A sensitive determination of acetaminophen in pharmaceutical preparations and biological samples using multi-walled carbon nanotube modified glassy carbon electrode

Abstracts

A chemically modified electrode is constructed based on multi-walled carbon nanotube modified glassy carbon electrode (MWCNTs/GCE). It was demonstrated that this sensor could be easily used for determination of acetaminophen (ACT) in aqueous buffered media. The measurements were carried out by application of differential pulse voltammetry (DPV), cyclic voltammetry (CV) and chronoamperometry (CA) methods. Application of DPV method showed two linear dynamic ranges. The first linear dynamic range was from 0.1 to 22 µmol L-1, with a calibration equation of Ip(µA) = 1.2782c (µmol L-1) + 0.2431 (R² = 0.9984) and the second linear dynamic range was between 26 to 340 µmol L-1 with a calibration equation of Ip(µA) = 0.7793c (µmol L-1) + 11.615 (R² = 0.9986). A detection limit of 0.029 µmol L-1 (S/N = 3) was obtained. The modified electrode showed electrochemical responses with high sensitivity, excellent selectivity and stability for ACT determination at optimal conditions, which makes it a suitable sensor for submicromolar detection of ACT in solutions. The analytical performance of this sensor has been evaluated for detection of ACT in human serum, human urine and a pharmaceutical preparation with satisfactory results.

acetaminophen; carbon nanotube; modified glassy carbon; electrochemical sensor


Um eletrodo quimicamente modificado foi construído, baseado em eletrodo de carbono vítreo modificado por nanotubos de carbono de paredes múltiplas (ECV-NTCPM). O uso deste sensor na determinação de acetaminofeno (ACT), em meio aquoso tamponado, foi demonstrado. As medidas foram realizadas com utilização dos métodos de voltametria de pulso diferencial (VPD), voltametria cíclica (VC) e cronoamperometria (CA). A aplicação de VPD mostrou dois intervalos dinâmicos lineares. O primeiro intervalo dinâmico linear foi de 0,1 a 22 µmol L-1, com equação de calibração Ip(µA) = 1,2782c (µmol L-1) + 0,2431 (R² = 0,9984) e o segundo intervalo dinâmico linear foi entre 26 e 340 µmol L-1 com equação de calibração Ip(µA) = 0,7793c (µmol L-1) + 11,615 (R² = 0,9986). O limite de detecção de 0,029 µmol L-1 (S/N = 3) foi obtido. O eletrodo modificado mostrou respostas eletroquímicas com alta sensitividade, excelente seletividade e estabilidade para determinação de ACT em condições ótimas, o que o torna um sensor adequado para detecção submicromolar de ACT em soluções. O desempenho analítico deste sensor foi avaliado para detecção de ACT em soro humano, urina humana e em preparações farmacêuticas com resultados satisfatórios.


ARTICLE

A sensitive determination of acetaminophen in pharmaceutical preparations and biological samples using multi-walled carbon nanotube modified glassy carbon electrode

Ali BabaeiI,II,* * e-mail: a-babaei@araku.ac.ir, a_babaei07@yahoo.com ; Mohammad AfrasiabiI; Somaieh MirzakhaniI; Ali Reza TaheriI

IDepartment of Chemistry and Research Center for Nanotechnology,University of Arak, Arak, P.O. Box 38156-879, Iran

IIResearch Center for Nanotechnology,University of Arak, Arak, P.O. Box 38156-879, Iran

ABSTRACT

A chemically modified electrode is constructed based on multi-walled carbon nanotube modified glassy carbon electrode (MWCNTs/GCE). It was demonstrated that this sensor could be easily used for determination of acetaminophen (ACT) in aqueous buffered media. The measurements were carried out by application of differential pulse voltammetry (DPV), cyclic voltammetry (CV) and chronoamperometry (CA) methods. Application of DPV method showed two linear dynamic ranges. The first linear dynamic range was from 0.1 to 22 µmol L-1, with a calibration equation of Ip(µA) = 1.2782c (µmol L-1) + 0.2431 (R2 = 0.9984) and the second linear dynamic range was between 26 to 340 µmol L-1 with a calibration equation of Ip(µA) = 0.7793c (µmol L-1) + 11.615 (R2 = 0.9986). A detection limit of 0.029 µmol L-1 (S/N = 3) was obtained. The modified electrode showed electrochemical responses with high sensitivity, excellent selectivity and stability for ACT determination at optimal conditions, which makes it a suitable sensor for submicromolar detection of ACT in solutions. The analytical performance of this sensor has been evaluated for detection of ACT in human serum, human urine and a pharmaceutical preparation with satisfactory results.

