A Fast Chromatographic Method for Determination of Daidzein and Genistein in Spiked Water River Samples Using Multivariate Curve Resolution

Abstract

This work reports the development of a fast chromatographic methodology for quantitation of two phytoestrogens: daidzein (DAI), and genistein (GEN), in river water samples. The proposed method is based on high performance liquid chromatography-diode array detection (HPLC-DAD) data, and multivariate curve resolution-alternative least square (MCR-ALS) second-order calibration. Initially, the method was evaluated analyzing a synthetic validation set; prepared based on a Taguchi design. Subsequently, the method was applied to predict the concentration of the phytoestrogens in spiked river water samples, previously pre-processed by solid phase extraction (SPE). By implementation of the present chromatographic methodology, a 50% reduction in operation time was achieved (from 7.00 to 3.25 min) when compared with previous work in the literature. Precision was achieved even in the presence of non-modeled constituents and strong background. Thus, the proposed method is a rapid and robust alternative for the quantitation of studied phytoestrogens.

Keywords
emerging contaminants; liquid chromatography; multiway calibration; second order advantage


Introduction

As a result of human activities, an increased production of waste has been noted, particularly in natural waters. Consequently, the level of pollutants such as heavy metals, (well known for their toxic effects on living beings), has also increased.1Pesavento, M.; Alberti, G.; Biesuz, R.; Anal. Chim. Acta 2009, 631, 129.,2Stankovic, S.; Kalaba, P.; Stankovic, A. R.; Environ. Chem. Lett. 2014, 12, 63. However, little is known about the impact of a class of compounds called emerging contaminants. These compounds are also released to the environment, and have recently become the object of a wide variety of studies.3Manciocco, A.; Calamandrei, G.; Alleva, E.; Chemosphere 2014, 100, 1.,4Pal, A.; He, Y.; Jekel, M.; Reinhard, M.; Gin, K. Y. H.; Environ. Int. 2014, 71, 46.

Within this latter group of contaminants, phytohormones have great importance, since it is known that this class of compounds is bioactive even at low concentrations.5Jefferson, W. N.; Padilla-Banks, E.; Newbold, R. R.; Reprod. Toxicol. 2007, 23, 308. Some phytohormones have been detected in environmental waters at alarming levels, as is the case of daidzein (DAI) and genistein (GEN).6Kiparissis, Y.; Hughes, R.; Metcalfe, C.; Ternes, T.; Environ. Sci. Technol. 2001, 35, 2423.,7Lundgren, M. S.; Novak, P. J.; Environ. Toxicol. Chem. 2009, 28, 2318.

Phytohormones are non-steroidal polyphenolic metabolites produced by plants,8Bednarek, P.; Franski, R.; Kerhoas, L.; Einhorn, J.; Wojtaszek, P.; Stobiecki, M.; Phytochemistry 2001, 56, 77. which can chemically bind to specific intracellular estrogenic receptors causing variations of endocrine signals in animals and humans, thus their characterization as phytoestrogens.9Coldham, N. G.; Sauer, M. J.; Food Chem. Toxicol. 2001, 39, 1211. In Figure 1 the structural similarity between the phytoestrogens DAI and GEN, to basic estrogen structures is shown. Recent results have shown that fish exposed to phytoestrogens may become more aggressive, due to testosterone reduction, and to immunosuppression.1010 Zhang, L.; Khan, I. A.; Foran, C. M.; Comp. Biochem. Physiol., Part C: Toxicol. Pharmacol. 2002, 132, 203.,1111 Clotfelter, E. D.; Rodriguez, A. C.; Environ. Pollut. 2006, 144, 833.

Figure 1
Molecular structure of (a) daidzein; (b) genistein; (c) estrone; (d) estradiol; (e) estriol and (f) progesterone.

Phytoestrogens originating in the soy-processing industries, (field grain cultivation, and the food industry) contaminate both rivers and lakes.1212 Kelly, M. M.; Arnold, W. A.; Environ. Sci. Technol. 2012, 46, 5396.,1313 Kawanishi, M.; Takamura-Enya, T.; Ermawati, R.; Shimohara, C.; Sakamoto, M.; Matsukawa, K.; Matsuda, T.; Murahashi, T.; Matsui, S.; Wakabayashi, K.; Environ. Sci. Technol. 2004, 38, 6424. Research shows that after consumption, phytoestrogens undergo numerous metabolic conversions. Both their metabolites, and their precursor compounds may be absorbed into the bloodstream, and then excreted in urine.1414 Bustamante-Rangel, M.; Delgado-Zamarreño, M. M.; Carabias-Martínez, R.; Domínguez-Álvarez, J.; Anal. Chim. Acta 2012, 709, 113.,1515 Dai, Q.; Franke, A. A.; Jin, F.; Shu, X. O.; Hebert, J. R.; Custer, L. J.; Cheng, J. R.; Gao, Y. T.; Zheng, W.; Cancer Epidem. Biomar. 2002, 11, 815. The increased production of soybeans and their derivatives for both eastern and western diets has been associated with a higher incidence of cancer in women.1616 He, F. J.; Chen, J. Q.; Food Sci. Hum. Wellness 2013, 2, 146.

