Influence of inter-root septum width on mini-implant stability

Abstracts

OBJECTIVE: The purpose of this study was to evaluate the influence of the inter-radicular septum width in the insertion site of self-drilling mini-implants on the stability degree of these anchorage devices. METHODS: The sample consisted of 40 mini-implants inserted in the inter-radicular septum between maxillary second premolars and first molars in 21 patients to provide skeletal anchorage for anterior retraction. The post-surgical radiographs were used to measure the septum width in the insertion site (ISW). In this regard, the mini-implants were divided in two groups: group 1 (critical areas, ISW<3 mm) and group 2 (non-critical areas, ISW>3 mm). The degree of mobility (DM) was monthly quantified to determine mini-implant stability, and the success rate of these devices was calculated. This study also evaluated the sensitivity degree during miniscrew load, amount of plaque around the miniscrew, insertion height, and total evaluation period. RESULTS: The results showed no significant difference in mobility degree and success rate between groups 1 and 2. The total success rate found was 90% and no variable was associated with the miniscrew failure. Nevertheless, the results showed that greater patient sensitivity degree was associated to the mini-implant mobility and the failure of these anchorage devices happened in a short time after their insertion. CONCLUSION: Septum width in the insertion site did not influence the self-drilling mini-implant stability evaluated in this study.

Orthodontic anchorage procedures; Dental implants; Dental radiography; Tooth root


OBJETIVO: este estudo teve como objetivo avaliar a influência da largura do septo inter-radicular no local de inserção de mini-implantes autoperfurantes sobre o grau de estabilidade desses dispositivos de ancoragem. MÉTODOS: a amostra consistiu de 40 mini-implantes inseridos entre as raízes do primeiro molar e segundo pré-molar superiores de 21 pacientes, com o intuito de fornecer ancoragem para retração anterior. A largura do septo no local de inserção (LSI) foi mensurada nas radiografias pós-cirúrgicas e, sob esse aspecto, os mini-implantes foram divididos em dois grupos: grupo 1 (áreas críticas, LSI<3mm) e grupo 2 (áreas não críticas, LSI>3mm). A estabilidade dos mini-implantes foi avaliada mensalmente pela quantificação do grau de mobilidade e a partir dessa variável foi calculada a proporção de sucesso. Avaliou-se também: a quantidade de placa, altura de inserção, grau de sensibilidade e período de observação. RESULTADOS: os resultados obtidos demonstraram que não houve diferença estatisticamente significativa para o grau de mobilidade e proporção de sucesso entre os mini-implantes inseridos em septos de largura mesiodistal crítica e não crítica. A proporção de sucesso total encontrada foi de 90% e nenhuma variável demonstrou estar relacionada ao insucesso dos mini-implantes. No entanto, observou-se maior sensibilidade nos pacientes cujos mini-implantes apresentavam mobilidade, e que a falha desses dispositivos de ancoragem ocorria logo após sua inserção. CONCLUSÃO: a largura do septo inter-radicular no local de inserção não interferiu na estabilidade dos mini-implantes autoperfurantes avaliados neste estudo.

Procedimentos de ancoragem ortodôntica; Implantes dentários; Radiografia dentária; Raiz dentária


ORIGINAL ARTICLE

Influence of inter-root septum width on mini-implant stability

Mariana Pracucio GigliottiI; Guilherme JansonII; Sérgio Estelita Cavalcante BarrosIII; Kelly ChiquetoIV; Marcos Roberto de FreitasV

IMSc in Orthodontics, Bauru Dental School (FOB) - University of São Paulo (USP)

IIProfessor and Head, Department of Pediatric Dentistry, Orthodontics and Public Health, FOB-USP. Coordinator of the Applied Dental Sciences Program, FOB-USP. Member of the “Royal College of Dentists of Canada"

IIIMaster, PhD and Postdoctoral in Orthodontics, FOB-USP

IVMSc and PhD in Orthodontics, FOB-USP

VProfessor, Department of Pediatric Dentistry, Orthodontics and Public Health, FOB-USP

Contact address

ABSTRACT

OBJECTIVE: The purpose of this study was to evaluate the influence of the inter-radicular septum width in the insertion site of self-drilling mini-implants on the stability degree of these anchorage devices.

