Fit-for-Purpose ICP-MS Procedure Supporting Brazilian Regulations on Inorganic Constituents in Fresh and Saline Waters

Abstract

This work aimed the development of an analytical procedure using inductively coupled plasma mass spectrometry (ICP-MS) in four modes of operation for simple and fast determination of Al, Ag, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, Hg, V, U, and Zn in fresh and sea water samples attending simultaneously four Brazilian regulations, Ordinance No. 2914, ANVISA No. 274 and CONAMA Nos. 357 and 396. High matrix introduction system enables the introduction of sea water (3.5% m v-1 of NaCl) only 2-fold diluted. Collision cell mode was efficient to overcome polyatomic interferences. The procedure presented limits of detection (LODs) lower than maximum concentrations allowed by regulations. The CRM NIST 1643e and addition-recovery experiments were used to check accuracy and recoveries varied from 84 to 115%. ICP-MS combined with collision cell mode and aerosol dilution led to accurate and simultaneous determinations of 25 elements at different concentration levels in samples containing high contents of total dissolved solids.

Keywords:
water; Brazilian regulations; HMI; collision cell; ICP-MS


Introduction

Essential micronutrients present in water, such as Cu and Se, have important roles in the body functioning. However, at excessive concentrations, they can negatively affect growth, reproduction and several biological functions. On the other hand, the presence of toxic elements, such as Cd and Pb, is extremely damaging, even at trace concentrations (µg L-1 and pg L-1). Thus, the accurate determination of toxic and essential elements in water is crucial for human health safety.11 Sela, H.; Karpas, Z.; Cohen, H.; Tal, A.; Zeiri, Y.; Biol. Trace Elem. Res. 2013, 155, 209.

In this context, four Brazilian regulations, Ordinance No. 2914/201122 Ministério da Saúde, Portaria No. 2914, de 12 de dezembro de 2011, Dispõe sobre os Procedimentos de Controle e de Vigilância da Qualidade da Água para Consumo Humano e seu Padrão de Potabilidade; DOU, Brasília, Brasil, 2011. Available at http://site.sabesp.com.br/site/uploads/file/asabesp_doctos/PortariaMS291412122011.pdf, accessed in March 2018.
http://site.sabesp.com.br/site/uploads/f...
of the Ministry of Health, Resolution RDC No. 274/200533 Agência Nacional de Vigilância Sanitária (ANVISA); Resolução RDC No. 274 de 22 de setembro de 2005; Regulamento Técnico para Águas Envasadas e Gelo; DOU, Brasil, 2005. Available at http://portal.anvisa.gov.br/documents/33916/394219/RDC_274_2005.pdf/19d98e61-fa3b-41df-9342-67e0167bf550, accessed in March 2018.
http://portal.anvisa.gov.br/documents/33...
and CONAMA Resolutions Nos. 357/200544 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
and 396/200855 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 396, de 3 de abril de 2008, Dispõe sobre a Classificação e Diretrizes Ambientais para o Enquadramento das Águas Subterrâneas e Dá Outras Providências; DOU, Brasília, Brasil, 2008, No. 066, seção 1, p. 64-68. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=562, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
establish maximum values for physico-chemical and chemical parameters in water samples. Ordinance No. 291422 Ministério da Saúde, Portaria No. 2914, de 12 de dezembro de 2011, Dispõe sobre os Procedimentos de Controle e de Vigilância da Qualidade da Água para Consumo Humano e seu Padrão de Potabilidade; DOU, Brasília, Brasil, 2011. Available at http://site.sabesp.com.br/site/uploads/file/asabesp_doctos/PortariaMS291412122011.pdf, accessed in March 2018.
http://site.sabesp.com.br/site/uploads/f...
establishes maximum values and procedures for controlling and monitoring of water quality. Resolution RDC No. 27433 Agência Nacional de Vigilância Sanitária (ANVISA); Resolução RDC No. 274 de 22 de setembro de 2005; Regulamento Técnico para Águas Envasadas e Gelo; DOU, Brasil, 2005. Available at http://portal.anvisa.gov.br/documents/33916/394219/RDC_274_2005.pdf/19d98e61-fa3b-41df-9342-67e0167bf550, accessed in March 2018.
http://portal.anvisa.gov.br/documents/33...
establishes secure identity and quality characteristics for natural mineral water, natural water and ice for human consumption. CONAMA Resolution No. 35744 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
establishes maximum values for natural waters (surface, brackish and saline water) and CONAMA Resolution No. 39655 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 396, de 3 de abril de 2008, Dispõe sobre a Classificação e Diretrizes Ambientais para o Enquadramento das Águas Subterrâneas e Dá Outras Providências; DOU, Brasília, Brasil, 2008, No. 066, seção 1, p. 64-68. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=562, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
establishes classification and maximum values for groundwater.