Keywords: acetaminophen, carbon nanotube, modified glassy carbon, electrochemical sensor

RESUMO

Um eletrodo quimicamente modificado foi construído, baseado em eletrodo de carbono vítreo modificado por nanotubos de carbono de paredes múltiplas (ECV-NTCPM). O uso deste sensor na determinação de acetaminofeno (ACT), em meio aquoso tamponado, foi demonstrado. As medidas foram realizadas com utilização dos métodos de voltametria de pulso diferencial (VPD), voltametria cíclica (VC) e cronoamperometria (CA). A aplicação de VPD mostrou dois intervalos dinâmicos lineares. O primeiro intervalo dinâmico linear foi de 0,1 a 22 µmol L-1, com equação de calibração Ip(µA) = 1,2782c (µmol L-1) + 0,2431 (R2 = 0,9984) e o segundo intervalo dinâmico linear foi entre 26 e 340 µmol L-1 com equação de calibração Ip(µA) = 0,7793c (µmol L-1) + 11,615 (R2 = 0,9986). O limite de detecção de 0,029 µmol L-1 (S/N = 3) foi obtido. O eletrodo modificado mostrou respostas eletroquímicas com alta sensitividade, excelente seletividade e estabilidade para determinação de ACT em condições ótimas, o que o torna um sensor adequado para detecção submicromolar de ACT em soluções. O desempenho analítico deste sensor foi avaliado para detecção de ACT em soro humano, urina humana e em preparações farmacêuticas com resultados satisfatórios.

Introduction

Acetaminophen (N-acetyl-p-aminophenol) (ACT), also known as paracetamol, is a drug with antipyretic and analgesic action, frequently used in therapy and which has low toxicity when used at the recommended doses. 1 Nevertheless, the ease with which ordinary people can access this drug makes it frequently associated with overdoses, in this situation causing serious or even fatal hepatic damage. 2 An overdose of ACT can lead to the accumulation of toxic metabolites, which may cause severe and some times fatal hepatoxicity and nephrotoxicity3-5 and it can also cause liver disorders, skin rashes and inflammation of the pancreas. 6 The product of hydrolytic degradation of ACT is 4-aminophenol which can be found in pharmaceutical preparations as a degradation product of ACT or as a synthetic intermediate which can be dangerous and causes teratogenic effect and nephrotoxicity. 7 Controlling the amount of ACT in a given pharmaceutical formulation is vital in the quality control of the pharmaceutical industry, and can have a significant impact in healthcare quality. Various methods such as capillary electrophoresis,8,9 amperometric batch injection analysis,10 colorimetry11 titrimetry,12 FTIR and Raman spectrometry,13 flow injection analysis (FIA) using different methods of detection,12,14,15 liquid chromatography,16-18 spectrophotometry19-22 and chemiluminescence23 have been employed in the determination of ACT in pharmaceutical preparations and biological fluids. However, the majority of these methods suffer from some disadvantages such as high costs, long analysis times and requirement for sample pretreatment, and in some cases low sensitivity and selectivity that makes them unsuitable for routine analysis. Therefore, development of a simple, inexpensive, sensitive and accurate analytical method for determination of ACT is of great importance. ACT is an electroactive compound which can be oxidized electrochemically. The development and application of electrochemical sensors for ACT analysis, with respect to its sensitivity, accuracy, and simplicity, has been of greater interest in recent years. 24-46 Careful studies of the reported articles revealed that apart from the variation of the electrochemical methods, there is a need to improve the electrochemical techniques with respect to their sensitivity, linear dynamic range or limit of detection.

Carbon nanotubes (CNT) represent an increasingly important group of nanomaterials with unique geometrical, mechanical, electronic and chemical properties. 47,48 Such properties of CNT make them also extremely attractive for the task of electrochemical detection.

