Test-Retest Reliability of Non-Invasive Cardiac Output Measurement during Exercise in Healthy Volunteers in Daily Clinical Routine

Abstract

Background:

Thoracic bioreactance (TB), a noninvasive method for the measurement of cardiac output (CO), shows good test-retest reliability in healthy adults examined under research and resting conditions.

Objective:

In this study, we evaluate the test-retest reliability of CO and cardiac power (CPO) output assessment during exercise assessed by TB in healthy adults under routine clinical conditions.

Methods:

25 test persons performed a symptom-limited graded cycling test in an outpatient office on two different days separated by one week. Cardiorespiratory (power output, VO2peak) and hemodynamic parameters (heart rate, stroke volume, CO, mean arterial pressure, CPO) were measured at rest and continuously under exercise using a spiroergometric system and bioreactance cardiograph (NICOM, Cheetah Medical).

Results:

After 8 participants were excluded due to measurement errors (outliers), there was no systematic bias in all parameters under all conditions (effect size: 0.2-0.6). We found that all noninvasively measured CO showed acceptable test-retest-reliability (intraclass correlation coefficient: 0.59-0.98; typical error: 0.3-1.8). Moreover, peak CPO showed better reliability (intraclass correlation coefficient: 0.80-0.85; effect size: 0.9-1.1) then the TB CO, thanks only to the superior reliability of MAP (intraclass correlation coefficient: 0.59-0.98; effect size: 0.3-1.8).

Conclusion:

Our findings preclude the clinical use of TB in healthy subject population when outliers are not identified.

Keywords:
Cardiac Output; Cardiography, Impedance/methods; Exercise; Exercise Test/methods; Echocardiography/methods; Reproducibility of Results; Adult

Resumo

Fundamento:

A biorreatância torácica (BT), um método não invasivo destinado à medição do débito cardíaco (DC), mostra boa confiabilidade teste-reteste em adultos saudáveis examinados em condições de pesquisa e repouso.

Objetivo:

No presente estudo, avaliamos a confiabilidade teste-reteste da avaliação do DC e trabalho cardíaco (TC) durante exercício, avaliado por BT em adultos saudáveis sob condições clínicas de rotina.

Métodos:

25 indivíduos realizaram teste ergométrico gradual sintoma-limitante em ambiente ambulatorial em dois dias diferentes, com intervalo de uma semana. Parâmetros cardiorrespiratórios (trabalho cardíaco, VO2máx) e hemodinâmicos (frequência cardíaca, volume sistólico, DC, pressão arterial média, TC) foram medidos em repouso e continuamente sob exercício utilizando sistema espiroergométrico e cardiógrafo de biorreatância (NICOM, Cheetah Medical).

Resultados:

Após 8 participantes terem sido excluídos devido a erros de medição (outliers), não houve viés sistemático em nenhum dos parâmetros em todas as condições (tamanho do efeito: 0,2-0,6). Observamos que todos os débitos cardíacos medidos de forma não invasiva apresentaram níveis aceitáveis de confiabilidade teste-reteste (coeficiente de correlação intraclasse: 0,59-0,98; erro típico: 0,3-1,8). Além disso, TC máximo apresentou melhor confiabilidade (coeficiente de correlação intraclasse: 0,80-0,85; tamanho do efeito: 0,9-1,1), seguido do DC pela BT, graças apenas à confiabilidade superior da PAM (coeficiente de correlação intraclasse: 0,59-0,98; tamanho do efeito: 0,3-1,8).

Conclusão:

Nossos achados impedem o uso clínico da BT em indivíduos saudáveis quando outliers não forem identificados.

Palavras-chave:
Débito Cardíaco; Cardiografia de Impedância/métodos; Exercício; Teste de Esforço/métodos; Ecocardiografia/métodos; Reprodutibilidade dos Testes; Adulto

Introduction

Cardiac output (CO) is an important physiological surrogate parameter, reflecting the hemodynamic demands of the organism. CO measuring has a wide application spectrum11 Jhanji S, Dawson J, Pearse RM. Cardiac output monitoring: basic science and clinical application. Anaesthesia. 2008;63(2):172-81. and can provide information on hemodynamic status in patients22 Myers J, Gujja P, Neelagaru S, Burkhoff D. Cardiac output and cardiopulmonary responses to exercise in heart failure: application of a new bio-reactance device. J Card Fail. 2007;13(8):629-36. as well as athletes.33 Klasnja AV, Jakovljevic DG, Barak OF, Popadic Gacesa JZ, Lukac DD, Grujic NG. Cardiac power output and its response to exercise in athletes and non-athletes. Clin Physiol Funct Imaging. 2013;33(3):201-5. In chronic heart failure, CO is decreased and patients suffer from exercise intolerance.44 Myers J, Froelicher VF. Hemodynamic determinants of exercise capacity in chronic heart failure. Ann Intern Med. 1991;115(5):377-86.,55 Sullivan MJ, Knight JD, Higginbotham MB, Cobb FR. Relation between central and peripheral hemodynamics during exercise in patients with chronic heart failure. Muscle blood flow is reduced with maintenance of arterial perfusion pressure. Circulation. 1989;80(4):769-81. In contrast, the athlete's heart shows structural and functional adaptations due to training66 Kovacs R, Baggish AL. Cardiovascular adaptation in athletes. Trends Cardiovasc Med. 2016;26(1):46-52. resulting in a higher CO.77 Rerych SK, Scholz PM, Sabiston DC, Jr., Jones RH. Effects of exercise training on left ventricular function in normal subjects: a longitudinal study by radionuclide angiography. Am J Cardiol. 1980;45(2):244-52. Interestingly altered cardiac structure and function do not predict exercise intolerance88 Franciosa JA, Park M, Levine TB. Lack of correlation between exercise capacity and indexes of resting left ventricular performance in heart failure. Am J Cardiol. 1981;47(1):33-9.,99 Wilson JR, Rayos G, Yeoh TK, Gothard P, Bak K. Dissociation between exertional symptoms and circulatory function in patients with heart failure. Circulation. 1995;92(1):47-53. or CO response33 Klasnja AV, Jakovljevic DG, Barak OF, Popadic Gacesa JZ, Lukac DD, Grujic NG. Cardiac power output and its response to exercise in athletes and non-athletes. Clin Physiol Funct Imaging. 2013;33(3):201-5. in both cases. Thus, cardiopulmonary exercise testing is necessary and peak oxygen consumption (VO2peak) is measured to determine exercise capacity.1010 Ponikowski P, Voors AA, Anker SD, Bueno H, Cleland JG, Coats AJ, et al. 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure: The Task Force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC). Developed with the special contribution of the Heart Failure Association (HFA) of the ESC. Eur J Heart Fail. 2016;18(8):891-975.,1111 Myers J. Applications of cardiopulmonary exercise testing in the management of cardiovascular and pulmonary disease. Int J Sports Med. 2005 Feb;26 (Suppl 1):S49-55. But, estimation of VO2 is influenced by several non-cardiac factors,44 Myers J, Froelicher VF. Hemodynamic determinants of exercise capacity in chronic heart failure. Ann Intern Med. 1991;115(5):377-86.,1212 Lang CC, Agostoni P, Mancini DM. Prognostic significance and measurement of exercise-derived hemodynamic variables in patients with heart failure. J Card Fail. 2007;13(8):672-9. and can, therefore, be misleading.99 Wilson JR, Rayos G, Yeoh TK, Gothard P, Bak K. Dissociation between exertional symptoms and circulatory function in patients with heart failure. Circulation. 1995;92(1):47-53.,1313 Becklake MR, Frank H, Dagenais GR, Ostiguy GL, Guzman CA. Influence of age and sex on exercise cardiac output. J Appl Physiol. 1965;20(5):938-47.,1414 Wilson JR, Rayos G, Yeoh TK, Gothard P. Dissociation between peak exercise oxygen consumption and hemodynamic dysfunction in potential heart transplant candidates. J Am Coll Cardiol. 1995;26(2):429-35. Furthermore, CO cannot be accurately predicted from cardiopulmonary exercise testing.44 Myers J, Froelicher VF. Hemodynamic determinants of exercise capacity in chronic heart failure. Ann Intern Med. 1991;115(5):377-86.,1515 Nicoletti I, Cicoira M, Zanolla L, Franceschini L, Brighetti G, Pilati M, et al. Skeletal muscle abnormalities in chronic heart failure patients: relation to exercise capacity and therapeutic implications. Congest Heart Fail. 2003;9(3):148-54.

