Acessibilidade / Reportar erro

Diagnosis of compliance of health care product processing in Primary Health Care1 1 Paper extracted from master's thesis "Diagnóstico de conformidade do processamento de produtos para saúde: análise de indicadores de qualidade na atenção básica", presented to Universidade Federal de São Carlos, São Carlos, SP, Brazil. Supported by Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES), Brazil.

ABSTRACT

Objective:

identify the compliance of health care product processing in Primary Health Care and assess possible differences in the compliance among the services characterized as Primary Health Care Service and Family Health Service.

Method:

quantitative, observational, descriptive and inferential study with the application of structure, process and outcome indicators of the health care product processing at ten services in an interior city of the State of São Paulo - Brazil.

Results:

for all indicators, the compliance indices were inferior to the ideal levels. No statistically significant difference was found in the indicators between the two types of services investigated. The health care product cleaning indicators obtained the lowest compliance index, while the indicator technical-operational resources for the preparation, conditioning, disinfection/sterilization, storage and distribution of health care products obtained the best index.

Conclusion:

the diagnosis of compliance of health care product processing at the services assessed indicates that the quality of the process is jeopardized, as no results close to ideal levels were obtained at any service. In addition, no statistically significant difference in these indicators was found between the two types of services studied.

Descriptors:
Primary Health Care; Sterilization; Indicators of Health Services; Nursing; Process Assessment (Health Care)

RESUMO

Objetivo:

identificar a conformidade do processamento de produtos para saúde na Atenção Primária à Saúde e avaliar possível diferença na conformidade entre as unidades caracterizadas como Unidade Básica de Saúde e Unidade Saúde da Família.

Método:

estudo quantitativo, observacional, descritivo e inferencial, com a aplicação de indicadores de estrutura, processo e resultado referentes ao processamento de produtos para a saúde em dez unidades, de um município do interior de São Paulo - Brasil.

Resultados:

todos os indicadores obtiveram índice de conformidade inferior ao ideal. Não houve diferença estatisticamente significante nos indicadores entre os dois tipos de unidades investigadas, sendo o indicador de limpeza de produtos para saúde o de menor índice de conformidade, enquanto indicador de recursos técnico-operacionais para preparo, acondicionamento, desinfecção/esterilização, guarda e distribuição de produtos para saúde, o de melhor índice.

Conclusão:

o diagnóstico de conformidade do processamento de produtos para saúde nas unidades avaliadas indica comprometimento da qualidade do processo, uma vez que não foram obtidos resultados próximos ao valor ideal em nenhuma unidade. Além disso, não houve diferença estatisticamente significante desses indicadores entre os dois tipos de unidades estudadas.

Descritores:
Atenção Primária à Saúde; Esterilização; Indicadores de Serviços; Enfermagem; Avaliação de Processos (Cuidados de Saúde)

RESUMEN

Objetivo:

identificar la conformidad del procesamiento de productos para la salud en la Atención Primaria a la Salud y evaluar una posible diferencia en la conformidad entre las unidades caracterizadas: Unidad Básica de Salud y Unidad Salud de la Familia.

Método:

estudio cuantitativo, observacional, descriptivo e inferencial; se aplicaron indicadores de estructura, proceso y resultado, referentes al procesamiento de productos para la salud en diez unidades de un municipio del interior de Sao Paulo, Brasil.

Resultados:

todos los indicadores obtuvieron índice de conformidad inferior a lo ideal. No hubo diferencia estadísticamente significativa de los indicadores entre los dos tipos de unidades investigadas, siendo el indicador de limpieza de productos para la salud el de menor índice de conformidad; por otro lado, como indicador de recursos técnico-operacionales para preparación, acondicionamiento, desinfección/esterilización, almacenamiento y distribución de productos para la salud, fue el de mejor índice.

Conclusión:

el diagnóstico de conformidad del procesamiento de productos para la salud en las unidades evaluadas indica que está comprometida la calidad del proceso, ya que no fueron obtenidos resultados próximos al valor ideal en ninguna unidad. Además de eso, no hubo diferencia estadísticamente significativa de esos indicadores entre los dos tipos de unidades estudiadas.

Descriptores:
Atención Primaria de Salud; Esterilización; Indicadores de Servicios; Evaluación de Proceso (Atención de Salud)

Introduction

Health care products manufactured from raw material that permits repeated cycles of cleaning, preparation, disinfection or sterilization can be processed until their functionality is lost. This process should be executed by qualified professionals and includes functionality and quality tests, guaranteeing that the transmission of microorganisms through this route is prevented11. Resolução n. 15, de 15 de março de 2012 (BR). Dispõe sobre requisitos de boas práticas para o processamento de produtos para saúde e dá outras providências. Diário Oficial da União. [Internet]. 19 mar 2012. [Acesso 6 ago 2016]. Disponível em: http://www.jusbrasil.com.br/diarios/35393450/dou-secao-1-19-03-2012-pg-43
http://www.jusbrasil.com.br/diarios/3539...
.