A well-documented phytoestrogens excretion in human urine1717 Tekel, J.; Daeseleire, A.; Heeremans, E.; van Peteghem, C.; J. Agric. Food Chem. 1999, 47, 3489. has now raised concerns about their presence in surface waters and domestic sewage.1818 Laganà, A.; Bacaloni, A.; De Leva, I.; Faberi, A.; Fago, G.; Marino, A.; Anal. Chim. Acta 2004, 501, 79.

19 Kuster, M.; Azevedo, D. A.; Alda, M. J. L.; Aquino Neto, F. R.; Barceló, D.; Environ. Int. 2009, 35, 997.
-2020 Comber, M. H. I.; Williams, T. D.; Stewart, K. M.; Water Res. 1993, 27, 273. A lack of proper water treatment implies that phytoestrogens (among other contaminants) reach the population.

Several methodologies have been proposed to quantify phytoestrogens, being that liquid (LC) or gas (GC) chromatography, coupled with mass spectrometry (MS),2121 Wang, C.; Prasain, J. K.; Barnes, S.; J. Chromatogr. B 2002, 777, 3.,2222 Bai, Y.; Du, F.; Bai, Y.; Liu, H.; Anal. Methods 2010, 2, 1867. are the most used methods. Other approaches have also been presented in the literature,2121 Wang, C.; Prasain, J. K.; Barnes, S.; J. Chromatogr. B 2002, 777, 3. like LC and capillary electrophoresis (CE) combined with absorbance (diode array detector, DAD),2323 Wang, C.; Wang, L.; Zhao, Q.; Chen, J.; Zheng, L.; Zheng, M.; Zhang, R.; Wang, Z.; Anal. Methods 2013, 5, 3688.,2424 Shihabi, Z. K.; Kute, T.; Garcia, L. L.; Hinsdale, M.; J. Chromatogr. A 1994, 680, 181. fluorescence and electrochemical detection,2525 Wang, G.; Kuan, S. S.; Francis, O. J.; Ware, G. M.; Carman, A. S.; J. Agric. Food Chem. 1990, 38, 185.,2626 Chen, G.; Zhang, J.; Ye, J.; J. Chromatogr. A 2001, 923, 255. immunoassays,2727 Lapcik, O.; Hampl, R.; Hill, M.; Wahala, K.; Al-Maharik, N.; Adlercreutz, H.; J. Steroid Biochem. 1998, 64, 261. and voltammetry.2828 Fogliatto, D. K.; Barbosa, A. M. J.; Ferreira, V. S.; Colloids Surf., B 2010, 78, 243.

The use of multiway data modeling with appropriate algorithms has been quite successful in quantifying a wide variety of chemical compounds in complex matrices.2929 Shenone, A. V.; Culzoni, M. J.; Marsili, N. R.; Goicoechea, H. C.; Food Chem. 2013, 138, 1928.

30 Borraccetti, M. D.; Damiani, P. C.; Olivieri, A. C.; Analyst 2009, 134, 1682.
-3131 Elcoroaristizabal, S.; Juan, A.; García, J. A.; Durana, N.; Alonso, L.; Chemom. Intell. Lab. Syst. 2014, 132, 63. Multiway data can be easily generated with the modern instrumentation available in research and/or industrial laboratories, as examples; LC-DAD, GC combined with mass spectroscopy (GC-MS), CE-DAD systems (which are known as hyphenated techniques), or even simple excitation emission matrices (EEM).3232 Escandar, G. M.; Goicoechea, H. C.; de la Peña, A. M.; Olivieri, A. C.; Anal. Chim. Acta 2014, 806, 8. In terms of modeling multiway data, in the specific case of LC-DAD systems, multivariate curve resolution-alternative least square (MCR-ALS) is an appropriate algorithm.3333 Jaumot, J.; Gargallo, R.; Juan, A.; Tauler, R.; Chemom. Intell. Lab. Syst. 2005, 76, 101.

The development and validation of chromatographic analytical methods involves several steps, among them, analysis of interferences, which involves costs and time consuming. The composition of real samples tends to be complex, and may vary from sample to sample. An interferent (unexpected) can arise at any time. This requires that the method be modified and validated. The second order advantage can ensure that a chromatographic method continues to work well in scenarios where conventional strategies have failed. Another aspect is that there is no need for full chromatographic resolution of the analytes, which allows the use of shorter runs with fewer solvent consuming.

Determinations of drugs and pesticides in blood, water, and/or food and drinks are examples of multiway calibration applications.3434 Lozano, V. A.; Escandar, G. M.; Anal. Chim. Acta 2013, 782, 37.

