NEUROMUSCULAR ELECTRICAL STIMULATION OF MEDIUM AND LOW FREQUENCY ON THE QUADRICEPS FEMORIS

ESTIMULAÇÃO ELÉTRICA NEUROMUSCULAR DE MÉDIA E BAIXA FREQUÊNCIA NO MÚSCULO QUADRÍCEPS FEMORAL

ABSTRACT

Objective:

The purpose of this study was to investigate the effects of neuromuscular electrical stimulation (NMES) in different frequencies on the quadriceps femoris. A randomized, controlled, blind cross-sectional study.

Methods:

Thirty subjects (12 men and 18 women), with an average age of 24.67 years, weight of 65.62 kg and height of 1.69 m were evaluated. Three random test conditions were applied: maximum voluntary isometric contraction (MVIC), maximum voluntary isometric contraction with medium frequency current (MVIC-MF) and maximum voluntary isometric contraction with low frequency current (MVIC-LF). Four MVICs were applied in each situation. The time between different isometric contraction types was 90 seconds while the time between the same conditions of contraction was 10 seconds.

Results:

Two-way ANOVA test showed that MVIC-MF had higher values for peak torque than MVIC-LF (p=0.02). Significant statistical results were found when comparing MVIC-MF and MVIC (p=0.03), but not for MVIC and MVIC-LF (p=0.52).

Conclusion:

Maximum voluntary isometric contraction associated with medium-frequency electrical stimulation was more effective than other NMES conditions. Level of Evidence II, Therapeutic Studies - Investigation of treatment results.

Keywords:
Quadriceps Muscles; Transcutaneous electrical nerve stimulation; Isometric contraction; Torque

RESUMO

Objetivo:

A finalidade deste estudo foi investigar os efeitos da estimulação elétrica neuromuscular (EENM) em diferentes frequências sobre o músculo quadríceps femoral. Estudo randomizado, transversal, controlado e cego.

Métodos:

Foram avaliados trinta indivíduos de ambos os sexos (12 homens e 18 mulheres) com média de idade de 24,67 anos, peso 65,62 kg e altura 1,69 m. Os indivíduos foram submetidos a três condições de testes randomizados: contração voluntária isométrica máxima (CVIM), contração voluntária isométrica máxima associada à corrente de média frequência (CVIMMF) e contração voluntária isométrica máxima associada à baixa frequência (CVIMBF), sendo realizadas quatro repetições de CVIM em cada situação. O tempo entre as diferentes condições de contrações isométricas foi de 90 segundos e o tempo entre as contrações isométricas das mesmas condições foi de 10 segundos.

Resultados:

O teste ANOVA mostrou que a CVIMMF obteve valores de pico de torque maiores do que a CVIMBF (p = 0,02), com diferença significativa. Foram encontrados resultados com significância estatística ao comparar CVIMMF e CVIM (p = 0,03), mas não entre CVIM e CVIMBF (p = 0,52).

Conclusões:

A contração voluntária isométrica máxima associada a EENM de média frequência foi mais efetiva do que as outras situações de EENM. Nível de evidência II, Estudos Terapêuticos – Investigação dos resultados do tratamento.

Descritores:
Músculo quadríceps; Estimulação elétrica nervosa transcutânea; Contração isométrica; Torque

INTRODUCTION

Neuromuscular stimulators can produce electrical pulses, which can cause excitation of peripheral nerves and, consequently contraction of muscular tissue. Therewith, the use of electrical stimulation by physical therapists to increase or maintain muscle performance has gained popularity.

Some factors like well-being of patient, location of electrodes, frequency and current intensity, encouraged researchers to investigate medium and low frequency currents in an attempt to minimize the discomforting as a result of the use of neuromuscular electrical stimulation (NMES).11. Sinacore DR, Delitto A, King DS, Rose SJ. Type II fiber activation with electrical stimulation: a preliminary report. Phys Ther. 1990;70(7):416-22.

Recruitment of motor units during voluntary contraction follows the principles of that the cell body size of motor neurons determines the order in which motor units are activated. Typically, type I (slow-twitch) fibers are recruited before the type II (fast-twitch) fibers according to the increasing in muscular contraction. Additionally, the order of recruitment of type II fibers occurs in the following order: First type IIa fibers and after type IIb fibers.

Studies involving electrical stimulation have suggested a selective recruitment and preferred the fast-twitch fibers (type II). This hypothesis, however, is based on the results of fatigue studies in humans, demonstrating limited scientific evidence.

