Gamma radiation effects on polydimethylsilane stationary phases for use in packed-column gas chromatographic analyses

Abstracts

Gamma radiation induced immobilization of several polydimethylsilane liquid stationary phases for use in packed-column gas chromatographic separations has been studied. Extraction tests show that moderate doses of gamma radiation (80-140 kGy) are sufficient to produce significant (90%) immobilization of most polydimethylsilanes onto Chromosorb supports, although a molecular mass effect is seen. Thermal stability also increases significantly with radiation dose, suggesting higher temperature use with smaller volatility losses. Infrared spectra confirm the continued presence of the stationary phase on the support after thermal stability tests and after exhaustive in-column washing. The column chromatographic behavior of the immobilized phases is equal to or better than that of the unirradiated phases, except for higher doses (300 kGy) of gamma-radiation. Columns prepared from gamma-immobilized polydimethylsilane have been used successfully in analyses where column contamination from high boiling materials requires frequent column recuperation.

gas chromatography; packed columns; immobilized phases; gamma radiation


Este trabalho estudou os efeitos da radiação gama como indutora da imobilização de fases estacionárias líquidas de polidimetilsilicones usadas para separações cromatográficas em colunas recheadas. Testes de extração em solventes mostram que doses moderadas de radiação gama (80-140 kGy) produzem imobilização (90%) da maioria dos polidimetilsilicones testados em suportes do tipo Chromosorb, apesar do efeito da massa molecular ser importante. A estabilidade térmica também aumenta significativamente, possibilitando o uso de altas temperaturas sem perdas por volatilização. Espectros no infravermelho confirmam a presença de fase estacionária no suporte após testes de estabilidade térmica e após lavagem exaustiva com solventes em coluna cromatográfica. O comportamento cromatográfico das fases imobilizadas é igual ou melhor do que o das fases não-irradiadas, exceto no caso de altas doses (300 kGy) de radiação gama. As colunas preparadas com polimetilsilicones imobilizados por radiação gama podem ser usadas com sucesso em análises onde a contaminação da coluna por compostos de alto ponto de ebulição exigem frequente recuperação da mesma.


Article

Gamma Radiation Effects on Polydimethylsilane Stationary Phases for Use in Packed-Column Gas Chromatographic Analyses

M.B. Adaime * * Present address: Departamento de Química, Universidade Federal de Santa Maria, Campus Universitário - Camobí, 97.119-900 Santa Maria, RS - Brazil , M.J.T.F. dos Santos, I. Facchin, K.E. Collins, and C.H. Collins

Instituto de Química , Universidade Estadual de Campinas, C.P. 6154, 13.083-970 Campinas - SP, Brazil

Este trabalho estudou os efeitos da radiação gama como indutora da imobilização de fases estacionárias líquidas de polidimetilsilicones usadas para separações cromatográficas em colunas recheadas. Testes de extração em solventes mostram que doses moderadas de radiação gama (80-140 kGy) produzem imobilização (90%) da maioria dos polidimetilsilicones testados em suportes do tipo Chromosorb, apesar do efeito da massa molecular ser importante. A estabilidade térmica também aumenta significativamente, possibilitando o uso de altas temperaturas sem perdas por volatilização. Espectros no infravermelho confirmam a presença de fase estacionária no suporte após testes de estabilidade térmica e após lavagem exaustiva com solventes em coluna cromatográfica. O comportamento cromatográfico das fases imobilizadas é igual ou melhor do que o das fases não-irradiadas, exceto no caso de altas doses (300 kGy) de radiação gama. As colunas preparadas com polimetilsilicones imobilizados por radiação gama podem ser usadas com sucesso em análises onde a contaminação da coluna por compostos de alto ponto de ebulição exigem frequente recuperação da mesma.