However, to evaluate hemodynamic status, catheter-based measuring (i.e., Fick method, thermodilution method) is considered as the clinical standard.1616 Lund-Johansen P. The dye dilution method for measurement of cardiac output. Eur Heart J. 1990 Dec;11(Suppl I):6-12.,1717 Warburton DE, Haykowsky MJ, Quinney HA, Humen DP, Teo KK. Reliability and validity of measures of cardiac output during incremental to maximal aerobic exercise. Part II: Novel techniques and new advances. Sports Med. 1999;27(4):241-60. Since such invasive methods are associated with high risk, their applicability is restricted.1818 Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, et al. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med. 2003;348(1):5-14.,1919 Harvey S, Stevens K, Harrison D, Young D, Brampton W, McCabe C, et al. An evaluation of the clinical and cost-effectiveness of pulmonary artery catheters in patient management in intensive care: a systematic review and a randomised controlled trial. Health Technol Assess. 2006;10(29):iii-iv, ix-xi, 1-133. Therefore, noninvasive measuring methods (i.e., transoesophageal echocardiography, lithium dilution CO, pulse contour CO, partial CO2 rebreathing, thoracic electrical bioimpedance) were developed.1717 Warburton DE, Haykowsky MJ, Quinney HA, Humen DP, Teo KK. Reliability and validity of measures of cardiac output during incremental to maximal aerobic exercise. Part II: Novel techniques and new advances. Sports Med. 1999;27(4):241-60. Of the noninvasive measuring methods, especially the thoracic electrical bioimpedance was frequently used in clinical studies and evaluated for its reliability.2020 Gordon N, Abbiss CR, Maiorana AJ, Marstron KJ, Peiffer JJ. Intrarater reliability and agreement of the physioflow bioimpedance cardiography device during rest, moderate and high-intensive exercise. Kinesiology. 2018;50(1 Suppl 1):140-9. However, thoracic bioreactance (TB) is a further promising technology to noninvasively monitor CO.2121 Maurer MM, Burkhoff D, Maybaum S, Franco V, Vittorio TJ, Williams P, et al. A multicenter study of noninvasive cardiac output by bioreactance during symptom-limited exercise. J Card Fail. 2009;15(8):689-99. TB is based on the measurement of blood flow-related phase shifts of transthoracic electric signals to monitor noninvasively and continuously CO. Therefore theoretically, TB is superior to other methods2222 Keren H, Burkhoff D, Squara P. Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. Am J Physiol Heart Circ Physiol. 2007;293(1):H583-9.,2323 Jakovljevic DG, Moore S, Hallsworth K, Fattakhova G, Thoma C, Trenell MI. Comparison of cardiac output determined by bioimpedance and bioreactance methods at rest and during exercise. J Clin Monit Comput. 2012;26(2):63-8. and has been used in several clinical settings.2121 Maurer MM, Burkhoff D, Maybaum S, Franco V, Vittorio TJ, Williams P, et al. A multicenter study of noninvasive cardiac output by bioreactance during symptom-limited exercise. J Card Fail. 2009;15(8):689-99.,2323 Jakovljevic DG, Moore S, Hallsworth K, Fattakhova G, Thoma C, Trenell MI. Comparison of cardiac output determined by bioimpedance and bioreactance methods at rest and during exercise. J Clin Monit Comput. 2012;26(2):63-8.

24 Marik PE, Levitov A, Young A, Andrews L. The use of bioreactance and carotid Doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest. 2013;143(2):364-70.
-2525 Elliott A, Hull JH, Nunan D, Jakovljevic DG, Brodie D, Ansley L. Application of bioreactance for cardiac output assessment during exercise in healthy individuals. Eur J Appl Physiol. 2010;109(5):945-51. But, before TB can be adopted for clinical and performance decision making, test's quality criteria, as the test-retest reliability, must be fulfilled.

Jones et al.2626 Jones TW, Houghton D, Cassidy S, MacGowan GA, Trenell MI, Jakovljevic DG. Bioreactance is a reliable method for estimating cardiac output at rest and during exercise. Br J Anaesth. 2015;115(3):386-91. tested the test-retest reliability in a healthy study population. 22 healthy adults performed twice a symptom-limited exercise test. Standard cardiorespiratory data were measured via spiroergometry and the hemodynamic response was monitored via TB using the NICOM® system. The authors state that TB allows good test-retest reliability for hemodynamic measurement at rest as well as under submaximal and maximal exertion. This particular study was the first to confirm that TB might be a feasible test method. Noteworthy, the study was performed under tightly controlled research conditions. Overall, three visits were necessary to determine individual cardiorespiratory capability and to perform both exercise tests. Furthermore, to exclude confounders, certain inclusion criteria had to be fulfilled (e.g., non-smokers, empty stomach for > 2 h, no vigorous exercise 24 h before testing, no alcohol or caffeine consumption). Such scientific testing conditions are often difficult to guarantee in daily clinical routine. Thus, it remains unclear, if TB is an appropriate examination procedure not only in a research setting but also in daily clinical routine.

