23 Qian, Y.; Ye, F.; Xu, J.; Le, Z. G.; Int. J. Electrochem. Sci. 2012, 7, 10063.

24 Luo, J.; Zhang, H. Y.; Jiang, S. S.; Jiangand, J. Q.; Liu, X. Y.; Microchim. Acta 2012, 177, 485.

25 Ritz, E.; Jia, W. Z.; Gentile, M.; Wang, Y.; Lei, Y.; Electroanalysis 2008, 20, 2482.

26 Luo, J.; Jiang, S.; Zhang, H.; Jiang, J.; Liu, X.; Anal. Chim. Acta 2012, 709, 53.

27 Hsua, Y. W.; Hsua, T. K.; Suna, C. L.; Nienc, Y. T.; Pud, N. W.; Gere, M. D.; Electrochim. Acta 2012, 82, 152.

28 Chen, Q.; Zhang, L.; Chen, G.; Anal. Chem. 2012, 84, 171.
-2929 Wang, X.; Liu, E.; Zhang, X.; Electrochim. Acta 2014, 130, 253. The electro catalytic oxidation of glucose with different electrodes like glassy carbon electrode, CuO nanobelt modified glassy carbon electrode, C rGO modified glassy carbon electrode and CuO nanobelt graphene composite modified glassy carbon electrode were investigated by CV studies (Figure 5). The glassy carbon exhibits very low oxidation current for glucose at 0.4 V and completed at 0.8 V for 0.4 mmol glucose at scan rate of 0.01 V s−1 (Figure 5a). Whereas with the CuO nanobelt modified glassy carbon electrode under same conditions, the oxidation current increases compared to that of the former system (Figure 5b). However, in case of graphene modified glassy carbon electrode the oxidation of glucose onsets at 0.38 V and the oxidation current was found to be greater compared to that of the two former electrode systems (Figure 5c). This may be due to the faster electron transfer as well as high surface area available on the graphene for electro oxidation of glucose. The electrocatalytic oxidation of glucose with CuO nanobelts graphene composite modified glassy carbon electrode was completely different from the studied above electrodes. The oxidation of glucose onsets at 0.2 V itself and goes to completion at 0.6 V and the well defined oxidation peak centered at 0.39 V was observed (Figure 5d). The oxidation potential for glucose by earlier reported sensors were found to be more than 0.4 V.2222 He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767.

23 Qian, Y.; Ye, F.; Xu, J.; Le, Z. G.; Int. J. Electrochem. Sci. 2012, 7, 10063.

24 Luo, J.; Zhang, H. Y.; Jiang, S. S.; Jiangand, J. Q.; Liu, X. Y.; Microchim. Acta 2012, 177, 485.
-2525 Ritz, E.; Jia, W. Z.; Gentile, M.; Wang, Y.; Lei, Y.; Electroanalysis 2008, 20, 2482. With our present sensor, the over potential for oxidation of glucose shows to be decreased by about 0.2 V. This decrease in over potential may be due to the synergetic effect of both CuO nanobelts and graphene present on the glassy carbon electrode. The CuO nanobelts assisted in electrocatalytic oxidation whereas the graphene provides larger surface area for nanobelts as well as for glucose adsorption and then assists in faster electrode kinetics. Figure 6 presents the CV responses obtained at the CuO nanobelts graphene composite modified glassy carbon electrode in 0.1 mol L−1 NaOH solution containing different concentrations of glucose. In the absence of glucose, a single small broad reduction peak at 0.6 V was observed (curve a). This peak might be due to the CuII/CuIII redox couple similar to the reported literature.3535 Huang, T. K.; Lin, K. W.; Tung, S. P.; Cheng, T. M.; Chang, I. C.; Hsieh, Y. Z.; Lee, C. Y.; Chiu, H. T.; J. Electroanal. Chem. 2009, 636, 123.

