Light-Assisted Batch Injection Analysis of Glucose Exploiting a p-n-Homojunction Based on Cu2O

Abstract

This work describes the development of a photoelectrochemical sensor for glucose quantification exploiting a light assisted-batch injection analysis (BIA) cell. A light-emitting diode (LED) lamp was employed to control the incidence of light on the p-Cu2O/n-Cu2O/fluorine-doped tin oxide (FTO) photoactive platform in BIA cell. The p-Cu2O/n-Cu2O/FTO platform was constructed by electrodepositing n-Cu2O and p-Cu2O and the characteristics of electrodeposited Cu2O films were investigated by X-ray diffraction (XRD), Raman spectroscopy, and electrochemical impedance spectroscopy. The light assisted-batch injection analysis of glucose based on the illuminated p-Cu2O/n-Cu2O/FTO photoelectrode presented a linear response of 10 μmol L-1-1 mmol L-1, a limit of detection of 4.0 μmol L-1, and sensitivity of 0.768 ± 0.011 μA L μmol-1 cm-2. The system presented an average recovery value of 96% when applied to the determination of glucose in an artificial saliva sample, which indicates that the incidence of light on photoelectroactive platforms is a promising approach for the determination and quantification of glucose.

Keywords:
batch injection analysis; copper(II) oxide; glucose; LED light


Introduction

Glucose is an essential monosaccharide for the proper functioning of living organisms.11 Galant, A. L.; Kaufman, R. C.; Wilson, J. D.; Food Chem. 2015, 188, 149. Nevertheless, high levels of glucose in the blood can result in the development of diabetes.22 American Diabetes Association; Diabetes Care 2014, 37, S81.,33 Chen, S.; Qian, D.; Burström, K.; Burström, B.; Patient Educ. Couns. 2006, 103, 1767. Diabetes is one of the most expressive causes of mortality nowadays, and it is linked to the development of several pathologies.44 Lin, K.-D.; Hsu, C.-C.; Ou, H.-Y.; Wang, C.-Y.; Chin, M.-C.; Shin, S.-J.; JFMA 2019, 118, S103. In this sense, the development of sensitive and reliable glucose detection devices is of high importance, and different approaches have been developed with this aim, including methods based on chromatography,55 Xie, W.-Q.; Gong, Y.-X.; You, K.-X.; J. Chromatogr. A 2017, 150, 143. colorimetry,66 Chen, X.; Chen, J.; Wang, F.; Xiang, X.; Luo, M.; Ji, X.; He, Z.; Biosens. Bioelectron. 2012, 35, 363. spectroscopy77 Hull, E. L.; Metter, N. I.; Olson, B. P.; Ediger, M. N.; Magee, A. J.; Way, J. F.; Vugrin, K. E.; Maynard, J. D.; J. Clin. Transl. Endocrinol. 2014, 1, 92. and sensors.88 Meng, W.; Wen, Y.; Dai, L.; He, Z.; Wang, L.; Sens. Actuators, B 2018, 260, 852.

9 Xu, F.; Wu, M.; Ma, G.; Xu, H.; Shang, W.; Microchem. J. 2020, 159, 105432.
-1010 Li, J.; Chen, Q.; Bai, J.; Zhou, B.; J. Hazard Mater. 2013, 262, 304.

Nowadays the development of enzymatic and non enzymatic sensors for glucose detection have received high attention due to its exquisite properties. On the one hand, enzymatic glucose sensors are mainly based on the immobilization of the enzyme glucose oxidase (GOx) on various substrates and currently they play an important role in the detection of blood glucose due to its high selectivity and sensitivity.1111 Liu, G.-Q.; Zhong, H.; Li, X.-R.; Yang, K.; Jia, F.-F.; Cheng, Z.-P.; Zhang, L.-L.; Yin, J.-Z.; Guo, L.-P.; Qian, H.-Y.; Sens. Actuators, B 2017, 242, 484.,1212 Mahmoud, A.; Echabaane, C.; Omri, K.; Mir, L. E.; Chaabane, R. B.; J. Alloys Compd. 2019, 786, 960. On the other hand, non-enzymatic sensors for glucose are based on the direct electro-oxidation of glucose on the surface of a catalyst. Therefore, non-enzymatic sensors for glucose do not require the use of enzymes which contribute to the stability, simplicity and reproducibility of the platforms developed for glucose detection.1313 Tian, K.; Prestgard, M.; Tiwari, A.; Mater. Sci. Eng., C 2014, 41, 100. Among the several classes of sensors recently proposed for glucose determination, the electrochemical methods appear as an inexpensive, fast, accurate, sensitive, and portable, being a low-cost and straightforward alternative to measure glucose levels in different types of samples.1414 Sehit, E.; Altintas, Z.; Biosens. Bioelectron. 2020, 159, 112165.

In the last decade, the photoelectrochemical effect in semiconductor materials has been widely explored as a transduction process for detecting and quantifying several species,1515 Zhang, Z.-X.; Zhao, C.-Z.; Chin. J. Anal. Chem. 2013, 41, 436. presenting one of the highest sensitivity among the methods based on electrochemical detection. The photoelectrochemical systems have opened new perspectives in sensor area since they can operate with low noise and background current due to the separation between the excitation light source and the photocurrent detector.1616 Devadoss, A.; Sudhagar, P.; Terashima, C.; Nakata, K.; Fujishima, A.; J. Photochem. Photobiol., C 2015, 24, 43.

Pioneering works in the development of photo­electrochemical platforms have opened new perspectives in determination of glucose, including photoelectrochemical enzymatic systems,1717 Zhao, W.-W.; Xu, J.-J.; Chen, H.-Y.; Biosens. Bioelectron. 2017, 92, 294. photoelectrochemical non-enzymatic systems,1818 Wang, S.; Li, S.; Wang, W.; Zhao, M.; Liu, J.; Feng, H.; Chen, Y.; Gu, Q.; Du, Y.; Hao, W.; Sens. Actuators, B 2019, 291, 34. photoelectrochemical systems based on flow injection analysis,1919 Saglam, O.; Dilgin, Y.; Electroanalysis 2017, 29, 1368. and recently photoelectrochemical systems based on batch injection analysis.2020 Soares, L. C.; Santos, C. C.; Luz, R. C. S.; Damos, F. S.; Electroanalysis 2020, 32, 1608.