Contrary to CO based on heart rate (HR) and stroke volume (SV), cardiac power output (CPO) indicates the overall function of the heart.2727 Tan LB. Evaluation of cardiac dysfunction, cardiac reserve and inotropic response. Postgrad Med J. 1991;67(Suppl 1):S10-20. CPO is the product of the CO and mean arterial pressure (MAP) and therefore is a measure of cardiac pumping.2828 Cotter G, Williams SG, Vered Z, Tan LB. Role of cardiac power in heart failure. Curr Opin Cardiol. 2003;18(3):215-22. Peak cardiac power output (CPOpeak), the CPO achieved during maximal stress, is a major determinant of exercise intolerance and performance in cardiac patients and healthy persons, respectively.2929 Tan LB. Clinical and research implications of new concepts in the assessment of cardiac pumping performance in heart failure. Cardiovasc Res. 1987;21(8):615-22.,3030 Cooke GA, Marshall P, al-Timman JK, Wright DJ, Riley R, Hainsworth R, et al. Physiological cardiac reserve: development of a non-invasive method and first estimates in man. Heart. 1998;79(3):289-94.

Worth mentioning, CPO measuring can improve medical management31,32 and risk stratification3333 Grodin JL, Mullens W, Dupont M, Wu Y, Taylor DO, Starling RC, et al. Prognostic role of cardiac power index in ambulatory patients with advanced heart failure. Eur J Heart Fail. 2015;17(7):689-96.

34 Williams SG, Cooke GA, Wright DJ, Parsons WJ, Riley RL, Marshall P, et al. Peak exercise cardiac power output; a direct indicator of cardiac function strongly predictive of prognosis in chronic heart failure. Eur Heart J. 2001;22(16):1496-503.
-3535 Lang CC, Karlin P, Haythe J, Lim TK, Mancini DM. Peak cardiac power output, measured noninvasively, is a powerful predictor of outcome in chronic heart failure. Circ Heart Fail. 2009;2(1):33-8. in cardiac patients. In chronic heart failure, CPO is a powerful and independent predictor of survival outcome.3535 Lang CC, Karlin P, Haythe J, Lim TK, Mancini DM. Peak cardiac power output, measured noninvasively, is a powerful predictor of outcome in chronic heart failure. Circ Heart Fail. 2009;2(1):33-8. CPO also reflects cardiovascular adaptions and training status in athletes.66 Kovacs R, Baggish AL. Cardiovascular adaptation in athletes. Trends Cardiovasc Med. 2016;26(1):46-52. In fact, compared to non-athletes,3636 Bromley PD, Hodges LD, Brodie DA. Physiological range of peak cardiac power output in healthy adults. Clin Physiol Funct Imaging. 2006;26(4):240-6. CPO is higher in athletes.33 Klasnja AV, Jakovljevic DG, Barak OF, Popadic Gacesa JZ, Lukac DD, Grujic NG. Cardiac power output and its response to exercise in athletes and non-athletes. Clin Physiol Funct Imaging. 2013;33(3):201-5.,3737 Schlader ZJ, Mundel T, Barnes MJ, Hodges LD. Peak cardiac power output in healthy, trained men. Clin Physiol Funct Imaging. 2010;30(6):480-4. Thus, CPO might be an additive performance diagnostic parameter, which could help to guide training modalities.3737 Schlader ZJ, Mundel T, Barnes MJ, Hodges LD. Peak cardiac power output in healthy, trained men. Clin Physiol Funct Imaging. 2010;30(6):480-4.,3838 Marshall P, Al-Timman J, Riley R, Wright J, Williams S, Hainsworth R, et al. Randomized controlled trial of home-based exercise training to evaluate cardiac functional gains. Clin Sci (Lond). 2001;101(5):477-83. Like other established measures of exercise capacity, CPO cannot be predicted from resting cardiac parameters.37 Rerych SK, Scholz PM, Sabiston DC, Jr., Jones RH. Effects of exercise training on left ventricular function in normal subjects: a longitudinal study by radionuclide angiography. Am J Cardiol. 1980;45(2):244-52.

Under this background, the aims of the present study were: 1) to evaluate the test-retest reliability of TB in healthy adults during the daily clinical routine, and 2) to assess the relationships between CPO and resting measures of cardiac structure and function as well as traditional cardiopulmonary exercise parameters. Here, we applied a progressive statistic approach to provide thresholds above which effects might be meaningful and to present CO and CPO values that may be used as reference values in future studies.

Methods

Participants

In the study, 25 test persons were included into the study. All participants had no history of cardiovascular or pulmonary diseases, no cardioactive medication, a blood pressure of ≤ 140/90 mmHg, a body mass index < 25, a normal electrocardiogram, and a normal echocardiogram at the time of inclusion.

Study design

This study is a prospective non-interventional diagnostic single-center study. Participants were recruited in a cardiologic and internal medicine facility. The study was approved by the ethics committee of the University Witten/Herdecke and written informed consent was obtained. A standard echocardiogram was performed to exclude structural heart diseases and to investigate the relationships between established echocardiographic parameters and cardiopulmonary and hemodynamic values. Heart size, wall thickness, systolic, and diastolic function were all in physiological limits. All participants underwent two cardiopulmonary exercise tests separated by on week. During testing, TB using the NICOMTM device was applied.

Transthoracic echocardiography

Echocardiography was performed to assess cardiac structure and function using a standard ultrasound system (Vivid 7, General Electric, Milwaukee, Wisconsin). A complete transthoracic study was performed, including 2D, M-mode, spectral, and color Doppler techniques according to current recommendations and guidelines.3939 Evangelista A, Gaudio C, De Castro S, Faletra F, Nesser HJ, Kuvin JT, et al. Three-dimensional echocardiography--state-of-the-art. Indian Heart J. 2008;60(3 Suppl C):C3-9.,4040 Lang RM, Bierig M, Devereux RB, Flachskampf FA, Foster E, Pellikka PA, et al. Recommendations for chamber quantification: a report from the American Society of Echocardiography's Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology. J Am Soc Echocardiogr. 2005;18(12):1440-63. Standard parameters were: interventricular septal wall thickness in diastole, left ventricle end-diastolic diameter, left ventricular posterior wall thickness in diastole, and fractional shortening. Left ventricular ejection fraction was measured by means of modified biplane Simpson's method. Doppler tissue imaging was performed at the junction of the septal and lateral mitral annulus in apical 4-chamber view to determine peak mitral annular velocity during early filling (E`) and the ratio between early mitral inflow velocity and mitral annular early diastolic velocity (V).