The practice of health care product processing is guided by regulatory policies that are based on risk management and public health safety22. Costa EAM, Costa EA. Risco e segurança sanitária de produtos em hospitais de Salvador, BA. Rev Saúde Pública. [Internet]. 2012 [Acesso 6 ago 2016];46(5):800-7. Disponível em: http://www.scielo.br/pdf/rsp/v46n5/06.pdf. doi: 10.1590/S0034-89102012000500006.
http://www.scielo.br/pdf/rsp/v46n5/06.pd...
-33. Passos IBD, Padoveze MC, Roseira CE, Figueiredo RM. Adaptation and validation of indicators concerning the sterilization process of supplies in Primary Health Care services. Rev. Latino-Am. Enfermagem. [Internet]. 2015 [Acesso 6 ago 2016];23(1):148-54. Disponível em: http://www.scielo.br/scielo.php?pid=S0104-11692015000100148&script=sci_arttext. doi: 10.1590/0104-1169.3518.2536.
http://www.scielo.br/scielo.php?pid=S010...
. These recommendations differ among countries though, being more or less restrictive, in view of technical-operational, economic, environmental, legal and political issues22. Costa EAM, Costa EA. Risco e segurança sanitária de produtos em hospitais de Salvador, BA. Rev Saúde Pública. [Internet]. 2012 [Acesso 6 ago 2016];46(5):800-7. Disponível em: http://www.scielo.br/pdf/rsp/v46n5/06.pdf. doi: 10.1590/S0034-89102012000500006.
http://www.scielo.br/pdf/rsp/v46n5/06.pd...
. In Brazil, this theme is regulated by RDC 15, a resolution by the Brazilian Health Surveillance Agency - ANVISA11. Resolução n. 15, de 15 de março de 2012 (BR). Dispõe sobre requisitos de boas práticas para o processamento de produtos para saúde e dá outras providências. Diário Oficial da União. [Internet]. 19 mar 2012. [Acesso 6 ago 2016]. Disponível em: http://www.jusbrasil.com.br/diarios/35393450/dou-secao-1-19-03-2012-pg-43
http://www.jusbrasil.com.br/diarios/3539...
.

Studies on the processing of health care products have identified that the quality of this procedure can interfere in patient safety, as it can take place ineffectively, whether in or outside the hospital44. Berlet LJ, Ascari RA, Silva OM, Trindade LL, Krauzer IM, Jacoby AM. Factors that influence the quality of the sterilisation process. Rev Enferm UFPE on line. 2014 [Acesso 6 ago 3016]; 8(7):1997-2003. Disponível em: http://www.revista.ufpe.br/revistaenfermagem/index.php/revista/article/view/5780/pdf_5484. doi: 10.5205/reuol.5963-51246-1-RV.0807201422.
http://www.revista.ufpe.br/revistaenferm...
-55. Buhr TL, Young AA, Johnson CA, Minter ZA, Wells CM. Decontamination of materials contaminated with Francisella philomiragia or MS2 bacteriophage using PES-Solid, a solid source of peracetic acid. J Appl Microbiol. [Internet]. 2014 [Acesso 6 ago 2016];117(2):397-404. Disponível em: http://www.ncbi.nlm.nih.gov/pubmed/24807242. doi: 10.1111/jam.12540.
http://www.ncbi.nlm.nih.gov/pubmed/24807...
, demonstrating infection outbreaks associated with the use of health care products, involving distinct microorganisms in different health care scenarios22. Costa EAM, Costa EA. Risco e segurança sanitária de produtos em hospitais de Salvador, BA. Rev Saúde Pública. [Internet]. 2012 [Acesso 6 ago 2016];46(5):800-7. Disponível em: http://www.scielo.br/pdf/rsp/v46n5/06.pdf. doi: 10.1590/S0034-89102012000500006.
http://www.scielo.br/pdf/rsp/v46n5/06.pd...
,66. Shuman EK, Chenoweth CE. Reuse of medical devices: implications for infection control. Infect Dis Clin N Am. [Internet]. 2012 [Acesso 6 ago 2016]; 26(1):165-72. Disponível em: http://www.sciencedirect.com/science/article/pii/S0891552011000857. doi: 10.1016/j.idc.2011.09.010.
http://www.sciencedirect.com/science/art...
.

Considering the expansion and diversification of extra-hospital care, including Primary Health Care (PHC), deeper knowledge is needed on health care product processing in this care context.

A study developed at rural health services in Nepal describes that 72% of the health professionals interviewed reported using unprocessed equipment and 50% did not have appropriate autoclaves at their services, illustrating the need to improve the infection control practices in PHC77. Timilshina N, Ansari MA, Dayal V. Risk of infection among primary health workers in the Western Development Region, Nepal: knowledge and compliance. J Infect Dev Ctries. [Internet]. 2011 [Acesso 6 go 2016];5(1):18-22. Disponível em: http://www.ncbi.nlm.nih.gov/pubmed/21330736. doi: 10.3855/jidc.782.
http://www.ncbi.nlm.nih.gov/pubmed/21330...
.