In addition, several photoelectroactive materials, such as semiconducting-metal-oxide-based materials, because of their interesting electrical and optical properties, have been exploited, being suitable candidates for the construction of photoelectrochemical sensors.2020 Soares, L. C.; Santos, C. C.; Luz, R. C. S.; Damos, F. S.; Electroanalysis 2020, 32, 1608. Among these materials, copper(I) oxide (Cu2O) has shown an impressive electrochemical catalytic activity for glucose oxidation.2121 Wang, M.; Song, X.; Song, B.; Liu, J.; Hu, C.; Wei, D.; Wong, C.-P.; Sens. Actuators, B 2017, 250, 333.

22 Lin, L.-Y.; Karakocak, B. B.; Kavadiya, S.; Soundappan, T.; Biswas, P.; Sens. Actuators, B 2018, 259, 745.

23 Li, M.; Wang, H.; Wang, X.; Lu, Q.; Li, H.; Zhang, Y.; Yao, S.; Biosens. Bioelectron. 2019, 142, 111535.

24 Praveen, R.; Ramaraj, R.; J. Electroanal. Chem. 2019, 851, 113454.
-2525 Jayasingha, L.; Jayanthilaka, C.; Kumara, R.; Ohara, K.; Kaumal, M.; Gunewardene, S.; Dissanayake, D.; Jayanetti, S.; Electrochim. Acta 2020, 329, 135177.

Cuprous(I) oxide (Cu2O) is commonly a p-type semiconductor, and it presents a direct bandgap of around 2 eV. In addition, Cu2O shows a high absorption coefficient, abundance on Earth, non-toxic behavior and low-cost manufacturing.2626 Xiong, L.; Huang, S.; Yang, X.; Qiu, M.; Chen, Z.; Yu, Y.; Electrochim. Acta 2011, 56, 2735. Although Cu2O is naturally a p-type semiconductor, due to copper vacancies in its structure, it is possible to obtain Cu2O as an n-type semiconductor, which occurs due to oxygen deficiency in its crystalline network.2727 Xu, M.; Liu, X.; Xu, W.; Hao, X.; Feng, X.; J. Alloys Compd. 2018, 769, 484.

The properties of cuprous oxide-based materials can be intensified by exploiting the homojunction based on p-type (p-Cu2O) and n-type (n-Cu2O) films. Considering that the electronic affinity of the cuprous oxides p-Cu2O and n-Cu2O are approximately equal, junctions based on these materials can decrease the discontinuity of the conduction band along the interface. Therefore, it is possible to form a high-quality interface through a homojunction p-n or n-p, considering that the position of the conduction and valence bands of p-Cu2O and n-Cu2O are close.2828 Wan, T.; Wei, Y.; Chang, X.; Li, C.; Li, A.; Liu, S.; Zhang, J.; Gong, J.; Appl. Catal., B 2018, 226, 31.

The search for fast, sensitive, simple, reproductive, versatile analytical methods motivated research groups to develop methods based on self-sampling,2020 Soares, L. C.; Santos, C. C.; Luz, R. C. S.; Damos, F. S.; Electroanalysis 2020, 32, 1608. such as batch injection analysis, for example, which shows attractive features, including high throughput analysis,2929 Gimenes, D. T.; Marra, M. C.; Freitas, J. M.; Muñoz, R. A. A.; Richter, E. M.; Sens. Actuators, B 2015, 212, 411. easy handling system,3030 Freitas, J. M.; Oliveira, T. C.; Gimenes, D. T.; Muñoz, R. A. A.; Richter, E. M.; Talanta 2016, 146, 670. precision,3131 Tormin, T. F.; Narciso, L. C. D.; Richter, E. M.; Munoz, R. A. A.; Electrochim. Acta 2015, 164, 90. accuracy,3232 Stefano, J. S.; Tormin, T. F.; da Silva, J. P.; Richter, E. M.; Munoz, R. A. A.; Microchem. J. 2017, 133, 398. use of small sample volumes, high sensitivity, low cost, simplicity and portability.3333 Pereira, P. F.; da Silva, W. P.; Munoz, R. A. A.; Richter, E. M.; J. Electroanal. Chem. 2016, 766, 87.

Therefore, the combined use of batch injection analysis to Cu2O based photoelectrochemical platforms can provide improvement of sensitivity, precision, detectability, and improve the throughput analysis. This work aims to develop, at first time, a photoelectrochemical assisted batch-injection analysis system based on a p-Cu2O/n-Cu2O/fluorine-doped tin oxide (FTO) photoelectrode for glucose determination.

Experimental

Reagents and materials

All chemicals used in this study were analytical grade, and they were used without further purification steps. Copper sulfate (CuSO4), tartaric acid (C4H6O6), copper acetate monohydrate (Cu(CH3COO)2.H2O), acetic acid (CH3COOH), glucose (C12H22O11), sodium chloride (NaCl), potassium chloride (KCl), calcium chloride dihydrate (CaCl2.2H2O), sodium bicarbonate (NaHCO3), sodium hydroxide (NaOH), sodium sulfate (Na2SO4) and ethyl alcohol (C2H6O) were purchased from ISOFAR (Duque de Caxias, Rio de Janeiro, Brazil). Glass plates with a surface covered with a transparent layer of fluorine-doped tin oxide (FTO), and potassium ferricyanide (K3[Fe(CN)6]) were purchased from Sigma-Aldrich (St. Louis, USA). The working solutions were prepared daily with water purified in an OS100LXE system from GEHAKA Company (São Paulo, Brazil).