Cardiopulmonary exercise testing

A symptom-limited incremental exercise test was performed in a seated position on a cycling ergometer (ec-3000, customed GmbH, Germany). The tests were performed by trained personal. After 5 min of rest, participants started at 0 W and the workload increased every 2 min by 25 W (standard WHO protocol). HR, blood pressure on the right arm using a sphygmomanometer, and a 12-lead electrocardiogram were obtained at rest and each stage as well as for 3 min post-exercise. The respiratory gas analysis was performed using a spiroergometry system (Cortex Metalyzer® 3B, Leipzig, Germany, software Metasoft studio 5.1.2 SR1). Ventilatory oxygen consumption and standard gas exchange data were measured breath-by-breath and averaged over 30 s. The following standard parameters were measured: Time to exhaustion, maximum workload, ventilatory anaerobic threshold (VAT) and peak oxygen uptake (VO2peak). The anaerobic threshold was determined using the V-slope method.4141 Hoppe MW, Sperlich B, Baumgart C, Janssen M, Freiwald J. Reliability of selected parameters of cycling ergospirometry from the powercube-ergo respiratory gas analyser. Sportverletz Sportschaden. 2015;29(3):173-9. The submaximal load was determined as the second last completed incremental. VO2peak was defined as the highest VO2 observed during testing.

Thoracic bioreactance

TB (NICOM®, Cheetah Medical, Portland, Oregon, USA) was added for noninvasive hemodynamic monitoring during rest and exercise. The examination was performed according to the manufacturer's protocol, as described previously.22 Myers J, Gujja P, Neelagaru S, Burkhoff D. Cardiac output and cardiopulmonary responses to exercise in heart failure: application of a new bio-reactance device. J Card Fail. 2007;13(8):629-36.,2121 Maurer MM, Burkhoff D, Maybaum S, Franco V, Vittorio TJ, Williams P, et al. A multicenter study of noninvasive cardiac output by bioreactance during symptom-limited exercise. J Card Fail. 2009;15(8):689-99.,4242 Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D. Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput. 2008;22(2):113-9. Table 1 shows the parameters that were calculated by the Cheetah NICOM® system. The NICOM® system uses four sensors applied to the right and left sides of the chest. Each sensor consists of an outer transmitting electrode and an inner receiving electrode. The outer electrodes transmit a low amplitude alternating electrical current with a frequency of 75 kHz to the thoracic cavity. The electrical properties of the thorax cyclically change due to the pulsatile volume of blood ejected from the heart. The pulsatile blood flow in the large thoracic arteries causes time delays (phase shifts) between the applied alternating electrical current and the thoracic voltage measured by the inner electrodes. Based on the measured phase shift the maximum aortic flow (dX/dtmax) and the ventricular ejection time (time from aortic valve opening to aortic valve closure, VET) were measured. Finally, the SV was obtained as SV = DX/DT × VET. Thereon, the CO, and finally the CPO, was derived.4343 Fincke R, Hochman JS, Lowe AM, Menon V, Slater JN, Webb JG, et al. Cardiac power is the strongest hemodynamic correlate of mortality in cardiogenic shock: a report from the SHOCK trial registry. J Am Coll Cardiol. 2004;44(2):340-8. The SV data were measured beat-by-beat and averaged over 60 s.

Table 1
Parameters calculated by the Cheetah NICOM® system

Statistical analysis

In a first step, participants were excluded from statistical analyses due to measurement errors (outliers), which were defined as ≥mean ± twofold pooled standard deviation.4444 Vickery WM, Dascombe BJ, Baker JD, Higham DG, Spratford WA, Duffield R. Accuracy and reliability of GPS devices for measurement of sports-specific movement patterns related to cricket, tennis, and field-based team sports. J Strength Cond Res. 2014;28(6):1697-705.

The test-retest reliability of cardiopulmonary and hemodynamic parameters was analyzed by (1) the difference in means to detect systematic bias, (2) intraclass correlation coefficients (ICC) to examine the relative reliability, and (3) typical error (TE) of measurements to quantify the absolute reliability.4545 Hopkins WG. Measures of reliability in sports medicine and science. Sports Med. 2000;30(1):1-15. To examine the difference in means, a progressive statistical approach using magnitude-based inferences for practical significance were computed.4646 Batterham AM, Hopkins WG. Making meaningful inferences about magnitudes. Int J Sports Physiol Perform. 2006;1(1):50-7. Compared to traditional null-hypothesis testing, that is influenced by the sample size, magnitude-based inferences ground an analysis, how big the observed effect is, and if the effect is lower, similar, or higher than the smallest worthwhile difference (SWD).4646 Batterham AM, Hopkins WG. Making meaningful inferences about magnitudes. Int J Sports Physiol Perform. 2006;1(1):50-7. Therefore, means and 90% confidence intervals (CIs) was computed first. Then, the disposition of the mean differences in relation to the SDWs were investigated. While the SDW for the maximal workload was calculated from the pooled standard deviation multiplied by 0.2, the SWD for all other physiological variables were calculated from the pooled standard deviations multiplied by 0.6, because it is well known that physiological variables showed a clearly higher spontaneous variability than biomechanical measures.4747 Hopkins WG, Marshall SW, Batterham AM, Hanin J. Progressive statistics for studies in sports medicine and exercise science. Med Sci Sports Exerc. 2009;41(1):3-13. Finally, the likelihoods for test 2 showing “true” higher, similar, or lower values than test 1 were determined and qualitatively described using the following probabilistic scale: <1%, most unlikely; 1 to <5%, very unlikely; 5 to <25%, unlikely; 25 to <75%, possibly; 75 to <95%, likely; 95 to <99%, very likely, and ≥99%, most likely. If the likelihoods for having both higher and lower values were ≥5%, the differences were described as unclear. Otherwise, the differences were interpreted according to the observed likelihoods. To clarify the meaningfulness of the differences, standardized differences labeled as effect sizes (ESs) were calculated and interpreted accordingly: 0.2 to <0.6, small; 0.6 to <1.2, moderate; 1.2 to <2.0, large; 2.0 to <4.0, very large; and ≥4.0, extreme large. To express the relative reliability, ICCS and 90% CIs were computed. The coefficients were described as follows: <0.20, very low; 0.20 to <0.50, low; 0.50 to <0.75, moderate; 0.75 to <0.90, high; 0.90 to <0.99, very high; and ≥0.99, extremely high. To quantify the absolute reliability, TEs and 90% CIs were calculated. The meaningfulness of the TEs was expressed via standardization for which the aforementioned scale for standardized differences was applied.4747 Hopkins WG, Marshall SW, Batterham AM, Hanin J. Progressive statistics for studies in sports medicine and exercise science. Med Sci Sports Exerc. 2009;41(1):3-13.