Thus, the development of sensitive and accurate analytical procedures for elemental analysis of water is required. Among the analytical instrumental methods, inductively coupled plasma mass spectrometry (ICP-MS) provides fast multi-elemental analysis, low detection limits (LOD), high sensitivity, precision and robustness. However, the determination of some isotopes is limited by the occurrence of spectral interferences.66 Thomas, R.; Practical Guide to ICP-MS, 3rd ed.; CRC Press: Boca Raton, New York, 2013, p. 73-90. In addition, sample dilution is often required to keep the total dissolved solids (TDS) below 0.1% m v-1 and acidity below 2% v v-1.

In order to eliminate spectral interferences, collision cell technology (CCT) features a cell introduced before the mass analyzer. Collision cell provides effective correction of polyatomic interferences, eliminating the need for reactive cell gases in routine analysis.77 D’ilio, S.; Violante, N.; Majorani, C.; Petrucci, F.; Anal. Chim. Acta 2011, 698, 6.

8 Wilbur, S. M.; Jones, L. C.; Open Chem. Biomed. Methods J. 2010, 3, 135.
-99 McCurdy, E.; Woods, G.; J. Anal. At. Spectrom. 2004, 5, 607. Additionally, high matrix introduction (HMI) technology uses auto-optimization of aerosol dilution to further improve matrix tolerance, reducing both aerosol density and water vapor loading in the plasma.88 Wilbur, S. M.; Jones, L. C.; Open Chem. Biomed. Methods J. 2010, 3, 135.,1010 Agilent Technologies; Agilent 7800 ICP-MS: Plasma Robustness in ICP-MS Benefits of a Low CeO/Ce Ratio, 5990-8060EN; Agilent Technologies: Santa Clara, USA, 2015. Available at https://www.agilent.com/cs/library/articlereprints/public/5990_8060EN.pdf, accessed in March 2018.
https://www.agilent.com/cs/library/artic...

11 Larivière, D.; Tremblay, M.; Durand-Jézéquel, M.; Tolmachev, S.; Anal Bioanal. Chem. 2012, 403, 409.

12 Novotnik, B.; Zuliani, T.; Martincic, A.; Scancar, J.; Milacic, R.; J. Anal. At. Spectrom. 2012, 27, 488.

13 Hein, C.; Sander, J. M.; Kautenburger, R.; Talanta 2017, 164, 477.

14 Celo, V.; Dabek-Zlotorzynska, E.; Zhao, J.; Okonskaia, I.; Bowman, D.; Anal. Chim. Acta 2011, 706, 89.

15 Döker, S.; Anal. Methods 2017, 9, 1710.
-1616 Döker, S.; Uslu, M.; Food Anal. Methods 2013, 7, 683.

Wilbur and Jones88 Wilbur, S. M.; Jones, L. C.; Open Chem. Biomed. Methods J. 2010, 3, 135. demonstrated the benefits of HMI system for determination of 26 elements in complex matrix certified reference materials (CRM), such as river sediment and soil samples. Aqueous samples were investigated using fast protein liquid chromatography (FPLC) coupled to ICP-MS with a high matrix introduction system. The HMI reduced the sample aerosol transport and enabled direct analysis of separated Cr species in the presence of up to 4% m v-1 NaCl.1313 Hein, C.; Sander, J. M.; Kautenburger, R.; Talanta 2017, 164, 477.

In this context, this study presents the development of an analytical procedure using a quadrupole ICP-MS operated in four acquisition modes: (i) No gas: standard mode; (ii) No gas (HMI): standard mode with HMI system; (iii) He: collision cell mode; (iv) He (HMI): collision cell mode with HMI system; for determination of Al, Ag, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, Hg, V, U and Zn in natural fresh water (Tables I, II and III of CONAMA 357);44 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
mineral and tap water samples in order to meet simultaneously Ordinance 2914/11,22 Ministério da Saúde, Portaria No. 2914, de 12 de dezembro de 2011, Dispõe sobre os Procedimentos de Controle e de Vigilância da Qualidade da Água para Consumo Humano e seu Padrão de Potabilidade; DOU, Brasília, Brasil, 2011. Available at http://site.sabesp.com.br/site/uploads/file/asabesp_doctos/PortariaMS291412122011.pdf, accessed in March 2018.
http://site.sabesp.com.br/site/uploads/f...
CONAMA 357/0544 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
and ANVISA 274/0533 Agência Nacional de Vigilância Sanitária (ANVISA); Resolução RDC No. 274 de 22 de setembro de 2005; Regulamento Técnico para Águas Envasadas e Gelo; DOU, Brasil, 2005. Available at http://portal.anvisa.gov.br/documents/33916/394219/RDC_274_2005.pdf/19d98e61-fa3b-41df-9342-67e0167bf550, accessed in March 2018.
http://portal.anvisa.gov.br/documents/33...
. Aluminum, Ag, B, As, Ba, Be, Cd, Cr, Cu, Fe, Mn, Ni, Pb, Se, Hg, U and Zn were determined in sea water (Tables IV and V of CONAMA 357)44 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
providing higher sample throughput and lower argon consumption by allowing the determination of all elements in a single run.