35 Alcaráz, M. R.; Shenone, A. V.; Culzoni, M. J.; Goicoechea, H. C.; Microchem. J. 2014, 112, 25.
-3636 Culzoni, M. J.; Goicoechea, H. C.; Pagani, A. P.; Cabezon, M. A.; Olivier, A. C.; Analyst 2006, 131, 718. These applications have the great advantage of being able to circumvent the presence of unexpected constituents in a single sample; this feature is known as the second-order advantage.3232 Escandar, G. M.; Goicoechea, H. C.; de la Peña, A. M.; Olivieri, A. C.; Anal. Chim. Acta 2014, 806, 8. Thus, steps for the removal of interferents are not required, thus permitting less costly methods to be developed.3737 Arancibia, J. A.; Damiani, P. C.; Escandar, G. M.; Ibañez, G. A.; Olivieri, A. C.; J. Chromatogr. B 2012, 910, 22.

This paper presents a method based on high performance liquid chromatography (HPLC) with molecular absorption detection in ultraviolet-visible region (UV-Vis) for simultaneous quantitation of two phytoestrogens in spiked river water samples (DAI and GEN), using multivariate curve resolution with alternating least squares. The predictive ability of the calibration models built, using pure standards of both analytes were evaluated for: (i) validation samples consisting of mixtures of DAI and GEN standards plus another phytoestrogen equol (EQL), which was added as a potential interferent, and (ii) river water samples. In addition, an isocratic elution mode was employed, with a run time of less than three and a half minutes, contributing to the development of green analytical methods.

Experimental

Reagents and solutions

All reagents used in this work were of high-purity (≥ 99%). Daidzein, Genistein and Equol were purchased from Sigma-Aldrich Co. Acetonitrile and water (both HPLC grade) were filtered through a cellulose filter of 0.45 µm. All stock solutions of DAI (100.0 mg L-1), GEN (200.0 mg L-1) and EQL (100.0 mg L-1) were prepared in volumetric flasks by dissolving appropriate amounts in acetonitrile.

Apparatus and HPLC procedure

The LC-DAD matrices were recorded in an Ultimate 3000 Dionex liquid chromatograph, equipped with a manual injector and a fixed loop of 20 µL. Detection was carried out with a diode array detector in the range of 229 to 349 nm with resolution of 1 nm. A C18 column (AcclaimTM120) of 150 mm × 4.6 mm, 5 μm particle size and pore size 120 Å was used. The LC-DAD matrices were recorded employing the Chromeleon 6.1 software (DIONEX CA). The elution of standards and samples was performed in isocratic mode with an acetonitrile:water (70:30, v/v) mixture at a flow rate of 1.0 mL min-1 while maintaining the temperature of the column constant at 30 °C.

Calibration and validation sets

The calibration set consists of nine pure standards in triplicate for each analyte in the range of 1000 to 5000 ng mL-1, with equally spaced increments of 500 ng mL-1. All calibration standards were prepared by dilution of an appropriate aliquot of the stock solution.

The validation set was designed to assess the predictive ability of the calibration models, and to exploit the second order advantage. To achieve this goal, the validation set consisted in sixteen mixtures (in triplicate) of DAI, GEN and EQL generated according to a Taguchi design3838 Wang, J.; Wa, W.; Int. J. Hydrogen Energy 2009, 34, 235. with three factors at four levels (1300, 2300, 3300, and 4300 ng mL-1), whose concentrations for each of the sixteen validation mixtures are shown in Table 1.

Table 1
Validation set built according to a Taguchi design

DAI and GEN are the analytes quantified in this work, and EQL was added to the validation mixtures to simulate a potential interferent in the quantification of DAI and GEN by the proposed methodology. EQL was selected because it co-elutes with DAI and GEN in the chromatographic conditions used in this work. Nevertheless, the latter compound is also detected in real water samples.

River water samples

Eight river water samples were collected at different points along the Cuiá River in Paraíba, Brazil. Each sample was stored in a 1 L amber bottle, and acidified to pH 3 with acetic acid. The samples were subsequently processed in triplicate, as explained in the solid phase extraction procedure section.

Recovery

To evaluate the accuracy of the method, the river water samples were spiked with appropriate amounts of each analyte, generating samples at a concentration of 1 ng mL-1. The spiked samples were also processed by solid phase extraction.

Solid phase extraction procedure

The samples (1.00 L spiked and non-spiked) were loaded on the C18 phase cartridge, pre-conditioned with 6 mL of the acetonitrile and 6 mL of water (both HPLC grade), at a flow rate of 3 mL min-1. The elution was collected and dried under a gentle stream of nitrogen gas. The residues were dissolved with 500 µL of acetonitrile in order to achieve an enrichment factor of 2000. After this process, the samples were stored in vials for further analysis.