Construction of the p-Cu2O/n-Cu2O/FTO photoelectro­chemical sensor

Initially, the glass slides coated with a layer of fluorine-doped tin oxide (FTO) were extensively rinsed with water and ethanol to remove any previously adsorbed species from the FTO surface. After that, n-type and p-type Cu2O films were deposited on the FTO surface. The electrodeposition of n-type Cu2O film was performed from bath solutions containing 0.02 mol L-1 Cu(CH3COO)2.H2O and 0.08 mol L-1 CH3COOH. The pH of the medium was adjusted to 4.9 with NaOH solution, and the FTO working electrode was biased at +0.02 V vs. Ag/AgCl for 1200 s at a temperature of about 70 ºC.3434 Mcshane, C. M.; Choi, K.-S.; Phys. Chem. Chem. Phys. 2012, 14, 6112. On the other hand, p-type Cu2O films were electrodeposited from a plating solution containing 0.1 mol L-1 CuSO4 and 0.1 mol L-1 tartaric acid. After adjusting the pH to 11 with NaOH solution, the temperature of the electrodeposition bath was maintained at about 60 °C with aid of a hotplate equipped with ceramic surface from Kasvi (São José dos Pinhais, Brazil) and monitored with a thermometer, and the electrochemical platform was biased at -0.4 V vs. Ag/AgCl during 300 s to produce the p-Cu2O modified photoelectrodes.3434 Mcshane, C. M.; Choi, K.-S.; Phys. Chem. Chem. Phys. 2012, 14, 6112.,3535 Nishikawa, M.; Fukuda, M.; Nakabayashi, Y.; Saito, N.; Appl. Surf. Sci. 2016, 363, 173.

Raman, X-ray diffraction and (photo)electrochemical measurements

Raman measurements were performed with a Horiba‐Jobin‐Yvon triple spectrometer, model T64000 (Kyoto, Japan), operating in the single-mode and equipped with a liquid‐N2‐cooled charge‐coupled device detector. A green 532 nm line from a solid-state laser (LAS 532 100 HREV) operating at 14 mW was employed for excitation, and the slit was adjusted to give a resolution of 2 cm-1. An Olympus microscope with an MPLN 50× objective lens was employed to adjust the light beam on the photoelectrochemical platforms. The intensity of the laser on the photoelectrochemical platform was adjusted with an ND 0.6 optical density filter (25% transmission). The spectra of the samples were obtained after 5 acquisitions of 30 s each at each range of the grade spectral dispersion.

The X-ray diffractograms (XRD) were performed with a Bruker D8 Advance diffractometer (Bruker AXS, Karlsruhe, Germany), equipped with the LynxEye linear detector, using Cu Kα (λ = 1.5406 Å) operating at 40 kV and a current of 40 mA. The X-ray diffractograms were performed in a 2θ ranging from 25-75º with a counting time of 0.5 s, step size of 0.02° and Bragg-Brentano geometry.

The photocurrent measurements were performed with a potentiostat/galvanostat model PGSTAT 128N from Metrohm Autolab BV (Utrecht, Netherlands). A light assisted batch injection analysis cell with a photoelectrochemical platform as working electrode, a platinum wire as counter electrode and an Ag/AgCl(sat) as reference electrode was employed for photoelectrochemical measurements. The batch injection analysis cell is composed of the glass cell and a cover with three orifices for the reference electrode, counter-electrode and micropipette, and a support for the positioning of the working electrode (Figure 1).

Figure 1
Photoelectrochemical batch injection analysis (PEC-BIA) cell for the electrochemical determination of glucose. (a) Auxiliary-electrode; (b) reference electrode; (c) micropipette tip; (d) photoelectrode; (e) LED lamp; (f) glass cell; and (g) 3D-printed cell cover.

The cell body was made of glass (5 mm thick), with dimensions of 7 cm × 6 cm. The 3D printed cover of the cell, the electrode holder and the injector holder were printed in a MakerBot Mini Plus 3D printer (Makerbot, Boston MA, USA) using the poly(acetic acid) biodegradable (PLA) filament as printing material while maintaining the extruder temperature at about 215 °C. The injections of the glucose solutions were performed with aid of an electronic micropipette model Pipetman from Gilson (Middleton, USA) operating with a fixed dispensing rate of 6 μL s-1 and an injection volume of 25 μL.

The p-Cu2O/n-Cu2O/FTO platform was employed for glucose determination in artificial saliva samples (Ringer’s solution) in order to evaluate the applicability of the light-assisted photoelectrochemical batch injection analysis (PEC-BIA) system.3636 Rinaldi, A. L.; Rodríguez-Castellón, E.; Sobral, S.; Carballo, R.; J. Electroanal. Chem. 2019, 832, 209. The artificial saliva samples (Ringer’s solution) were prepared in water with the following constituents and final concentrations: 9.0 g L-1 NaCl (or ca. 0.150 mol L-1); 0.4 g L-1 KCl (or ca. 0.005 mol L-1); 0.2 g L-1 CaCl2.6H2O (or ca. 0.001 mol L-1), and 0.2 g L-1 NaHCO3 (or 0.002 mol L-1). The medium’s pH was adjusted to 7.4 with 10% (v/v) HCl solution.3636 Rinaldi, A. L.; Rodríguez-Castellón, E.; Sobral, S.; Carballo, R.; J. Electroanal. Chem. 2019, 832, 209. After preparation, the solution was stored under refrigeration (ca. 4 °C). The artificial saliva samples were diluted to 10% (v/v) in 0.1 mol L-1 NaOH electrolyte and fortified with appropriate amounts of glucose before the photoelectrochemically-assisted batch injection analysis (PEC-BIA) measurements.