Experimental

Instrumentation

Analysis was performed using an Agilent 7800 Quadrupole ICP-MS (Agilent Technologies, Tokyo, JHS, Japan) operated in four acquisition modes: (i) No gas; (ii) No gas (HMI); (iii) He and (iv) He (HMI). No gas mode means not using the collision cell and He mode means that collision cell is pressurized with pure He (99.999%, White Martins-Praxair, Sertãozinho, SP, Brazil). Plasma operating conditions are described in Table 1. HMI mode implies that aerosol was diluted with argon at the optimized HMI gas flow rate of 0.62 L min-1 and carrier gas flow rate of 0.40 L min-1, thus 1.02 L min-1 of total flow rate. Argon (99.9996%, White Martins-Praxair, Sertãozinho, SP, Brazil) was used for all measurements. Mira Mist nebulizer (Burgener Research, Mississauga, ON, Canada) was used due to the eventual presence of solid particles in natural water samples. The reading time, considering these four modes of operation, plus sample collection and rinse times, totalized approximately 4 min per sample at 1.0 mL min-1 of sample uptake rate.

Table 1
Operating parameters used in Agilent 7800 Quadrupole ICP-MS

Reagents, standard solutions and samples

Experiments were performed using HNO3 (Synth, Diadema, SP, Brazil) purified in a sub-boiling distillation system Distillacid™ BSB-939-IR (Berghof, Eningen, Germany) and ultrapure water, resistivity higher than 18.2 MΩ cm (Milli-Q®, Millipore, Bedford, MA, USA). Standard solutions used for ICP-MS calibration and for addition and recovery experiments were prepared by dilution of 1000 mg L-1 of Al, Ag, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, Hg, V, U and Zn (Qhemis, São Paulo, Brazil) in 0.14 mol L-1 HNO3 medium, as well as the internal standards (ISs) evaluated: Ge, Li, In, Ir and Pt.

For fresh water analysis, the concentrations of the solutions used for obtaining analytical calibration curves were 0, 0.05, 0.1, 0.5, 1.0, 2.0, 5.0, 10, 20 and 50 µg L-1 for Al, Ag, As, B, Be, Cd, Co, Cr, Cu, Fe, Li, Mn, Ni, Pb, Sb, Se, Tl, V, U and Zn. Barium concentrations were 0, 1.0, 2.0, 5.0, 10, 50, 100, 200, 500 and 700 µg L-1. For Hg, the concentrations were 0, 0.01, 0.05, 0.1, 0.2, 0.5, 1.0, 1.5, 3.0 and 5.0 µg L-1. For elements present at high concentrations, such as Na, calibration points were 0, 10, 50, 100, 150 and 200 mg L-1 and for Ca and Mg were 0, 1.0, 2.5, 5.0, 10 and 20 mg L-1.

For analysis of sea water, the concentrations of the solutions used for analytical calibration curves were 0, 1.0, 2.0, 5.0, 10, 20, 50 and 100 µg L-1 for Al, Ag, As, Ba, Be, Cd, Cr, Cu, Fe, Mn, Ni, Pb, Se, U and Zn and for Hg were 0, 0.05, 0.1, 0.2, 0.5, 1.0 and 1.5 µg L-1, prepared by matrix-matching with 3.5% m v-1, 1.75% m v-1 NaCl (Merck, Rio de Janeiro, RJ, Brazil) and only in 0.14 mol L-1 HNO3 medium (without matrix-matching).

The certified reference material of trace elements in water, NIST SRM 1643e (National Institute of Standard and Technology, Gaithersburg, MD, USA), was used for optimization of experimental procedures and evaluation of accuracy. Addition and recovery experiments were performed for Hg and U, both not present in the SRM 1643e. Analogously, for sea water analysis, addition and recovery experiments were performed at two addition levels 0.2 and 0.5 µg L-1 for Hg and 10 and 25 µg L-1 for the other elements investigated. Internal standards were added at 10 µg L-1 to the analytical calibration solutions, analytical blanks and samples. The cleaning of the sample introduction system with 0.12 mol L-1 HCl after measurement of each solution is mandatory for avoiding memory effects caused by mercury.1717 Amaral, C. D. B.; Amais, R. S.; Fialho, L. L.; Schiavo, D.; Morim, T. A.; Nogueira, A. R. A.; Rocha, F. R. P.; Nóbrega, J. A.; Anal. Methods 2015, 7, 1215. Afterwards, a fast cleaning step with 0.14 mol L-1 HNO3 solution was made for avoiding any undesirable effect caused by chloride.