The relationships between the CPO and measures of cardiac structure and function as well as traditional cardiopulmonary exercise parameters were investigated using Pearson correlation coefficients (r) that were interpreted accordingly: <0.1, trivial; 0.1 to <0.3, small, 0.3 to <0.5, moderate; 0.5 to <0.7, large; 0.7 to <0.9, very large; 0.9 to 1.0, almost perfect.4747 Hopkins WG, Marshall SW, Batterham AM, Hanin J. Progressive statistics for studies in sports medicine and exercise science. Med Sci Sports Exerc. 2009;41(1):3-13. Lastly, common variances from coefficients of determinations (R2) were computed. Thereby, a cutting-off value of 50% was defined to clarify, if two variables are dependent or independent from each other.4848 Thomas JR, Nelson JK, Silverman SJ. Research methods in physical activity. Champaign: Human Kinectcs; 2005.

Results

25 participants completed both exercise tests. 17 participants (10 male, 7 female) were finally included. 8 participants were excluded due to measurement errors (outliers). Anthropometric, echocardiographic, and spiroergometric data of the participants are presented in Table 2.

Table 2
Anthropometric, echocardiographic, and maximal exercise characteristics of the participants (male: n = 10; female: n = 7)

Reliability

Data concerning systematic bias are presented in Table 3. It shows the differences in means between test 1 and test 2 for all hemodynamic and cardiopulmonary parameters measured at rest and during submaximal and peak exercise conditions. For all parameters, there were unclear to very likely trivial differences with small to moderate ESs (ES: 0.2-0.6).

Table 3
Changes in means of the resting, submaximal and maximal hemodynamic and cardiorespiratory characteristics

Table 4 summarizes the relative and absolute reliability expressed by ICCs and TEs, respectively, for all measured parameters. The ICCS ranged from moderate (ICC: 0.59) to very high (ICC: 0.98), whereas the TEs ranged from small (ES: 0.3) to large (ES: 1.8). CPO demonstrated superior relative and absolute reliability under all measurement conditions (ICC: 0.80-0.85; ES: 0.9-1.1) than its underlying parameters (ICC: 0.59-0.98; ES: 0.3-1.8).

Table 4
Relative (ICC) and absolute reliability (TE) of the resting, submaximal, and maximal cardiorespiratory and hemodynamic characteristics

Relationships

Figure 1 shows the relationships between echocardiographic measures of cardiac structure and function, traditional cardiopulmonary exercise parameters, and peak CPO. The peak CPO correlated moderately with VO2peak (Figure 1A: r = 0.68; R2= 0.47) and VAT (Figure 1B: r = 0.55; R2= 0.31), but only small with left ventricular wall thickness (Figure 1E: r = 0.33; R2= 0.11), left ventricular end-diastolic diameter (Figure 1F: r = 0.38; R2= 0.14), and systolic (Figure 1C: r = -0.32; R2= 0.11) as well as trivial with diastolic function (Figure 1D: r = 0.20; R2= 0.04).

Figure 1
Relationships between the CPO and established cardiorespiratory and echocardiographic characteristics. CPO: cardiac power output; VO2peak: peak oxygen uptake; VAT: ventilatory anaerobic threshold; EF: ejection fraction; E/E‘: ratio between early mitral inflow velocity and mitral annular early diastolic velocity; r: Pearson correlation coefficient; R2: coefficient of determination.

Discussion

Our main findings were: (1) there were no systematic bias for all measured parameters during all conditions, (2) all noninvasively measured hemodynamic parameters showed small to large test-retest-reliability, whereas the CPOpeak demonstrated a superior reliability than its underlying parameters, and (3) CPO was independent of measures of cardiac structure and function as well as traditional cardiopulmonary exercise parameters.

Our first finding was that there was no systematic bias during all examination conditions. These outcomes are in line with further studies, investigating hemodynamic and cardiopulmonary exercise parameters.2626 Jones TW, Houghton D, Cassidy S, MacGowan GA, Trenell MI, Jakovljevic DG. Bioreactance is a reliable method for estimating cardiac output at rest and during exercise. Br J Anaesth. 2015;115(3):386-91.,4949 Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C. Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med. 2007;33(7):1191-4. Overall, in our study, systematic bias due to learning, subject motivation, and fatiguing effects as well as errors in calibration procedures can be excluded.4545 Hopkins WG. Measures of reliability in sports medicine and science. Sports Med. 2000;30(1):1-15.,5050 Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med. 1998;26(4):217-38. This assumption supports our research design.

The second major finding was that all noninvasively measured hemodynamic parameters showed an acceptable test-retest reliability during rest, submaximal, and maximal exertion. Jones et al.2626 Jones TW, Houghton D, Cassidy S, MacGowan GA, Trenell MI, Jakovljevic DG. Bioreactance is a reliable method for estimating cardiac output at rest and during exercise. Br J Anaesth. 2015;115(3):386-91. first showed a good test-retest reliability of TB in a healthy population at rest as well as during submaximal and maximal exertion. However, acceptable test-retest reliability was impacted by the fact that we have previously excluded a significant number of outliers (n = 8) due to measurement errors. It is further noteworthy that the reliability of our TB measurements was to some degree inferior compared to a previous study in which the reliability of a comparable technology (beat by beat signal morphology impedance cardiography) to evaluate the hemodynamic response was assessed.2020 Gordon N, Abbiss CR, Maiorana AJ, Marstron KJ, Peiffer JJ. Intrarater reliability and agreement of the physioflow bioimpedance cardiography device during rest, moderate and high-intensive exercise. Kinesiology. 2018;50(1 Suppl 1):140-9. One possible explanation for the differences may be that we investigated the reliability under less standardized conditions, another one could be related to significant technological differences.

Overall, when outliers are excluded, TB can be considered as an appropriate technology to not only assess hemodynamic status in a research setting but also in everyday practice.