The electrochemical impedance spectroscopy (EIS) measurements were carried out in order to evaluate the electrochemical properties of the photoelectrodes. The EIS measurements were performed with the aid of an Autolab PGSTAT 128N Potentiostat/Galvanostat (Metrohm Autolab B. V. Utrecht, Netherlands) equipped with an FRA 2 module controlled by FRA software (version 4.9). The Mott‐Schottky (M-S) analysis and Bode‐phase measurements were performed in 0.1 mol L-1 sodium sulfate. The effect of the light on the electrochemical behavior of the photoelectrochemical platforms was evaluated in 0.1 mol L-1 NaOH containing 1000 µmol L-1 of glucose. The Nyquist and Bode plots were recorded from 10-1 to 105 Hz under an AC amplitude of 0.01 V. The M‐S measurements were performed changing the applied potential to the photoelectrode, maintaining the frequency constant. The electrochemical characterization of the photoelectrochemical platforms was also performed by cyclic voltammetry.

Results and Discussion

Raman and XRD characterization of the photoelectrochemical platforms

Figure 2 shows the Raman spectra for electrodeposited n-Cu2O/FTO, p-Cu2O/FTO, and p-Cu2O/n-Cu2O/FTO. As can be seen, similar spectra were obtained for all samples. According to the literature,3737 Ha, T.; Park, I.; Sim, K. I.; Lee, H.; Bae, J.-S.; Kim, S. J.; Kim, J. P.; Kim, T.-T.; Kim, J. H.; Jang, J. I.; Jeong, S.-Y.; APL Mater. 2019, 7, 031115.

38 Sahaia, A.; Goswamia, N.; Kaushikb, S. D.; Tripathic, S.; Appl. Surf. Sci. 2016, 390, 974.

39 Huang, Q.; Li, J.; Bi, X.; J. Alloys Compd. 2015, 647, 585.

40 Huang, M.-C.; Wang, T. H.; Chang, W.-S.; Lin, J.-C.; Wu, C.-C.; Chen, I.-C.; Peng, K.-C.; Lee, S.-W.; Appl. Surf. Sci. 2014, 301, 369.

41 Mao, Y.; He, J.; Sun, X.; Li, W.; Lu, X.; Gan, J.; Liu, Z.; Gong, L.; Chen, J.; Liu, P.; Tong, Y.; Electrochim. Acta 2012, 62, 1.
-4242 Powell, D.; Compaan, A.; Macdonald, J. R.; Phys. Rev. B 1975, 12, 20. all the observed phonon modes are characteristic of the crystalline Cu2O, with cubic symmetry belonging to the space group (Oh4). Although just one Raman-active mode with symmetry Γ25+ is expected in the Raman spectrum for this structure, infrared (with symmetries Γ15-(1) and Γ15-(2)) and selection-ruler-forbidden modes, such as the modes with symmetries Γ12- and Γ25-, can also be allowed due to local defects and intrinsic selection rule violation mechanism, which explains the presence of the extra modes. Therefore, we attributed the peak at about 149 cm-1 as being from an infrared-allowed phonon mode with symmetry Γ15-(1), the peak at 198 cm-1 as due to a two-phonon combination mode of Γ12- and Γ25-, and the intense line at about 218 cm-1 from the second-order overtone of the phonon with symmetry Γ12- (2Γ12-).3737 Ha, T.; Park, I.; Sim, K. I.; Lee, H.; Bae, J.-S.; Kim, S. J.; Kim, J. P.; Kim, T.-T.; Kim, J. H.; Jang, J. I.; Jeong, S.-Y.; APL Mater. 2019, 7, 031115.

38 Sahaia, A.; Goswamia, N.; Kaushikb, S. D.; Tripathic, S.; Appl. Surf. Sci. 2016, 390, 974.

39 Huang, Q.; Li, J.; Bi, X.; J. Alloys Compd. 2015, 647, 585.
-4040 Huang, M.-C.; Wang, T. H.; Chang, W.-S.; Lin, J.-C.; Wu, C.-C.; Chen, I.-C.; Peng, K.-C.; Lee, S.-W.; Appl. Surf. Sci. 2014, 301, 369.

Figure 2
Raman spectra for the n-Cu2O/FTO (a), p-Cu2O/FTO (b) and p-Cu2O/n-Cu2O/FTO (c) platforms.

In addition, the weak peaks at about 308 cm-1 correspond to the second-order overtone mode 2Γ15-(1),3434 Mcshane, C. M.; Choi, K.-S.; Phys. Chem. Chem. Phys. 2012, 14, 6112. the peak at 414 cm-1 is assigned to a four-phonon mode 3Γ12- + Γ25-,4141 Mao, Y.; He, J.; Sun, X.; Li, W.; Lu, X.; Gan, J.; Liu, Z.; Gong, L.; Chen, J.; Liu, P.; Tong, Y.; Electrochim. Acta 2012, 62, 1. the peak at about 516 cm-1 is the Raman-active mode (Γ25+),4141 Mao, Y.; He, J.; Sun, X.; Li, W.; Lu, X.; Gan, J.; Liu, Z.; Gong, L.; Chen, J.; Liu, P.; Tong, Y.; Electrochim. Acta 2012, 62, 1.,4242 Powell, D.; Compaan, A.; Macdonald, J. R.; Phys. Rev. B 1975, 12, 20. and the signal at 626 cm-1 is attributed to the infrared-allowed mode Γ15-(2).4141 Mao, Y.; He, J.; Sun, X.; Li, W.; Lu, X.; Gan, J.; Liu, Z.; Gong, L.; Chen, J.; Liu, P.; Tong, Y.; Electrochim. Acta 2012, 62, 1.

The X-ray diffraction patterns of all the Cu2O samples electrodeposited on FTO substrate presented peaks from the cubic (space group, PDF:00-005-0667) structure of the Cu2O as well as peaks of the FTO substrate (Figure 3). The XRD patterns presented peaks (2θ) at 29.6°, 36.5°, 42.4°, 61.4°, and 73.6° indicative of the (110), (111), (200), (220) and (311) planes, respectively, which correspond to the cubic structure of Cu2O. In addition, the intensity of the peak (111) is higher than the others indicating a preferential growth of the Cu2O films along the (111) direction.4040 Huang, M.-C.; Wang, T. H.; Chang, W.-S.; Lin, J.-C.; Wu, C.-C.; Chen, I.-C.; Peng, K.-C.; Lee, S.-W.; Appl. Surf. Sci. 2014, 301, 369.