Gamma radiation induced immobilization of several polydimethylsilane liquid stationary phases for use in packed-column gas chromatographic separations has been studied. Extraction tests show that moderate doses of gamma radiation (80-140 kGy) are sufficient to produce significant (90%) immobilization of most polydimethylsilanes onto Chromosorb supports, although a molecular mass effect is seen. Thermal stability also increases significantly with radiation dose, suggesting higher temperature use with smaller volatility losses. Infrared spectra confirm the continued presence of the stationary phase on the support after thermal stability tests and after exhaustive in-column washing. The column chromatographic behavior of the immobilized phases is equal to or better than that of the unirradiated phases, except for higher doses (300 kGy) of gamma-radiation. Columns prepared from gamma-immobilized polydimethylsilane have been used successfully in analyses where column contamination from high boiling materials requires frequent column recuperation.

Keywords: gas chromatography, packed columns, immobilized phases, gamma radiation

Introduction

Liquid polymeric stationary phases used in gas chromatography often present limitations due to volatility or thermal stability, which impose upper temperature limits on their use. One way to alleviate this problem is through chemical bonding of the liquid stationary phase to the support1, as is done for most HPLC stationary phases. Another way to improve stationary phase stability is by crosslinking reactions which increase the polymer molecular mass. These immobilization processes, which involve bonding between the chains with possible bonding to the support2, are usually initiated by free radicals produced by heat3,4, chemical initiators such as peroxides5,6, azo compounds7,8 or ozone9, low temperature plasmas10, or by ionizing radiation from accelerated electrons11 or from gamma irradiation12-17.

These free radical initiated immobilization methods have been very successful in producing modern high temperature capillary columns. However, very little information exists about immobilization of supported liquid stationary phases for use in packed-column gas chromatography18. This investigation was undertaken to investigate the properties of several polydimethylsilanes on Chromosorb supports after immobilization by different doses of gamma irradiation.

Experimental

Materials

The polydimethylsilane liquid phases were dissolved in chloroform for preparing different loadings on Chromosorb W (80-100 mesh) or Chromosorb W-HP (80-100 mesh). Portions of the prepared packings were placed in glass ampoules and sealed under air for irradiation in an industrial irradiation facility operated by IBRAS-CBO (Campinas, Brazil). All solvents and most of the test compounds were p. a. and used without further treatment. When necessary, reagents were purified by conventional methods19.

Equipment

Two different gas chromatographic systems were used: a Varian 1700 and an Instrumentos Científicos CG 500, both equipped with flame ionization detectors. Packed stainless steel columns were positioned to permit on-column injection.

Infrared spectra were taken on a Perkin-Elmer 399B spectrophotometer while thermogravimetric analyses were carried out on a DuPont model 1090B thermal analyzer with a model 951'thermogravimetric analyzer.

Testing procedures

Solvent extraction of the packings was carried out in a modified Soxhlet apparatus20 with three solvents (methanol, hexane and chloroform), each for six hours.

A number of compounds with different polarities were used to evaluate the efficiency of each stationary phase, as determined by the retention factor (k), plate number (N), resolution (Rs) and asymmetry factor (As) at 10% of peak height. The chromatographic conditions for each separation, isothermal or with temperature programming, were determined by optimizing resolution of each mixture on an unirradiated column. The mobile phase (N2) retention time was determined by injecting a mixture of propane and butane (cooking gas) at three different temperatures (60, 110 and 160 °C).

For the thermal stability tests, the column was heated in the chromatograph oven, without the detector connection, for one hour each at successively higher temperature, measuring the efficiency (plate number) of a standard compound (2-butanone) at 60 °C after each step.

The columns were also subjected to successive rinsings at 200 °C15, injecting one column volume each of acetone, 1:1 acetone-dichloromethane, 4:1 dichloromethane-pentane, 1:1 dichloromethane-pentane and pentane, with the detector disconnected. Following this, the efficiencies of the columns were determined (n-decane and benzyl alcohol) at 110 °C.

Results and Discussion

Solvent extraction removes all of the polydimethylsilane liquid stationary phase (SP) from unirradiated packings. Extraction removes a small amount of the SP from the irradiated packings, quantities which appear directly related to the irradiation dose received (Table 1). Immobilization by radiation was less effective with SP-2100 and DC-200, although even with these packings some immobilization occurred, as shown by the continued presence of the 2960 cm-1 band in the infrared spectra (Figs. 1 and 2) after extraction of the irradiated samples. Extraction of unirradiated packings removes the liquid SP and this band disappears.