The central task of the heart is to produce a sufficient CO and maintain an adequate MAP. Therefore, cardiac performance can be best explained by CPO, because it takes both the flow- and pressure-generating capacities of the heart into account.2929 Tan LB. Clinical and research implications of new concepts in the assessment of cardiac pumping performance in heart failure. Cardiovasc Res. 1987;21(8):615-22.

In chronic heart failure, the application of hemodynamic measuring to standard cardiopulmonary exercise testing may help to explain the underlying mechanism of exercise intolerance with impact on clinical decision making,3131 Chomsky DB, Lang CC, Rayos GH, Shyr Y, Yeoh TK, Pierson RN 3rd, et al. Hemodynamic exercise testing. A valuable tool in the selection of cardiac transplantation candidates. Circulation. 1996;94(12):3176-83. therapy planning, and performance3232 Wilson JR, Groves J, Rayos G. Circulatory status and response to cardiac rehabilitation in patients with heart failure. Circulation. 1996;94(7):1567-72. as well as risk stratification.5151 Metra M, Faggiano P, D'Aloia A, Nodari S, Gualeni A, Raccagni D, et al. Use of cardiopulmonary exercise testing with hemodynamic monitoring in the prognostic assessment of ambulatory patients with chronic heart failure. J Am Coll Cardiol. 1999;33(4):943-50. Chomsky et al.3131 Chomsky DB, Lang CC, Rayos GH, Shyr Y, Yeoh TK, Pierson RN 3rd, et al. Hemodynamic exercise testing. A valuable tool in the selection of cardiac transplantation candidates. Circulation. 1996;94(12):3176-83. showed that CO respond to exercise is a strong predictor of mortality in cardiac transplantation candidates. In addition, Lang et al.3535 Lang CC, Karlin P, Haythe J, Lim TK, Mancini DM. Peak cardiac power output, measured noninvasively, is a powerful predictor of outcome in chronic heart failure. Circ Heart Fail. 2009;2(1):33-8. demonstrated CPO as the most powerful and independent predictor of survival chronic heart failure outcome in patients with chronic heart failure and that may enhance the prognostic power of traditional cardiopulmonary exercise testing.

In sports medicine, monitoring of the training status is essential to guide the training process. Training leads to significant structural and functional changes of the cardiovascular system.66 Kovacs R, Baggish AL. Cardiovascular adaptation in athletes. Trends Cardiovasc Med. 2016;26(1):46-52. In a randomized cross-over study, Marshall et al.3838 Marshall P, Al-Timman J, Riley R, Wright J, Williams S, Hainsworth R, et al. Randomized controlled trial of home-based exercise training to evaluate cardiac functional gains. Clin Sci (Lond). 2001;101(5):477-83. assessed the effect of moderate exercise training on cardiac performance in non-athletic adults. Due to training, the CPOpeak increased by 16%, whereas the CPO at rest remained unchanged. In highly trained endurance athletes, Schlader et al.3737 Schlader ZJ, Mundel T, Barnes MJ, Hodges LD. Peak cardiac power output in healthy, trained men. Clin Physiol Funct Imaging. 2010;30(6):480-4. found double CPOpeak values compared to non-athletes. These results have been confirmed by Klasnja et al.33 Klasnja AV, Jakovljevic DG, Barak OF, Popadic Gacesa JZ, Lukac DD, Grujic NG. Cardiac power output and its response to exercise in athletes and non-athletes. Clin Physiol Funct Imaging. 2013;33(3):201-5. in football and basketball players.

In our study, the CPOpeak showed superior reliability than the underlying physiological single parameters. However, it should be noted that the reliability of the CPO was potentially influenced by the reliability of the MAP (which was higher) rather than that of the SV and CO (which were lower). Thus, CPO measuring by TB seems to be feasible due to its surrogate character. It is, however, important to mention that we have averaged all our beat-by-beat measured TB data, including the CPO, over 60 s, which might have also artificially improved our statistical outcomes. The reason for our data processing method was that we aimed to investigate the global cardiac performance. Such a data processing proceed is of course inadequately, when aiming to assess transient cardiac abnormalities during exercise like ischemia. Since the beat-by-beat reliability of our TB based measures remains unknown, we recommend other impedance-based technologies, which offer a reliable beat-by-beat analysis of hemodynamic parameters during exercise.2020 Gordon N, Abbiss CR, Maiorana AJ, Marstron KJ, Peiffer JJ. Intrarater reliability and agreement of the physioflow bioimpedance cardiography device during rest, moderate and high-intensive exercise. Kinesiology. 2018;50(1 Suppl 1):140-9.

The third major finding was that CPO was found to be independent of cardiac structure and function at rest as well as to traditional cardiopulmonary exercise parameters. Klasnja et al.33 Klasnja AV, Jakovljevic DG, Barak OF, Popadic Gacesa JZ, Lukac DD, Grujic NG. Cardiac power output and its response to exercise in athletes and non-athletes. Clin Physiol Funct Imaging. 2013;33(3):201-5. previously demonstrated a weak correlation between CPOpeak and resting parameters of left ventricular morphology and function.33 Klasnja AV, Jakovljevic DG, Barak OF, Popadic Gacesa JZ, Lukac DD, Grujic NG. Cardiac power output and its response to exercise in athletes and non-athletes. Clin Physiol Funct Imaging. 2013;33(3):201-5. We also did not find a strong relation between CPOpeak and echocardiographic findings at rest. Our findings show once again that resting parameters cannot be used to estimate maximal cardiovascular performance.

For the first time driven by our progressive statistics,4747 Hopkins WG, Marshall SW, Batterham AM, Hanin J. Progressive statistics for studies in sports medicine and exercise science. Med Sci Sports Exerc. 2009;41(1):3-13. we report the SWDs for all investigated TB parameters. From a practical point of view, the provided thresholds can be used as a framework to judge in healthy adults, whether observed differences in the analyzed parameters should be interpreted or not in a daily medical routine. Further, it is promising to use these thresholds as cutting-off values for minimal required effects detected by longitudinal or cross-sectional studies using the here investigated TB measures in the future. For example, in healthy adults, the calculated SWD of the CPO was 0.7 W, meaning that longitudinal or cross-sectional differences should only be interpreted, when this cut-off value is exceeded.

The major limitation of our study is the high dropout rate (n=8). However, to detect outliers, we objectively defined them as those values, which were greater than the pooled standard deviation. Based on this approach and our recruited healthy adults, it can be assumed that the detected outliers had not a physiological cause. Contrary, it is more likely that the identified outliers had rather an underlying technical reason. Therefore, further improvements in TB, for example, regarding the application and quality of electrodes are required. Consequently, technical errors must be executed by proprietary algorithms, before valid decisions are possible. When taken these aspects together, our findings indicate that TB can only be considered as a reliable technology for measuring hemodynamic parameters after outliers have been excluded.