41 Mao, Y.; He, J.; Sun, X.; Li, W.; Lu, X.; Gan, J.; Liu, Z.; Gong, L.; Chen, J.; Liu, P.; Tong, Y.; Electrochim. Acta 2012, 62, 1.

42 Powell, D.; Compaan, A.; Macdonald, J. R.; Phys. Rev. B 1975, 12, 20.
-4343 Hossain, M. A.; Al-Gaashani, R.; Hamoudi, H.; Marri, M. J. A.; Hussein, I. A.; Belaidi, A.; Merzougui, B. A.; Alharbi, F. H.; Tabet, N.; Mater. Sci. Semicond. Process. 2017, 63, 203.

Figure 3
X-ray diffraction of the n-Cu2O/FTO (black line), p-Cu2O/FTO (red line) and p-Cu2O/n-Cu2O/FTO (blue line) platforms.

Electrochemical impedance spectroscopy investigation of the photoelectrochemical platforms

Electrochemical impedance spectroscopy measurements were performed to evaluate the effects of the type of Cu2O semiconductor on the electrochemical properties of the platform. Figure 4 shows the Nyquist plots obtained for n-Cu2O/FTO (Figure 4a), p-Cu2O/FTO (Figure 4b), and p-Cu2O/n-Cu2O/FTO (Figure 4c) in 0.1 mol L-1 NaOH containing 1 mmol L-1 of glucose under dark (blue circle) and light (red square) conditions.

Figure 4
Nyquist plots for n-Cu2O/FTO (a), p-Cu2O/FTO (b) and p-Cu2O/n-Cu2O/FTO (c) in 0.1 mol L-1 NaOH aqueous solution containing 1 mmol L-1 of glucose under dark (blue circles) and light (red square) conditions. The red full lines represent the Nyquist semicircle fit.

As can be seen in Figure 4, the p-Cu2O/n-Cu2O/FTO presented a semicircle diameter slightly lower than that observed to the n-Cu2O/FTO, and p-Cu2O/FTO photoelectrodes. In addition, the Figure 4 shows the effects of the incidence of light on the p-Cu2O/FTO/n-Cu2O/FTO platform. The Nyquist plots of p-Cu2O/n-Cu2O/FTO platform were affected by the incidence of light indicating that the harvested photons can promote electrons from valence band to conduction band of the semiconductor, which can contribute to a lower charge transfer resistance at the photoelectrode-electrolyte interface. The lower diameter semicircle observed for p-Cu2O/FTO/n-Cu2O/FTO under the incidence of light can be attributed to the more favorable electron transfer from the analyte to the holes in the p-Cu2O/FTO/n-Cu2O/FTO platform due to a better band alignment formed in the homojunction which facilitates the action of glucose as a hole scavenger. The electrons can diffuse from the n-Cu2O film to the p-Cu2O probably due to the higher Fermi level of the n-type film than the Fermi level of the p-Cu2O film. Similarly, the photogenerated holes can diffuse from the p-type film to the n-type film. Thus, the flow of charges carriers between the films can establish an internal electrical field at the p-n interface. The internal potential developed in the interface can improve the charge separation and charge flow at the interface, which can also contribute to a low charge resistance.

Mott-Schottky plots were performed in 0.1 mol L-1 Na2SO4 in order to evaluate the semiconductor type (p-type or n-type) and the flat-band potential (Efb) (Figure 5). As can be seen from Figure 5a, the slope of the Mott-Schottky plot for n-Cu2O/FTO platform was positive while the slope of the p-Cu2O/FTO platform was negative indicating an n-type and p-type conductivity, respectively. On the other hand, the Mott-Schottky plot of the p-Cu2O/n-Cu2O/FTO platform (Figure 5b) presented co-existence of positive and negative slopes indicating that the photoelectrochemical platform presents the behavior of a p-n-junction, which is helpful for effective separation of electrons and holes.4444 Su, T.; Yang, Y.; Dong, G.; Ye, T.; Jiang, Y.; Fan, R.; RSC Adv. 2016, 6, 65125.

Figure 5
Mott-Schottky plots (C-2 vs. E) for n-Cu2O/FTO and p-Cu2O/FTO (a) and p-Cu2O/n-Cu2O/FTO (b) were obtained in 0.1 mol L-1 Na2SO4.

The Mott-Schottky plot (CSC-2 vs. Eappl) were constructed according to the following equation:4545 Santos, G. K. C.; Silva, F. G. S.; Ferreira, A. R.; Luz, R. C. S.; Damos, F. S.; Microchem. J. 2019, 148, 424.

(1)(1CSC)-2=±2e0Na,dε0εr(Eappl-Efb-kBTe0)
where e0 is the elementary charge of the electron, Na,d is the acceptor or donor charge density, ε0 is the vacuum permittivity, εr is the relative permittivity of Cu2O, CSC is the capacitance of the space charge layer, Eappl is the applied potential, kB is the Boltzmann’s constant, T is the absolute temperature, and Efb is the flat band potential.

The flat band potentials of the n-Cu2O type and p-Cu2O type branches of Mott-Schottky plot obtained for the p-Cu2O/n-Cu2O/FTO platform were determined from the intercept of the Mott-Schottky plot under flat band condition considering that Efb=Eappl-kBTe0 and (1CSC)-2=0. The flat band potentials for n-Cu2O and p-Cu2O were about -0.269 V and +0.579 V, respectively. In addition, the Mott-Schottky plots of p-Cu2O/n-Cu2O/FTO platform presented an inverted “V-shape” as shown in the Figure 5b, indicating the formation of p-n junctions in the p-Cu2O/n-Cu2O/FTO platform.4646 Meng, F.; Li, J.; Cushing, S. K.; Zhi, M.; Wu, N.; J. Am. Chem. Soc. 2013, 135, 10286.