In this work we present the application of the multi-walled carbon nanotube modified glassy carbon electrode (MWCNTs/GCE) as a sensor for determination of ACT in buffer solutions at the biological pH. This study has led to the development of a voltammetric method with good characteristics, such as simplicity of electrode preparation, high selectivity, wide linear dynamic range, and low detection limit. Finally the analytical performance of this sensor for determination of ACT in human serum, human urine and a pharmaceutical preparation samples is evaluated.

Experimental

Reagents and solutions

All chemicals were analytical grade and used without further purification. ACT and multi-walled carbon nanotubes (MWCNTs ) (> 95 wt. %, 5-20 nm) were obtained from Merck and PlasmaChem GmbH companies, respectively. Stock standard solutions of 10 µmol L-1 ACT were freshly prepared in 0.1 mol L-1 phosphate buffers of pH 7. All subsequent ACT solutions were prepared by diluting the stock standard solutions using 0.1 mol L-1 phosphate buffer (pH 7). 0.1 mol L-1 phosphate buffer solution (PBS) was prepared by dissolving appropriate amounts of sodium hydrogen phosphate and sodium dihyrogen phosphate in 250 mL volumetric flask. Electrochemical experiments on the ACT were carried out in 0.1 mol L-1 PBS at pH 7.

Fresh human serum samples were obtained from the Razi Institute of Vaccine and Serum Company (Tehran, Iran). The serum and urine sample were filtered and diluted 100 times with 0.1 mol L-1 PBS of pH 7 and checked for the determination of the recovery by spiking the ACT. Ten tablets of Tylenol (McNeil-PPC Inc., USA); (each labeled as containing 500 mg of ACT ) were accurately weighed and powdered in a mortar. An amount equivalent to one tablet was dissolved in 70 mL of 0.1 mol L-1 PBS (pH 7). After sonication for 10 min, the solution was filtered, the residue was washed three times with 10 mL of the buffer solvent, and the filtered volume was adjusted to 100 mL with the same solvent. This solution was diluted 1000 times for ACT determination with 0.1 mol L-1 PBS of pH 7. The solution was used for the determination of the recovery efficiency.

Instrumentation

All the voltammetric measurements were carried out using a nanotube modified glassy carbon electrode (MWCNTs/GCE) as a working electrode, an Ag/AgCl/(3 mol L-1) KCl as a reference electrode and platinum wire as an auxiliary electrode. DPV, CV and CA experiments were carried out using an Autolab PGSTAT 30 Potentiostat Galvanostat (EcoChemie, The Netherlands) coupled with a 663 VA stand (Metrohm Switzerland). All potentials given are with respect to the potential of the reference electrode. The pH measurements were made with a Metrohm 744 pH meter using a combination glass electrode.

Modification of the electrodes

A glassy carbon electrode (GCE, 3 µm diameter, Metrohm) was polished with 0.3 and 0.05 µm aluminum slurry and rinsed thoroughly with triply distilled water. The GC electrode was individually cleaned by ultrasonic agitation for 5 min in ethanol and then distilled water. The electrode was then dried under nitrogen gas flow.

A stock solution of 1 mg mL-1 MWCNTs-DMF was prepared by dispersing 1mg of MWNTs in 1 mL DMF. 30 µL of MWCNTs-DMF solution was casted on electrode surface. The electrode was dried at room temperature to obtain MWCNTs/GCE.

The fabricated MWCNTs/GCE was placed in the electrochemical cell containing 0.1 mol L-1 PBS (pH 7) and several cycles in the potential range of 0.1 to 0.7 V were applied using CV method to obtain stable responses. The electrochemical surface area of the modified MWCNTs/GCE and bare GCE were determined by cyclic voltammogram measured between -0.1 to 0.6 V in ferricyanide solution (0.1 mol L-1 phosphate buffer, pH 7.0) at different scan rates (not shown). The modified MWCNTs/GCE showed surface area of 9.8 times of GCE.

General procedure

Solutions (10 mL) containing the appropriate amounts of ACT in 0.1 mol L-1 PBS at pH 7 were transferred into the voltammetric cell. The voltammograms were recorded by applying positive-going potential from 0.2 to 0.6 V. The voltammogram showed an anodic peak around 0.32 V corresponding to the oxidation of ACT compound with height proportional to its concentration in solution. The calibration curves were obtained by plotting anodic peak currents of ACT versus the corresponding concentrations. All experiments were carried out under open circuit conditions.