Software and chemometric analysis

The data modeling by MCR-ALS was carried out in environmental MatLab® using the graphical interface MVC2 developed by Olivieri et al.,3939 Olivieri, A. C.; Wu, H.; Yu, R.; Chemom. Intell. Lab. Syst. 2009, 96, 246. and available on line.4040 http://www.iquir-conicet.gov.ar/descargas/mvc2.rar accessed on May 03, 2015.
http://www.iquir-conicet.gov.ar/descarga...

Briefly, MCR is a bilinear decomposition method which assumes that the responses of each constituent of the system are additive3737 Arancibia, J. A.; Damiani, P. C.; Escandar, G. M.; Ibañez, G. A.; Olivieri, A. C.; J. Chromatogr. B 2012, 910, 22.as shown in equation 1.

where D contains information collected experimentally as dimensions J × K. The information contained in D is the signal recorded in a system LC-DAD, E(J × K) is the residue matrix,4141 Tauler, R.; Chemom. Intell. Lab. Syst. 1995, 30, 133. where J represents the elution times recorded at K wavelengths. C and ST are matrices truncated to N factors. For well-behaved systems, N is the number of chemical compounds which contain information on the pure concentration and spectral profiles.

Initially, it is necessary to know N, which may come from prior knowledge of the samples, or by use of principal component analysis (PCA).4242 Maeder, M.; Zilian, A.; Chemom. Intell. Lab. Syst. 1998, 3, 205. Here the N value was accessed by inspection of the explained variance by each PCA component.

Starting from an initial estimative of C or ST, available by evolving factor analysis,4343 Maeder, M.; Anal. Chem. 1987, 57, 527. or by determination of the purest variables;4444 Windig, W.; Guilment, J.; Anal. Chem. 1991, 63, 1425. C (J × N), and ST (N × K) are estimated and optimized using alternating least squares.3333 Jaumot, J.; Gargallo, R.; Juan, A.; Tauler, R.; Chemom. Intell. Lab. Syst. 2005, 76, 101. In this work, for all cases were used initial estimative of the spectral profile obtained by the method SIMPLISMA with 10% noise.

The bilinear decomposition described in equation 1 suffers from rotational freedom, i.e., there are many possible solutions for C and ST. However, the desirable solution is the “chemical solution”, in this case pure chromatograms and spectra. Fortunately, the drawback of rotational freedom can be circumvented by applying restrictions for the solutions (C and ST), obtained in each ALS iteration. Non-negativity, and uni-modality are the most common restrictions.3232 Escandar, G. M.; Goicoechea, H. C.; de la Peña, A. M.; Olivieri, A. C.; Anal. Chim. Acta 2014, 806, 8.

In the case of I samples being analyzed on the LC-DAD system, I matrices (J × K) are generated. When using traditional algorithms for multiway data (like parallel factor analysis; PARAFAC), the data are arranged in a cube structure (I × J × K). Such methods are based on the principle of trilinearity,3232 Escandar, G. M.; Goicoechea, H. C.; de la Peña, A. M.; Olivieri, A. C.; Anal. Chim. Acta 2014, 806, 8. commonly not obeyed by LC-DAD data, wherein elution time can be shifted for analytes from sample to sample.

The MCR-ALS algorithm is capable of handling tri-linear data deviation. The three way array (I × J × K) is arranged in the form of an augmented matrix (Daug) of size (IJ × K) (column-wise augmented matrix), or size (IK×J) (row-wise augmented matrix). It is recommended that the augmented way is the same as that which breaks the tri-linearity. The decomposition presented in equation 1 can be applied on Daug, generating the matrices Caug, ST, and Eaug for the case of the column-wise augmented matrix.3232 Escandar, G. M.; Goicoechea, H. C.; de la Peña, A. M.; Olivieri, A. C.; Anal. Chim. Acta 2014, 806, 8.

Finally, the scores obtained for the sum of the elements of the corresponding profile for each of the sub-matrices of Caug is used to construct a pseudo-univariate model in function of the concentrations of the calibration standards. The analyte concentration n is obtained by interpolation in the pseudo-univariate calibration line.3232 Escandar, G. M.; Goicoechea, H. C.; de la Peña, A. M.; Olivieri, A. C.; Anal. Chim. Acta 2014, 806, 8.

Results and Discussion

Calibration set: general considerations

Figure 2 displays the chromatograms obtained for each of the calibration standards with absorbance recorded at 280 nm. A partial separation of the analytes (DAI and GEN) can be observed, although the resolution is less than one.

Figure 2
Chromatograms corresponding to pure calibration standards daizein (solid line) and genistein (dashed line) at 280 nm. The black vertical line is the average retention time for each calibration set.

Another interesting observation can be made by visual inspection of Figure 2; significant peak shifts are produced with different runs. This problem can be exacerbated by the presence of interferents (in real samples), which typically vary from sample to sample, making the use of peak alignment algorithms unfeasible, as was reported by Boeris et al.4545 Boeris, V.; Arancibia, J. A.; Olivieri, A. C.; Anal. Chim. Acta 2014, 814, 23. Given the above, MCR-ALS was chosen for modeling the data, because of its advantages, as was discussed in the introduction. A previous analysis carried out with the calibration matrices for a single analyte using extended MCR-ALS showed that two contributions (analyte and background profile) were retrieved.