The force of contraction produced by NMES has been reported, in some studies, as being higher than the maximum voluntary contraction.22. Kots JM, Hvilon VA. The training of muscular power by method of electrical stimulation. 1st ed. Moscú: State Central Institute of Physical Culture, 1971. However; some studies report that use of NMES produces a significantly lower contraction force or equal force to the maximum voluntary contraction (MVC). Therefore, in untrained individuals can generate greater force than the MVC by electrical stimulation.33. Maffiuletti NA, Zory R, Miotti D, Pellegrino MA, Jubeau M, Bottinelli R. Neuromuscular adaptations to electrostimulation resistance training. Am J Phys Med Rehabil. 2006;85(2):167-75.

The low-frequency electric current is that less than 1000Hz and are generally used for muscle contraction below 100Hz. The stimulation frequency of 50Hz is fast enough to cause muscular contraction even in motor units of rapidly contraction. Nevertheless, some studies using lower frequencies (30- 35Hz) achieved positive results.44. Robinson AJ, Snder-Mackler L. Clinical electrophysiology: electrotherapy and electrophysiologic testing. 2 ed. Baltimore: Williams & Wilkins Press, 1995.

In other investigated the influence of NMES in producing muscular force using low-frequency current (70Hz, 250μs).55. Brasileiro JS, Villar AFS. Comparação dos torques gerados por estimulação elétrica a contração muscular voluntária no músculo. Rev Bras Fisio. 2000;4(2):75-81. The torque generated by the electrical current was 45.05% of the maximum voluntary contraction and superposition of electrical stimulation to the MVC showed no significant increase in the knee extensor torque in the individuals analyzed.

The medium-frequency alternating current is defined as current frequency between 1 and 10,000Hz. Interferential current (4.000Hz) and Russian current (2.500 Hz) are the most popular NMES utilized. The reason to use higher frequency than 1000Hz is that the skin works as a barrier to capacitive current flow. As the current frequency applied increases, the skin impedance decreases.66. Robinson AJ, Snyder-Mackler L. Clinical Electrophysiology: Electrotherapy and Electrophysiologic Testing. 2ᵃ ed. Baltimore: Williams & Wilkins Press, 1995. Above 1.000 Hz skin impedance is low, less energy is dissipated to the surface of epidermis and a larger proportion of energy is intended to stimulate deeper tissues.77. Ward AR, Robertson VJ, Makowski RJ. Optimal frequencies for electric stimulation using medium-frequency alternating current. Arch Phys Med Rehabil. 2002;83(7):1024-7. Thus, it is expected that the devices provide medium frequency current (1-10.000Hz) may be able to produce higher levels of muscle contraction with less discomfort than the low frequency stimulators.

It is important to considered well-being of patient during training with electrical stimulation because the discomfort is considered one of the limiting factors in the use of NMES in promoting increased muscle strength.88. Lyons GM, Leane GE, Clarke-Moloney M, O'Brien JV, Grace PA. An investigation of the effect of electrode size and electrode location on comfort during stimulation of the gastrocnemius muscles. Med Eng Phys. 2004;26(10):873-8. An ideal electrical stimulator should be able to increase muscle strength with the least possible discomfort during stimulation.99. Delitto A, Rose SJ. Comparative confort of three waveforms used in electrically eliciting quadriceps femoris muscle contractions. Phys Ther. 1986;66(11):1704-7. A Russian physiologist1010. Kots YM. Electrostimulation. (Canadian-Soviet exchange symposium on electrostimulation of skeletal muscles, Concordia University, Montreal, Quebec, Canada; December 6–15,1977). Quoted in: Kramer J, Mendryk SW. Electrical stimulation as a strength improvement technique. J Orthop Sports Phys Ther. 1982; 4:91–98. indicated that using frequency of 2.500Hz; he was able to increase the strength of the subjects by 30-40% in 20 stimulation sessions. Each session is consisted of 10 contractions in 10 seconds each with 50 seconds rest.1111. Ward AR, Shkuratova N. Russian electrical stimulation: the early experiments. Phys Ther. 2002;82(10):1019-30.,1212. Rich NC. Strength training via high frequency electrical stimulation. J Sports Med Phys Fitness. 1992;32(1):19-25.

Taking into account the physical and physiological differences of electric current in low and medium frequency and the lack of studies, this research aimed to verify the effect of neuromuscular electrical stimulation of medium and low frequency in the generation of torque in quadriceps femoral muscle in healthy subjects.