Six mineral bottled waters purchased in local supermarkets, one sample of artesian well water (groundwater), one sample of drinking water from the Federal University of São Carlos, were directly analyzed without any pretreatment. One sea water sample from Caraguatatuba Beach (São Paulo State, Brazil) was analyzed after a 2-fold dilution and filtration on 0.45 µm filter paper.

Optimization of the procedure

The effect of HMI flow rate on sodium chloride solutions containing 0, 0.1, 0.5, 1.7 and 3.5% m v-1 by dissolution of the pure salt in water and doped at 1 µg L-1 of all analytes were investigated by varying the dilution gas flow rate for HMI from 0.30 to 0.80 L min-1, also with the variation of the nebulization gas flow rate from 0.72 to 0.22 L min-1, with increments of 0.10 L min-1. In all experiments, the flow rate summation of both was kept at 1.02 L min-1. Isotopes were evaluated, as well as the CeO+/Ce+ ratio.

For analysis of sea water, in order to optimize the acquisition mode (No gas (HMI) or He (HMI)) and aerosol dilution conditions, three analytical curves were prepared by matrix-matched with 3.5 and 1.75% m v-1 NaCl, and HNO3 1% v v-1 for sea water samples analysis without dilution, 2-fold dilution and 5-fold dilution, respectively. The best working conditions were established based on addition and recovery experiments for each condition. For analysis of fresh water, the acquisition mode (No gas, No gas (HMI), He and He (HMI)) were established according to recoveries of the certified reference material of trace elements in water, NIST SRM 1643e. After optimization, the monitored isotopes and the acquisition mode were selected for each analyte.

Results and Discussion

Effect of the HMI system for analysis of saline solutions

Sea water is a high TDS matrix for plasma. Additionally, elemental detection in an NaCl matrix by ICP-MS is also affected by polyatomic interferences formed by chlorine ions, such as 35Cl16O+ affecting 51V+; 35Cl16OH+ affecting 52Cr+; 37Cl17OH+ affecting 55Mn+; 37Cl18OH+ affecting 56Fe+; 23Na35Cl+ affecting 58Ni+; 24Mg35Cl+ affecting 59Co+ and 40Ar35Cl+ affecting 75As+ in quadrupole based mass spectrometers.6,7

The effect of HMI use on chlorine interferences and oxide formation was evaluated by signal intensities monitoring for 75As+ 1.0 µg L-1 (Figure 1) and for CeO+/Ce+ ratio (Figure 2). These investigations were performed at 0, 0.1, 0.5, 1.7 and 3% m v-1 NaCl and different HMI gas flow rates (0.30, 0.40, 0.50, 0.60, 0.70 and 0.82 L min-1) in He (HMI) and No gas (HMI) acquisition modes.

Figure 1
Effect of HMI (high matrix introduction) gas flow rate (L min-1) on 75As+ net signal (1 µg L-1) spiked in solutions containing different concentrations of NaCl (%) in He (HMI) acquisition mode. HMI gas flow rate: (■) 0.30 L min-1; (○) 0.40 L min-1; (▲) 0.50 L min-1; (□) 0.60 L min-1; (♦) 0.70 L min-1; (◊) 0.80 L min-1.

Figure 2
CeO+/Ce+ ratio at different concentrations of NaCl (%) and HMI (high matrix introduction) gas flow rate (L min-1) by ICP-MS (inductively coupled plasma mass spectrometry) in No gas (HMI) acquisition mode. HMI gas flow rate: (■) 0.30 L min-1; (○) 0.40 L min-1; (▲) 0.50 L min-1; (□) 0.60 L min-1; (♦) 0.70 L min-1; (◊) 0.80 L min-1.