METHODS: The sample consisted of 40 mini-implants inserted in the inter-radicular septum between maxillary second premolars and first molars in 21 patients to provide skeletal anchorage for anterior retraction. The post-surgical radiographs were used to measure the septum width in the insertion site (ISW). In this regard, the mini-implants were divided in two groups: group 1 (critical areas, ISW≤3 mm) and group 2 (non-critical areas, ISW>3 mm). The degree of mobility (DM) was monthly quantified to determine mini-implant stability, and the success rate of these devices was calculated. This study also evaluated the sensitivity degree during miniscrew load, amount of plaque around the miniscrew, insertion height, and total evaluation period.

RESULTS: The results showed no significant difference in mobility degree and success rate between groups 1 and 2. The total success rate found was 90% and no variable was associated with the miniscrew failure. Nevertheless, the results showed that greater patient sensitivity degree was associated to the mini-implant mobility and the failure of these anchorage devices happened in a short time after their insertion.

CONCLUSION: Septum width in the insertion site did not influence the self-drilling mini-implant stability evaluated in this study.

Keywords: Orthodontic anchorage procedures. Dental implants. Dental radiography. Tooth root.

Introduction

Mini-implants have become a routine anchorage method in orthodontic practice given their high predictability and scientifically proven benefits.1 The success rate of these devices ranges from 70% to 95%.2-7 On the other hand, considerable failure rates have prompted several studies that seek to determine the risk factors which may compromise mini-implant success.

According to the literature, location, angle of insertion, cortical bone thickness and bone tissue quality, presence of attached gingiva, mini-implant features, degree of primary stability and load intensity,10,14 hygiene and degree of peri-implant tissue inflammation3,5,11 are all factors associated with mini-implant stability. Moreover, some recent studies have found that proximity and contact between mini-implant and tooth root are significant risk factors in the failure of this anchorage system.2,15,16,17 However, contradictory results regarding the degree of influence of these various factors on the success rate of mini-implants are often found in the literature, given sample heterogeneity and the wide range of variables.

It is known that self-drilling mini-implants are the state-of-the-art device for orthodontic anchorage. Furthermore, studies have shown a larger contact area between the surface of mini-implants and bone tissue, thereby enhancing stability.19

One should also consider that the surgical risk of injury to tooth roots adjacent to the mini-implant has been significantly reduced by advances in this anchorage system, especially when three-dimensional surgical orientation guides are employed.19,20,21 Although risks inherent in the insertion procedure have been reduced, thereby facilitating the placement of these mini-implants in areas with critical dimensions, few studies have evaluated whether mini-implants placed in close proximity to the periodontal ligament may have their stability compromised.2,15,16,17

This study, therefore, aimed to compare the stability and success rate of self-drilling mini-implants placed in interradicular septa with critical and non-critical mesiodistal dimensions, i.e., septa with width equal to 3 mm and greater than 3 mm, respectively.

MATERIAL AND METHODS

Material

Twenty-one patients were selected (9 males, 12 females, mean age 16.99 ± 5.08 years) from the Clinic of Orthodontics, School of Dentistry of Bauru (FOB-USP) who were undergoing orthodontic treatment involving premolar extractions and requiring maximum anchorage for anterior retraction. The selection criteria used in this study were: Mini-implants located in the interradicular septum, between second premolars and maxillary first molars, self-drilling mini-implants (length = 7 mm and diameter = 1.5 mm, Absoanchor, Dentos®), inserted by the same dentist (Fig 1). The surgical protocol recommended and described by Barros20,21,22 was employed in this study as it makes use of a three-dimensional graded radiographic-surgical guide (GRSG), allowing a perpendicular insertion path with reasonable predictability of final mini-implant positioning (Fig 2).



However, some exclusion criteria were used: Absence of any local or systemic condition that could influence mini-implant stability, such as Active periodontal disease, smoking and diabetes, mini-implants placed in the mandible and with indications other than to provide anchorage for anterior retraction.

Thus, this study utilized 40 mini-implants, which were divided according to the width of the interradicular septum where they were inserted: ≤3.0 mm (group 1, critical areas) and >3.0 mm (group 2, non-critical areas).