Actions undertaken in PHC and the availability of health care products for single use or processing vary according to the country's level of economic development, making it difficult to elaborate general recommendations for their processing.

Concerning the Brazilian standards, the recommendations are hardly specific for PHC, except for the classification of the Material Storage Centers (MSC), which in this case can be classified as class 1 MSC33. Passos IBD, Padoveze MC, Roseira CE, Figueiredo RM. Adaptation and validation of indicators concerning the sterilization process of supplies in Primary Health Care services. Rev. Latino-Am. Enfermagem. [Internet]. 2015 [Acesso 6 ago 2016];23(1):148-54. Disponível em: http://www.scielo.br/scielo.php?pid=S0104-11692015000100148&script=sci_arttext. doi: 10.1590/0104-1169.3518.2536.
http://www.scielo.br/scielo.php?pid=S010...
. The standardized analysis of data on the situation of health care product processing in Brazilian PHC is undoubtedly highly relevant to expand the knowledge on the theme, guide educational and supervisory policies and to serve as a reference for countries with similar development levels.

Another aspect to be taken into account in this study is a possible difference in the quality of the product processing, depending on the type of service analyzed. In Brazil, Primary Health Care Services working in the traditional model (UBS) and Services working in the Family Health Strategy (USF) constitute PHC88. Castro RCL, Knauth DR, Harzheim E, Hauser L, Duncan BB. Avaliação da qualidade da atenção primária pelos profissionais de saúde: comparação entre diferentes tipos de serviços. Cad Saúde Pública. [Internet]. 2012 [Acesso 6 ago 2016];28(9):1772-84. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0102-311X2012000900015. doi: 10.1590/S0102-311X2012000900015.
http://www.scielo.br/scielo.php?script=s...
.

Both have particular characteristics, related to human and structural resources. Studies that analyze these services99. Moura BLA, Cunha RC, Fonseca ACF, Aquino R, Medina MG, Vilasbôas ALQ et al. Atenção primária à saúde: estrutura das unidades como componente da atenção à saúde. Rev Bras Saude Mater Infantil. [Internet]. 2010 [Acesso 6 ago 2016];10(Suppl 1):S69-81. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1519-38292010000500007. doi: 10.1590/S1519-38292010000500007.
http://www.scielo.br/scielo.php?script=s...
-1010. Cesar M C, Campos GWS, Montebelo MIL, Sarmento G. Avaliação da atenção primária no município de Piracicaba, SP, Brasil. Saúde Debate. [Internet]. 2014 [Acesso 6 ago 2015];38(spe):296-306. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0103-11042014000600296&lng=en. doi: 10.5935/0103-1104.2014S022.
http://www.scielo.br/scielo.php?script=s...
suggest that, in terms of functionality, the USF are better assessed, despite problems in terms of infrastructure.

This study aimed to identify the compliance of the health care product processing in a sample of PHC services, using a specific validated tool33. Passos IBD, Padoveze MC, Roseira CE, Figueiredo RM. Adaptation and validation of indicators concerning the sterilization process of supplies in Primary Health Care services. Rev. Latino-Am. Enfermagem. [Internet]. 2015 [Acesso 6 ago 2016];23(1):148-54. Disponível em: http://www.scielo.br/scielo.php?pid=S0104-11692015000100148&script=sci_arttext. doi: 10.1590/0104-1169.3518.2536.
http://www.scielo.br/scielo.php?pid=S010...
, and to assess possible differences in the compliance rates observed between the services characterized as UBS and USF.

Method

Quantitative, observational, descriptive and inferential study. The objective was to identify the compliance index (CI) of health care product processing at ten PHC services in an interior city in the state of São Paulo.

The city under analysis has 221,950 inhabitants1111. IBGE (BR). Evolução da Divisão Territorial do Brasil (1872-2010): População por Municípios (1872 a 2010). Rio de Janeiro: IBGE; 2010. Disponível em: http://www.ibge.gov.br/home/geociencias/geografia/default_evolucao.shtm.
http://www.ibge.gov.br/home/geociencias/...
, offering 29 health services in PHC, with 15 USF and 14 UBS, distributed across five Regional Health Administrations (ARES), which coordinate the services within their area.

Convenience sampling was used (randomly using Microsoft Excel(r) 2010) to define the number of services, with a view to including two representatives (one USF and one UBS) from each of the five ARES in the city, resulting in a sample of ten health services (34.5% of all PHC services in the city).