Conclusion

In conclusion, at this stage, our results preclude the clinical use of TB in healthy subject population when outliers are not identified even if a previous study seem to show its possible application in a strictly controlled research setting.

  • Sources of Funding
    This study was funded by Bayer Pharma.
  • Study Association
    This article is part of the thesis of Doctoral submitted by Christian Kiefer, from University Hospital Witten/Herdecke.
  • Ethics approval and consent to participate
    This study was approved by the Ethics Committee of the University of Witten/Herdecke under the protocol number 131/2914. All the procedures in this study were in accordance with the 1975 Helsinki Declaration, updated in 2013. Informed consent was obtained from all participants included in the study.

References

  • 1
    Jhanji S, Dawson J, Pearse RM. Cardiac output monitoring: basic science and clinical application. Anaesthesia. 2008;63(2):172-81.
  • 2
    Myers J, Gujja P, Neelagaru S, Burkhoff D. Cardiac output and cardiopulmonary responses to exercise in heart failure: application of a new bio-reactance device. J Card Fail. 2007;13(8):629-36.
  • 3
    Klasnja AV, Jakovljevic DG, Barak OF, Popadic Gacesa JZ, Lukac DD, Grujic NG. Cardiac power output and its response to exercise in athletes and non-athletes. Clin Physiol Funct Imaging. 2013;33(3):201-5.
  • 4
    Myers J, Froelicher VF. Hemodynamic determinants of exercise capacity in chronic heart failure. Ann Intern Med. 1991;115(5):377-86.
  • 5
    Sullivan MJ, Knight JD, Higginbotham MB, Cobb FR. Relation between central and peripheral hemodynamics during exercise in patients with chronic heart failure. Muscle blood flow is reduced with maintenance of arterial perfusion pressure. Circulation. 1989;80(4):769-81.
  • 6
    Kovacs R, Baggish AL. Cardiovascular adaptation in athletes. Trends Cardiovasc Med. 2016;26(1):46-52.
  • 7
    Rerych SK, Scholz PM, Sabiston DC, Jr., Jones RH. Effects of exercise training on left ventricular function in normal subjects: a longitudinal study by radionuclide angiography. Am J Cardiol. 1980;45(2):244-52.
  • 8
    Franciosa JA, Park M, Levine TB. Lack of correlation between exercise capacity and indexes of resting left ventricular performance in heart failure. Am J Cardiol. 1981;47(1):33-9.
  • 9
    Wilson JR, Rayos G, Yeoh TK, Gothard P, Bak K. Dissociation between exertional symptoms and circulatory function in patients with heart failure. Circulation. 1995;92(1):47-53.
  • 10
    Ponikowski P, Voors AA, Anker SD, Bueno H, Cleland JG, Coats AJ, et al. 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure: The Task Force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC). Developed with the special contribution of the Heart Failure Association (HFA) of the ESC. Eur J Heart Fail. 2016;18(8):891-975.
  • 11
    Myers J. Applications of cardiopulmonary exercise testing in the management of cardiovascular and pulmonary disease. Int J Sports Med. 2005 Feb;26 (Suppl 1):S49-55.
  • 12
    Lang CC, Agostoni P, Mancini DM. Prognostic significance and measurement of exercise-derived hemodynamic variables in patients with heart failure. J Card Fail. 2007;13(8):672-9.
  • 13
    Becklake MR, Frank H, Dagenais GR, Ostiguy GL, Guzman CA. Influence of age and sex on exercise cardiac output. J Appl Physiol. 1965;20(5):938-47.
  • 14
    Wilson JR, Rayos G, Yeoh TK, Gothard P. Dissociation between peak exercise oxygen consumption and hemodynamic dysfunction in potential heart transplant candidates. J Am Coll Cardiol. 1995;26(2):429-35.
  • 15
    Nicoletti I, Cicoira M, Zanolla L, Franceschini L, Brighetti G, Pilati M, et al. Skeletal muscle abnormalities in chronic heart failure patients: relation to exercise capacity and therapeutic implications. Congest Heart Fail. 2003;9(3):148-54.
  • 16
    Lund-Johansen P. The dye dilution method for measurement of cardiac output. Eur Heart J. 1990 Dec;11(Suppl I):6-12.
  • 17
    Warburton DE, Haykowsky MJ, Quinney HA, Humen DP, Teo KK. Reliability and validity of measures of cardiac output during incremental to maximal aerobic exercise. Part II: Novel techniques and new advances. Sports Med. 1999;27(4):241-60.
  • 18
    Sandham JD, Hull RD, Brant RF, Knox L, Pineo GF, Doig CJ, et al. A randomized, controlled trial of the use of pulmonary-artery catheters in high-risk surgical patients. N Engl J Med. 2003;348(1):5-14.
  • 19
    Harvey S, Stevens K, Harrison D, Young D, Brampton W, McCabe C, et al. An evaluation of the clinical and cost-effectiveness of pulmonary artery catheters in patient management in intensive care: a systematic review and a randomised controlled trial. Health Technol Assess. 2006;10(29):iii-iv, ix-xi, 1-133.
  • 20
    Gordon N, Abbiss CR, Maiorana AJ, Marstron KJ, Peiffer JJ. Intrarater reliability and agreement of the physioflow bioimpedance cardiography device during rest, moderate and high-intensive exercise. Kinesiology. 2018;50(1 Suppl 1):140-9.
  • 21
    Maurer MM, Burkhoff D, Maybaum S, Franco V, Vittorio TJ, Williams P, et al. A multicenter study of noninvasive cardiac output by bioreactance during symptom-limited exercise. J Card Fail. 2009;15(8):689-99.
  • 22
    Keren H, Burkhoff D, Squara P. Evaluation of a noninvasive continuous cardiac output monitoring system based on thoracic bioreactance. Am J Physiol Heart Circ Physiol. 2007;293(1):H583-9.
  • 23
    Jakovljevic DG, Moore S, Hallsworth K, Fattakhova G, Thoma C, Trenell MI. Comparison of cardiac output determined by bioimpedance and bioreactance methods at rest and during exercise. J Clin Monit Comput. 2012;26(2):63-8.
  • 24