Validation set

MCR-ALS was applied to predict concentration for the validation set composed of sixteen mixtures of DAI and GEN in the presence of EQL (un-modeled). As an example of the modeled data, Figure 3a presents a typical LC-DAD surface for the samples validation set (sample No. 1, see Table 1).

Figure 3
(a) Landscape obtained by LC-DAD for validation sample number 1; (b) chromatograms registered at 280 nm for pure standard solutions of (solid line) DAI, (dashed line) GEN and (circle line) EQL, all of them at 5000 ng L-1 and (c) the corresponding spectra of the latter compounds.

On inspection of Figure 3a, it is possible to visualize the strong overlapping among the analyte signals, as well as the one corresponding to the interferent. EQL shows a retention time between that of DAI and GEN, overlapping simultaneously the signals of both analytes in both instrumental modes. It is important to remember that for traditional chromatographic methods, with complete resolution of all peaks, an unexpected constituent in a single sample co-eluting with the analytes requires revalidation of the method, a highly time consuming and laborious task. Figures 3b and 3c show the chromatograms and spectra corresponding to the pure standards of both analytes, and the interfering compound (all of them at 5000 ng mL-1). In these figures, the high degree of overlapping among the signals in both modes is evident.

As was previously mentioned, the first step when modeling data via MCR is to estimate the number of components N by principal component analysis. In ideal conditions, N must be equal to the chemical rank, i.e., equal to the number of chemical compounds in the validation mixtures, in this case three. However, sometimes a larger N value should be considered for better model fit. The increased number of components can be attributed to the complexity of the mixture, and to the presence of a strong background, which can lead to models with rank deficiency, as has been well discussed elsewhere.4646 Amrhein, M.; Srinivasan, B.; Bonvin, D.; Schumacher, M. M.; Chemom. Intell. Lab. Syst. 1996, 33, 17. In this study, analysis of the augmented matrices was carried out by PCA, and the explained variance by each PC for the calibration and validation sets is shown in Table 2.

Table 2
Explained variance in PCA for calibration and validation sets

As can be seen, the analysis of PCA suggests two factors for both calibration sets, and four factors for the validation set, this amount of factors was used in the decomposition of data by MCR.

The initial estimatives of the pure spectral profiles was conducted using a method based on detection of pure variables (SIMPLISMA) with 10% noise. The estimated initial profiles retrieved for the validation samples can be assigned to DAI, GEN, the interferent EQL, and to the background.

The column-wise augmented matrix was then subjected to MCR-ALS decomposition, applying the non-negativity constraint in both modes and uni-modality in the time mode, except for background profile. In all cases ALS converged with a number of iterations less than or equal to thirty, and a residual fit of less than 0.51, which is in agreement with the typical noise of a DAD detector.

Figure 4 displays the optimized spectral and concentration profiles retrieved by MCR-ALS for the validation set samples. In this figure we observe a very close resemblance between the experimental profiles of both instrumental modes (see Figures 3a and 3b).

Figure 4
Profiles retrieved by MCR-ALS: (a) concentration, and (b) spectral. Daizein (solid line), Genistein (dashed line), Equol (circle line) and background (diamond line).

Beyond simple visual comparison of the spectra recovered by MCR, they may also be evaluated for degree of overlap (S12) between the experimental normalized pure spectrum (s1), and the retrieved spectrum (s2).4747 Culzoni, M. J.; Goicoechea, H. C.; Ibañez, G. A.; Lozano, V. A.; Marsili, N. R.; Olivieri, A. C.; Pagani, A. P.; Anal. Chim. Acta 2008, 614, 46. The S12 value is calculated according to equation 2.

The S12 value can vary in a range from 0 to 1, indicating extreme conditions, wherein 1 represents complete overlap between the real and retrieved profiles. On the other hand, 0 indicates that the profiles are very different. By applying equation 2 for S12, values of 1.0000 and 0.9998 were respectively obtained for DAI and GEN, indicating a satisfactory fit, and a high similarity for each spectrum and profile couple.

After the fitting, a pseudo-univariate calibration line was constructed through linear regression by ordinary least squares, between the value of the concentration of the calibration standards and the MCR scores. The concentration of the test samples was calculated according to equation 3.

where atest,n is the area under the concentration profile for the analyte n of a matrix test sample; b1 and b0 are the slope and intercept of the pseudo-univariate curve, respectively; and cn is the concentration of the analyte n in the test sample obtained by interpolation of the curve. The statistical parameters corresponding to the validation of the MCR models for DAI and GEN quantification are summarized in Table 3. In the latter table, observes that the MCR models were able to predict the concentration of GEN and DAI in validation samples, even in the presence of the interferent EQL. Root mean square error validation (RMSEV) (equation 4) values of 177 (DAI) and 144 ng mL-1 (GEN), and relative error prediction (REP) (equation 5) values of 5.9 (DAI) and 4.8% (GEN) were obtained.

