Evaluation of the influence of patient age and spacer device volume on aerosol lung deposition

Abstracts

OBJECTIVE: To evaluate the efficacy of three frequently used spacer devices to deliver aerosol to the lung, and to compare radioaerosol deposition with each device in different age groups. METHODS: Nine healthy, non-smoking volunteers were recruited: three adults and six children, including three toddlers and three school age children. Qualitative and semi-quantitative analysis of radioaerosol deposition in the lung were carried out. Yet, two small-volume devices (Aerochamber® and Inal-Air®) and one large-volume device (Flumax®) were compared. Each patient inhaled 99m technetium-phytate. The device was filled during 30 seconds with radioaerosol. Oxygen was used as the driving gas. During 10 seconds, the patients inhaled the radioaerosol. The radiation emitted at the front and back of the chest was measured. The radiation inside the device was also measured. RESULTS: The quantitative evaluation of lung deposition revealed that the younger the patient, the less aerosol was deposited in the lung with the large-volume spacer device (Flumax®). The difference between small-volume devices (Aerochamber® and Inal-Air®) was not significant. CONCLUSION: Small-volume spacers are the most appropriate for children. Large-volume devices should only be used by adolescents and adults.

Inhalation therapy; spacers; 99m technetium-fitate


OBJETIVO: Avaliar, através de estudo da deposição pulmonar de 99m tecnécio-fitato, a eficácia de três espaçadores freqüentemente utilizados na prática clínica e comparar esta deposição de acordo com diferentes faixas etárias. MÉTODOS: Nove pacientes voluntários sadios, sem história de doença pulmonar, foram estudados, sendo três adultos e seis crianças de diferentes faixas etárias. Através de verificação qualitativa e semiquantitativa da deposição de 99m tecnécio-fitato em vias aéreas, procedeu-se a análise comparativa entre dois espaçadores de pequeno volume (Aerochamber® e Inal-Air®) e um espaçador de grande volume (Flumax®). Cada paciente recebeu inalação de 99m tecnécio-fitato. O espaçador foi preenchido, durante 30 segundos, por radioaerossóis impulsionados por fluxo de oxigênio. Por 10 segundos, os pacientes procederam à inalação com máscara acoplada ao espaçador. Por 4 minutos, realizou-se contagem de radiação emitida nas faces anterior e posterior do tórax. Realizou-se, ainda, contagem de radiação presente no interior de cada espaçador. RESULTADOS: Na avaliação semiquantitativa da deposição pulmonar, verificou-se que, em todas as faixas etárias, houve uma menor deposição percentual quando se utilizou o espaçador de grande volume (Flumax®), sendo esta diferença relativa inversamente proporcional à faixa etária. Em outras palavras, quanto menor a criança, menor a deposição de aerossol radioativo ao se utilizar espaçadores de grande volume. Por outro lado, não se verificou diferença significativa entre os espaçadores de pequeno volume (Inal-Air® e Aerochamber®) nas faixas etárias estudadas. CONCLUSÃO: Os dados obtidos indicam que espaçadores de pequeno volume são os mais indicados para utilização na faixa pediátrica, devendo-se restringir o uso de espaçadores de grande volume a adolescentes e adultos.

Terapia inalatória; espaçadores; 99m tecnécio-fitato


ORIGINAL ARTICLE

Evaluation of the influence of patient age and spacer device volume on aerosol lung deposition

Wilson Rocha FilhoI; Virgínia X. NoronhaII; Simone N. SennaIII; Carlos Jorge R. SimalIV; Wawrwyk B. MendonçaV

ICoordinator of the Service of Pediatric Pneumology and Allergy, Hospital Felício Rocho (HFR), Belo Horizonte, MG, Brazil

II Graduate student in Pediatric Pneumology and Allergy, Hospital Felício Rocho (HFR), Belo Horizonte, MG, Brazil

III Preceptor of Pediatric Pneumology and Allergy, Hospital Felício Rocho (HFR), Belo Horizonte, MG, Brazil

IV Coordinator of the Service of Nuclear Medicine, Hospital Felício Rocho (HFR), Belo Horizonte, MG, Brazil

V Graduate student in Nuclear Medicine, Hospital Felício Rocho (HFR), Belo Horizonte, MG, Brazil

Correspondence Correspondence to Wilson Rocha Filho Avenida do Contorno, 9530/2º andar - Pulmolab CEP 30110-130 - Belo Horizonte, MG Brazil E-mail: wrochajr@felicoop.org.br

ABSTRACT

OBJECTIVE: To evaluate the efficacy of three frequently used spacer devices to deliver aerosol to the lung, and to compare radioaerosol deposition with each device in different age groups.