MATERIALS AND METHODS

This is a cross-sectional randomized controlled blind study. Thirty healthy individuals of both sexes (12 male, 18 female) where selected, average age of 24.67 year (SD 3.70), weight 65.62Kg (SD 15.21) and height 1.69m (SD 0.11). Potential participants were excluded if they presented muscular recent lesion, knee injury, sensory changes, neuromuscular disease, morbid obesity and pregnancy. The project was cleared by the Institution's Ethics Committee (CEP 2.178.833) and all patients signed an Informed Consent Form. The choice of dominant lower limb was randomized with sealed envelopes. Thus, 16 right lower limbs and 14 left were selected (53.3% and 46.7%, respectively). Before being submitted to isokinetic evaluation, the volunteers performed the following protocol: 10 minutes by bicycle without load for warming, quadriceps and hamstring stretching muscles with 3 repetitions series of 30 seconds each. All individuals where familiarized with the protocol test that consisted of 4 seconds of submaximal isometric contraction (50-75%), 4 seconds of isometric contraction associated with electrical stimulation (medium and low frequency). The aim of this familiarization session was to minimize any misunderstanding with the test, especially in relation to contraction associated with electrical stimulation.

This study was conducted at Cohen Orthopedic Institute, Rehabilitation and Sports Medicine - São Paulo - Brazil. KLD (Endophasis R) and Compex were used for NMES with medium and low frequency, respectively. We used two self-adhesive electrodes with dimensions of 4.5cm X 9.3cm. The evaluation of muscle strength was performed using the isokinetic dynamometer REV 9000 Technogym, Gambettola - Italy. The patient was placed in a sitting position, hip at 90° flexion, knee at 60° of flexion. The upper body was supported by the backrest with 2 crossed belts to stabilize the trunk and one pelvic belt. Final adjustments were made by alignment of the dynamometer rotating axis with knee joint axis. The resistance of lever arm was placed in the distal leg above the malleolus (Figure 1).

Figure 1
The patient is positioning in the isokinetic dynamometer with 60° of knee flexion to evaluate isometric torque of femoral quadriceps.

Individuals were submitted to the following 3 randomized test conditions:

  • Four maximal voluntary isometric contractions (MVIC) of the femoral quadriceps muscle: 4 seconds each.

  • Four MIVC of femoral quadriceps muscle (4 seconds each), associated with low frequency electrical stimulation (MIVC-LF). 75Hz frequency, 400 μs pulse width; time on = 4s; time off 10s.

  • Four MVIC of the femoral quadriceps muscle (4 seconds each), associated with medium frequency electrical stimulation (MVIC-MF), with the following characteristics: 75 burst 2500Hz/s; time on=4s; time off=10 seconds.

Current intensity modulation took into account the maximum tolerance of the patient without any pain. Cathode electrodes were placed on the vastus medialis muscle and the anode in the rectus femoris muscle (Figure 2).

Figure 2
Placement of electrodes for neuromuscular electrical stimulation. Cathode in the vastus medialis muscle and anode in the rectus femoris muscle.

The time between different conditions of isometric contractions was 90 seconds for adequate replacement of ATP1313. Fleck S. Interval training: physiological basis. J Strength Cond Res. 1983;5(5):40-57. and the time between isometric contractions in same conditions was 10 seconds. Randomized contractions sequence was applied with sealed envelopes and the volunteers were not aware by which electrical stimulation was being used. The examiner knew about the study because it required practice and knowledge with electrical stimulator.

RESULTS

The measures of the variables of interest (MVIC) were obtained from the same person in three evaluations. This fact highlights the association between data collected in different frequencies. In view of this, to study the behavior of the MVIC variable in each situation considered, we apply analysis of variance (ANOVA).

It was found that there is difference between three measurements performed (Table 1). Statistically significant difference was found when compared MVIC-MF measures with MVIC-LF (p=0.02) measures. When compare MVIC-MF with MVIC, the values for the average frequency was higher (p=0.03), resulting statistically significant difference. Finally, comparing MVIC-LF with MVIC, there were no statistically significant difference (p=0.52).

DISCUSSION

The use of NMES to gain muscle strength or performance in high-level athletes has been employed with frequencies by physiotherapists who work in sports. It was found that the medium frequency current is more comfortable and is able to generate a peak torque significantly higher than the low-frequency current.

Our findings may be based on the explanation that the higher the frequency, the lower the impedance to current flow and therefore more comfortable perception thereof. Thus, a larger current magnitude is permitted using the medium frequency current.