One of the authors collected the data between January 22nd and July 23rd 2013, using a previously validated tool33. Passos IBD, Padoveze MC, Roseira CE, Figueiredo RM. Adaptation and validation of indicators concerning the sterilization process of supplies in Primary Health Care services. Rev. Latino-Am. Enfermagem. [Internet]. 2015 [Acesso 6 ago 2016];23(1):148-54. Disponível em: http://www.scielo.br/scielo.php?pid=S0104-11692015000100148&script=sci_arttext. doi: 10.1590/0104-1169.3518.2536.
http://www.scielo.br/scielo.php?pid=S010...
to assess the health care product processing in PHC. The tool33. Passos IBD, Padoveze MC, Roseira CE, Figueiredo RM. Adaptation and validation of indicators concerning the sterilization process of supplies in Primary Health Care services. Rev. Latino-Am. Enfermagem. [Internet]. 2015 [Acesso 6 ago 2016];23(1):148-54. Disponível em: http://www.scielo.br/scielo.php?pid=S0104-11692015000100148&script=sci_arttext. doi: 10.1590/0104-1169.3518.2536.
http://www.scielo.br/scielo.php?pid=S010...
assesses the structure, process and outcome and is organized by indicators in these three dimensions.

Departing from these three dimensions, the assessment was based on Donabedian's model, which is widely used in health, concerning the assessment of service quality based on health indicators in different contexts1212. Lachance J, Douville F, Dallaire C, Padilha KG, Gallani MC. The use of the Nursing Activities Score in clinical settings: an integrative review. Rev Esc Enferm USP. [Internet]. 2015 [Acesso 6 ago 2016]; 49(Esp):147-56. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0080-62342015000700147. doi: 10.1590/S0080-623420150000700021.
http://www.scielo.br/scielo.php?script=s...

13. Augusto MC, Nitsche MJT, Parada CMGL, Zanetti ML, Carvalhaes MABL. Evaluation of the Capillary Blood Glucose Self-monitoring Program. Rev. Latino-Am. Enfermagem [Internet]. 2014 [Acesso 6 ago 2016];22(5):801-9. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0104-11692014000500801&lng=en. doi: 10.1590/0104-1169.3282.2483.
http://www.scielo.br/scielo.php?script=s...
-1414. Gardner G, Gardner A, O'Connell J. Using the Donabedian framework to examine the quality and safety of nursing service innovation. J Clin Nurs. [Internet]. 2014 [Acesso 6 ago 2016];23(1-2):145-55. Disponível em: http://www.ncbi.nlm.nih.gov/pubmed/23834585. doi: 10.1111/jocn.12146.
http://www.ncbi.nlm.nih.gov/pubmed/23834...
.

As structure indicators, the assessment indicator of the technical-operational resources for the cleaning of health care products (L.1) was used, consisting of 22 items, and the assessment indicator of the technical-operational resources for the preparation, conditioning, disinfection/sterilization, storage and distribution of health care products (PE.5), including 21 items. To assess the process, the assessment indicator of the cleaning process of health care products was used (L.2), with 13 items, and the assessment indicator of the preparation, conditioning, disinfection/sterilization, storage and distribution process of health care products (PE.6), consisting of 36 items. For the outcomes dimension, the assessment indicator of the preserved packing of sterilized health care products (PE.9).

Each indicator consists of different components, whose information can be obtained by inspection, registration or interview with the responsible professional, depending on their relevance according to the indicator instructions.

The compliance index of each indicator is obtained by calculating the number of components in compliance/components assessed, divided by the total number of components, expressed in percentage, with 100% as the ideal score.

For PE. 9, the compliance index is obtained by the number of packages of sterilized health care products with preservation problems divided by the total number of packages, expressed in percentages, with 0% as the ideal compliance index.

Some components required the inspection of a sample of processed products. To calculate this sample, the software OpenEpi(r) was used, totaling 384 items. Therefore, 38 items should be observed at each of the ten health services studied. Due to the varying frequency of the products' use at the health services, this number could not be reached in all situations though, as presented in Table 1.

Table 1
Sample of health care products analyzed to assess the compliance index of the health care product processing quality, according to the processing steps and service type. São Carlos, SP, Brazil, 2013

The data were analyzed in the Statistical Package for the Social Sciences (SPSS(r)) version 22.0. The descriptive analysis was developed according to the type of service (UBS/UFS). The Mann-Whitney test was applied to compare the quality indicators in the structure, process and outcome indicators between the UBS and USF, considering a 5% significance level (p-value ≤ 0.05).

Approval for the research was obtained from the Ethics Committee for Research involving Human Beings at Universidade Federal de São Carlos, Opinion 112.528. To apply the indicators that demanded an interview, the person responsible for the processing at the MSC of each service was invited to participate in the study, receiving explanations on the research objective and the Informed Consent Form for signing. All professionals invited agreed to participate in the study.

Results

During the period observed, the processing at the services studied involved 46 professionals at the UBS (21.7% oral health aids and 78.3% auxiliary nurses or nursing technicians) and 18 professionals at the USF (33.3% oral health aids and 66.7% auxiliary nurses or nursing technicians).

The results concerning the CI of the indicators assessed are displayed in Table 2. In this table, the behavior of the structure indicators L.1 and PE.5, of the process indicators L.2 and PE.6 and of the outcome indicator PE.9 can be observed comparatively in the PHC contexts, based on the compliance index of each.

