Flow injection spectrofluorimetric determination of cystine and cysteine

Abstracts

A relatively simple and sensitive procedure with spectrofluorimetric detection was developed for the determination of cystine and cysteine by flow injection system with sequential determination. This method is based on the reduction of Tl(III) with cysteine in acidic media, producing a fluorescence reagent, TlCl3(2-) (λex = 227 nm, λem = 419 nm). Before injection, the sample solution was divided into two streams. The first stream was treated with Cd reduction column and then joined with the carrier to react with Tl(III) at pH 5.0 and then passed through a 100 cm reaction coil to the flow cell of the spectrofluorimeter, where the fluorescence intensity was measured (λex = 227 nm, λem = 419 nm). This signal is related to cystine and cysteine concentrations. The second stream of sample solution was injected directly into the carrier stream to react with the reagents and then passed through the reaction coil and detector for measuring the fluorescence intensity. The signal in this step is related only to cysteine. Thus, the cystine content was determined directly from difference of the two signals. Cystine and cysteine can be determined in the range of 0.10 to 5.50 µmol L-1 and 0.20 to 8.0 µmol L-1, respectively, at a rate of 20 samples per hour. The limit of detection (3s/k) was 0.10 µmol L-1 for both analytes. The relative standard deviations for ten replicates determination of 4.0 and 3.5 µmol L-1 cystine or cysteine were 1.1% and 1.8%, respectively. The influence of potential interfering substances was studied. The proposed method was successfully applied to the sequential determination of both analytes in pharmaceutical samples.

cysteine; cystine; sequential determination; spectrofluorimetry


Um procedimento relativamente simples e sensível com detecção espectrofluorimétrica foi desenvolvido para a determinação de cistina e cisteína por sistema de injeção em fluxo com determinação seqüencial. Esse método é fundamentado na redução de Tl(III) com cisteína em meio ácido, produzindo o reagente fluorescente TlCl3(2-) (λex = 227 nm, λem = 419 nm). Antes da injeção, a solução da amostra foi dividida em dois fluxos. O primeiro fluxo foi tratado com coluna de redução de Cd e então refluxado com a solução do carregador para reagir em pH 5,0 com Tl(III), passado através de uma cela de reação de 100 cm e posteriormente para a cela fluxo do espectrofluorímetro, onde a intensidade de fluorescência foi medida (λex = 227 nm, λem = 419 nm). Esse sinal está relacionado às concentrações de cistina e cisteína. O segundo fluxo da solução de amostra foi injetado diretamente no fluxo carregador para reagir e, então, pela cela de reação e detector para medida da intensidade de fluorescência. O sinal nessa etapa é relacionado apenas à cisteína. Assim, o conteúdo de cistina foi determinado diretamente da diferença entre os dois sinais. Cistina e cisteína podem ser determinadas no intervalo de 0,10 a 5,50 µmol L-1 e 0,20 a 8,0 µmol L-1, respectivamente, em uma razão de 20 amostras por hora. O limite de detecção (3s/k) foi 0,10 µmol L-1 para os dois analitos. Os desvios padrões relativos para a determinação de dez replicatas de 4,0 e 3,5 µmol L-1 de cistina ou cisteína foram 1,1% e 1,8%, respectivamente. A influência de substâncias interferentes foi estudada. O método proposto foi aplicado com sucesso na determinação seqüencial de ambos analitos em amostras farmacêuticas.