Methods

Measurement of postoperative radiographs

Postoperative radiographs were obtained with a surgical radiographic positioner (SRP) in conjunction with GRSG and followed the technique of parallelism/bitewing radiographs (Fig 3). In these radiographs, the following variables were assessed: Width of the interradicular septum at the insertion site (ISW; Fig 4) and mini-implant insertion height (IH; Fig 5).




A Spectro II X-ray machine (Dabi Atlante, Ribeirão Preto, Brazil) was used with 50 kVp voltage, 10 mA current, and exposure time from 0.5 to 0.7 seconds. After processing, the radiographs were scanned in a 35 mm slide scanner (Sprint Scan 35 Plus, Version 2.7.2, Polaroid Corporation) with 675 dpi resolution and 1:1 ratio. Subsequently, the images were measured using Adobe Photoshop software (version 7.0, Adobe Corporation) and manipulated by the same examiner with an accuracy of 0.1 mm.21 High resolution scanning allowed up to 300% magnification without any loss in quality.

Assessment of mini-implant stability

Mini-implant stability was assessed through monthly measurements made from the time of insertion (primary stability) until removal using a method that evaluates horizontal mobility with the aid of an adjustable telescopic rod (ATR) developed by Barros (Fig 6). Stability measuring was performed with a digital caliper (Mitutoyo 500-144B, Mitutoyo South American) and an orthodontic tension gauge (Correx series 040-712-00, Dentaurum Orthodontics), with the aim of assigning numerical values to the degree of mobility of mini-implants.


Figure 7 illustrates the method used to evaluate the horizontal mobility of mini-implants, which will be described in the following steps:


» Step 1: First, it was necessary to define a random reference point such as the distal tie-wing of the canine bracket. ATR length was then adjusted according to the distance between the head of the mini-implant and the desired point. To this end, part "c" of the ATR was then connected to the head of the mini-implant (Fig 7A), while the tip of part "b" touched the reference point (Fig 7B). The device was then locked within this dimension (distance from the mini-implant to the distal tie-wing of the canine bracket) through the self-threading lock (a). This length was defined as the initial measurement and was measured with a digital caliper (Fig 7C).

» Step 2: With the device in position, i.e. part "c" connected to the head of the mini-implant (Fig 7D), the tension gauge was supported at part "d" and pulled mesially to apply 400 g of force (Fig 7E). At this point, it was checked whether any movement occurred at the end of the device's removable rod (part b) relative to the reference point.

» Step 3: To define the degree of mobility the amount of movement experienced by the end of the removable rod relative to the reference point was evaluated (Figs 7F and 7G). If no movement occurred the mini-implant was considered stable, i.e., the difference between initial and final ATR measurements was zero. However, if mesial movement of the removable rod end (b) was observed during evaluation, the mini-implant was deemed to have mobility. Thus, the device had its length reduced (final measurement) to make sure that the end of the removable rod (b) would once again coincide with the reference point during force application. After this procedure, the final measurement was taken with a digital caliper. Measurement of the degree of mobility was therefore considered representative of mini-implant stability, and equal to the difference between the final and initial measurements of ATR length.

The average degree of mobility of each mini-implant was obtained by calculating the means of monthly measurements, and the success rate was defined by the number of mini-implants that remained clinically stable divided by the total number of mini-implants evaluated in the study.

Evaluation of stability-related factors

Some factors that could affect mini-implant stability in this study were clinically evaluated:

  • Insertion site (IS): Divided between (1) region of attached gingiva, (2) region of alveolar mucosa or (3) mucogingival line.

  • Degree of sensitivity (DS) measured monthly during force application and analyzed by means of scores: (0) When the patient reported no discomfort during force application, (1) when the patient reported slight discomfort, (2) when the patient reported pain, but it was bearable, and (3) when sensitivity was considered unbearable by the patient.

  • Evaluation of peri-implant biofilm using the Modified Plaque Index (MPI) for dental implants since no specific index for mini-implants was found in the literature. This index uses a zero (0) score when there is no detectable plaque, (1) when plaque is detected by sliding a probe, (2) when plaque is visible to the naked eye, and (3) when soft matter is abundant.