Another hypothesis for the increasing of peak torque with the medium frequency current is that current has been generated by a clinical device, as portable devices (battery) are not displayed when the goal is the strength of recovery femoral quadriceps muscle in patients undergoing reconstruction of the anterior cruciate ligament.1414. Snyder-Mackler L, De Luca PF, Williams PR, Eastlack ME, Bartolozzi AR. Reflex inhibition of the quadriceps femoris muscle after injury or reconstruction of the anterior cruciate ligament. J Bone Joint Surg Am. 1994;76(4):555-60.

Difficulties were found when comparing the results of this study to another works in literature because of differences in isokinetic evaluation protocols. In this study we choose the angle 60° of knee flexion, because in this angulation the isometric muscle contraction reaches its maximum peak torque.

The sequence of NMES application was randomized to minimize the bias of muscle fatigue when generated by the medium current, low frequency or MVIC.

The electrodes positioning was based on the direction of quadriceps muscle fibers, with the focus on the rectus femoral and vastus medialis oblique. The placement of the electrodes can interfere with the extensor torque of the quadriceps muscle and that this type of placement is effective in producing a significant peak torque.1515. Hartssell HD, Kramer JF. A comparison of the effects of electrode placement muscle tension, and isometric torque of the knee extensors. J Orthop Sports Phys Ther. 1992;15(4):168-74. Moreover, this parameter must be considered when using electrical stimulation or during interpretation studies.

The fact of the medium frequency current generate a peak torque average greater than the low frequency does not necessarily imply its effectiveness when introduced into a strengthening muscular program. In other study evaluated the isokinetic strength of the quadriceps muscle in healthy subjects after 3 weeks of training with medium and low-frequency current, finding a significant increase after training in both groups, but there was no difference between them.1616. Bircan C, Senocak O, Peker O, Kaya A, Tamci SA, Gulbahar S, et al. Efficacy of two forms of electrical stimulation in increasing quadriceps strength: a randomized controlled trial. Clin Rehabil. 2002;16(2):194-9. However, medium frequency current can lead to increasing muscle fatigue when compared with the low frequency current.1717. Laufer Y. Ries JD, Leininger PM, Alon G. Quadriceps femoris muscle torque and fatigue generated by neuromuscular electrical stimulation with three different waveforms. Phys Ther. 2001;81(7):1307-16.

This study was conducted in healthy and young adults. Thus, we cannot extrapolate our results to individuals who have a muscular deficit due to some involvement of the lower limb, as well as in patients undergoing knee surgery. A systematic review1818. Bax L, Staes F, Verhagen A. Does Neuromuscular Electrical Stimulation Strenghten the Quadriceps Femoris? A systematic review of randomized controlled trials. Sports Med. 2005;35(3):191-212. concluded that in healthy volunteers, exercises can be equally or more effective than electrical stimulation and the realization of an associated contraction appears to be relevant to the effectiveness of neuromuscular electrical stimulation. However, reports that more research is needed to quantify and establish the optimal parameters of electric current.

Authors investigated muscle biopsy before and immediately after medium frequency electrical stimulation (2500Hz, 50Hz to 80% of maximum isometric torque).11. Sinacore DR, Delitto A, King DS, Rose SJ. Type II fiber activation with electrical stimulation: a preliminary report. Phys Ther. 1990;70(7):416-22. A decrease in glycogen of type II fibers was found, which according to the authors support the use of medium frequency electrical stimulation as a viable technique to be used in patients with impairment of type II fibers.

In the other study, author reports that electrical stimulation of 2500Hz was applied to the triceps and biceps, with a 10-second contraction time and standby time of 50 seconds.1212. Rich NC. Strength training via high frequency electrical stimulation. J Sports Med Phys Fitness. 1992;32(1):19-25. This study evaluated strength, endurance, perimetry and fat percentage. These findings suggest that electrical stimulation alone is not an appropriate form of strength training in healthy subjects. In addition, to gain strength voluntary exercise should be worked with.

In other study, authors evaluated that electrical stimulation (2500Hz) is able to increase strength compared to an isometric exercise protocol.1919. Laughman RK, Youdas JW, Garrett TR, Chao EY. Strength changes in the normal quadriceps femoris muscle as a result of electrical stimulation. Phys Ther. 1983;63(4):494-9. After 25 sessions, there was no significant difference in increasing strength in the group of isometric exercises and the group stimulated electrically.

Although the present study did not have done any training to gain muscle strength, it may be suggested that the electrical stimulation with medium frequency promotes higher muscle fiber recruitment to gain muscle strength, because of the significant peak torque measurements increase. Although the present study did not have done any training to gain muscle strength, it may be suggested that the electrical stimulation with medium frequency promotes higher muscle fiber recruitment to gain muscle strength, because of the significant peak torque measurements increase.