ARTICLE

Flow injection spectrofluorimetric determination of cystine and cysteine

Ali A. Ensafi* * e-mail: ensafi@cc.iut.ac.ir ; B. Rezaei; S. Nouroozi

Department of Chemistry, Isfahan University of Technology, Isfahan 84156, Iran

ABSTRACT

A relatively simple and sensitive procedure with spectrofluorimetric detection was developed for the determination of cystine and cysteine by flow injection system with sequential determination. This method is based on the reduction of Tl(III) with cysteine in acidic media, producing a fluorescence reagent, TlCl32-ex = 227 nm, λem = 419 nm). Before injection, the sample solution was divided into two streams. The first stream was treated with Cd reduction column and then joined with the carrier to react with Tl(III) at pH 5.0 and then passed through a 100 cm reaction coil to the flow cell of the spectrofluorimeter, where the fluorescence intensity was measured (λex = 227 nm, λem = 419 nm). This signal is related to cystine and cysteine concentrations. The second stream of sample solution was injected directly into the carrier stream to react with the reagents and then passed through the reaction coil and detector for measuring the fluorescence intensity. The signal in this step is related only to cysteine. Thus, the cystine content was determined directly from difference of the two signals. Cystine and cysteine can be determined in the range of 0.10 to 5.50 µmol L-1 and 0.20 to 8.0 µmol L-1, respectively, at a rate of 20 samples per hour. The limit of detection (3s/k) was 0.10 µmol L-1 for both analytes. The relative standard deviations for ten replicates determination of 4.0 and 3.5 µmol L-1 cystine or cysteine were 1.1% and 1.8%, respectively. The influence of potential interfering substances was studied. The proposed method was successfully applied to the sequential determination of both analytes in pharmaceutical samples.

Keywords: cysteine, cystine, sequential determination, spectrofluorimetry

RESUMO

Um procedimento relativamente simples e sensível com detecção espectrofluorimétrica foi desenvolvido para a determinação de cistina e cisteína por sistema de injeção em fluxo com determinação seqüencial. Esse método é fundamentado na redução de Tl(III) com cisteína em meio ácido, produzindo o reagente fluorescente TlCl32-ex = 227 nm, λem = 419 nm). Antes da injeção, a solução da amostra foi dividida em dois fluxos. O primeiro fluxo foi tratado com coluna de redução de Cd e então refluxado com a solução do carregador para reagir em pH 5,0 com Tl(III), passado através de uma cela de reação de 100 cm e posteriormente para a cela fluxo do espectrofluorímetro, onde a intensidade de fluorescência foi medida (λex = 227 nm, λem = 419 nm). Esse sinal está relacionado às concentrações de cistina e cisteína. O segundo fluxo da solução de amostra foi injetado diretamente no fluxo carregador para reagir e, então, pela cela de reação e detector para medida da intensidade de fluorescência. O sinal nessa etapa é relacionado apenas à cisteína. Assim, o conteúdo de cistina foi determinado diretamente da diferença entre os dois sinais. Cistina e cisteína podem ser determinadas no intervalo de 0,10 a 5,50 µmol L-1 e 0,20 a 8,0 µmol L-1, respectivamente, em uma razão de 20 amostras por hora. O limite de detecção (3s/k) foi 0,10 µmol L-1 para os dois analitos. Os desvios padrões relativos para a determinação de dez replicatas de 4,0 e 3,5 µmol L-1 de cistina ou cisteína foram 1,1% e 1,8%, respectivamente. A influência de substâncias interferentes foi estudada. O método proposto foi aplicado com sucesso na determinação seqüencial de ambos analitos em amostras farmacêuticas.