After each measurement, the MWCNTs/GCE was regenerated by thoroughly washing the electrode successively with triply distilled water and then 5% sodium hydroxide solution. Finally the electrode was rinsed carefully with distilled water to remove any adsorbate from electrode surface and to provide a fresh one for subsequent experiments.

Results and Discussion

Cyclic voltammetric investigation of ACT on MWCNTs /GCE

The recorded cyclic voltammogram for 50 µmol L-1 ACT using the MWCNTs/GCE is shown in Figure 1. The ACT showed a reversible redox peak with ΔEp equal to 0.028 V, which is similar to a two electron redox reversible system. This phenomenon is in accordance with electrocatalytic behavior of MWCNTs.


The effect of potential scan rate on redox response of ACT was investigated in phosphate buffer (pH 7.0) solution containing 50 µmol L-1 ACT. Figure 2 shows the cyclic voltammograms of the modified electrode at different scan rates in potential range 0 to 0.65 V. The anodic and cathodic peak currents of ACT were proportional to the scan rate over the range 10-280 mV s-1 (Figure 2A) indicating adsorptive properties of the electrochemical process. Moreover, the anodic peak currents were almost the same as the corresponding cathodic peak currents.







At sweep rates between 280 mV s-1 and 1000 mV s-1, the plot of peak currents vs. scan rate plot deviates from linearity and the peak current becomes proportional to the square root of the scan rate (Figure 2B), indicating a diffusion controlled process. At sweep rates higher than 100 mV s-1 peak separations (DEp) begin to increase (Figure 2C), indicating the limitation due to charge transfer kinetics. Based on Laviron theory49 the electron transfer rate constant (ks) and charge transfer coefficient (α) can be determined by measuring the variation of ΔEp vs. log scan rate. The slope of the ΔEp vs. log(n), was about, 131.1 mV. Using the equation of:

Ep = K - 2.3030 (RT/αnF) log(ν)

By considering two electrons transferred for ACT, charge transfer coefficient (α) of 0.562 was obtained. Introducing αvalue in the following equation, an apparent surface electron transfer rate constant, ks = 2.558 s-1, was estimated.

The large value of electron transfer rate constant indicates high ability of MWCNTs for promoting electron transfer between ACT and the electrode surface. The surfaces of MWCNTs contain a large number of defects as well as special nanostructure of MWCNTs may act as molecular wires, enhance the direct electron transfer of ACT at MWCNTs.

Electroxidation of ACT on MWCNTs /GCE

The differential pulse voltammograms (DPV) recorded for ACT at bare GCE, and MWCNTs/GCE are shown in Figure 3. Curve a shows the voltammogram of 40 µmol L-1 ACT in PBS (pH 7) at GC. Curve b is a voltammogram of ACT obtained under the same conditions as curve a, at a MWCNTs /GCE. As can be seen, for the GC the oxidation peak for ACT is very small. However, the DPVs of ACT at MWCNTs/GCE (curve b) showed considerable increases in its oxidation peak current. The presence of MWNTs could increase the surface area of the electrode and facilitate electron transfer between electrode and the analyte. Therefore the enhancements in the corresponding electrochemical oxidation peak currents were observed. Our electrochemical data showed that surface area of modified MWCNTs/GCE is 9.8 times of GCE. However, the oxidation peak current of ACT was increased upto 50 times, therefore the catalytic effect of MWCNTs is also responsible for increasing current.


Effects of solution pH

The effect of pH of solutions on the electrochemical response of the MWCNTs/GCE towards the determination of 50 µmol L-1 ACT was investigated using CV method in 0.1 mol L-1 phosphate buffer solution. Variations of peak current with respect to pH of the electrolyte in the pH range from 3 to 10 are shown in Figure 4. It can be seen that the anodic peak currents of ACT increases with raising the solution pH until it reaches 7 (Figure 4, inset A). However at higher pH the ACT oxidation peak current starts to diminish. Therefore the pH value of 7, which is close to biological pH value, was chosen as an optimum solution pH for further experiments. Variation of ACT oxidation peak potential with pH is in accordance with equation of Ep = -0.0542 pH + 0.704 (Figure 4, inset B). For a Nernstian process which numbers of transferred electrons are equal to number of transferred proton, the slope would be expected to be -59 mV pH-1 unit. The -54 mV pH-1 unit slope is in good agreement with other previously reported results. 50-52 This suggests that the number of electrons and protons transferred in the redox reaction of ACT is equal.