METHODS: Nine healthy, non-smoking volunteers were recruited: three adults and six children, including three toddlers and three school age children. Qualitative and semi-quantitative analysis of radioaerosol deposition in the lung were carried out. Yet, two small-volume devices (Aerochamber® and Inal-Air®) and one large-volume device (Flumax®) were compared. Each patient inhaled 99mtechnetium-phytate. The device was filled during 30 seconds with radioaerosol. Oxygen was used as the driving gas. During 10 seconds, the patients inhaled the radioaerosol. The radiation emitted at the front and back of the chest was measured. The radiation inside the device was also measured.

RESULTS: The quantitative evaluation of lung deposition revealed that the younger the patient, the less aerosol was deposited in the lung with the large-volume spacer device (Flumax®). The difference between small-volume devices (Aerochamber® and Inal-Air®) was not significant.

CONCLUSION: Small-volume spacers are the most appropriate for children. Large-volume devices should only be used by adolescents and adults.

Key words: Inhalation therapy, spacers, 99mtechnetium-fitate.

Introduction

Nowadays inhalation therapy is one of the principal therapeutic weapons in the arsenal of respiratory diseases treatment. Although it was used even before the Christian era, its use has only become widespread over the last 30 to 40 years with the advent of ever more efficient devices in terms of pulmonary deposition. Added to this is a growing number of available medications, highly effective and with low incidence of collateral effects. Inhalation therapy by metered-dose aerosols is quick and easy to administrate. In addition to this it offers better pulmonary deposition with a lower cost than conventional nebulizers. One obstacle that is encountered when metered-dose aerosols are employed is the need for good patient coordination, in order that the medication is adequately deposited within the lungs.

Spacing devices are valved devices capable of overcoming these difficulties. The correct choice and adequate utilization of spacing devices is a pre-requisite for the successful treatment of asthma in children. Spacing devices facilitate medication deposition within the lungs and eliminates the need for precise coordination, which is hard to achieve with certain patients.1-3 Over recent years, a variety of different spacing devices have become available on the market, with bivalve versions being most effective in the attempt to maximize the availability of the medication to the patient. It is known that their physical characteristics, such as size, format, volume and electrostatic charge directly affect this availability.4

One of the main dilemmas in inhalation therapy is choosing a spacer that unites the ideal characteristics that lead to the greatest efficacy in terms of the pulmonary deposition of the medication inhaled from among those most employed within the health services. Nowadays we have innumerable spacing devices available in Brazil, of varying shapes and sizes. However, we lack studies testing their efficacy. Rubim et al., employed scintigraphic techniques with the objective of evaluating a large volume spacer.5 Studies using radioisotopes for the comparative evaluation of spacing device efficacy in terms of pulmonary deposition have not yet been performed in our country. Starting from these assumptions, the study in question has the objective of comparing, by means of clinical trial, three of the most often prescribed spacing devices at health services: Flumax®, a large volume, plastic spacer (700 ml); Aerochamber®, a small volume, acrylic space (190 ml); and the Inal-Air®, small volume, aluminum, and, therefore, electrostatically uncharged, spacer (230 ml).

Methods

Nine healthy, non-smoking patients with no history of respiratory diseases were studied. Three adults and six children were chosen, three pre-school age (3 and 4 years) and three school-age (6 and 7 years), who voluntarily offered to participate in the study, with the signed authorization given by the candidate themselves or their legal guardian. The study and consent form were approved by the Ethics Committee at our institution.

Aerosol particles containing 99mtechnetium-phytate were impelled into the spacing devices for 30 seconds by an oxygen flow at 10 l/min. Images were made of the spacing devices filled with radioaerosol using scintigraphy (Figure 1). The patient then inhaled the 99mtechnetium-phytate particles. Inhalation was sustained for 10 seconds, smoothly and by mouth, using a mask coupled to the spacer, a similar technique to that employed by Rubim et al.5 After inhalation, with the patient within the gamma-chamber, the radiation count was begun, of emissions from both the anterior and posterior thorax. The count was sustained for 2 minutes for each projection. Gamma radiation was detected and counted by the standardized technique for pulmonary inhalation scintigraphy, modified for this study. To perform this an Aerogama Medical nebulization system was used with a Siemens gamma-chamber, model Orbiter, connected to an Alfanuclear, IM512P image acquisition and processing system.6 The pulmonary deposition of the radioaerosol was estimated as a percentage of the total radioactive activity seen in the spacer.