Statistical Analysis

Method error and compatibility

between groups

Calculation of method error was performed for variables ISW and IH on 15 mini-implants distributed in both groups of this study. The formula proposed by Dahlberg24 (Se2=Σd2/2n) and the paired t-test were used to perform the calculation of random and systematic errors, respectively.

As for the monthly assessment of degree of mobility (DM) using ATR, parameters were used to ensure that measurements were reproducible: ATR's position relative to the reference point was checked by removing the device and putting it back in position. ATR positioning on the digital caliper had to be parallel to the caliper ruler. The value obtained for ATR length was checked by repositioning the device on the caliper. ATR locking was checked prior to application of force by the tension gauge. The visual analysis of presence or absence of movement at the ATR's removable rod end relative to the reference point was performed with utmost thoroughness.

T-test was used between groups 1 and 2 to check sample homogeneity for variables IH, MPI and OP.

Comparison between groups and variables

The Kolmogorov-Smirnov test was applied and showed absence of normal distribution for variables DM and DS, indicating the application of nonparametric tests for these variables and parametric tests for the others.

For the comparative analyses the following statistical tests were conducted:

  • Descriptive statistics: Means, standard deviations, maximum and minimum value of variables ISW, IH, DM, DS, MPI and OP.

  • Nonparametric Mann-Whitney test: To compare groups 1 and 2 for differences in the degree of mobility of mini-implants.

  • Fisher's exact test: To check for any association between the success rates of mini-implants in groups 1 and 2.

  • T-test and chi-square test: To compare all variables between the success and failure groups, and determine the risk factors associated with mini-implant failures.

All statistical tests were performed with Statistica software (Version 7.0, StatSoft Inc., Tulsa, OK, USA), adopting a significance level of p<0.05.

RESULTS

Method error results yielded very small values for random errors (0.0577 to 0.0912) and no significant systematic errors (Table 1). In addition, both groups showed compatibility regarding insertion height, amount of plaque and mini-implant observation period (Table 2).

Table 3 shows the results yielded by descriptive statistics of the 40 mini-implants for variables ISW, IH, DM, MPI, DS and OP. Table 4 shows a significant difference between groups 1 and 2 in terms of variable ISW.

Results demonstrated that the degree of mobility and success rate of groups 1 and 2 were similar (Tables 5 and 6).

An analysis of the risk factors involved in mini-implant stability in this investigation revealed that none of the variables could be associated with mini-implant failure. However, differences were observed in degree of sensitivity and mini-implant observation period in the success and failure groups (Table 7).

DISCUSSION

Sample

The sample utilized in this study (40 mini-implants) may be relatively small compared to previous studies that sought to determine risk factors for mini-implant success.1,3,5-8,11

In fact, sample size is a very important factor, but no less important are the selection criteria.26 Since this study followed strict selection criteria in determining its sample, the number of mini-implants was understandably reduced. Moreover, the results were influenced by a smaller number of uncontrolled variables, thereby contributing to the veracity of the inferences, which might not occur if mini-implants with different characteristics were compared, inserted and used under uncontrolled conditions, which would excessively increase the number of variables involved in the stability of these anchorage devices.1,7

Sample selection included only mini-implants inserted with the aid of GRSG, since the use of surgical guides for mini-implant installation in areas of interradicular septum is mandatory to prevent injury to adjacent structures,20,27 and when guides are not used the success of this procedure depends almost exclusively on the skill and experience of the surgeon.

Factors age and gender were not standardized since several studies have shown that these characteristics are not directly associated with decreased stability or mini-implant success rate.3-8,11

Results

Periapical/bitewing radiographs are often employed to evaluate the final positioning of mini-implants.2,7 Their accuracy, however, depends on standardized implementation and/or the aid of radiographic and surgical guides.

An examination of postoperative radiographs showed that septum width at mini-implant insertion exhibited a mean value of 3.05±0.82 mm. It is noteworthy that the mini-implants were inserted in a more coronal region of the septum (3.11 ± 0.72 mm, Table 3), which usually has a shorter interradicular distance compared to a more apical region.28 The results of this study were similar to those reported by Hernández et al28 and Hu et al29 but slightly higher than those achieved by Deguchi et al8 and Poggio et al.9 It should be emphasized that, unlike the methodology used in this study, the measuring method employed in these investigations made use of computed tomography and cross-sections of human maxillas, which may explain the different findings.