36 Zhuang, Z. J.; Su, X. D.; Yuan, H. Y.; Sun, Q.; Xiao, D.; Chio, M. M. F.; Analyst 2008, 33, 126.
-3737 Casella, I. G.; Gatta, M.; Guascito, M. R.; Cataldi, T. R. I.; Anal. Chim. Acta 1997, 357, 63. Upon the addition of 0.2 mmol L−1glucose, notable enhancement of oxidative peak current corresponding to the irreversible oxidation of glucose is observed and the voltammetric response increased with a rising concentration of glucose (0.4 mmol L−1) at about 0.4 V (curves b and c). The oxidation peak for glucose was observed at about 0.37 V. The effect of potential scan rate on oxidation peak current at the CuO nanobelts graphene composite modified glassy carbon electrode has been investigated in the range 0.1 to 0.5 V s−1 in 0.1 mol L−1 NaOH containing 0.4 mmol L−1 glucose (Figure 7). From Figure 7 it is clear that anodic peak current linearly increased with scan rates, implying that the electrochemical oxidation of glucose on CuO nanobelts graphene composite modified glassy carbon electrode is a surface controlled electrochemical process. The mechanism of electrocatalytic oxidation of glucose by CuO in alkaline medium is as given below (Scheme 1).3838 Safavi, A.; Maleki, N.; Farjami, E.; Biosens. Bioelectron. 2009, 24, 1655. In the first step, the CuO is electrochemically oxidized to strong oxidizing agent CuIII species such as CuOOH or [Cu(OH)4].3939 Wang, J.; Zhang, W. D.; Electrochim. Acta 2011, 56, 7510. This species oxidizes the glucose to give hydrolyzed gluconic acid.

Figure 4
Cyclic voltammograms obtained in 0.1 mol L−1 NaOH solution of pH 7 at scan rate of 0.01V s−1 of (a) CuO nanobelts modified glassy carbon electrode and (b) CuO nanobelts graphene composite modified glassy carbon electrode.
Figure 5
Cyclic voltammograms obtained in 0.1 mol L−1 NaOH solution containing 0.4 mmol L−1 glucose at a scan rate of 0.01 V s−1 of (a) glassy carbon electrode; (b) CuO nanobelts modified glassy carbon electrode; (c) (C rGO) modified glassy carbon electrode and (d) CuO nanobelts graphene composite modified glassy carbon electrode.
Figure 6
Cyclic voltammograms of CuO nanobelt graphene modified glassy carbon electrode in 0.1 mmol L−1 NaOH solution containing different concentrations of glucose: (a) 0; (b) 0.2 and (c) 0.4 mmol L−1 at a scan rate of 0.01 V s−1.
Figure 7
Cyclic voltammograms obtained in 0.1 mol L−1 NaOH solution containing 0.4 mmol L−1 glucose at different scan rates (inner to outer 0.1, 0.2, 0.3, 0.4 and 0.5 V s−1) at CuO nanobelts graphene composite modified glassy carbon electrode. Inset shows current as a function of different scan rate.
Scheme 1
Mechanism for electro catalytic oxidation of glucose by CuO.

Optimization of reaction parameters

Various reaction parameters that affect the current response during glucose determination have been optimized. In non-enzymatic sensors, pH of the sensing medium plays a vital role in the electrocatalytic activity of the catalysts towards glucose. Hence, we systematically investigated various mediums like 0.1 mol L−1solutions of NaOH, phosphate and carbonate buffers. Under fixed reaction, conditions (applied oxidation potential 0.5 V, 1.0 µmol glucose) the amperometric responses were good in the 0.1 mol L−1 NaOH compared to other physiological mediums mentioned above (Figure S1 in the Supplementary Information (SI) section). This may be due to the less glucose anion obtained or may be the less electrocatalytic oxidation of glucose by CuO nanobelts graphene composite modified glassy carbon electrode in those media.4040 Zhou, X.; Nie, H.; Yao, Z.; Dong, Y.; Yang, Z.; Huang, S.; Sens. Actuators, B 2012, 168, 1. Then, the other important parameter studied is applied oxidation potential which was investigated in the range 0.4 to 0.75 V. The amperometric response increases in potential up to 0.55 V and then decreases. So, 0.55 V has been set as optimum oxidation potential in all further investigations (Figure S2). The electrocatalytic activity of the material also depends on the amount of composite material deposited on to the electrode. To investigate this parameter under fixed conditions (applied oxidation potential 0.55 V, 1.0 µmol L−1glucose and 0.1 mol NaOH), different aliquots of the composite dispersed in ethanol (0.01 mg L−1) were deposited and their current response was plotted against the amount of composite deposited. From Figure S3 it is clear that 20 µL of composite dispersed in ethanol-water (1 mg mL−1) was sufficient to generate enough current.