Effects of accumulation time

Figure 5 shows plots of the anodic peak currents, obtained from CV experiments, versus accumulation time for 15 µmol L-1 ACT. Initially, peak currents for this compound increase with accumulation time up to 60 s. However after 60 s the rate of increase falls away and it eventually reaches plateau. As a consequence, the accumulation time of 60 s was chosen as an optimum time for further experiments.


Linear dynamic range and limit of detection of the method

The electrochemical responses for consecutive addition of ACT to a 0.1 mol L-1 PBS pH 7 using MWCNTs/GCE are depicted in Figures 6 and 7. Figure 6 show differential pulse voltammograms and corresponding calibration curves obtained at MWCNTs/GCE in various concentrations of ACT. By application of DPV method two linear ranges were obtained. The first linear dynamic range was from 0.1 to 22 µmol L-1, with a calibration equation of Ip(µA) = 1.2782c (µmol L-1) + 0.2431 (R2 = 0.9984) and the second linear dynamic range was between 26 to 340 µmol L-1 with a calibration equation of Ip(µA) = 0.7793c (µmol L-1) + 11.615 (R2 = 0.9986). A limit of detection of 0.029 µmol L-1 (S/N = 3) was obtained.


Figure 7

Repeatability and long-term stability of the electrode

The repeatability of the analytical method for determination of ACT has been studied using DPV method. Indeed, the relative standard deviations (RSD) of 2.06% and 1.82% for 10.0 and 50.0 µmol L-1 ACT over ten consecutive determinations have been obtained, respectively.

Another advantage of the proposed modified electrode is that the resulting modified electrode has a good long-term stability. Stability of the proposed electrode was tested by measuring the decrease in voltammetric current during repetitive DPV measurements of ACT solutions with MWCNTs/GCE stored in solution or air for certain period of time. For example, in the determination of 50 µmol L-1 ACT in 0.1 mol L-1 PBS (pH 7), subjecting the modified electrode to an experiment every 30 min, led to a less than 7% decrease in the voltammetric currents after 24 h. When the electrode was stored in the atmosphere for 5 days, the corresponding current responses fell less than 11% when the electrode subjected to a solution containing 50 µmol L-1 ACT.

Interference studies

The effects of common interfering species solution of 100 µmol L-1 ACT under the optimum conditions were investigated. The results showed that concentrations of L-dopa(500), dopamine(500), L-alanin(1000), L-glutamic acid(1400), uric acid(200), ascorbic acid(700) and aspartic acid(3500) did not significantly influence the height of the oxidation peak current for ACT. The tolerance limit was defined as the concentrations which give an error of < 10% in the determination of ACT compound. The data in the brackets are concentration of the interfering species in µmol L-1. The results confirm that the proposed method is free from interferences from the most common interferants.

Analytical applications

The applicability of the method to the determination of ACT content in real samples such as pharmaceutical preparations, or in solutions in human serum or human urine was investigated. The differential pulse voltammograms were obtained by spiking appropriate amounts of ACT in the prepared real solutions using MWCNTs/GCE at optimum conditions as described earlier. The ACT concentrations were obtained by applying the corresponding calibration plot. The recoveries were acceptable, showing that the proposed methods could be efficiently used for the determination of trace amounts of ACT in biological systems and pharmaceutical preparations. The results are presented in Table 1. The linear dynamic range, detection limit, RSD, sensitivity and some real sample analysis of the proposed method were compared with other voltammetric methods for determination of ACT (Table 2). The results show that the proposed method has higher sensitivity, lower detection limit and wide linear dynamic range.