In order to eliminate the effect of this polyatomic interference on As signal, CCT was combined with HMI mode. It was observed (Figure 1) that the increase in concentration of chloride ions did not cause positive errors in 75As+ monitoring, however, for some flow rates it was noticed a mild suppression of the signal. In HMI gas flow rates higher than 0.60 L min-1, it was not observed significant signal difference when increasing NaCl concentration, which is a positive aspect of this procedure. The use of HMI system (in the same dilution gas flow rate of 0.60 L min-1) enabled the direct introduction of urine samples and dialysis solutions 2-fold diluted.1818 Barros, A. I.; Pinheiro, F. C.; Lorençatto, R.; Amaral, C. D. B.; Nóbrega, J. A.; Talanta 2018, 178, 805. Figure 2 shows an increase of oxide formation when increasing NaCl concentration, probably this is due to the decrease of plasma energy caused by plasma loading, in consequence of the presence of sample matrix. However, by increasing HMI gas flow rate (> 0.50 L min-1) it was observed that CeO+/Ce+ ratio did not vary significantly in most NaCl concentrations. Previously, Hein et al.1313 Hein, C.; Sander, J. M.; Kautenburger, R.; Talanta 2017, 164, 477. proposed an analytical procedure for analysis of highly saline samples by ICP-MS with an ultra-high matrix introduction system (UHMI). This procedure was suitable for determining metal ions, such as Eu and U in samples with salinities up to 5 mol L-1 NaCl. Accurate analysis of lanthanoids in environmental samples by ICP-MS equipped with HMI system was investigated by Celo et al.1414 Celo, V.; Dabek-Zlotorzynska, E.; Zhao, J.; Okonskaia, I.; Bowman, D.; Anal. Chim. Acta 2011, 706, 89. The amounts of oxide and hydroxide molecular species formed in the plasma were reduced by up to 5 times. In other studies, hydroxide formation was lower than 0.02% when HMI system was used.88 Wilbur, S. M.; Jones, L. C.; Open Chem. Biomed. Methods J. 2010, 3, 135.,1313 Hein, C.; Sander, J. M.; Kautenburger, R.; Talanta 2017, 164, 477.,1515 Döker, S.; Anal. Methods 2017, 9, 1710.

Evaluation of accuracy and internal standardization

The accuracy of the analytical procedure was evaluated by addition and recovery experiments and by analysis of the certified reference material of water (NIST SRM 1643e). Table 2 shows the recoveries and the relative standard deviations (RSDs) obtained for Al, Ag, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, V and Zn in the SRM 1643e.

Table 2
Determination of Al, Ag, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, V and Zn (mean ± standard deviation, n = 3) and recoveries in NIST SRM 1643e using ICP-MS

According to the recoveries obtained by NIST SRM 1643e analysis, it is possible to select three modes of acquisition for fresh water analysis: No gas, He and He (HMI). Recoveries for the CRM varied from 87 (107Ag) to 115% (64Zn) and suitable recoveries, i.e., 87% for Hg and 107% for U, were also obtained in addition and recovery experiments.

Calcium, Na and Mg were accurately determined in fresh water in He (HMI) mode, since this approach combines CCT, which controls spectral interferences, and HMI technology, which provides sample dilution by diluting the aerosol with argon gas at constant flow rate. This mode of operation improved matrix effects tolerance and provided the determination of elements at high concentrations simultaneously with elements at lower concentrations, thereby minimizing the need for recalibrations or even the use of other instrumental method with lower sensitivity.

Sea water was analyzed in HMI mode (Table 3), since when the instrument is operated in standard mode, analysis of samples with high contents of dissolved solids (> 0.2% m v-1) can cause salt deposition at the interfaces of the cones and ion lenses, and transport effects. The HMI system allows the introduction of samples with TDS around 3% m v-1. For sample analysis without dilution, recoveries for 27Al, 64Zn, 137Ba, 202Hg and 238U were not quantitative. Significant improvement of recoveries was observed when 193Ir was used as IS for Ba, the recoveries were from 4782 and 5612% without IS to 96 and 117%, respectively, when using IS. On the other hand, significant improvements were not observed for Al, Zn, Hg and U when using internal standardization. Good recoveries were obtained for all analytes in samples 2-fold diluted, using matrix matching without internal standardization. Recoveries ranged from 84 (55Mn) to 115% (137Ba).

Table 3
Evaluation of combined use of HMI system, collision cell and IS for Al, Ag, As, Ba, Be, Cd, Cr, Cu, Fe, Hg, Mn, Ni, Pb, Se, Tl, U and Zn determination in sea water by ICP MS with and without matrix-matching. Mean (relative standard deviation, n = 3)

For sea water 5-fold diluted, even without matrix matching, appropriate recoveries were obtained in addition and recovery experiments, ranging from 80 to 118%. IS was not required for accurate determination of 9Be, 52Cr, 56Fe, 58Ni, 65Cu, 75As, 78Se, 202Hg, 205Tl, 208Pb and 238U. However, significant improvement of recoveries was observed when 72Ge was used as IS for 27Al and 55Mn, 195Pt was used as IS for 107Ag and 114Cd and when 193Ir was used as IS for 137Ba, 205Tl, 208Pb and 238U. The improvement of recoveries with 193Ir as IS for 137Ba, 205Tl, 208Pb, 238U; 195Pt as IS for 107Ag, 114Cd; and 72Ge as IS for 27Al and 55Mn probably may be due to the similarity of the mass to charge ratios, m/z, of the IS evaluated with the analytes.1919 Amaral, C. D. B.; Machado, R. C.; Virgilio, A.; Schiavo, D.; Nogueira, A. R. A.; Nóbrega, J. A.; J. Anal. At. Spectrom. 2016, 31, 1179. However, as demonstrated by Barros et al.1818 Barros, A. I.; Pinheiro, F. C.; Lorençatto, R.; Amaral, C. D. B.; Nóbrega, J. A.; Talanta 2018, 178, 805. and also obtained here, significant improvement of recoveries (110 and 118%) were also reached with 193Ir as IS for 27Al.