Amperometric response of the (CuO nanobelts@G) composite modified glassy carbon electrode

The amperometric response of the CuO nanobelts@G composite modified glassy carbon electrode for glucose was investigated under optimized conditions. Figure 8 illustrates that after each addition of 0.5 µmol L−1 glucose, the enhancement in the current at 0.55 V in 0.1 mol L−1 NaOH solutions was enormous. The response time was also found to be very quick and the steady state current was obtained within 5 s. These results confirm that the CuO@G nano belt hybrid exhibits excellent sensitivity and rapid response time. This may be due to the high electro conductivity and good electrocatalytic activity of the CuO nanobelts@G composite materials. These electrodes displayed a good linear range from 0.5 to 6.5 µmol L−1with a correlation coefficient of 0.9735. The limit of detection (LOD) was found to be 0.05 µmol L−1 at a signal to noise ratio of 3. The analytical parameters of the CuO nanobelts@G composite materials were compared with some of the reported CuO nano material carbon composites (Table 1).7Wang, W.; Zhang, L. L.; Tong, S. F.; Li, X.; Song, W. B.; Biosens. Bioelectron. 2005, 25, 708.,1111 He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767.,1616 Zhang, X.; Wang, G.; Zhang, W.; Wei, Y.; Fang, B.; Biosens. Bioelectron. 2009, 24, 3395.

17 Wu, H. X.; Cao, W. M.; Li, Y.; Liu, G.; Wen, Y.; Yang, H. F.; Yang, S. P.; Electrochim. Acta 2010, 55, 3734.

18 Yang, J.; Zhang, W. D.; Gunashekaran, S.; Biosens. Bioelectron. 2010, 26, 279.
-1919 Kang, X. H.; Mai, Z. B.; Zou, X. Y.; Cai, P. X.; Mo, J. Y.; Anal. Biochem. 2007, 363, 143.,2222 He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767.,2727 Hsua, Y. W.; Hsua, T. K.; Suna, C. L.; Nienc, Y. T.; Pud, N. W.; Gere, M. D.; Electrochim. Acta 2012, 82, 152.,4040 Zhou, X.; Nie, H.; Yao, Z.; Dong, Y.; Yang, Z.; Huang, S.; Sens. Actuators, B 2012, 168, 1.

41 Zhu, L.; Wang, Z.; Bioelectrochemistry 2012, 88, 156.
-4242 Male, K. B.; Hrapovic, S.; Liu, Y.; Wang, D.; Luong, J. H. T.; Anal. Chim. Acta 2004, 516, 35. The present composite material exhibits better sensitivity, selectivity, response time and linear calibration range than some of the reported methods. This better performance, which includes high electrocatalytic activity of CuO nanobelts and large active surface area originating from the electrical network formed by CuO nanobelts and graphene is attributed to the synergetic effect of CuO nanobelts and graphene.