Introduction

L-cystine and L-cysteine are two nonessential amino acids with significant biological importance. L-cysteine participates in a variety of important cellular functions, among others protein synthesis, detoxification and metabolism. Disorders of cysteine metabolism include cystinosis, an antosomal recessive disease produced by a defect in lysosomal transport and cystinuria, a common heritable disorder of amino acids transport. Altered levels of cysteine have been implicated in hyperhomocysteinemia and in a number of pathological conditions, including Alzheimer's and Parkinson's disease. The oxidized derivative of L-cysteine (L-cystine) has additional metabolic roles such as stabilization of the tertiary structure of proteins (via disulfide bonds). Thereby, cysteine and cystine are widely used in many pharmaceutical products and as supplement in some foods.1-3 As cysteine is readily oxidized to cystine in air, and cystine is hydrolyzed to cysteine in water, they are often presented together (in hair, liver, horn, etc.). So, the determination of cysteine and cystine in biological fluids, clinical investigations and pharmaceutical preparations are of great importance. Several different methods have been developed for the determination of cysteine and/or cystine.4-6 However, the analysis of mixtures of cysteine and cystine has rarely been performed by non-chromatographic techniques.1,3,7,8 Perez-Ruiz et al.1 introduced a flow injection chemiluminescence method for just L-cysteine based on the reaction of L-cysteine with tris(2,2'-bipyridyl) ruthenium(II) and peroxydisulphate under UV irradiation to produce chemiluminescence, with limit of determination of 0.5 µmol L-1. The proposed method suffers from many amino acids interference. Pérez-Ruiz et al.3 introduced a flow injection spectrophotometric method based on the inhibitory effect of cysteine/cystine on the oxidation of thiamine by mercury(II). The linear calibration range was 10-100 µmol L-1, and the method suffers from many interferences. Chrastil7 used a flow injection spectrophotometric method based on the formation of a colored complex with osmic acid for determination of cysteine and/or cystine with low limit of detection of 10 µmol L-1. Tütem and Apak8 introduced a conventional spectrophotometric method for determination of cysteine/cystine with copper(II)-neocuproin system. The method has low precision. The other reported papers introduced a relatively sensitive spectrofluorimetric method for determination of only L-cysteine,6,9 with limit of detection of 0.1 µmol L-1, and 0.04 µmol L-1, respectively. No any of the reported method is able to measure both of L-cysteine and cystine together.

A spectrophotometric enzymatic determination of L-cysteine has been described by following the decrease in the absorbance of NADH through two-enzymatic reactions; L-cysteine desulfhydrase from streptococcus anginosus bacterial cells to give pyruvate and lactate dehydogenase in the presence of NADH to convert pyruvate into lactate.10 Enzymatic colorimetric method for determining L-cysteine has been also described based on the use of recombinant enzymes methionine α,γ-lyase and adenosylhomocysteine hydrolase cloned from pseudomonas putida and trichomonas vaginalis, respectively.11 However, these methods involved two enzymatic reaction steps and suffer from severe interferences from methionine and homocysteine which need prior removal.10,11 Therefore, the development of a simple and selective method without tedious separation for detecting cysteine and cystine is obviously desirable.

Thallium(I) can be determined by measuring its fluorescence as TlCl3, whereas Tl(III) does not show any fluorescent characteristics. Tl(III)/Tl(I) system has been used as an analytical tools with flow injection analysis (FIA) manifold for speciation of Tl(III) and Tl(I)12 and also for determination of ascorbic acid.13 In this paper, a new, fast and relatively simple flow injection spectrofluorimetric method is proposed for sequential determination of cysteine and cystine based on reduction of cystine to cysteine and then reaction of cysteine with Tl(III) in chloride media to produce a fluorescence agent, TlCl3, by use of a suitable manifold design.

Experimental

Reagents

All solutions were prepared using reagent grade chemicals and doubly distilled water was used throughout.

Cysteine and cystine stock solutions (0.0010 mol L-1) were prepared by dissolving 0.0175 g of L-cysteine hydrochloride (Merck) and 0.0240 g of L-cystine (Merck) in doubly distilled water in 100 mL volumetric flasks, and kept in refrigerator. Those solutions were stable up to 7 days. Working standard solutions were freshly prepared by diluting the stock solutions with water.

Tl(III) stock solution (0.010 µmol L-1) was prepared by dissolving 0.4444 g of Tl(NO3)3.3H2O (Merck) in water and diluting to 100 mL.

Potassium chloride standard solution (1.0 mol L-1) was prepared by dissolving 7.456 g of KCl (Merck) in water and diluting to 100 mL.

Buffer solutions (pH 2 to 10) were prepared from universal buffer systems (phosphoric, acetic and boric acid) from 0.04 mol L-1 solution of each acid and 0.20 mol L-1 NaOH or 0.50 mol L-1 HCl solution using a pH-meter.