The isotopes 7Li, 9Be, 11B and 137Ba were not affected by spectral or matrix interferences. Thus, no interference correction strategy was required and such elements were determined in No gas and No gas (HMI) modes in fresh water and sea water samples, respectively.

Best results were obtained for 23Na, 24Mg, 27Al, 44Ca, 51V, 52Cr, 55Mn, 56Fe, 58Ni, 59Co, 65Cu, 64Zn, 75As, 78Se, 107Ag, 114Cd, 123Sb, 137Ba, 202Hg, 205Tl, 208Pb, 238U using collision cell acquisition mode. Although He mode caused a significant reduction in net signals (from 2 to 3 times), the monitoring of some isotopes can be hampered by spectral interferences (polyatomic ions and double charge ions) formed by elements present in the sample or from species present in the plasma, such as argon, oxygen and nitrogen.66 Thomas, R.; Practical Guide to ICP-MS, 3rd ed.; CRC Press: Boca Raton, New York, 2013, p. 73-90.,77 D’ilio, S.; Violante, N.; Majorani, C.; Petrucci, F.; Anal. Chim. Acta 2011, 698, 6.

It is possible to conclude that the developed procedure using specific operation mode for different analytes is accurate. For sea water, dilution gas flow rate of 0.60 L min-1 and 2-fold sample dilution were adopted. However, sea water 5-fold diluted, even without matrix matching, was also accurately analyzed for all analytes (IS required for Al, Ag, Ba, Cd, Mn, Pb, Tl and U) except for Zn, allowing 24 elements determination in a single ICP-MS run to satisfy four Brazilian regulations.

Method detection limits

The focus here was the simultaneous determination of 25 elements in a single ICP-MS run to satisfy four Brazilian regulations.22 Ministério da Saúde, Portaria No. 2914, de 12 de dezembro de 2011, Dispõe sobre os Procedimentos de Controle e de Vigilância da Qualidade da Água para Consumo Humano e seu Padrão de Potabilidade; DOU, Brasília, Brasil, 2011. Available at http://site.sabesp.com.br/site/uploads/file/asabesp_doctos/PortariaMS291412122011.pdf, accessed in March 2018.
http://site.sabesp.com.br/site/uploads/f...

3 Agência Nacional de Vigilância Sanitária (ANVISA); Resolução RDC No. 274 de 22 de setembro de 2005; Regulamento Técnico para Águas Envasadas e Gelo; DOU, Brasil, 2005. Available at http://portal.anvisa.gov.br/documents/33916/394219/RDC_274_2005.pdf/19d98e61-fa3b-41df-9342-67e0167bf550, accessed in March 2018.
http://portal.anvisa.gov.br/documents/33...

4 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
-55 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 396, de 3 de abril de 2008, Dispõe sobre a Classificação e Diretrizes Ambientais para o Enquadramento das Águas Subterrâneas e Dá Outras Providências; DOU, Brasília, Brasil, 2008, No. 066, seção 1, p. 64-68. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=562, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
Accuracy and sensitivity were evaluated for each element according to the respective mode of operation. Limits of detection (LOD) and quantification (LOQ) were calculated considering standard deviation (SD) for 10 measurements of a blank solution divided by slope of analytical curve multiplied by 3 (LOD) and 10 (LOQ), and then multiplied by the dilution factor.2020 Thomsen, V.; Schatzlein, D.; Mercuro, D.; Spectroscopy 2003, 18, 112. The isotopes and their best mode of acquisition, linear correlation coefficient, slope of analytical curves and LOD obtained for all analytes are shown for fresh and sea water samples in Table 4.