Figure 8
Amperometric responses obtained at CuO nanobelts@G composite modified glassy carbon electrode in 0.1 mol L−1 NaOH solution by successive addition of 0.5 µmol L−1 glucose at 0.55 V vs. Ag/AgCl electrode. Inset showing corresponding calibration plot.
Table 1
Analytical parameters of some of the CuO nanoparticle composites

Interference study

In real physiological samples, for example, human blood, interfering species such as chloride ion, AA, UA and DA normally co-exist with glucose. The concentration of chloride ions is about ten times as that of glucose concentration. It is known that, many non-enzyme glucose sensors based on metal oxides, usually prevent them from being poisoned by chloride ions.4343 Chen, J.; Zhang, W. D.; Ye, J. S.; Electrochem. Commun. 2008, 10, 1268. The concentration of ascorbic acid and uric acid are about one-tenth of the glucose concentration. In order to investigate the antifouling ability of the CuO nanobelts@G composite modified glassy carbon electrode, 50 and 100 µmol L−1 of glucose were added to 0.1 mol L−1 NaOH containing 0.2 mol L−1NaCl. The electrochemical response was investigated by recording change in current. Figure 9 illustrates that the current response of the glucose was not altered even in the presence of excess of chloride ions, indicating that the proposed electrode exhibits good resistance to surface fouling. Then in the presence of 0.2 mol L−1 NaCl, the recovery of spiked samples was about 99%, which indicates that, the effect of the NaCl on the sensor material towards glucose analysis were negligible. The current response of glucose (100 µmol L−1) in 0.1 mol NaOH at 0.55 V was not disturbed even in the presence of other oxidisable species like ascorbic acid (100 µmol L−1), uric acid (100 µmol L−1) and dopamine (100 µmol L−1) and other carbohydrates (galactose, lactose and mannose (100 µmol L−1 each). Then the recovery studies of the spiked samples in the presence of these species were found to be nearly 100%. The above results indicate that the proposed electrode exhibits good selectivity towards glucose in the presence of common interferents present in blood and is promising for development of non-enzymatic glucose sensor.

Figure 9
Amperometric responses obtained at CuO nanobelts@G composite modified glassy carbon electrode in 0.1 mol L−1 NaOH solution containing 50 and 100 μmol L−1 glucose and in the presence of other oxidants like AA, UA, DA, mannose (Man), galactose (Gal), lactose (Lac) (100 μmol L−1) and NaCl (0.2 mol L−1) (applied potential 0.55 V vs. Ag/AgCl electrode).

Reproducibility and stability

Next, the reproducibility of the electrode was investigated. A set of ten different amperometric measurements were performed for 10 µmol L−1 glucose in 0.1 mol L−1 NaOH at 0.55 V with an Ag/AgCl electrode. Similar current responses were obtained in all the cases with relative standard deviation (RSD) of 2.8%. Further, 5 different electrodes were fabricated under similar conditions and RSD for the current responses for glucose (10 µmol L−1) was obtained for each measurement with a relative standard deviation of 5.0%. This confirms that the present electrode material exhibits excellent reproducibility. The stability of the fabricated electrodes stored in vacuum desiccator was determined. The electrode retained about 95% responses even after 2 months (Figure 10) suggesting that the present electrode is very stable.

Figure 10
Stability of the CuO nanobelts@G composites stored in vacuum desiccator over 60 days (amperometric currents obtained in 0.1 mol L−1 NaOH solution containing 1.0 μmol glucose, applied potential 0.55 V vs. Ag/AgCl electrode).

Application study

The applicability of the proposed sensor in real samples have been evaluated. The sensor was utilized for glucose monitoring in human blood serum samples. The human blood serum samples were diluted to 100 folds by 0.1 mol L−1 NaOH and glucose was monitored under optimized conditions in five different serum samples. The results obtained by the proposed method are in good agreement with certified values (measured by hospital staff using CA-958 H Rapid Semi-Automatic Biochemical Analyzer) which are depicted in Table 2.