Apparatus

A diagram of the flow system employed is shown in Figure 1. The fluorescence intensity was measured on a Shimadzu Model RF-535 spectrofluorimeter equipped with a flow through cell (20 µL volume and 1.0 cm optical path length) equipped with a Shimadzu signal processing (Model C-R4AX Chromatopac). A 12-channels peristaltic pump (Desaga, Model PLG) with silicon rubber tubes (1.0 mm i.d.) was used. PTFE mixing joints and PTFE tubing (1.0 mm i.d.) was used for the connections and for the reaction coil. A cadmium reduction column, home-made, (1.5 mm i.d., 10 cm length, with 10 g of cadmium granule, about 0.2 mm particle size) was used for on-line reduction of cystine to cysteine. A thermostat water bath (Gallen-Kamp, BLG) was used and adjusted at temperature of 25.0 ± 0.1ºC. Sample solutions were injected using a six port rotary Supelco valve. A Jasco spectrofluorimeter, Model FP-750, was used for recording the spectra of the reaction solution (Tl(III) plus cysteine in acidic media). A pH-meter, Corning Model 140, equipped with a combined Metrohm electrode, Model 6.0202.100, was used for pH determination of the buffer solutions.


Recommended procedure

As shown in Figure 1, solutions containing carrier (H2O, (R1)), Tl(III) solution at pH 5.0 (R2) and sample solution (A), are pumped at 0.48 mL min-1 (for each channel) via a peristaltic pump. The standard solution containing 0.2-8.0 µmol L-1 cysteine and 0.1-5.5 µmol L-1 cystine was injected into the carrier stream via sample injection valve. The sample solution was delivered to the sample injection valve using a selection valve (SV). The first stream of the sample solution was injected, and after with a reaction mixture of Tl(III)-KCl solution at pH 5.0, it was passed to the sample flow cell of the spectrofluorimeter, where the fluorescence at 419 nm was measured as the initial cysteine content (first peak). The second stream of the sample solution was passed through Cd-reduction column and then injected into the carrier stream and after reaction with Tl(III)-KCl solution at pH 5.0 it was passed to the sample flow cell of the spectrofluorimeter, where the fluorescence was measured at 419 nm. This signal is related to the concentration of both cysteine and cystine (second peak). The concentration of cysteine and cystine were evaluated from the peak height measurements by using a calibration curve prepared from the results obtained after the analysis of standards.

Real sample analysis

For the NAC (N-acetyl cysteine) tablet, twenty tablets of each brand were grind and mixed well. Then 0.3550 g of the sample (equal to weight of one tablet) was dissolved in water in a 50 mL beaker. The mixture was filtered using a filter paper (Whatman No. 1). The filtrate was collected and diluted with water in a 50 mL volumetric flask. River water samples were passed through a filter paper (Whatman No. 1) and then used directly for analysis.

Results and Discussion

Cysteine is oxidized by Tl(III) in weak acidic media (pH 5) to produce Tl(I), which forms a stable anionic complex, TlCl3, in the presence of chloride ions. This complex has a violet fluorescence,14 whereas Tl(III) does not have any fluoresce. Figure 2 shows the fluorescence signal for cysteine recorded with time. Cystine does not have any effect on the Tl(III)-KCl mixture in acidic media, indicating that cystine could not reduce Tl(III) to Tl(I). On the other hand, cysteine reduces Tl(III) to Tl(I). Therefore, by on-line reducing of cystine to cysteine and designing a suitable FIA manifold it is possible to measure both of the two compounds.


Influence of chemical variables

The influence of pH and Tl(III) and KCl concentrations on the peak height (sensitivity) were investigated using a 200 µL 1.0 × 10-5 mol L-1 cysteine, a reaction coil length of 170 cm and a pump flow rate of 0.23 mL min-1 (for each channel) at 25 °C.

The influence of pH on the sensitivity was studied with 2.0 × 10-5 mol L-1 Tl(III) and 0.150 mol L-1 KCl solution, in the range 2.0-10.0. The results showed that the fluorescence signal increased slowly with increasing pH from 2.0 to 5.0 and then decreased sharply up to 7.0 and finally reached to very low intensity at pH 8.0. This is due to the fact that in higher acidity condition, H+ ions cause the decreasing of the cysteine oxidation rate to cystine by Tl(III), because H+ ion is one of the products of the oxidation reaction of cysteine to cystine. Further, the oxidation ability of Tl(III) decreases in alkaline solution. Therefore, a pH of 5.0 was selected as the optimum pH.