Table 4
Permissible maximum limits (PML), linear correlation coefficients and limits of quantification obtained for Al, Ag, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, Hg, V, U and Zn in fresh and sea water samples

The developed procedure for water analysis by ICP-MS presented limits of detection and of quantification in compliance with Ordinance No. 291422 Ministério da Saúde, Portaria No. 2914, de 12 de dezembro de 2011, Dispõe sobre os Procedimentos de Controle e de Vigilância da Qualidade da Água para Consumo Humano e seu Padrão de Potabilidade; DOU, Brasília, Brasil, 2011. Available at http://site.sabesp.com.br/site/uploads/file/asabesp_doctos/PortariaMS291412122011.pdf, accessed in March 2018.
http://site.sabesp.com.br/site/uploads/f...
and Resolutions Nos. 274,33 Agência Nacional de Vigilância Sanitária (ANVISA); Resolução RDC No. 274 de 22 de setembro de 2005; Regulamento Técnico para Águas Envasadas e Gelo; DOU, Brasil, 2005. Available at http://portal.anvisa.gov.br/documents/33916/394219/RDC_274_2005.pdf/19d98e61-fa3b-41df-9342-67e0167bf550, accessed in March 2018.
http://portal.anvisa.gov.br/documents/33...
35744 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
(including sea water, Tables IV and V) and 396,55 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 396, de 3 de abril de 2008, Dispõe sobre a Classificação e Diretrizes Ambientais para o Enquadramento das Águas Subterrâneas e Dá Outras Providências; DOU, Brasília, Brasil, 2008, No. 066, seção 1, p. 64-68. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=562, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
allowing the determination of all elements simultaneously using each optimized acquisition mode.

Sample analysis

Table 5 shows the concentrations of Al, Ag, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, Hg, V, U and Zn found in six bottled mineral water, one drinking water, one artesian well water and one sea water samples. All fresh and sea water samples analyzed presented concentrations below the maximum allowed limits imposed by Ordinance No. 291422 Ministério da Saúde, Portaria No. 2914, de 12 de dezembro de 2011, Dispõe sobre os Procedimentos de Controle e de Vigilância da Qualidade da Água para Consumo Humano e seu Padrão de Potabilidade; DOU, Brasília, Brasil, 2011. Available at http://site.sabesp.com.br/site/uploads/file/asabesp_doctos/PortariaMS291412122011.pdf, accessed in March 2018.
http://site.sabesp.com.br/site/uploads/f...
and Resolutions Nos. 274,33 Agência Nacional de Vigilância Sanitária (ANVISA); Resolução RDC No. 274 de 22 de setembro de 2005; Regulamento Técnico para Águas Envasadas e Gelo; DOU, Brasil, 2005. Available at http://portal.anvisa.gov.br/documents/33916/394219/RDC_274_2005.pdf/19d98e61-fa3b-41df-9342-67e0167bf550, accessed in March 2018.
http://portal.anvisa.gov.br/documents/33...
35744 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...
and 396.55 Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 396, de 3 de abril de 2008, Dispõe sobre a Classificação e Diretrizes Ambientais para o Enquadramento das Águas Subterrâneas e Dá Outras Providências; DOU, Brasília, Brasil, 2008, No. 066, seção 1, p. 64-68. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=562, accessed in March 2018.
http://www.mma.gov.br/port/conama/legiab...

Table 5
Determination of Al, Ag, As, B, Ba, Be, Ca, Cd, Co, Cr, Cu, Fe, Li, Mg, Mn, Na, Ni, Pb, Sb, Se, Tl, Hg, V, U and Zn (mean ± standard deviation, n = 3) in samples of mineral water (A-F), groundwater, drinking water and sea water by ICP-MS

Conclusions

A procedure for analysis of elemental composition of water samples by ICP-MS was developed, in order to answer the Ordinance No. 2914 of the Ministry of Health, the Resolutions Nos. 357 and 396 of CONAMA and the Resolution RDC No. 274 of ANVISA. The use of ICP-MS operated in different acquisition modes allowed precise, accurate and simultaneous determinations of elements at large concentration ranges and samples with high salts concentration in a single run. The measurements performed in He mode allowed accurate determinations of elements whose isotopes were affected by spectral interferences. Additionally, HMI system allowed the determination of Na, Ca and Mg in fresh water samples without requiring any further dilution and sea water analysis with only 2-fold dilution, demonstrating the improvement in plasma robustness when operating ICP-MS in HMI mode. Thus, the determination of trace elements can be made simultaneously to elements in higher concentrations. These benefits translate into simplicity, high sample throughput, time saving and reliable analysis.

Acknowledgments

The authors are grateful to the Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES/PNPD, Department of Chemistry, Graduate Program in Chemistry) and to the Conselho Nacional de Desenvolvimento Científico e Tecnológico (CNPq, 141634/2017-0 and 303107/2013-8) for fellowships and financial support. The authors also would like to thank Agilent Technologies for technical support and by providing the ICP-MS instrument.