Conclusions

In this paper we introduced a simple method to make a sensor based on multi-walled carbon nanotube modified glassy carbon electrode. We have shown that the application of MWCNTs can increase anodic peak currents by enhancement of electron transfers of ACT compound on the electrode surface. The results indicated that modification of GCE with MWCNT allows the determination of ACT with good sensitivity and selectivity. The electrode showed high stability in repetitive experiments due to high water stability and high mechanical strength of MWCNTs. The effects of potential interferant were studied, and it was found that the proposed procedure is free from interferences of most common interfering compounds. When the procedure was used for the determination of ACT in some real samples like human serum, urine and some drugs, satisfactory results were obtained The simple fabrication procedure, high speed, reproducibility, high stability, wide linear dynamic range, low detection limit and high sensitivity, all suggest that the proposed sensor is an attractive candidate for practical applications.

Acknowledgments

The authors gratefully acknowledge the support of this work by the Research Council of Arak University. Thanks to Dr. Alan Happer from Canterbury University in New Zealand for his valuable comments.

Submitted: April 5, 2010

Published online: October 14, 2010

  • 1. Sweetman, S. C.; The Complete Drug Reference, 33rd ed., Pharmaceutical Press: London, 2002.
  • 2. Anker, A. In Clinical Toxicology; Marx, J., ed.; W.B. Saunders Company: Philadelphia, 2001.
  • 3. Martin, F. L.; McLean, A. E.; Drug Chem. Toxicol. 1998, 21, 477.
  • 4. Mugford, C. A.; Tarloff, J. B.; Toxicol. Lett. 1997, 93, 15.
  • 5. Nagasawa, H. T.; Shoeman, D. W.; Cohen, J. F.; Rathbun, W. B.; J. Biochem. Toxicol. 1996, 11, 289.
  • 6. Prabakar, S. J. R.; Narayanan, S. S.; Talanta 2007, 72, 1818.
  • 7. Yesilada, A.; Erdogan, H.; Ertan, M.; Anal. Lett 1991, 24, 129.
  • 8. Zhao, S.; Bai, W.; Yuan, H.; Xiao, D.; Anal. Chim. Acta 2006, 559, 195.
  • 9. Capella-Peiro, M. E.; Bose, D.; Rubert, M. F.; Esteve-Romero, J.; J. Chromatogr., B 2006, 839, 95.
  • 10. Quintino, M. S. M.; Araki, K.; Toma, H. E.; Angnes, L.; Electroanalysis 2002, 14, 1629.
  • 11. Usifoh, C. O.; Adelusi, S. A.; Adebambo, R. F.; Pak. J. Sci. Ind. Res. 2002, 45, 7.
  • 12. Knochen, M.; Giglio, J.; Reis, B. F.; J. Pharm. Biomed. Anal. 2003, 33, 191.
  • 13. Al-Zoubi, N.; Koundourellis, J. E.; Malamataris, S.; J. Pharm. Biomed. Anal. 2002, 29, 459.
  • 14. Xiong, X. Y.; Tang, Y. H.; Wang, N. N.; Wu, S. J.; Fenxi Shiyanshi 2007, 26, 80.
  • 15. Silva, M. L. S.; Garcia, M. B. Q.; Lima, J. L. F. C.; Barrado, E.; Portugaliae Electrochim. Acta 2006, 24, 261.
  • 16. Ravisankar, S.; Vasudevan, M.; Gandhimathi, M.; Suresh, B.; Talanta 1998, 46, 1577.
  • 17. Goyal, A.; Jain, S.; Acta Pharm. Sci. 2007, 49, 147.
  • 18. Selvan, P. S.; Gopinath, R.; Saravanan, V. S.; Gopal, N.; Kumar, S. A.; Periyasamy, K.; Asian J. Chem. 2007, 19, 1004.
  • 19. Bouhsain, Z.; Garrigues, S.; de la Guardia, M.; Fresenius J. Anal.Chem. 1997, 357, 973.