Table 2
Descriptive analysis and p-value of compliance indices according to the structure, process and outcome indicators of health care product processing and type of service (UBS and USF). São Carlos, SP, Brazil, 2013

Concerning the structure indicators L.1 and PE.5, it is added that most health services (60% of UBS and 80% of USF) did not have an exclusive room for expurgation or preparation, conditioning, disinfection, sterilization, storage and distribution of health care products (40% for UBS and 80% for USF). The technical barrier concept was applied at three UBS and two USF.

Regarding L.1, as for the material resources available at the expurgation room, it was observed that 40% of the UBS and 80% of the USF had an appropriate recipient for the disposal of piercing and cutting material; 40% of the UBS and 20% of the USF had deep sinks for product cleaning and 20% of the UBS had a soft brush for this purpose, which was not found at the USF. It is highlighted that, at the other places, steel sponges or toothbrushes were found for this processing step. No individual protection equipment (IPE) was found, such as masks, impermeable gowns and glasses, at 100% of the services studied.

During the assessment, at 100% of the services studied, no product cleaning standards and routines were available, not fully complying with L.1. As for the indicator PE.5, the standards for the preparation, conditioning, disinfection/sterilization, storage and distribution of health care products were present at 40% of the UBS but outdated.

The following were unavailable at the health services studied: forced drying devices, validation documents of steam autoclaves, reports of the quality of the water for the autoclaves (PE.5) or documentation on preventive maintenance of sterilization equipment and records on the efficacy of the sterilization process through chemical, physical or biological tests. At all services, sterilized health products are monitored using an indicator tape of the process. Nevertheless, there is no control on the sterilization by autoclaves through biological indicators (PE.6).

Sinks for hand washing at the room for the preparation, conditioning, disinfection/sterilization, storage and distribution of health care products, recommended by indicator PE.5, were found at 40% of the units.

Concerning indicator L.2, at all services, cleaning was done manually using enzyme detergents, without any standardization in the dilution, immersion time or change of product being used. The professionals reported that, in the previous year, no type of training or educative action had been offered on this step of the processing and that they did not participate in the definition of the substances to be purchased.

As verified by indicator PE.6, at the preparation, conditioning, disinfection/sterilization, storage and distribution rooms of health care products, no magnifying glasses were found to inspect the cleaning of the products, no registers on the sterilization/disinfection and no appropriate packing for the health care products.

The same indicator mentioned revealed that sodium hypochlorite was the product used to disinfect nebulization material. Its concentration was not standardized though, with values ranging between 0.001% and 1% at the different services studied. As for the preliminary drying of the items to be immersed for disinfection, at two USF, this procedure was done; one UBS and two USF fully immersed the products in the disinfecting solution.

Based on PE.6, the use of damp heat was verified as the sterilization method at 100% of the services, using gravitational autoclaves with less than 100 liters of capacity. It was observed that 60% of the UBS and no USF stored the items for sterilization in the autoclave vertically with space between the packages. Except for one UBS, at the other health services, the professionals awaited the drying and cooling of the products sterilized before removing them from the autoclaves.

According to the data obtained based on the application of PE.6, it was verified that no appropriate product storage places were found, in accordance with current recommendations, which can negatively affect the maintenance of their sterility.

As regards the preservations of packages of sterilized health care products (indicator PE.9), 462 packages wrapped in kraft paper were inspected. Although this packing is not recommended11. Resolução n. 15, de 15 de março de 2012 (BR). Dispõe sobre requisitos de boas práticas para o processamento de produtos para saúde e dá outras providências. Diário Oficial da União. [Internet]. 19 mar 2012. [Acesso 6 ago 2016]. Disponível em: http://www.jusbrasil.com.br/diarios/35393450/dou-secao-1-19-03-2012-pg-43
http://www.jusbrasil.com.br/diarios/3539...
, the conservation condition of the packing was observed. It was verified that 28 packages were stained, 17 ruptured, 11 dirty, six open, four with the tape that kept them closed detached, two stained and ruptured.

Based on the analysis of the p-value, no statistically significant difference was observed between the types of health services concerning any of the indicators used. Overall, the structure analysis (CI< 50%) was found closer to an appropriate CI when compared to the process indicator L.2, which obtained CI < 20%.

Discussion

The structure of the health care product processing services

The professionals engaged in the processing of health care products at the services studied had profiles similar to what was found in the literature, which appoints the nursing team (auxiliary nurses and nursing technicians) as the main responsible for this practice, followed by the dental aid in PHC44. Berlet LJ, Ascari RA, Silva OM, Trindade LL, Krauzer IM, Jacoby AM. Factors that influence the quality of the sterilisation process. Rev Enferm UFPE on line. 2014 [Acesso 6 ago 3016]; 8(7):1997-2003. Disponível em: http://www.revista.ufpe.br/revistaenfermagem/index.php/revista/article/view/5780/pdf_5484. doi: 10.5205/reuol.5963-51246-1-RV.0807201422.
http://www.revista.ufpe.br/revistaenferm...
.