References

  • 1
    Sela, H.; Karpas, Z.; Cohen, H.; Tal, A.; Zeiri, Y.; Biol. Trace Elem. Res 2013, 155, 209.
  • 2
    Ministério da Saúde, Portaria No. 2914, de 12 de dezembro de 2011, Dispõe sobre os Procedimentos de Controle e de Vigilância da Qualidade da Água para Consumo Humano e seu Padrão de Potabilidade; DOU, Brasília, Brasil, 2011. Available at http://site.sabesp.com.br/site/uploads/file/asabesp_doctos/PortariaMS291412122011.pdf, accessed in March 2018.
    » http://site.sabesp.com.br/site/uploads/file/asabesp_doctos/PortariaMS291412122011.pdf
  • 3
    Agência Nacional de Vigilância Sanitária (ANVISA); Resolução RDC No. 274 de 22 de setembro de 2005; Regulamento Técnico para Águas Envasadas e Gelo; DOU, Brasil, 2005. Available at http://portal.anvisa.gov.br/documents/33916/394219/RDC_274_2005.pdf/19d98e61-fa3b-41df-9342-67e0167bf550, accessed in March 2018.
    » http://portal.anvisa.gov.br/documents/33916/394219/RDC_274_2005.pdf/19d98e61-fa3b-41df-9342-67e0167bf550
  • 4
    Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 357, de 17 de março de 2005, Dispõe sobre a Classificação dos Corpos de Água e Diretrizes Ambientais para o seu Enquadramento, bem como Estabelece as Condições e Padrões de Lançamento de Efluentes, e Dá Outras Providências; DOU, Brasília, Brasil, 2005, No. 053, p. 58-63. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459, accessed in March 2018.
    » http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=459
  • 5
    Conselho Nacional do Meio Ambiente (CONAMA); Resolução No. 396, de 3 de abril de 2008, Dispõe sobre a Classificação e Diretrizes Ambientais para o Enquadramento das Águas Subterrâneas e Dá Outras Providências; DOU, Brasília, Brasil, 2008, No. 066, seção 1, p. 64-68. Available at http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=562, accessed in March 2018.
    » http://www.mma.gov.br/port/conama/legiabre.cfm?codlegi=562
  • 6
    Thomas, R.; Practical Guide to ICP-MS, 3rd ed.; CRC Press: Boca Raton, New York, 2013, p. 73-90.
  • 7
    D’ilio, S.; Violante, N.; Majorani, C.; Petrucci, F.; Anal. Chim. Acta 2011, 698, 6.
  • 8
    Wilbur, S. M.; Jones, L. C.; Open Chem. Biomed. Methods J 2010, 3, 135.
  • 9
    McCurdy, E.; Woods, G.; J. Anal. At. Spectrom 2004, 5, 607.
  • 10
    Agilent Technologies; Agilent 7800 ICP-MS: Plasma Robustness in ICP-MS Benefits of a Low CeO/Ce Ratio, 5990-8060EN; Agilent Technologies: Santa Clara, USA, 2015. Available at https://www.agilent.com/cs/library/articlereprints/public/5990_8060EN.pdf, accessed in March 2018.
    » https://www.agilent.com/cs/library/articlereprints/public/5990_8060EN.pdf
  • 11
    Larivière, D.; Tremblay, M.; Durand-Jézéquel, M.; Tolmachev, S.; Anal Bioanal. Chem 2012, 403, 409.
  • 12
    Novotnik, B.; Zuliani, T.; Martincic, A.; Scancar, J.; Milacic, R.; J. Anal. At. Spectrom 2012, 27, 488.
  • 13
    Hein, C.; Sander, J. M.; Kautenburger, R.; Talanta 2017, 164, 477.
  • 14
    Celo, V.; Dabek-Zlotorzynska, E.; Zhao, J.; Okonskaia, I.; Bowman, D.; Anal. Chim. Acta 2011, 706, 89.
  • 15
    Döker, S.; Anal. Methods 2017, 9, 1710.
  • 16
    Döker, S.; Uslu, M.; Food Anal. Methods 2013, 7, 683.
  • 17
    Amaral, C. D. B.; Amais, R. S.; Fialho, L. L.; Schiavo, D.; Morim, T. A.; Nogueira, A. R. A.; Rocha, F. R. P.; Nóbrega, J. A.; Anal. Methods 2015, 7, 1215.
  • 18
    Barros, A. I.; Pinheiro, F. C.; Lorençatto, R.; Amaral, C. D. B.; Nóbrega, J. A.; Talanta 2018, 178, 805.
  • 19
    Amaral, C. D. B.; Machado, R. C.; Virgilio, A.; Schiavo, D.; Nogueira, A. R. A.; Nóbrega, J. A.; J. Anal. At. Spectrom 2016, 31, 1179.
  • 20
    Thomsen, V.; Schatzlein, D.; Mercuro, D.; Spectroscopy 2003, 18, 112.

Publication Dates

  • Publication in this collection
    July 2018

History

  • Received
    29 Jan 2018
  • Accepted
    02 Apr 2018
Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
E-mail: office@jbcs.sbq.org.br