In this study, septum width was defined as the distance between the inner boundaries of the lamina dura of the roots adjacent to the mini-implants, not including the periodontal ligament space, as it is not advisable to encroach upon it during mini-implant insertion. However, some studies use the cementum as boundary6,9,28 and include the periodontal ligament space in their measurements, which inevitably yields values that are different from those attained in this study.

As yet, the literature has not reached a consensus on the minimum distance required between mini-implants and tooth roots. Most studies merely speculate on the ideal "safety margin," but fail to show accurate values for such distance. Only a small number of studies have examined the proximity of mini-implants to tooth roots and the influence of such proximity on the stability of this anchorage system.2,6,16

Poggio et al9 and Schnelle et al30 recommend a minimum distance of 1 mm between mini-implant and adjacent root. In fact, this distance reduces the potential for injury to tooth roots, but the distance between the roots of maxillary second premolars and maxillary first molars should be at least 3.5 mm to accommodate a mini-implant with 1.5 mm length. This distance is seldom found in a more coronal region of the septum, where an increased amount of attached gingiva is present.

Accurate assessment of mini-implant stability is usually reported only in animal experiments.19 Few authors report a more categorical assessment of mini-implant mobility in humans,14 and most of them only check for the presence or absence of mobility with clinical tweezers.5,11 Furthermore, this assessment is never evaluated monthly but only a few months after mini-implant installation.

The results of this study demonstrate that mini-implants in groups 1 and 2 had a similar degree of mobility (Table 5) and no association was observed between the success rate of mini-implants and septum width at the insertion site (Table 6). Corroborating these results, Motoyoshi et al6 also found no differences when evaluating a sample of standardized mini-implants. The accuracy of mini-implant insertion found in the present study—even in septa with critical dimensions—was due to the use of GRSG. In the study by Motoyoshi et al6 no guide was used, however, the authors achieved a certain degree of precision with their surgical technique using CT scans to define the ideal insertion height.

The total success rate found in the sample was 90% (Table 7), similar to the work of Motoyoshi et al,6 who also used Absoanchor® mini-implants. An analysis of the factors associated with mini-implant failure showed that none of the variables assessed in this study has any bearing on the success rate of mini-implants.

The similarity found between the success rate of mini-implants placed in different insertion sites had no precedent in previous studies, in which the absence of keratinized mucosa significantly increased the risk of infection and mini-implant failure.11 However, all mini-implants in this study were included in a more coronal region of the septum, i.e., in a region of alveolar mucosa near the mucogingival junction. Since in the studies cited above the authors provide no information regarding mini-implant insertion height, one can speculate that in these samples some mini-implants may have been placed in apical regions where the cortical bone is thinner and proximity to frenum areas facilitate the process of inflammation and tissue hyperplasia around the mini-implant.3 Additionally, since few mini-implants were lost (4 mini-implants in the failure group), the odds of finding significant values to support comparison analyses and thereby determine risk factors with reliability are reduced.

Generally, studies only assess the quality of patient toothbrushing5,11 and/or the presence of inflammation around the implant.3,5,11 The results showed no significant difference between the plaque index found in successful and failed mini-implants (Table 7). Inflammation around the mini-implant was not evaluated since it was hardly noticed during treatment given that patients were constantly instructed on oral hygiene. This fact is reflected in the mean value of MPI (1.52±0.70, Table 7), which demonstrates that most of the mini-implants did not show abundant plaque.

During the monthly evaluation of the mini-implants it was noted that sensitivity was nonexistent in devices with no mobility. However, degree of sensitivity increased significantly as mini-implants lost stability (Table 7). This sensitivity is probably related to compression of the surrounding soft tissues caused by the movement of mini-implants with a high degree of mobility since in general there is no such thing as spontaneous pain, but only pain caused by movement. Moreover, this traumatic condition favors peri-implant tissue inflammation, progressively increasing sensitivity in the region.