Low irradiation doses are sufficient to immobilize significant quantities of SE-30, with no dependence on either the initial quantity placed on the support or on whether the support was, or was not, silanized (Table 1). Higher doses increase the quantity immobilized. The fact that the same percentage of SE-30 appears to be immobilized onto the support, independent of the initial quantity of liquid18 or of the presence of methyl groups from the silanizing reaction on Chromosorb W-HP, suggests that the immobilization is mainly related to crosslinking of the polydimethylsilane chains rather than to formation of chemical bonds to the support.

The different percentages of immobilization observed, comparing SE-30 and DC-200, may be related to their different molecular masses, since SE-30 has a mass of about 1.0 - 2.5 x 106 daltons while DC-200 has a range of only 1.6 - 6.7 x 104 daltons21. Thus, to fully immobilize DC-200 would require much more crosslinking than is needed with SE-30.

In contrast to the results obtained by extraction, thermogravimetry on the several packings did not indicate significant differences between unirradiated and irradiated packings (Table 2). On the other hand, thermal stability tests carried out on packed columns indicate significant increases in the temperature maximum of the irradiated (immobilized) packings, in comparison to unirradiated packings. The temperature maximum usually indicated22 for SE-30 is 300 °C, for DC-200, 250 °C and for SP-2100, 350 °C. As shown in Table 3, the immobilized packings have useful temperature ranges up to 50 °C higher than these values.

In order to see how much liquid stationary phase remained after this heat treatment, a portion of the packing was removed from the column after the 400 °C step for extraction. For the unirradiated packing, 62% of the quantity initially present was extracted after the heat treatment, indicating that part of the liquid SP had decomposed and/or volatilized, since infrared spectra taken after the extraction did not show the presence of residual polydimethylsilane. No evidence of volatilization was obtained for the irradiated phases after the heat treatment; both the extraction tests and the infrared spectra were the same as before the 400 °C thermal treatment.

Extensive column rinsing of radiation immobilized SE-30 packings did not significantly change their efficiencies, nor did it modify the infrared spectra (Fig. 3). Extraction of the unirradiated packing after rinsing removed 78% of the quantity originally present, indicating that up to 22% had been washed off in the rinsing test. By contrast, no change was observed in the amount extracted (and in the subsequent infrared spectra) for irradiated phases after the rinsing test. This enhanced stability property, which permits extensive rinsing of a packed column, has proved very useful in column recuperation when injecting reaction product mixtures having a wide range of properties and including high boiling materials where frequent column recuperation is required23.


Several different synthetic mixtures were analyzed on the packed columns. In the packing containing DC-200, both the plate number and the asymmetry factor were poorer after irradiation, although no change in resolution was observed (Fig. 4). On the other hand, the immobilization of SP-2100 produced a packing with better chromatographic characteristics (Fig. 5). With SE-30 the chromatographic behavior appears to depend on the test compound, as summarized in Table 4.



As shown in Figs. 6 and 7, resolution improved after irradiation for SE-30 on Chromosorb W-HP, even through neither the retention times nor the plate numbers changed significantly. This improvement is related to the irradiation dose received. However, as indicated in Fig. 7c, both retention time and resolution decreased after a high dose. This may be related to a higher degree of crosslinking, which has produced a much more viscous phase, or to radiation decomposition of the SP.



It thus appears that, for each liquid stationary phase-support combination, there is an optimal dose which results in a high percentage of liquid stationary phase being immobilized without causing so much crosslinking that the phase becomes too viscous to show good chromatographic behavior, or promoting polymer degradation. Irradiation immobilization of liquid stationary phases on supports can be carried out before packing the column, permitting an extraction test to evaluate the degree of immobilization obtained. Such immobilized packings show enhanced stability at higher temperatures and permit rinsing the column for recuperation, when it becomes contaminated, without losing characteristic chromatographic properties.