Table 3
Summary of the figures of merit obtained when analyzing the validation set samples

where ypred is the concentration predicted by the MCR model for the ith validation sample, ynom is the nominal concentration, I is the number of validation samples. In equation 5, ycal is the average of the nominal concentration values of the calibration set samples.

Table 3 also presents several figures of merit: selectivity (SEL), analytical sensitivity (γ), limit of detection (LOD), and limit of quantitation (LOQ), which were calculated according to Bauza et al.,4848 Bauza, M. C.; Ibañez, G. A.; Tauler, R.; Olivieri, A. C.; Anal. Chem. 2012, 84, 8697. and which complement the validation of the MCR models. As can be seen, the analytical sensitivity and selectivity achieved is high enough to make the method even more attractive, considering the values of LOD and LOQ which were reached, computing them through the formulas presented by Olivieri.4949 Olivieri, A. C.; Chem. Rev. 2014, 114, 5358.Although the proposed method achieved a LOD in parts per billion, for real samples a pre-concentration step is needed.

Real samples

The eight river water samples were processed in triplicate using solid phase extraction, and run in an HPLC-DAD system under the same calibration conditions and validation set. The typical LC-DAD landscape sample (No. 1) being representative, is shown in Figure 5a. For all real samples, only one peak 1 was observed with an elution time around 2.60 min. We observed significant background presence (peak 2). The real samples data were analyzed similarly to the validation set samples by MCR-ALS, and for both analytes the concentration was below the limit of detection of the propose method. The real samples were then spiked as indicated in the recovery section. The surface landscape, corresponding to a typical sample (No. 1) LC-DAD matrix registered for a river water sample, and spiked with DAI and GEN is presented in Figure 5b. In this figure, the complexity of the analytical problem involved can be appreciated: the presence of a heavy organic load in samples of river water generates an extremely overlapped signal. The analyte signals are identified by the numbers 3 and 4. Behavior similar to that which is commented for the first real sample (No. 1), in Table 3, was observed for all samples.

Figure 5
(a) Typical LC-DAD landscape after SPE for river water sample and (b) landscape obtained for the typical (No. 1) real spiked water river sample with 1.0 ng mL-1 for both analytes.

Four factors were suggested from the PCA results for the LC-DAD data of the spiked real samples. A number of factors were chosen based on the results of Table 4.

Table 4
Explained variance in PCA for test set

The second column of Table 4 shows the explained variance for the first 10 principal components for non-spiked samples. After the second factor there were no appreciable variations in the explained variance. After the samples were spiked with DAI and GEN, PCA analysis indicated four factors (see third column of Table 4).

MCR was conducted using initial estimatives of the spectra (by SIMPLISMA with 10% noise), under the non-negativity restriction in both modes, and mean unimodality only in the chromatographic mode. Unimodality was not applied to the background profile. In Figure 6a (concentration profile), and Figure 6b (spectral profile), the profiles retrieved by MCR-ALS (using four factors) are shown. The profiles corresponding to DAI and GEN are displayed as solid blue and dashed green lines, respectively, and show good agreement with the experimentally recorded spectra (both shown with black circles and squares in Figure 6b), suggesting a good model fit with four factors.

Figure 6
Representative profiles retrieved by MCR-ALS (from sample No. 1) (a) concentration and (b) spectral. Daizein (solid line), Genistein (dashed line), unknown (circle line), and background (diamond line).

Regarding Figure 6, the recovered profile can be also seen corresponding to the contribution of the interfering (circle red line). The concentration profile (Figure 6a) shows a wide peak, suggesting co-elution of an organic load signal, superimposed with both analytes. The spectral profile (Figure 6b) displays a spectrum with a maximum around 235 nm. It should be noted that in conventional chromatographic method, co-elution may require revalidation of the method. However, the use of mathematic signal separation overcomes this type of drawback in the analysis of complex samples. The background profile (diamond line), recorded for real samples appears much more intense than in the validation samples (see Figure 4).

Finally, concentration values of GEN and DAI were estimated by interpolation of the MCR areas of the test samples in the pseudo univariate calibration curve (the typical linear fit is displayed in the Figure 7). The pseudo univariate calibration curve was obtained by linear fit of the MCR areas for calibration samples vs. nominal concentration.

Figure 7
Representative pseudo univariate calibration curve for (a) DAI and (b) GEN.

The results obtained are presented in Table 5. It is important to notice that the results for the DAI and GEN quantitations refer to real sample concentrations before the SPE step; in other words the figures of merit take into account the enrichment factor of the samples. Thus, a different magnitude in values can be seen when comparing the later values with those presented in Table 3 (validation set).