The influence of Tl(III) concentration on the fluorescence signal was studied in the range of 5.0 × 10-6 to 1.0 × 10-3 mol L-1 at pH 5.0 and 25 °C. The results showed that the sensitivity is increased with increasing Tl(III) concentration up to 1.0 × 10-5 mol L-1, because increasing the amount of reagents results in increasing formation of the fluorescence product, TlCl3. However, the sensitivity decreases for Tl(III) concentrations over 1.0 × 10-5 mol L-1, which can be due to self quenching. Thus, 1.0 × 10-5 mol L-1 Tl(III) was selected as optimum.

A chloride ion has an important role in the fluorescence signal of Tl(I), due to formation of TlCl3. The influence of chloride ion concentration on the sensitivity (pH 5.0) was studied in the range 0.05-0.35 mol L-1, using KCl solution. The results showed that addition of KCl up to 0.30 mol L-1 had a considerable effect on increasing the sensitivity and, beyond that, the signal leveled off. Therefore, 0.30 mol L-1 KCl was selected for the study.

Influence of manifold variables and reduction column

The effects of reaction coil length, pump flow rate and sample volume on the sensitivity were studied with the optimum reagents concentration at 25 °C and with 1.0 × 10-5 mol L-1 cysteine. The peak height depends on the residence time of the sample zone in the system (tube length and flow rate). Reaction coils of 50 to 250 cm were tested. A little increasing in sensitivity was observed when the length of the reaction coil was increased from 50 to 100 cm, above which the signal decreased slightly due to increasing the dispersion. Therefore, a 100 cm coil length was selected for the study.

The influence of flow rate on the sensitivity was studied in the range 0.17-0.58 mL min-1 for each channel. At low flow rates the fluorescence signal is higher, but the peaks are broadened and the analysis time is longer. Therefore, for a compromise between reaction time and sensitivity, a flow rate of 0.48 mL min-1 was selected for each channel.

The influence of sample loop volume on the sensitivity was studied under the optimum reagents and manifold conditions at 25 °C. The sample volume was varied between 145 to 300 µL. The results showed that the peak height increased with increasing injected volume up to 200 µL, above which they remained almost constant. Thus, a sample volume of 200 µL was chosen for further experiments.

The influence of temperature on the sensitivity was studied in the range of 5-50 °C. 25 °C was selected because it is more comfort.

The influence of Cd reduction column length on the measurement sensitivity of cystine was studied in the range 2.5-12.5 cm. The results showed that the sensitivity was increased by increasing the column length up to 10 cm and after that the signal leveled off. Thus, column length of 10 cm was selected for the reduction.

The reduction of cystine takes place in acidic media. So, the influence of HCl concentration on the efficiency of the reduction column was studied in the range 1.0 × 10-4-0.03 mol L-1 (Figure 3). The results showed that by increasing HCl concentration up to 0.02 mol L-1, the efficiency of the reduction was increased, whereas at higher concentration it remains almost constant. This is due to the fact that, the reduction of cystine takes place in acidic media. On the other hand, regeneration of column can be done in acidic solution successfully. For these reasons, 0.02 mol L-1 HCl was selected sample solution in this study.


Calibration graph and analysis of mixtures

Under the optimum reagents concentration and manifold variables, the calibration equation for the dependence of fluorescence signal on cysteine concentration in the range 0.2-8.0 µmol L-1 was: IFl = 5.9313 + 17.449Ccysteine (r2 = 0.9992), and for cystine in the range 0.1 to 5.5 µmol L-1 with calibration equation of IFl = 6.3312 + 28.019Ccystine (r2 = 0.9986), where C is concentration of cysteine or cystine in µmol L-1. A typical FI signals is shown in Figure 4.