CONCLUSION

Individuals when submitted to MVIC-MF showed higher peak torque than when subjected to MVIC-BF and MVIC alone. Thus it can be concluded that maximal voluntary isometric contraction associated with medium frequency NMES was more effectiveness than other situations of MVIC.

  • Work conducted at the Cohen Orthopedic Institute, Rehabilitation and Sports Medicine - São Paulo - Brazil.

REFERENCES

  • 1
    Sinacore DR, Delitto A, King DS, Rose SJ. Type II fiber activation with electrical stimulation: a preliminary report. Phys Ther. 1990;70(7):416-22.
  • 2
    Kots JM, Hvilon VA. The training of muscular power by method of electrical stimulation. 1st ed. Moscú: State Central Institute of Physical Culture, 1971.
  • 3
    Maffiuletti NA, Zory R, Miotti D, Pellegrino MA, Jubeau M, Bottinelli R. Neuromuscular adaptations to electrostimulation resistance training. Am J Phys Med Rehabil. 2006;85(2):167-75.
  • 4
    Robinson AJ, Snder-Mackler L. Clinical electrophysiology: electrotherapy and electrophysiologic testing. 2 ed. Baltimore: Williams & Wilkins Press, 1995.
  • 5
    Brasileiro JS, Villar AFS. Comparação dos torques gerados por estimulação elétrica a contração muscular voluntária no músculo. Rev Bras Fisio. 2000;4(2):75-81.
  • 6
    Robinson AJ, Snyder-Mackler L. Clinical Electrophysiology: Electrotherapy and Electrophysiologic Testing. 2ᵃ ed. Baltimore: Williams & Wilkins Press, 1995.
  • 7
    Ward AR, Robertson VJ, Makowski RJ. Optimal frequencies for electric stimulation using medium-frequency alternating current. Arch Phys Med Rehabil. 2002;83(7):1024-7.
  • 8
    Lyons GM, Leane GE, Clarke-Moloney M, O'Brien JV, Grace PA. An investigation of the effect of electrode size and electrode location on comfort during stimulation of the gastrocnemius muscles. Med Eng Phys. 2004;26(10):873-8.
  • 9
    Delitto A, Rose SJ. Comparative confort of three waveforms used in electrically eliciting quadriceps femoris muscle contractions. Phys Ther. 1986;66(11):1704-7.
  • 10
    Kots YM. Electrostimulation. (Canadian-Soviet exchange symposium on electrostimulation of skeletal muscles, Concordia University, Montreal, Quebec, Canada; December 6–15,1977). Quoted in: Kramer J, Mendryk SW. Electrical stimulation as a strength improvement technique. J Orthop Sports Phys Ther. 1982; 4:91–98.
  • 11
    Ward AR, Shkuratova N. Russian electrical stimulation: the early experiments. Phys Ther. 2002;82(10):1019-30.
  • 12
    Rich NC. Strength training via high frequency electrical stimulation. J Sports Med Phys Fitness. 1992;32(1):19-25.
  • 13
    Fleck S. Interval training: physiological basis. J Strength Cond Res. 1983;5(5):40-57.
  • 14
    Snyder-Mackler L, De Luca PF, Williams PR, Eastlack ME, Bartolozzi AR. Reflex inhibition of the quadriceps femoris muscle after injury or reconstruction of the anterior cruciate ligament. J Bone Joint Surg Am. 1994;76(4):555-60.
  • 15
    Hartssell HD, Kramer JF. A comparison of the effects of electrode placement muscle tension, and isometric torque of the knee extensors. J Orthop Sports Phys Ther. 1992;15(4):168-74.
  • 16
    Bircan C, Senocak O, Peker O, Kaya A, Tamci SA, Gulbahar S, et al. Efficacy of two forms of electrical stimulation in increasing quadriceps strength: a randomized controlled trial. Clin Rehabil. 2002;16(2):194-9.
  • 17
    Laufer Y. Ries JD, Leininger PM, Alon G. Quadriceps femoris muscle torque and fatigue generated by neuromuscular electrical stimulation with three different waveforms. Phys Ther. 2001;81(7):1307-16.
  • 18
    Bax L, Staes F, Verhagen A. Does Neuromuscular Electrical Stimulation Strenghten the Quadriceps Femoris? A systematic review of randomized controlled trials. Sports Med. 2005;35(3):191-212.
  • 19
    Laughman RK, Youdas JW, Garrett TR, Chao EY. Strength changes in the normal quadriceps femoris muscle as a result of electrical stimulation. Phys Ther. 1983;63(4):494-9.