Each patient performed the procedure three times, one for each spacer, with the order being randomized by simple lottery. The procedures were performed on different days, at a minimum of 72-hour intervals apart, in order to permit complete elimination of the radioactive material inhaled on the previous occasion. On each occasion the patients were requested to inhale deeply for approximately 10 seconds, which is the equivalent of three or four breaths. These breaths should be by mouth, smooth and homogeneous. The total radiation received by each patient was less than they would have been subjected to during chest tomography.

Data were analyzed using exact inference and the Friedman test and the Wilcoxon-Mann-Whitney test were applied, with a significance level 5%, i.e. p value of less than 0.05, for an adequate analysis of data (Table 1). Taking into account the 5% significance level, the sample size9 and the mean values and standard deviation between treatments, we obtained a statistical power of approximately 80%.

Results

Employing the methods described above, the aerosol pulmonary deposition index was calculated for each spacing device. Thus, each individual was analyzed three times, each time with a different spacer, making a total sample of 27 events.

Significant variation was observed in terms of pulmonary deposition, with a large standard deviation, from patient to patient and from spacer to spacer (Table 1). In eight of the nine patients, irrespective of age group, there was less pulmonary deposition when the large volume spacer was used (Flumax®), compared with the small volume ones (Aerochamber® and Inal-Air®). The children exhibited significant differences in pulmonary deposition between small and large volume spacing devices, with significantly greater deposition occurring when the Aerochamber® and Inal-Air® small volume spacing devices were used. In these children, pulmonary deposition was statistically greater using the Inal-Air small volume, uncharged spacer when compared with the Flumax® large volume spacer with electrostatic charge, with a p value of 0.0313 (Figure 2). Nevertheless, pulmonary deposition was similar for the adults, with no significant difference being found between the three spacing devices used for this age group. Comparing the Inal-Air and Aerochamber small volume spacing devices, we found that there was no statistically significant difference in pulmonary deposition (p = 0.2188).


Discussion

In world literature studies using radioisotopes to assess the efficacy of spacers remain scarce and with disparate results.5 Pedersen et al. showed deposition of 4 to 8% of aerosols inhaled via spacing devices and nebulizers.7 In adult patients, studied by Dolovich et al., deposition by metered-dose aerosols varied from 7 to 14%.4 It is important to note that in all these studies, including ours, there is underestimation of pulmonary aerosol deposition when radioactively marked material is employed, due to radiation absorption by thoracic structures.

Our study has a number of limitations. Among these is the small number of patients studied. Despite this the statistical power was calculated for around 80%. Pulmonary deposition may be different with chronic respiratory disease patients. Asthmatic children, during an acute crisis, or even during the inter-crisis period, may exhibit a pulmonary deposition pattern that is different from normal patients. We would also point out that the observed differences in pulmonary deposition do not necessarily indicate greater clinical efficacy. In order to determine this, studies with greater patient numbers which assess the clinical effectiveness of each spacer.

Spacer volume may impact on the availability of medication for inhalation, which itself may vary depending on the medication employed. Clinicians should be aware that deposition data for a given spacer, originating from studies of a particular drug may not be applicable to others.4,8,9 It is estimated that the tidal volume of children is around 8 to 10 ml/kg.4 A four year-old child weighing 20 kg would therefore have tidal volume of between 160 and 200 ml. It is not difficult to imagine that such a child, using a large volume spacer (500 to 800 ml), would not manage to inhale its entire contents. The child's respiratory capacity is insufficient and does not permit the inhalation of the whole contents. Furthermore,, after a few seconds, progressive decantation of the medication occurs within the spacer reservoir and this portion is no longer available to the patient.9 We had the opportunity to document this finding when analyzing the images of the spacing devices filled with radioisotopes (Figure 1). Greater deposition is clearly observable within the larger volume spacer, due to the decantation of this aerosol. Another important factor to take into account is the electrostatic charge present on the inner surface of each spacer.2,10 Synthetic spacing devices present a negatively charged surface. Aerosols have a positive charge, which fact facilitates their adherence to the wall of the reservoir. The electrostatic charge, therefore, reduces aerosol deposition in the lungs. This is inversely proportional to air humidity and has a greater effect in small volume spacers. While washing with neutral detergent reduces the charge a dry cloth will increase it considerably. On the other hand, the electrostatic charge is avoided when metal spacers are used. Wildhaber et al. demonstrated that in the absence of the charge, plastic spacing devices attain the same level of efficacy as metal ones.10 Our study was not capable of discerning any statistical difference in pulmonary deposition when we compared spacers with and without electrostatic charges.