Table 5
Summary of the results for application of the validated for MCR-ALS models in analysis of the spiked (1 ng mL-1) real river water samples

As can be seen, the proposed method was able to predict the concentration of both analytes in spiked river water samples even in the presence of interferents and a strong background. This was achieved with an acceptable accuracy, which was verified through the recovery values that ranged between 99 and 125 for DAI, and from 77 to 111 for GEN, with REP% values of 9.38 and 8.51.

Comparing the method reported in this work with the one published by Wang et al.,2323 Wang, C.; Wang, L.; Zhao, Q.; Chen, J.; Zheng, L.; Zheng, M.; Zhang, R.; Wang, Z.; Anal. Methods 2013, 5, 3688. in which the quantitation of DAI and GEN in river water samples using HPLC-DAD is reported, we noted a 50% reduction in the running time. The REP values obtained by application of both methods are quite similar. However, the new method shows higher sensitivity, with 0.17 ng mL-1 of LOD for DAI, and 0.20 for GEN. Moreover, the proposed method makes use of the second order advantage, making it more robust when unexpected constituents occur in the sample, thus avoiding re-validations.

Conclusions

In this work a rapid method using high performance liquid chromatography with molecular absorption in the ultraviolet-visible region of detection for simultaneous quantification of daidzein, and genistein, (two potential emerging water contaminants) was presented. The strategy successfully models higher order data with MCR-ALS to overcome the disadvantages of a complex matrix which has a wide variety of interferents, which change from sample to sample. The proposed methodology was first evaluated for a synthetic validation set and then in spiked real water samples from the Cuiá River in João Pessoa, Brazil. Both cases showed good results.

Acknowledgements

The authors are grateful to CNPq, CAPES and INCTAA for scholarships and financial support. Hector C. Goicoechea thanks UNL, CONICET, and ANPCyT for financial support.