Table 7 shows that evaluation of failed mini-implants lasted approximately 1.75 months, and successful mini-implants were assessed for almost 10 months, which is consistent with the findings of Moon et al7 and Cheng et al.11 A somewhat different result was found by Park, Jeong and Kwon5 (3.40 months). It is noteworthy that, in most studies, mini-implant failure occurs almost immediately after installation. This short-term failure is therefore directly related to the primary stability of mini-implants.10

Even when measured in isolation, ISW appeared not to be associated with failures in this anchorage system (Table 7). It was observed that the widths of interradicular septa with critical dimensions are not considered a risk factor for the stability and success of self-drilling mini-implants. It is speculated that this lack of correlation between septum width and mini-implant success rate is directly linked to the use of three-dimensional radiographic-surgical guides, which enabled highly accurate and safe mini-implant insertion.21,22

Given the small number of studies using similar methodologies—since it was only recently that a surge of interest in these risk factors began to occur—further studies are warranted if the relevant literature is to reach any lasting agreement.

CONCLUSIONS

Based on the results attained for the study sample and in accordance with the methodology applied, it was concluded that:

1. No statistically significant difference was found in the degree of mobility and success rate of self-drilling mini-implants placed in septa with critical (≤3 mm) and noncritical (>3 mm) mesiodistal width.

2. None of the variables studied proved to be related to mini-implant failure. However, greater sensitivity was noted in patients whose mini-implants exhibited some degree of mobility, with failure of these devices occurring shortly after insertion.