47 Li, J.; Meng, F.; Suri, S.; Ding, W.; Huang, F.; Wu, N.; Chem. Commun. 2012, 48, 8213.
-4848 Wen, X.-J.; Niu, C.-G.; Zhang, L.; Zeng, G.-M.; Dalton Trans. 2017, 46, 4982.

Analytical performance and application of p-Cu2O/n Cu2O/FTO photoelectrochemical platform for glucose determination in a PEC-BIA cell

In order to evaluate the analytical efficiency of the p-Cu2O/n-Cu2O/FTO photoelectrochemical platform to the glucose determination were performed successive batch injections of different concentrations of glucose with the aid of an electronic micropipette in the absence and the presence of light (Figure 6a).

Figure 6
(a) Amperometric responses obtained after injection of glucose in the batch injection analysis system in the absence (red line) and presence of light (black line). (b) Analytical curve for glucose determination under the incidence of light (black circle) and under dark conditions (red square). Dispensing rate: 6 μL s-1; injection volume: 25 μL; and applied potential: 0.6 V vs. Ag/AgCl.

As can be seen in the Figure 6b, the peak current increases linearly with the increase of the concentration of glucose concentrations ranging from 10 up to 1000 μmol L-1 under the incidence of visible light biasing the platform at 0.6 V vs. Ag/AgCl. The calibration curve obtained for glucose determination by PEC-BIA based on p-Cu2O/n-Cu2O/FTO photoelectrochemical platform presented a linear behavior according to the following equation: Ipeak / µA = 14.643 + 0.768 ([glucose] / µmol-1), for n = 7 and a correlation coefficient of 0.999.

The light-assisted determination of glucose with PEC BIA system presented a limit of detection (LOD) of about 4.0 µmol L-1 (LOD = 3sB/S, in which sB is the standard deviation of the blank and S is the slope of the calibration curve) and sensitivity of 0.768 µA L µmol1 cm-2 which is competitive when compared to some recent works (Table 1)4949 Tang, L.; Lv, J.; Kong, C.; Yang, Z.; Li, J.; New J. Chem. 2016, 40, 6573.5454 Su, Y.; Guo, H.; Wang, Z.; Long, Y.; Li, W.; Tu, Y.; Sens. Actuators, B 2018, 255, 2510. for glucose detection based on copper-based platforms. On the other hand, the sensitivity of the platform in the absence of light was 0.546 µA L µmol-1 cm-2 (Figure 6b).

Table 1
Comparison of some non-enzymatic sensors for the determination of glucose

As can be seen in the Figure 6b, the incidence of light on the p-Cu2O/n-Cu2O/FTO platform has resulted in an increase of the sensitivity of the system. The calibration curves for the determination of glucose with the proposed p-Cu2O/n-Cu2O/FTO platform shows a higher slope under the incidence of light than under dark conditions (Figure 6b). The p-Cu2O/n-Cu2O/FTO can generate electron-hole couples under the incidence of light in order to increase the sensitivity of the photoelectrochemical platform for glucose detection. Thus, the p-Cu2O/n Cu2O/FTO photoelectrochemical platform acts as a photoelectrochemical-assisted detector for the batch injection analysis since the electron-hole couples produced can improve the glucose oxidation.

Figure 7 shows a proposed scheme for the light assisted batch-injection analysis of glucose employing a p-Cu2O/n-Cu2O/FTO photoelectrochemical platform biased at 0.6 V vs. Ag/AgCl. The p-Cu2O/n-Cu2O/FTO photoelectrochemical platform harvests photons from incident light in order to produce electron-hole couples in p-Cu2O and n-Cu2O (equations 2 and 3).

Figure 7
Schematic representation of the proposed mechanism for the oxidation of glucose on p-Cu2O/n-Cu2O/FTO photoelectrode surface under illumination.

The electrons photogenerated in the conduction band of p-Cu2O are driven to the conduction band of n-Cu2O to become further dragged by the FTO platform. Simultaneously, the holes photogenerated can be harvested by the glucose molecules at the interface photoelectrode-solution, and the holes photogenerated in n-Cu2O can be transferred to the valence band of p-Cu2O. Considering that the p-Cu2O/n-Cu2O/FTO photoelectrochemical platform was biased at 0.6 V, the Cu2O at the semiconductor/electrolyte interface can be oxidized to form CuO. The photogenerated holes in CuO can then form ion pairs with OH- with sufficient energy to promote the glucose oxidation. The following reactions are proposed for glucose oxidation on the p-Cu2O/n-Cu2O/FTO photoelectrochemical platform under the incidence of light:

(2)n-Cu2O+hven-Cu2O-+hn-Cu2O+
(3)p-Cu2O+hvep-Cu2O-+hp-Cu2O+
(4)p-Cu2O+2 h+p-Cu2O+2OH-2CuO+H2O
(5)CuO+hve-(CuO)+CuO(h+(CuO))
(6)CuO(h+(CuO))+OH-CuO(h+(CuO))(OH-ads)
(7)CuO(h(CuO)+)(OHads-)+glucose H2Oads+ gluconolactone 

The repeatability of the p-Cu2O/n-Cu2O/FTO photoelectrochemical platform biased at +0.6 V vs. Ag/AgCl was evaluated performing successive photoelectrochemical-assisted batch injection analysis of 50 µmol L-1 of glucose. The relative standard deviation (RSD) of the response of the p-Cu2O/n-Cu2O/FTO photoelectrochemical platform for three successive injections of glucose was about 2.9%. On the other hand, the reproducibility of the p-Cu2O/n-Cu2O/FTO photoelectrochemical platform for detecting glucose was about 3.5% for the photocurrents. The long-term stability of the proposed p-Cu2O/n-Cu2O/FTO photoelectrochemical platform was evaluated monitoring the response of the system for one month. It was observed that the p-Cu2O/n Cu2O/FTO photoelectrochemical platform retained more than 95% of its initial response after one month.