    Marik PE, Levitov A, Young A, Andrews L. The use of bioreactance and carotid Doppler to determine volume responsiveness and blood flow redistribution following passive leg raising in hemodynamically unstable patients. Chest. 2013;143(2):364-70.
  • 25
    Elliott A, Hull JH, Nunan D, Jakovljevic DG, Brodie D, Ansley L. Application of bioreactance for cardiac output assessment during exercise in healthy individuals. Eur J Appl Physiol. 2010;109(5):945-51.
  • 26
    Jones TW, Houghton D, Cassidy S, MacGowan GA, Trenell MI, Jakovljevic DG. Bioreactance is a reliable method for estimating cardiac output at rest and during exercise. Br J Anaesth. 2015;115(3):386-91.
  • 27
    Tan LB. Evaluation of cardiac dysfunction, cardiac reserve and inotropic response. Postgrad Med J. 1991;67(Suppl 1):S10-20.
  • 28
    Cotter G, Williams SG, Vered Z, Tan LB. Role of cardiac power in heart failure. Curr Opin Cardiol. 2003;18(3):215-22.
  • 29
    Tan LB. Clinical and research implications of new concepts in the assessment of cardiac pumping performance in heart failure. Cardiovasc Res. 1987;21(8):615-22.
  • 30
    Cooke GA, Marshall P, al-Timman JK, Wright DJ, Riley R, Hainsworth R, et al. Physiological cardiac reserve: development of a non-invasive method and first estimates in man. Heart. 1998;79(3):289-94.
  • 31
    Chomsky DB, Lang CC, Rayos GH, Shyr Y, Yeoh TK, Pierson RN 3rd, et al. Hemodynamic exercise testing. A valuable tool in the selection of cardiac transplantation candidates. Circulation. 1996;94(12):3176-83.
  • 32
    Wilson JR, Groves J, Rayos G. Circulatory status and response to cardiac rehabilitation in patients with heart failure. Circulation. 1996;94(7):1567-72.
  • 33
    Grodin JL, Mullens W, Dupont M, Wu Y, Taylor DO, Starling RC, et al. Prognostic role of cardiac power index in ambulatory patients with advanced heart failure. Eur J Heart Fail. 2015;17(7):689-96.
  • 34
    Williams SG, Cooke GA, Wright DJ, Parsons WJ, Riley RL, Marshall P, et al. Peak exercise cardiac power output; a direct indicator of cardiac function strongly predictive of prognosis in chronic heart failure. Eur Heart J. 2001;22(16):1496-503.
  • 35
    Lang CC, Karlin P, Haythe J, Lim TK, Mancini DM. Peak cardiac power output, measured noninvasively, is a powerful predictor of outcome in chronic heart failure. Circ Heart Fail. 2009;2(1):33-8.
  • 36
    Bromley PD, Hodges LD, Brodie DA. Physiological range of peak cardiac power output in healthy adults. Clin Physiol Funct Imaging. 2006;26(4):240-6.
  • 37
    Schlader ZJ, Mundel T, Barnes MJ, Hodges LD. Peak cardiac power output in healthy, trained men. Clin Physiol Funct Imaging. 2010;30(6):480-4.
  • 38
    Marshall P, Al-Timman J, Riley R, Wright J, Williams S, Hainsworth R, et al. Randomized controlled trial of home-based exercise training to evaluate cardiac functional gains. Clin Sci (Lond). 2001;101(5):477-83.
  • 39
    Evangelista A, Gaudio C, De Castro S, Faletra F, Nesser HJ, Kuvin JT, et al. Three-dimensional echocardiography--state-of-the-art. Indian Heart J. 2008;60(3 Suppl C):C3-9.
  • 40
    Lang RM, Bierig M, Devereux RB, Flachskampf FA, Foster E, Pellikka PA, et al. Recommendations for chamber quantification: a report from the American Society of Echocardiography's Guidelines and Standards Committee and the Chamber Quantification Writing Group, developed in conjunction with the European Association of Echocardiography, a branch of the European Society of Cardiology. J Am Soc Echocardiogr. 2005;18(12):1440-63.
  • 41
    Hoppe MW, Sperlich B, Baumgart C, Janssen M, Freiwald J. Reliability of selected parameters of cycling ergospirometry from the powercube-ergo respiratory gas analyser. Sportverletz Sportschaden. 2015;29(3):173-9.
  • 42
    Raval NY, Squara P, Cleman M, Yalamanchili K, Winklmaier M, Burkhoff D. Multicenter evaluation of noninvasive cardiac output measurement by bioreactance technique. J Clin Monit Comput. 2008;22(2):113-9.
  • 43
    Fincke R, Hochman JS, Lowe AM, Menon V, Slater JN, Webb JG, et al. Cardiac power is the strongest hemodynamic correlate of mortality in cardiogenic shock: a report from the SHOCK trial registry. J Am Coll Cardiol. 2004;44(2):340-8.
  • 44
    Vickery WM, Dascombe BJ, Baker JD, Higham DG, Spratford WA, Duffield R. Accuracy and reliability of GPS devices for measurement of sports-specific movement patterns related to cricket, tennis, and field-based team sports. J Strength Cond Res. 2014;28(6):1697-705.
  • 45
    Hopkins WG. Measures of reliability in sports medicine and science. Sports Med. 2000;30(1):1-15.
  • 46
    Batterham AM, Hopkins WG. Making meaningful inferences about magnitudes. Int J Sports Physiol Perform. 2006;1(1):50-7.
  • 47
    Hopkins WG, Marshall SW, Batterham AM, Hanin J. Progressive statistics for studies in sports medicine and exercise science. Med Sci Sports Exerc. 2009;41(1):3-13.
  • 48
    Thomas JR, Nelson JK, Silverman SJ. Research methods in physical activity. Champaign: Human Kinectcs; 2005.
  • 49
    Squara P, Denjean D, Estagnasie P, Brusset A, Dib JC, Dubois C. Noninvasive cardiac output monitoring (NICOM): a clinical validation. Intensive Care Med. 2007;33(7):1191-4.
  • 50
    Atkinson G, Nevill AM. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med. 1998;26(4):217-38.
  • 51
    Metra M, Faggiano P, D'Aloia A, Nodari S, Gualeni A, Raccagni D, et al. Use of cardiopulmonary exercise testing with hemodynamic monitoring in the prognostic assessment of ambulatory patients with chronic heart failure. J Am Coll Cardiol. 1999;33(4):943-50.

Publication Dates

  • Publication in this collection
    10 July 2019
  • Date of issue
    Aug 2019

History

  • Received
    19 July 2018
  • Reviewed
    31 Oct 2018
  • Accepted
    14 Nov 2018
Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
E-mail: revista@cardiol.br