References

  • 1. Janssen KI, Raghoebar GM, Vissink A, Sandham A. Skeletal anchorage in orthodontics: a review of various systems in animal and human studies. Int J Oral Maxillofac Implants. 2008 Jan-Feb;23(1):75-88.
  • 2. Kuroda S, Yamada K, Deguchi T, Hashimoto T, Kyung HM, Takano-Yamamoto T. Root proximity is a major factor for screw failure in orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S68-73.
  • 3. Miyawaki S, Koyama I, Inoue M, Mishima K, Sugahara T, Takano-Yamamoto T. Factors associated with the stability of titanium screws placed in the posterior region for orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2003 Oct;124(4):373-8.
  • 4. Motoyoshi M, Hirabayashi M, Uemura M, Shimizu N. Recommended placement torque when tightening an orthodontic mini-implant. Clin Oral Implants Res. 2006 Feb;17(1):109-14.
  • 5. Park HS, Jeong SH, Kwon OW. Factors affecting the clinical success of screw implants used as orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2006 Jul;130(1):18-25.
  • 6. Motoyoshi M, Yoshida T, Ono A, Shimizu N. Effect of cortical bone thickness and implant placement torque on stability of orthodontic mini-implants. Int J Oral Maxillofac Implants. 2007 Sep-Oct;22(5):779-84.
  • 7. Moon CH, Lee DG, Lee HS, Im JS, Baek SH. Factors associated with the success rate of orthodontic miniscrews placed in the upper and lower posterior buccal region. Angle Orthod. 2008 Jan;78(1):101-6.
  • 8. Deguchi T, Nasu M, Murakami K, Yabuuchi T, Kamioka H, Takano-Yamamoto T. Quantitative evaluation of cortical bone thickness with computed tomographic scanning for orthodontic implants. Am J Orthod Dentofacial Orthop. 2006 Jun;129(6):721.e7-12.
  • 9. Poggio PM, Incorvati C, Velo S, Carano A. "Safe zones": a guide for miniscrew positioning in the maxillary and mandibular arch. Angle Orthod. 2006 Mar;76(2):191-7.
  • 10. Wilmes B, Rademacher C, Olthoff G, Drescher D. Parameters affecting primary stability of orthodontic mini-implants. J Orofac Orthop. 2006 May;67(3):162-74.
  • 11. Cheng SJ, Tseng IY, Lee JJ, Kok SH. A prospective study of the risk factors associated with failure of mini-implants used for orthodontic anchorage. Int J Oral Maxillofac Implants. 2004 Jan-Feb;19(1):100-6.
  • 12. Wilmes B, Su YY, Drescher D. Insertion angle impact on primary stability of orthodontic mini-implants. Angle Orthod. 2008 Nov;78(6):1065-70.
  • 13. Chun YS, Lim WH. Bone density at interradicular sites: implications for orthodontic mini-implant placement. Orthod Craniofac Res. 2009 Feb;12(1):25-32.
  • 14. Liou EJ, Pai BC, Lin JC. Do miniscrews remain stationary under orthodontic forces? Am J Orthod Dentofacial Orthop. 2004 Jul;126(1):42-7.
  • 15. Chen YH, Chang HH, Chen YJ, Lee D, Chiang HH, Yao CC. Root contact during insertion of miniscrews for orthodontic anchorage increases the failure rate: an animal study. Clin Oral Implants Res. 2008 Jan;19(1):99-106.
  • 16. Asscherickx K, Van de Vannet B, Wehrbein H, Sabzevar MM. Success rate of miniscrews relative to their position to adjacent roots. Eur J Orthod. 2008 Aug;30(4):330-5.
  • 17. Kang YG, Kim JY, Lee YJ, Chung KR, Park YG. Stability of mini-screws invading dental roots and their impact on the paradental tissues in Beagles. Angle Orthod. 2009 Mar;79(2):248-55.
  • 18. Heidemann W, Terheyden H, Louis GK. Analysis of the osseous/metal interface of drill free screws and self-tapping screws. J Maxillofac Surg. 2001 Apr;29(2):69-74.
  • 19. Kim JW, Ahn SJ, Chang YI. Histomorphometric and mechanical analyses of the drill-free screw as orthodontic anchorage. Am J Orthod Dentofacial Orthop. 2005 Aug;128(2):190-4.
  • 20. Barros SEC, Janson G, Chiqueto K, Freitas MR, Henriques JF, Pinzan A. A three-dimensional radiographic-surgical guide for mini-implant placement. J Clin Orthod. 2006 Sep;40(9):548-54.
  • 21. Barros SEC, Janson G, Chiqueto K, Janson M, Freitas MR. Predictable drill-free screw positioning with a graduated 3-D radiographic-surgical guide: a preliminary report. Am J Orthod Dentofacial Orthop. 2009 Nov;136(5):722-35.
  • 22. Barros SEC. Avaliação da precisão de um guia radiográfico-cirúrgico para inserção de mini-implantes [dissertação]. Bauru (SP): Universidade de São Paulo; 2008.
  • 23. Mombelli A, Van Oosten MA, Schurch E Jr, Land NP. The microbiota associated with successful or failing osseointegrated titanium implants. Oral Microbiol Immunol. 1987 Dec;2(4):145-51.
  • 24. Dahlberg G. Statistical methods for medical and biological students. New York: Interscience; 1940.
  • 25. Houston WJ. The analysis of errors in orthodontic measurements. Am J Orthod. 1983 May;83(5):382-90.
  • 26. Padovani CR. Estatística na metodologia da investigação científica. Botucatu (SP): UNESP; 1995.
  • 27. Kravitz ND, Kusnoto B. Risks and complications of orthodontic miniscrews. Am J Orthod Dentofacial Orthop. 2007 Apr;131(4 Suppl):S43-51.
  • 28. Hernandez LC, Montoto G, Puente Rodriguez M, Galban L, Martinez V. Bone map for a safe placement of miniscrews generated by computed tomography. Clin Oral Implants Res. 2008 Jun;19(6):576-81.
  • 29. Hu KS, Kang MK, Kim TW, Kim KH, Kim HJ. Relationships between dental roots and surrounding tissues for orthodontic miniscrew installation. Angle Orthod. 2009 Jan;79(1):37-45.
  • 30. Schnelle MA, Beck FM, Jaynes RM, Huja SS. A radiographic evaluation of the availability of bone for placement of miniscrews. Angle Orthod. 2004 Dec;74(6):832-7.
  • Endereço para correspondência

    Mariana Pracucio Gigliotti
    Rua José Lúcio de Carvalho, 558 Centro
    CEP: 17.201-150 - Jaú / SP
    E-mail:
  • Publication Dates

    • Publication in this collection
      18 July 2011
    • Date of issue
      Apr 2011

    History

    • Received
      June 2009
    • Accepted
      Feb 2010
    Dental Press International Av. Luís Teixeira Mendes, 2712 , 87015-001 - Maringá - PR, Tel: (55 44) 3033-9818 - Maringá - PR - Brazil
    E-mail: artigos@dentalpress.com.br