As regards the physical structure, no statistically significant difference was found in the CI between UBS and USF (p-value equal to 0.3 and 0.9 for L.1 and PE.5, respectively), and the mean CI were inferior to 50% for both structure indicators applied. As opposed to the expected difference, considering studies that assessed the structure of health services in PHC and identified that the USF had larger problems related to the physical structure, as these services are frequently domestic adaptations99. Moura BLA, Cunha RC, Fonseca ACF, Aquino R, Medina MG, Vilasbôas ALQ et al. Atenção primária à saúde: estrutura das unidades como componente da atenção à saúde. Rev Bras Saude Mater Infantil. [Internet]. 2010 [Acesso 6 ago 2016];10(Suppl 1):S69-81. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S1519-38292010000500007. doi: 10.1590/S1519-38292010000500007.
http://www.scielo.br/scielo.php?script=s...
,1515. Oliveira MM, Pinto IC, Cruz VD, Coimbra VCC, Oliveira WEA, Lopes TS, et al. Análise da estrutura de uma unidade de saúde da família sob a perspectiva da ambiência. Rev APS [Internet]. 2014 [Acesso 6 ago 2016);17(4):423-8. Disponível em: https://aps.ufjf.emnuvens.com.br/aps/article/view/2046/836.
https://aps.ufjf.emnuvens.com.br/aps/art...
.

Structurally, according to the legislation in force33. Passos IBD, Padoveze MC, Roseira CE, Figueiredo RM. Adaptation and validation of indicators concerning the sterilization process of supplies in Primary Health Care services. Rev. Latino-Am. Enfermagem. [Internet]. 2015 [Acesso 6 ago 2016];23(1):148-54. Disponível em: http://www.scielo.br/scielo.php?pid=S0104-11692015000100148&script=sci_arttext. doi: 10.1590/0104-1169.3518.2536.
http://www.scielo.br/scielo.php?pid=S010...
, MSC in PHC can be classified as class I MSC, so that there is no need for physical separation between the clean and dirty areas of MSC, the use of the technical barrier should be established to impede the contact between the health care products from the different areas. Technical barrier are considered to be behavioral measures by health professionals to prevent cross-contamination between both areas11. Resolução n. 15, de 15 de março de 2012 (BR). Dispõe sobre requisitos de boas práticas para o processamento de produtos para saúde e dá outras providências. Diário Oficial da União. [Internet]. 19 mar 2012. [Acesso 6 ago 2016]. Disponível em: http://www.jusbrasil.com.br/diarios/35393450/dou-secao-1-19-03-2012-pg-43
http://www.jusbrasil.com.br/diarios/3539...
. The application of this concept was not identified at most of the services studied, whose flow of health care products did not follow a one-way sense.

As observed in this study, a research developed at four MSC in hospitals from Salvador - BA found no resources on site for hand washing22. Costa EAM, Costa EA. Risco e segurança sanitária de produtos em hospitais de Salvador, BA. Rev Saúde Pública. [Internet]. 2012 [Acesso 6 ago 2016];46(5):800-7. Disponível em: http://www.scielo.br/pdf/rsp/v46n5/06.pdf. doi: 10.1590/S0034-89102012000500006.
http://www.scielo.br/pdf/rsp/v46n5/06.pd...
, influencing the prevention of recontamination of already processed products1616. Association of Operating Room Nurses. Perioperative Standards and Recommended Practices. Denver: AORN Inc; 2012. by collecting them from the autoclave before the storage.

Compliance of health care product processing as a process

Departing from the manual execution of product cleaning at all services that participated in the research, the use of steel sponges was observed for this practice, differently from the standards in force in the country11. Resolução n. 15, de 15 de março de 2012 (BR). Dispõe sobre requisitos de boas práticas para o processamento de produtos para saúde e dá outras providências. Diário Oficial da União. [Internet]. 19 mar 2012. [Acesso 6 ago 2016]. Disponível em: http://www.jusbrasil.com.br/diarios/35393450/dou-secao-1-19-03-2012-pg-43
http://www.jusbrasil.com.br/diarios/3539...
. The indicator related to cleaning obtained the lowest CI (Table 2), which can deeply compromise the final quality of the process1717. Pinto MB, Vilas-Boas VA, Freitas MIP. Medical devices and cleaning validation: an integrative review. Rev SOBECC. [Internet]. 2013;18(1):64-72. Disponível em: http://sobecc.org.br/arquivos/artigos/2012/pdf/Artigos-Cientificos/Ano18_n1_jan_mar2013_validacao-do-processo-de-limpeza-de-artigos-odonto-medico-hospitalares.pdf.
http://sobecc.org.br/arquivos/artigos/20...
.