Table 2
Amperometric determination of glucose in blood serum samples

Conclusions

We have successfully demonstrated for the first time a simple and efficient sonochemical route for in situ reduction of GO to graphene and growing CuO nanobelts in the presence of CTAB and NaOH. As prepared CuO@G composite were deposited on glassy carbon electrode and used as non-enzymatic sensor for glucose. Our sensor exhibits better electrocatalytic oxidation properties compared to traditional graphene and CuO nanobelt alone. The CuO-graphene sensor also exhibits low detection limits (0.05 µmol L−1), good sensitivity, selectivity, fast response time, good reproducibility and stability. This study would open a new avenue to design metal oxide graphene composites by simple sonochemical route for variety of electrochemical devices.

  • Supplementary Information
    Supplementary data are available free of charge at http://jbcs.sbq.org.br as PDF file.

Acknowledgements

The authors acknowledge Jain University for financial support.

References

  • 1
    Hellers, A.; Feldman, B.; Chem. Rev. 2008, 108, 2482.
  • 2
    Wilson, R.; Turner, A. P. F.; Biosens. Bioelectron. 1992, 7, 165.
  • 3
    Si, P.; Huang, Y.; Wanga, T.; Ma, J.; RSC Adv. 2013, 3, 3487.
  • 4
    Umar, A.; Rahman, M. M.; Hajry, A. A.; Hahn, Y. B.; Electrochem. Commun. 2009, 11, 278.
  • 5
    Wang, X. J.; Wang, G. F.; Liu, X. W.; Wu, J. J.; Li, M.; Gu, J.; Liu, H.; Fang, B.; J. Phys. Chem. A 2008, 112, 16845.
  • 6
    Zhang, L.; Li, H.; Ni, Y. H.; Li, J.; Liao, M.; Zhao, G. C.; Electrochem. Commun. 2009, 11, 812.
  • 7
    Wang, W.; Zhang, L. L.; Tong, S. F.; Li, X.; Song, W. B.; Biosens. Bioelectron. 2005, 25, 708.
  • 8
    Wang, X.; Hu, C. G.; Liu, H.; Du, G. J.; Xi, X. S.; Sens. Actuators, B 2010, 144, 220.
  • 9
    Li, C. I.; Su, Y.; Zhang, S. W.; Lv, X. Y.; Xia, H. L.; Wang, Y. Z.; Biosens. Bioelectron. 2010, 26, 903.
  • 10
    Sun, S.; Zhang, X.; Sun, Y.; Yang, S.; Song, X.; Yang, Z.; ACS Appl. Mater. Interfaces 2013, 5, 4429.
  • 11
    He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767.
  • 12
    Meng, L.; Jin, J.; Yang, G. X.; Lu, T. H.; Zhangand, H.; Cai, C. X.; Anal. Chem. 2009, 81, 7271.
  • 13
    Xiao, F.; Zhao, F. Q.; Mei, D. P.; Mo, Z. R.; Zeng, B. Z.; Biosens. Bioelectron. 2009, 24, 3481.
  • 14
    Yang, Z.; Feng, J.; Qiao, J.; Yan, Y.; Yu, Q.; Sun, K.; Anal. Methods 2012, 4, 1924.
  • 15
    Myung, Y.; Jang, D. M.; Cho, Y. J.; Kim, H. S.; Park, J.; J. Phys. Chem. C 2009, 113, 1251.
  • 16
    Zhang, X.; Wang, G.; Zhang, W.; Wei, Y.; Fang, B.; Biosens. Bioelectron. 2009, 24, 3395.
  • 17
    Wu, H. X.; Cao, W. M.; Li, Y.; Liu, G.; Wen, Y.; Yang, H. F.; Yang, S. P.; Electrochim. Acta 2010, 55, 3734.