Acknowledgements

The authors thank CNPq for fellowships, FAPESP, FAEP (UNICAMP) and Dow Química do Brasil for financial support and IBRAS-CBO for invaluable cooperation.

References

Received: June 23, 1998

FAPESP helped in meeting the publication costs of this article

  • 1.Abel, E.W.; Polland, F.W.; Uden, P.C.; Nickless, G.D. J. Chromatogr. 1966, 22, 23.
  • 2.Grob, K.; Grob, G. J. High Resol. Chromatogr. Chromatogr. Commun. 1981, 4, 491.
  • 3.Petsev, N.D.; Pekov, G.I.; Alexandrova, M.D.; Dimitrov, C. Chromatographia 1985, 20, 228.
  • 4.Lai, G.; Mühlek, U.; Nicholson, G.J.; Schmid, J.; Bayer, E. Chromatographia 1991, 32, 241.
  • 5.Schneider, R.C.S.; Pizzutti, I.R.; Adaime, M. B. J. Braz. Chem. Soc. 1997, 8, 3.
  • 6.Cigánek, M.; Dressler, M.; Teply, J. J. Chromatogr. 1991, 588, 225.
  • 7.Janák, K.; Horká, M.; Tesarík, K. J. Chromatogr. 1989, 471, 237.
  • 8.Horká, M.; Kahle, V.; Krejá, M. J. Chromatogr. 1993, 637, 96.
  • 9.Chuang, C.H.; Shanfield, H.; Zlatkis, A. Chromatographia 1987, 3, 169.
  • 10.Springston, S.R.; Dezaro, D.A. J. Chromatogr. 1989, 473, 79.
  • 11.Markides, K.; Blomberg, L.; Buijten, J.; Wämmam, T. J. Chromatogr. 1983, 267, 29.
  • 12.Bertsch, W.; Pretorius, V.; Pearce, M.; Thompson, J.C.; Schnautz, N.G. J. High Resol. Chromatogr. Chromatogr. Commun. 1982, 5, 432.
  • 13.Schomburg, G.; Husmann, H.; Ruthe, S.; Herraiz, M. Chromatographia 1982, 15, 599.
  • 14.Barry, E.F.; Hubball, J.A.; DiMauro, P.R.; Chabot, G.E. J. High Resol. Chromatogr. Chromatogr. Commun. 1983, 6, 300.
  • 15.Hubball, J.A.; DiMauro, P.R.; Barry, E.F.; Lyons, E.A.; George, W.A. J. Chromatogr. Sci. 1984, 22, 185.
  • 16.Vigh, G.; Etler, O. J. High Resol. Chromatogr. Chromatogr. Commun. 1984, 7, 620.
  • 17.Tatar, V.; Mopl, M.; Matucha, M.; Pesek, M. J. Chromatogr. 1985, 328, 337.
  • 18.Basso, M.A.; dos Santos, M.J.T.F.; Collins, K.E.; Collins, C.H. J. High Resol. Chromatogr. 1989, 12, 500.
  • 19.Perrin, D.D.; Armarege, W.L.F.; Perrin, D.R. "Purification of Laboratory Chemicals", 2nd. ed., Pergamon Press, Oxford, 1980.
  • 20.Sanchez, E.F.; Dominguez, J.A.G.; Muńoz, J.G.; Molera, M.J. J. Chromatogr. 1984, 299, 151.
  • 21.Haken, J.K. J. Chromatogr. 1984, 300, 1.
  • 22.Yancey, J.A. J. Chromatogr. 1985, 23, 161.
  • 23.Carvalho, W.A.; Krähembühl, C.G.Z.; Collins, C.H.; Schuchardt, U.F. Anais of the 6th Brazilian Seminar on Catalysis, Instituto Brasileiro de Petróleo, 260, 1991.
  • *
    Present address: Departamento de Química, Universidade Federal de Santa Maria, Campus Universitário - Camobí, 97.119-900 Santa Maria, RS - Brazil
  • Publication Dates

    • Publication in this collection
      01 June 2001
    • Date of issue
      Dec 1999

    History

    • Received
      23 June 1998
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br