As regards the inputs used for cleaning, the inappropriate use of enzymatic detergents was observed. The range of products of this type that exist in the market and the different usage orientations can justify difficulties in training professionals in the area1717. Pinto MB, Vilas-Boas VA, Freitas MIP. Medical devices and cleaning validation: an integrative review. Rev SOBECC. [Internet]. 2013;18(1):64-72. Disponível em: http://sobecc.org.br/arquivos/artigos/2012/pdf/Artigos-Cientificos/Ano18_n1_jan_mar2013_validacao-do-processo-de-limpeza-de-artigos-odonto-medico-hospitalares.pdf.
http://sobecc.org.br/arquivos/artigos/20...
.

Concerning the solutions used for chemical disinfection, the use of sodium hypochlorite was observed, a substance summarily used to disinfect nebulization products in PHC in Brazil55. Buhr TL, Young AA, Johnson CA, Minter ZA, Wells CM. Decontamination of materials contaminated with Francisella philomiragia or MS2 bacteriophage using PES-Solid, a solid source of peracetic acid. J Appl Microbiol. [Internet]. 2014 [Acesso 6 ago 2016];117(2):397-404. Disponível em: http://www.ncbi.nlm.nih.gov/pubmed/24807242. doi: 10.1111/jam.12540.
http://www.ncbi.nlm.nih.gov/pubmed/24807...
,1818. Martins RJS, Bogéa RLN, Silva EL, Viana SCSL, Azevedo PR. Desinfecção de nebulizadores nas unidades básicas de saúde de São Luís, Maranhão. Rev Pesq Saúde. [Internet]. 2013[Acesso 6 ago 2016];14(2):101-4. Disponível em: http://www.periodicoseletronicos.ufma.br/index.php/revistahuufma/article/view/2302/385.
http://www.periodicoseletronicos.ufma.br...
.

Aspects of concentration and immersion time are linked to the quality of the disinfection and product integrity, such as the presence of toxic residues on the product that was disinfected, particularly inhalation material33. Passos IBD, Padoveze MC, Roseira CE, Figueiredo RM. Adaptation and validation of indicators concerning the sterilization process of supplies in Primary Health Care services. Rev. Latino-Am. Enfermagem. [Internet]. 2015 [Acesso 6 ago 2016];23(1):148-54. Disponível em: http://www.scielo.br/scielo.php?pid=S0104-11692015000100148&script=sci_arttext. doi: 10.1590/0104-1169.3518.2536.
http://www.scielo.br/scielo.php?pid=S010...
. The Brazilian Association of Surgical Center, Anesthetic Recovery and Material Sterilization Central Nurses recommends that it should correspond to 10.000ppm (1%), with a product immersion time of 30 minutes or 200 ppm (0.02%) for 60 minutes of exposure1818. Martins RJS, Bogéa RLN, Silva EL, Viana SCSL, Azevedo PR. Desinfecção de nebulizadores nas unidades básicas de saúde de São Luís, Maranhão. Rev Pesq Saúde. [Internet]. 2013[Acesso 6 ago 2016];14(2):101-4. Disponível em: http://www.periodicoseletronicos.ufma.br/index.php/revistahuufma/article/view/2302/385.
http://www.periodicoseletronicos.ufma.br...
.

Besides the concentration, another aspect related to the disinfection of health products that should be observed is its full immersion, so that the entire surface and all channels have contact with the solution and are then carefully rinsed, removing all toxins and irritating residues1616. Association of Operating Room Nurses. Perioperative Standards and Recommended Practices. Denver: AORN Inc; 2012..

It is highlighted that similar results were found in a study developed at a UBS in São Luís (MA), where the nebulizers were not subject to preliminary cleaning before the disinfection (89.90% of the services) and sodium hypochlorite for disinfection (78.50%)1818. Martins RJS, Bogéa RLN, Silva EL, Viana SCSL, Azevedo PR. Desinfecção de nebulizadores nas unidades básicas de saúde de São Luís, Maranhão. Rev Pesq Saúde. [Internet]. 2013[Acesso 6 ago 2016];14(2):101-4. Disponível em: http://www.periodicoseletronicos.ufma.br/index.php/revistahuufma/article/view/2302/385.
http://www.periodicoseletronicos.ufma.br...
.

According to the legislation in force, the gravitational autoclaves are permitted when their capacity is inferior to 100 l, the situation found in the study context. The same legislation is not complied with, due to the absence of reports to support the water quality and records of preventive maintenance of the autoclaves11. Resolução n. 15, de 15 de março de 2012 (BR). Dispõe sobre requisitos de boas práticas para o processamento de produtos para saúde e dá outras providências. Diário Oficial da União. [Internet]. 19 mar 2012. [Acesso 6 ago 2016]. Disponível em: http://www.jusbrasil.com.br/diarios/35393450/dou-secao-1-19-03-2012-pg-43
http://www.jusbrasil.com.br/diarios/3539...
.

What the inappropriate arrangement of the products inside the autoclaves is concerned and the lack of awaiting the cooling time, this practice compromises the circulation of the steam and lukewarm or hot packages should not be placed on surfaces with a temperature inferior to their own, as this can cause humidity in- and outside the packages, compromising the protection barrier properties of the packing1616. Association of Operating Room Nurses. Perioperative Standards and Recommended Practices. Denver: AORN Inc; 2012..