  • 20. Hanaee, J.; Pharm. Acta Helv. 1997, 72, 239.
  • 21. Mohamed, F. A.; Abdallah, M. A.; Shammat, S. M.; Talanta 1997, 44, 61.
  • 22. Khaskheli, A. R.; Shah, A.; Bhange, M. I.; Niazr, A.; Mahesar, S.; Spectrochim. Acta, Part A 2007, 68, 747.
  • 23. Easwaramoorthy, D.; Yu, Y. C.; Huang, H. J.; Anal. Chim. Acta 2001, 439, 95.
  • 24. Quintino, M. S. M.; Araki, K.; Toma, H. E.; Angnes, L.; Electroanalysis 2002, 14, 1629.
  • 25. Wangfuengkanagul, N.; Chailapakul, O.; J. Pharm. Biomed. Anal 2002, 28, 841.
  • 26. Masawat, P.; Liawruangrath, S.; Vaneesorn, Y.; Liawruangrath, B.; Talanta 2002, 58, 1221.
  • 27. Vieira, I. C.; Lupetti, K. O.; Fatibello-Filho, O.; Quim. Nova 2003, 26, 39.
  • 28. Boopathi, M.; Won, M. S.; Shim, Y. B.; Anal. Chim. Acta 2004, 512, 191.
  • 29. Goyal, R. N.; Gupta, V. K.; Oyama, M.; Bachheti, N.; Electrochem. Commun 2005, 7, 803.
  • 30. Gomez-Caballero, A.; Goicolea, M. A.; Barrio, R. J.; Analyst 2005, 130, 1012.
  • 31. Wang, C.; Li, C.; Wang, F.; Wang, C.; Microchim. Acta 2006, 155, 365.
  • 32. Goyal Rajendra, N.; Singh Sudhanshu, P.; Electrochim. Acta 2006, 51, 3008.
  • 33. Messina, G. A.; De Vito, I. E.; Raba, J.; Anal. Chim. Acta 2006, 559, 152.
  • 34. Silva, M. L. S.; Garcia, M. B. Q.; Lima, J. L. F. C.; Barrado, E.; Anal. Chim. Acta 2006, 573, 383.
  • 35. Wang, C.; Li, C.; Wei, L.; Wang, C.; Microchim. Acta 2007, 158, 307.
  • 36. Sun, D.; Zhang, H.; Microchim. Acta 2007, 158, 131.
  • 37. Jia, L.; Zhang, X. H.; Li, Q.; Wang, S. F.; Anal. Chem 2007, 62, 266.
  • 38. Wang, S. F.; Xie, F.; Hu, R. F.; Sens. Actuators, B 2007, 123, 495.
  • 39. Ozcan, L.; Sahin, Y.; Sens. Actuators, B 2007, 127, 362.
  • 40. Felix, F. S.; Brett, C. M. A.; Angnes, L.; J. Pharm. Biomed. Anal. 2007, 43, 1622.
  • 41. Sima, V.; Cristea, C.; Lapadus, F.; Marian, I. O.; Marian, A.; Sandulescu, R.; J. Pharm. Biomed. Anal. 2008, 48, 1195.
  • 42. Ashok kumar, S.; Chun-Fang Tang; Shen-Ming Chen; Talanta 2008, 76, 997.
  • 43. ShangGuan, X.; Zhang, H.; Zheng, J.; Anal. Bioanal. Chem 2008, 391, 1049.
  • 44. Kachoosangi, R. T.; Wildgoose, G. G.; Compton, R. G.; Anal. Chim. Acta 2008, 618, 54.
  • 45. Lourencao, B. C.; Medeiros, R. A.; Rocha-Filho, R. C.; Mazo, L. H.; Fatibello-Filho, O.; Talanta 2009, 78, 748.
  • 46. Fanjul-Bolado, P.; Lamas-Ardisana, P. J.; Hernandez-Santos, D.; Costa-Garcia, A.; Anal. Chim. Acta 2009, 638, 133.
  • 47. Rao, C. N. R.; Satishkumar, B. C.; Govindaraj, A.; Nath, M.; Chem. Phys. Chem. 2001, 2, 78.
  • 48. Baughman, R. H.; Zakhidov, A. A.; de Heer, W. A.; Science 2002, 297, 787.
  • 49. Laviron, E.; J. Electroanal. Chem 1979, 101, 19.
  • 50. Miner, D. J.; Rice, J. R.; Riggin, R. M.; Kissinger, P. T.; Anal. Chem. 1981, 53, 2258.
  • 51. Gilmartin, M. A. T.; Hart, J. P.; Analyst 1994, 119, 2431.
  • 52. Felix, F. S.; Brett, C. M. A.; Angnes, L.; J. Pharm. Biomed. Anal. 2007, 43, 1622.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      14 Feb 2011
    • Date of issue
      Feb 2011

    History

    • Received
      05 Apr 2010
    • Accepted
      14 Oct 2010
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br