The effects of some species frequently observed in formulations of artificial saliva samples on the response of the photoelectrochemical platform were investigated in order to evaluate the selectivity of the photoelectrochemical-assisted batch injection analysis of glucose exploiting the p-Cu2O/n Cu2O/FTO photoelectrode. In this sense, the effects of potassium, sodium, calcium, magnesium, dihydrogen phosphate, ammonium, and chloride ions, as well as the interference of urea and amylase molecules on the response of the p-Cu2O/n-Cu2O/FTO photoelectrode were tested in a glucose/foreign specie ratio of 1/1, respectively. Recovery tests were performed applying the p-Cu2O/n-Cu2O/FTO photoelectrode in artificial saliva samples spiked with a known amount of glucose. The glucose concentration found in spiked samples was calculated using the standard addition method, and the concentration value found of 67.2 µmol L-1 was in good agreement to spiked in sample of 70 µmol L-1, having a detection recovery of 96%.

Conclusions

The present work describes at the first time the development of a photoelectrochemical-assisted batch injection analysis system exploiting a homojunction based on a p-Cu2O/n-Cu2O/FTO photoelectrode for glucose detection. The Raman spectra of n-Cu2O and p-Cu2O, electrodeposited on the FTO surface, indicated that they have the phonon frequencies characteristic of crystalline Cu2O. The XRD patterns confirmed the crystalline character of the electrodeposits and that they preferentially grow along the (111). The Mott-Schottky analysis indicated that the n-Cu2O shows the behavior of an n-type semiconductor while the p-Cu2O electrodeposited sample presented the behavior of a p-type semiconductor. The p-Cu2O/n Cu2O/FTO photoelectrode was successfully applied in photoelectrochemical assisted batch-injection analysis system for the determination of glucose in artificial saliva sample being a promising alternative method for glucose determination with good sensitivity, wide linear-response range, low limit of detection, and good sensitivity.

Acknowledgments

The authors are grateful to FAPEMA (INFRA-03186/18; UNIVERSAL-01057/19; UNIVERSAL-01194/17), CNPq (308204/2018-2; 309828/2020-1), Instituto Nacional de Ciência e Tecnologia em Bioanalítica (465389/2014-7), and FINEP for financial support. The authors gratefully acknowledge the Analytical Center of UFMA that provided the FTIR spectra.