Nevertheless, an experimental study to identify the maintained sterility of health care products in case of presence of moisture due to steam showed that, when the conditioning and storage conditions were appropriate, the inside of the sterilized boxes was not contaminated after they were withdrawn while not cooled yet and stored for 30 days1919. Moriya GAA, Graziano KU. Sterility Maintenance Assessment of Moist/Wet Material After Steam Sterilization and 30-day Storage. Rev. Latino-Am. Enfermagem. [Internet]. 2010 [Acesso 6 ago 2016];18(4):786-91. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0104-11692010000400018. doi: 10.1590/S0104-11692010000400018.
http://www.scielo.br/scielo.php?script=s...
. Nevertheless, this cannot be extrapolated to other types of packaging nor to a lack of inappropriate manipulation, like without hand washing for example.

Biological indicators were not used as frequently as recommended by the assessment tools used and chemical indicators were the most used, similar to a study developed in cities in Goiás, in which the chemical indicator was the most used test in 83.8%2020. Tipple AFV, Pires FV, Guadagnin SVT, Melo D. O monitoramento de processos físicos de esterilização em hospitais do interior do estado de Goiás. Rev Esc Enferm USP. [Internet]. 2011 [Acesso 6 ago 2016] ;45(3):751-7. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0080-62342011000300029. doi: 10.1590/S0080-62342011000300029.
http://www.scielo.br/scielo.php?script=s...
. This scenario compromises the confidence that the autoclaves available for the processing of health care products are used for the sterilization at these places.

Preservation of packing of sterilized products

For the storage of the products processed, a consensus exists that it should at least guarantee the integrity of the packing, avoiding tears, dirt or wetting, which was not found in this study. More controlled situations such as air humidity, temperature and specific storage places (such as cupboards or plastic boxes) do not seem to interfere in the maintenance of their sterility2121. Serratine ACP, Gonçalves CS, Lucolli IC. Influência do armazenamento e da embalagem na manutenção da esterilidade do instrumental odontológico. Rev Eletrônica Enferm. 2009 [Acesso 6 ago 2016];11(1):158-64. Disponível em: https://www.fen.ufg.br/fen_revista/v11/n1/v11n1a20.htm.
https://www.fen.ufg.br/fen_revista/v11/n...
-2222. Bruna CQM, Graziano KU. Temperatura e umidade no armazenamento de materiais autoclavados: revisão integrativa. Rev Esc Enferm USP. 2012 [Acesso 3 jun 2016];46(5):1215-20. Disponível em: http://www.scielo.br/scielo.php?script=sci_arttext&pid=S0080-62342012000500025. doi: 10.1590/S0080-62342012000500025.
http://www.scielo.br/scielo.php?script=s...
.

Although inappropriate, the use of Kraft paper is still a reality in PHC44. Berlet LJ, Ascari RA, Silva OM, Trindade LL, Krauzer IM, Jacoby AM. Factors that influence the quality of the sterilisation process. Rev Enferm UFPE on line. 2014 [Acesso 6 ago 3016]; 8(7):1997-2003. Disponível em: http://www.revista.ufpe.br/revistaenfermagem/index.php/revista/article/view/5780/pdf_5484. doi: 10.5205/reuol.5963-51246-1-RV.0807201422.
http://www.revista.ufpe.br/revistaenferm...
, as verified in this study.

Although the study comes with limitations, as it does not permit the establishment of cause-and-effect relations among the findings or the generalization of results, it is extremely relevant, as it presents measurable and standardized data.

Conclusion

The diagnosis of compliance of health care product processing at the services assessed indicates the commitment of the process quality, as the CI obtained was not close to ideal levels at any service. In addition, there was no statistically significant difference in the quality indicators of structure, process and outcome between the UBS and the USF investigated, as opposed to the initial research hypothesis.

Cleaning indicators obtained worse CI, for structure as well as for process (L.1 < 40% L.2 CI 20%, respectively), which is undoubtedly a source of concern as it seriously compromises the subsequent steps.

This study joins important systemized information on the panorama of health care product processing in PHC, contributing to the expansion of knowledge on one of the pillars of healthcare-related infection control in this environment.

References

  • 1
    Paper extracted from master's thesis "Diagnóstico de conformidade do processamento de produtos para saúde: análise de indicadores de qualidade na atenção básica", presented to Universidade Federal de São Carlos, São Carlos, SP, Brazil. Supported by Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES), Brazil.

Publication Dates

  • Publication in this collection
    2016

History

  • Received
    13 Jan 2016
  • Accepted
    21 July 2016
Escola de Enfermagem de Ribeirão Preto / Universidade de São Paulo Av. Bandeirantes, 3900, 14040-902 Ribeirão Preto SP Brazil, Tel.: +55 (16) 3315-3451 / 3315-4407 - Ribeirão Preto - SP - Brazil
E-mail: rlae@eerp.usp.br