The limit of detection, 3s/k (three times blank standard deviation/slope of analytical curve) was 0.1 µmol L-1 for both species. The limit of quantification was 0.12 µmol L-1 for the both species.

The relative standard deviations for ten replicates determination of 4.0 and 3.5 µmol L-1 cystine or cysteine were 1.1% and 1.8%, respectively. The sampling rate was about 20 samples per hour.

Influence of foreign substances

In order to validate of the possible analytical applications of the method, the effect of various compounds on the determination of 5.0 µmol L-1 cysteine were investigated. (Table 1). The potential interfering substances are added to the sample solution and the cysteine was determined as the proposed method. The tolerance limit was defined as the concentration of added species causing less than 3% relative error. The results showed that many substances especially amino acids did not interfere in the determination of cysteine and cystine.

Applications

To evaluate the applicability of the proposed method for real samples analysis, water and pharmaceutical samples were selected for determination of cysteine and/or cystine. The sample solutions were prepared as described before. The standard addition method was used by spiking of the samples with cysteine and/or cystine. Then the cysteine and/or cystine were determined using the recommended procedure. Tables 2 and 3 summarize the analytical results for the determination of cysteine and cystine in pharmaceutical formulations and river water by the proposed method. The good recoveries of the spiked samples confirm the ability of the proposed method to measure cysteine and/or cystine. The results show that the flow injection allows us to determine rapidly the same and different concentration levels of cysteine and cystine with good recoveries.

Conclusions

The new method developed for flow injection determination of cysteine and cystine is relatively simple, sensitive and rapid in real sample analyses. The sensitivity and reproducibility are coupled with a high speed and safety analysis with the proposed FIA technique. The resulting time analysis from the chosen conditions was equal to 3 min per analysis.

Acknowledgments

The authors acknowledge the University of Arkansas in Fayetteville (Foundation, CHE-0719097) and Isfahan University of Technology Research Council, and Center of Excellence in Sensors (IUT) for support of this work.

Received: February 21, 2008

Web Release Date: December 18, 2008

  • 1. Pérez-Ruiz, T.; Martínez-Lozano, C.; Tomás, V.; Martín, J.; Talanta 2002, 58, 987.
  • 2. Zhu, X.; Li, Y.; Zhu, G.; Anal. Sci. 2003, 19, 575.
  • 3. Pérez-Ruiz, T.; Martínez-Lozano, C.; Tomás, V.; Lambertos, G.; Talanta 1991, 38, 1235,
  • 4. Robinson, E. A.; Anal. Chim. Acta 1995, 21, 190.
  • 5. Vinas, P.; Lopez Garcia, I.; Martinez Gil, J. A.; J. Pharm. Biomed. Anal. 1993, 11, 15.
  • 6. Yahong, C.; Ruxiu, C.; Spectrochim. Acta, Part A 2003, 59, 3357.
  • 7. Chrastil, J.; Analyst 1989, 114, 1133.
  • 8. Tütem, E.; Apak, R.; Anal. Chim. Acta 1991, 255, 121.
  • 9. Wang, H.; Wang, W. S.; Zhang, H. S.; Spectrochimica Acta, Part A 2001, 57, 2403.
  • 10. Nishiya, Y.; Yoshida, Y.; Yoshimura, M.; Fukamachi, H.; Nakano, Y.; Biosci. Biotechnol. Biochem. 2005, 69, 2244.
  • 11. Han, Q.; Xu, M.; Tang, L.; Sun, X.; Zhang, N.; Tan, X.; Tan, X.; Tan, Y.; Hoffman, R. M.; Clin. Chem. 2004, 50, 290.
  • 12. Ensafi, A. A.; Rezaei, B.; Microchemical J. 1998, 60, 75.
  • 13. Ensafi, A. A.; Rezaei, B.; Anal. Lett. 1998, 31, 333.
  • 14. Kirkbright, G. F.; West, T. S.; Woodward, C.; Talanta 1965, 12, 517.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      27 Feb 2009
    • Date of issue
      2009

    History

    • Accepted
      18 Dec 2008
    • Received
      21 Feb 2008
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br