Reference

  • 1
    Pesavento, M.; Alberti, G.; Biesuz, R.; Anal. Chim. Acta 2009, 631, 129.
  • 2
    Stankovic, S.; Kalaba, P.; Stankovic, A. R.; Environ. Chem. Lett 2014, 12, 63.
  • 3
    Manciocco, A.; Calamandrei, G.; Alleva, E.; Chemosphere 2014, 100, 1.
  • 4
    Pal, A.; He, Y.; Jekel, M.; Reinhard, M.; Gin, K. Y. H.; Environ. Int. 2014, 71, 46.
  • 5
    Jefferson, W. N.; Padilla-Banks, E.; Newbold, R. R.; Reprod. Toxicol 2007, 23, 308.
  • 6
    Kiparissis, Y.; Hughes, R.; Metcalfe, C.; Ternes, T.; Environ. Sci. Technol 2001, 35, 2423.
  • 7
    Lundgren, M. S.; Novak, P. J.; Environ. Toxicol. Chem 2009, 28, 2318.
  • 8
    Bednarek, P.; Franski, R.; Kerhoas, L.; Einhorn, J.; Wojtaszek, P.; Stobiecki, M.; Phytochemistry 2001, 56, 77.
  • 9
    Coldham, N. G.; Sauer, M. J.; Food Chem. Toxicol 2001, 39, 1211.
  • 10
    Zhang, L.; Khan, I. A.; Foran, C. M.; Comp. Biochem. Physiol., Part C: Toxicol. Pharmacol. 2002, 132, 203.
  • 11
    Clotfelter, E. D.; Rodriguez, A. C.; Environ. Pollut. 2006, 144, 833.
  • 12
    Kelly, M. M.; Arnold, W. A.; Environ. Sci. Technol 2012, 46, 5396.
  • 13
    Kawanishi, M.; Takamura-Enya, T.; Ermawati, R.; Shimohara, C.; Sakamoto, M.; Matsukawa, K.; Matsuda, T.; Murahashi, T.; Matsui, S.; Wakabayashi, K.; Environ. Sci. Technol 2004, 38, 6424.
  • 14
    Bustamante-Rangel, M.; Delgado-Zamarreño, M. M.; Carabias-Martínez, R.; Domínguez-Álvarez, J.; Anal. Chim. Acta 2012, 709, 113.
  • 15
    Dai, Q.; Franke, A. A.; Jin, F.; Shu, X. O.; Hebert, J. R.; Custer, L. J.; Cheng, J. R.; Gao, Y. T.; Zheng, W.; Cancer Epidem. Biomar. 2002, 11, 815.
  • 16
    He, F. J.; Chen, J. Q.; Food Sci. Hum. Wellness 2013, 2, 146.
  • 17
    Tekel, J.; Daeseleire, A.; Heeremans, E.; van Peteghem, C.; J. Agric. Food Chem 1999, 47, 3489.
  • 18
    Laganà, A.; Bacaloni, A.; De Leva, I.; Faberi, A.; Fago, G.; Marino, A.; Anal. Chim. Acta 2004, 501, 79.
  • 19
    Kuster, M.; Azevedo, D. A.; Alda, M. J. L.; Aquino Neto, F. R.; Barceló, D.; Environ. Int. 2009, 35, 997.
  • 20
    Comber, M. H. I.; Williams, T. D.; Stewart, K. M.; Water Res 1993, 27, 273.
  • 21
    Wang, C.; Prasain, J. K.; Barnes, S.; J. Chromatogr. B 2002, 777, 3.
  • 22
    Bai, Y.; Du, F.; Bai, Y.; Liu, H.; Anal. Methods 2010, 2, 1867.
  • 23
    Wang, C.; Wang, L.; Zhao, Q.; Chen, J.; Zheng, L.; Zheng, M.; Zhang, R.; Wang, Z.; Anal. Methods 2013, 5, 3688.
  • 24
    Shihabi, Z. K.; Kute, T.; Garcia, L. L.; Hinsdale, M.; J. Chromatogr. A 1994, 680, 181.
  • 25
    Wang, G.; Kuan, S. S.; Francis, O. J.; Ware, G. M.; Carman, A. S.; J. Agric. Food Chem 1990, 38, 185.
  • 26
    Chen, G.; Zhang, J.; Ye, J.; J. Chromatogr. A 2001, 923, 255.
  • 27
    Lapcik, O.; Hampl, R.; Hill, M.; Wahala, K.; Al-Maharik, N.; Adlercreutz, H.; J. Steroid Biochem. 1998, 64, 261.
  • 28
    Fogliatto, D. K.; Barbosa, A. M. J.; Ferreira, V. S.; Colloids Surf., B 2010, 78, 243.
  • 29
    Shenone, A. V.; Culzoni, M. J.; Marsili, N. R.; Goicoechea, H. C.; Food Chem. 2013, 138, 1928.
  • 30
    Borraccetti, M. D.; Damiani, P. C.; Olivieri, A. C.; Analyst 2009, 134, 1682.
  • 31
    Elcoroaristizabal, S.; Juan, A.; García, J. A.; Durana, N.; Alonso, L.; Chemom. Intell. Lab. Syst 2014, 132, 63.
  • 32
    Escandar, G. M.; Goicoechea, H. C.; de la Peña, A. M.; Olivieri, A. C.; Anal. Chim. Acta 2014, 806, 8.
  • 33
    Jaumot, J.; Gargallo, R.; Juan, A.; Tauler, R.; Chemom. Intell. Lab. Syst 2005, 76, 101.
  • 34
    Lozano, V. A.; Escandar, G. M.; Anal. Chim. Acta 2013, 782, 37.
  • 35
    Alcaráz, M. R.; Shenone, A. V.; Culzoni, M. J.; Goicoechea, H. C.; Microchem. J 2014, 112, 25.
  • 36
    Culzoni, M. J.; Goicoechea, H. C.; Pagani, A. P.; Cabezon, M. A.; Olivier, A. C.; Analyst 2006, 131, 718.
  • 37
    Arancibia, J. A.; Damiani, P. C.; Escandar, G. M.; Ibañez, G. A.; Olivieri, A. C.; J. Chromatogr. B 2012, 910, 22.
  • 38
    Wang, J.; Wa, W.; Int. J. Hydrogen Energy 2009, 34, 235.
  • 39
    Olivieri, A. C.; Wu, H.; Yu, R.; Chemom. Intell. Lab. Syst. 2009, 96, 246.
  • 40
    http://www.iquir-conicet.gov.ar/descargas/mvc2.rar accessed on May 03, 2015.
    » http://www.iquir-conicet.gov.ar/descargas/mvc2.rar
  • 41
    Tauler, R.; Chemom. Intell. Lab. Syst. 1995, 30, 133.
  • 42
    Maeder, M.; Zilian, A.; Chemom. Intell. Lab. Syst 1998, 3, 205.
  • 43
    Maeder, M.; Anal. Chem. 1987, 57, 527.
  • 44
    Windig, W.; Guilment, J.; Anal. Chem. 1991, 63, 1425.
  • 45
    Boeris, V.; Arancibia, J. A.; Olivieri, A. C.; Anal. Chim. Acta 2014, 814, 23.
  • 46
    Amrhein, M.; Srinivasan, B.; Bonvin, D.; Schumacher, M. M.; Chemom. Intell. Lab. Syst. 1996, 33, 17.
  • 47
    Culzoni, M. J.; Goicoechea, H. C.; Ibañez, G. A.; Lozano, V. A.; Marsili, N. R.; Olivieri, A. C.; Pagani, A. P.; Anal. Chim. Acta 2008, 614, 46.
  • 48
    Bauza, M. C.; Ibañez, G. A.; Tauler, R.; Olivieri, A. C.; Anal. Chem 2012, 84, 8697.
  • 49
    Olivieri, A. C.; Chem. Rev. 2014, 114, 5358.

Publication Dates

  • Publication in this collection
    Aug 2015

History

  • Received
    03 Mar 2015
  • Accepted
    15 May 2015
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br