References

  • 1
    Galant, A. L.; Kaufman, R. C.; Wilson, J. D.; Food Chem. 2015, 188, 149.
  • 2
    American Diabetes Association; Diabetes Care 2014, 37, S81.
  • 3
    Chen, S.; Qian, D.; Burström, K.; Burström, B.; Patient Educ. Couns. 2006, 103, 1767.
  • 4
    Lin, K.-D.; Hsu, C.-C.; Ou, H.-Y.; Wang, C.-Y.; Chin, M.-C.; Shin, S.-J.; JFMA 2019, 118, S103.
  • 5
    Xie, W.-Q.; Gong, Y.-X.; You, K.-X.; J. Chromatogr. A 2017, 150, 143.
  • 6
    Chen, X.; Chen, J.; Wang, F.; Xiang, X.; Luo, M.; Ji, X.; He, Z.; Biosens. Bioelectron. 2012, 35, 363.
  • 7
    Hull, E. L.; Metter, N. I.; Olson, B. P.; Ediger, M. N.; Magee, A. J.; Way, J. F.; Vugrin, K. E.; Maynard, J. D.; J. Clin. Transl. Endocrinol. 2014, 1, 92.
  • 8
    Meng, W.; Wen, Y.; Dai, L.; He, Z.; Wang, L.; Sens. Actuators, B 2018, 260, 852.
  • 9
    Xu, F.; Wu, M.; Ma, G.; Xu, H.; Shang, W.; Microchem. J. 2020, 159, 105432.
  • 10
    Li, J.; Chen, Q.; Bai, J.; Zhou, B.; J. Hazard Mater. 2013, 262, 304.
  • 11
    Liu, G.-Q.; Zhong, H.; Li, X.-R.; Yang, K.; Jia, F.-F.; Cheng, Z.-P.; Zhang, L.-L.; Yin, J.-Z.; Guo, L.-P.; Qian, H.-Y.; Sens. Actuators, B 2017, 242, 484.
  • 12
    Mahmoud, A.; Echabaane, C.; Omri, K.; Mir, L. E.; Chaabane, R. B.; J. Alloys Compd. 2019, 786, 960.
  • 13
    Tian, K.; Prestgard, M.; Tiwari, A.; Mater. Sci. Eng., C 2014, 41, 100.
  • 14
    Sehit, E.; Altintas, Z.; Biosens. Bioelectron. 2020, 159, 112165.
  • 15
    Zhang, Z.-X.; Zhao, C.-Z.; Chin. J. Anal. Chem. 2013, 41, 436.
  • 16
    Devadoss, A.; Sudhagar, P.; Terashima, C.; Nakata, K.; Fujishima, A.; J. Photochem. Photobiol., C 2015, 24, 43.
  • 17
    Zhao, W.-W.; Xu, J.-J.; Chen, H.-Y.; Biosens. Bioelectron. 2017, 92, 294.
  • 18
    Wang, S.; Li, S.; Wang, W.; Zhao, M.; Liu, J.; Feng, H.; Chen, Y.; Gu, Q.; Du, Y.; Hao, W.; Sens. Actuators, B 2019, 291, 34.
  • 19
    Saglam, O.; Dilgin, Y.; Electroanalysis 2017, 29, 1368.
  • 20
    Soares, L. C.; Santos, C. C.; Luz, R. C. S.; Damos, F. S.; Electroanalysis 2020, 32, 1608.
  • 21
    Wang, M.; Song, X.; Song, B.; Liu, J.; Hu, C.; Wei, D.; Wong, C.-P.; Sens. Actuators, B 2017, 250, 333.
  • 22
    Lin, L.-Y.; Karakocak, B. B.; Kavadiya, S.; Soundappan, T.; Biswas, P.; Sens. Actuators, B 2018, 259, 745.
  • 23
    Li, M.; Wang, H.; Wang, X.; Lu, Q.; Li, H.; Zhang, Y.; Yao, S.; Biosens. Bioelectron. 2019, 142, 111535.
  • 24
    Praveen, R.; Ramaraj, R.; J. Electroanal. Chem. 2019, 851, 113454.
  • 25
    Jayasingha, L.; Jayanthilaka, C.; Kumara, R.; Ohara, K.; Kaumal, M.; Gunewardene, S.; Dissanayake, D.; Jayanetti, S.; Electrochim. Acta 2020, 329, 135177.
  • 26
    Xiong, L.; Huang, S.; Yang, X.; Qiu, M.; Chen, Z.; Yu, Y.; Electrochim. Acta 2011, 56, 2735.
  • 27
    Xu, M.; Liu, X.; Xu, W.; Hao, X.; Feng, X.; J. Alloys Compd. 2018, 769, 484.
  • 28
    Wan, T.; Wei, Y.; Chang, X.; Li, C.; Li, A.; Liu, S.; Zhang, J.; Gong, J.; Appl. Catal., B 2018, 226, 31.
  • 29
    Gimenes, D. T.; Marra, M. C.; Freitas, J. M.; Muñoz, R. A. A.; Richter, E. M.; Sens. Actuators, B 2015, 212, 411.
  • 30
    Freitas, J. M.; Oliveira, T. C.; Gimenes, D. T.; Muñoz, R. A. A.; Richter, E. M.; Talanta 2016, 146, 670.
  • 31
    Tormin, T. F.; Narciso, L. C. D.; Richter, E. M.; Munoz, R. A. A.; Electrochim. Acta 2015, 164, 90.
  • 32
    Stefano, J. S.; Tormin, T. F.; da Silva, J. P.; Richter, E. M.; Munoz, R. A. A.; Microchem. J. 2017, 133, 398.
  • 33
    Pereira, P. F.; da Silva, W. P.; Munoz, R. A. A.; Richter, E. M.; J. Electroanal. Chem. 2016, 766, 87.
  • 34
    Mcshane, C. M.; Choi, K.-S.; Phys. Chem. Chem. Phys. 2012, 14, 6112.
  • 35
    Nishikawa, M.; Fukuda, M.; Nakabayashi, Y.; Saito, N.; Appl. Surf. Sci. 2016, 363, 173.
  • 36
    Rinaldi, A. L.; Rodríguez-Castellón, E.; Sobral, S.; Carballo, R.; J. Electroanal. Chem. 2019, 832, 209.
  • 37
    Ha, T.; Park, I.; Sim, K. I.; Lee, H.; Bae, J.-S.; Kim, S. J.; Kim, J. P.; Kim, T.-T.; Kim, J. H.; Jang, J. I.; Jeong, S.-Y.; APL Mater. 2019, 7, 031115.
  • 38
    Sahaia, A.; Goswamia, N.; Kaushikb, S. D.; Tripathic, S.; Appl. Surf. Sci. 2016, 390, 974.
  • 39
    Huang, Q.; Li, J.; Bi, X.; J. Alloys Compd. 2015, 647, 585.
  • 40
    Huang, M.-C.; Wang, T. H.; Chang, W.-S.; Lin, J.-C.; Wu, C.-C.; Chen, I.-C.; Peng, K.-C.; Lee, S.-W.; Appl. Surf. Sci. 2014, 301, 369.
  • 41
    Mao, Y.; He, J.; Sun, X.; Li, W.; Lu, X.; Gan, J.; Liu, Z.; Gong, L.; Chen, J.; Liu, P.; Tong, Y.; Electrochim. Acta 2012, 62, 1.
  • 42
    Powell, D.; Compaan, A.; Macdonald, J. R.; Phys. Rev. B 1975, 12, 20.
  • 43
    Hossain, M. A.; Al-Gaashani, R.; Hamoudi, H.; Marri, M. J. A.; Hussein, I. A.; Belaidi, A.; Merzougui, B. A.; Alharbi, F. H.; Tabet, N.; Mater. Sci. Semicond. Process. 2017, 63, 203.
  • 44
    Su, T.; Yang, Y.; Dong, G.; Ye, T.; Jiang, Y.; Fan, R.; RSC Adv. 2016, 6, 65125.
  • 45
    Santos, G. K. C.; Silva, F. G. S.; Ferreira, A. R.; Luz, R. C. S.; Damos, F. S.; Microchem. J 2019, 148, 424.
  • 46
    Meng, F.; Li, J.; Cushing, S. K.; Zhi, M.; Wu, N.; J. Am. Chem. Soc. 2013, 135, 10286.
  • 47
    Li, J.; Meng, F.; Suri, S.; Ding, W.; Huang, F.; Wu, N.; Chem. Commun. 2012, 48, 8213.
  • 48
    Wen, X.-J.; Niu, C.-G.; Zhang, L.; Zeng, G.-M.; Dalton Trans. 2017, 46, 4982.
  • 49
    Tang, L.; Lv, J.; Kong, C.; Yang, Z.; Li, J.; New J. Chem. 2016, 40, 6573.
  • 50
    Li, Y.; Zhong, Y.; Zhang, Y.; Weng, W.; Li, S.; Sens. Actuators, B 2015, 206, 735.
  • 51
    Liu, M.; Liu, R.; Chen, W.; Biosens. Bioelectron. 2013, 45, 206.
  • 52
    Long, M.; Tan, L.; Liu, H.; He, Z.; Tang, A.; Biosens. Bioelectron. 2014, 59, 243.
  • 53
    Yin, H.; Cui, Z.; Wang, L.; Nie, Q.; Sens. Actuators, B 2016, 222, 1018.
  • 54
    Su, Y.; Guo, H.; Wang, Z.; Long, Y.; Li, W.; Tu, Y.; Sens. Actuators, B 2018, 255, 2510.

Edited by

Editor handled this article: Rodrigo A. A. Muñoz (Associate)