  • 18
    Yang, J.; Zhang, W. D.; Gunashekaran, S.; Biosens. Bioelectron. 2010, 26, 279.
  • 19
    Kang, X. H.; Mai, Z. B.; Zou, X. Y.; Cai, P. X.; Mo, J. Y.; Anal. Biochem. 2007, 363, 143.
  • 20
    Ziang, L. C.; Zhang, W. D.; Biosens. Bioelectron. 2010, 25, 1402.
  • 21
    Jiang, F.; Wang, S.; Lin, J. J.; Jin, H. L.; Zhang, L. J.; Huang, S. M.; Wang, J. C.; Electrochem. Commun. 2011, 13, 363.
  • 22
    He, Y.; Zheng, J.; Anal. Methods 2013, 5, 767.
  • 23
    Qian, Y.; Ye, F.; Xu, J.; Le, Z. G.; Int. J. Electrochem. Sci. 2012, 7, 10063.
  • 24
    Luo, J.; Zhang, H. Y.; Jiang, S. S.; Jiangand, J. Q.; Liu, X. Y.; Microchim. Acta 2012, 177, 485.
  • 25
    Ritz, E.; Jia, W. Z.; Gentile, M.; Wang, Y.; Lei, Y.; Electroanalysis 2008, 20, 2482.
  • 26
    Luo, J.; Jiang, S.; Zhang, H.; Jiang, J.; Liu, X.; Anal. Chim. Acta 2012, 709, 53.
  • 27
    Hsua, Y. W.; Hsua, T. K.; Suna, C. L.; Nienc, Y. T.; Pud, N. W.; Gere, M. D.; Electrochim. Acta 2012, 82, 152.
  • 28
    Chen, Q.; Zhang, L.; Chen, G.; Anal. Chem. 2012, 84, 171.
  • 29
    Wang, X.; Liu, E.; Zhang, X.; Electrochim. Acta 2014, 130, 253.
  • 30
    Zhou, H.; Wang, X.; Yu, P.; Chen, X.; Mao, L.; Analyst 2012, 137, 305.
  • 31
    Chen, Y. J.; Meng, F.; Yu, H.; Zhu, C.; Wang, T.; Gao, P.; Ouyang, Q.; Sens. Actuators, B 2013, 176, 15.
  • 32
    Hummers, W. S.; Offeman, R. E.; J. Am. Chem. Soc. 1958, 80, 1339.
  • 33
    Muralikrishna, S.; Sureshkumar, K.; Varley, T. S.; Nagaraju, D. H.; Ramakrishnappa, T.; Anal. Methods 2014, 6, 8698.
  • 34
    Nethravathi, C.; Viswanath, B.; Michael, J.; Rajamath, M.; Carbon 2012, 50, 4839.
  • 35
    Huang, T. K.; Lin, K. W.; Tung, S. P.; Cheng, T. M.; Chang, I. C.; Hsieh, Y. Z.; Lee, C. Y.; Chiu, H. T.; J. Electroanal. Chem. 2009, 636, 123.
  • 36
    Zhuang, Z. J.; Su, X. D.; Yuan, H. Y.; Sun, Q.; Xiao, D.; Chio, M. M. F.; Analyst 2008, 33, 126.
  • 37
    Casella, I. G.; Gatta, M.; Guascito, M. R.; Cataldi, T. R. I.; Anal. Chim. Acta 1997, 357, 63.
  • 38
    Safavi, A.; Maleki, N.; Farjami, E.; Biosens. Bioelectron. 2009, 24, 1655.
  • 39
    Wang, J.; Zhang, W. D.; Electrochim. Acta 2011, 56, 7510.
  • 40
    Zhou, X.; Nie, H.; Yao, Z.; Dong, Y.; Yang, Z.; Huang, S.; Sens. Actuators, B 2012, 168, 1.
  • 41
    Zhu, L.; Wang, Z.; Bioelectrochemistry 2012, 88, 156.
  • 42
    Male, K. B.; Hrapovic, S.; Liu, Y.; Wang, D.; Luong, J. H. T.; Anal. Chim. Acta 2004, 516, 35.
  • 43
    Chen, J.; Zhang, W. D.; Ye, J. S.; Electrochem. Commun. 2008, 10, 1268.

Data availability

Publication Dates

  • Publication in this collection
    Aug 2015

History

  • Received
    28 Nov 2014
  • Accepted
    26 May 2015
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br