We assessed two spacing devices from the domestic market (Flumax® and Inal-Air®) and one imported model (Aerochamber®), internationally recognized for its pulmonary deposition adequacy. We found that domestically sourced spacing devices present deposition indices that are comparable with international standards, with the Inal-Air® small volume uncharged spacer exhibiting superior results to those of the Aerochamber®, particularly in children. Our study clearly demonstrates that small volume spacing devices are superior to those of large volume. We demonstrate that the benefits of small volume spacing devices are greater the smaller the child and that spacing devices with no electrostatic charge exhibit superior pulmonary deposition than the others. These findings are in agreement with Tal et al. who studied radioactively marked salbutamol delivered via a plastic spacer without reducing the electrostatic charge.9

The ideal spacer should permit the greatest possible pulmonary deposition of the medication being administered. To this end, small volume spacing devices with the lowest possible internal surface charge are to be considered the most adequate. When we compared the Inal-Air® and Aerochamber® small volume spacing devices we found better deposition with the former, particularly with pre-school aged children due to the absence of electrostatic charge. The results obtained in this study confirm data from earlier studies. Additionally, we documented the efficacy of low cost, Brazilian spacing devices which offer pulmonary deposition similar to internationally recognized devices. Large volume spacing devices should be restricted to adolescents and adults.

Acknowledgements

The authors would like to thank Yáskara Ferreira, Dra. Marisa Lages, Dra. Juliana Barros, Dra. Rosiléia Alves, Julienne (statistician), Patrícia Lapertosa and the Felicoop team.

References

Manuscript received Dec 12 2003, accepted for publication Jun 30 2004.

  • 1. Forte R Jr, Dibble C. The role of international environmental agreements in metered-dose inhaler technology changes. J Allergy Clin Immunol 1999;104:217-20.
  • 2. Kamps A, Ewijk B, Roorda RJ, Brand PLP. Poor inhalation technique, even after inhalation instructions, in children with asthma. Ped Pulmonol. 2000;29:39-42.
  • 3. Coates AL, Ho SL. Drug administration by jet nebulization. Ped Pulmonol. 1998;26:412-23.
  • 4. Closa RM, Ceballos JM, Gomes-Papí A, Galiana AS, Gutierrez C, Martí-Henneber C. Efficacy of bronchodilators administered by nebulizers versus spacer devices in infants with acute wheezing. Ped Pulmonol. 1998;26:344-8.
  • 5. Reinsner C, Katial RK, Bartelson BB, Buchmeir A, Rosenwasser LJ, Nelson HS. Characterization of aerosol output from various nebulizer/compressor combinations. Ann Allergy Asthma Immunol. 2001;86:566-74.
  • 6. Duarte M, Camargos P. Efficacy and safety of a home-made non-valved spacer for bronchodilator therapy in acute asthma. Acta Paediatr. 2002;91:909-13.
  • 7. O'Callaghan C, Barry P. How to choose delivery devices for asthma. Arch Dis Child. 2000;82:185-91.
  • 8. Karpel JP, Aldrich TK, Prezant DJ, Guguchev K, Gaitan-Salas A, Pathiparti R. Emergency treatment of acute asthma with albuterol metered-dose inhaler plus holding chamber. Chest. 1997;112:348-56.
  • 9. Dolovich MB. Aerosols. In: Barnes PJ, Grunstein MM, editors. Asthma. Philadelphia: Lippincott-Raven; 1997. p. 1349-65.
  • 10. Rubim JA, Simal CJR, Lasmar LMLBF, Camargos PAM. Deposição pulmonar de radioaerossol e desempenho clínico verificado com espaçador fabricado no Brasil. J Pediatr (Rio J). 2000;76:434-42.
  • 11. Mallol J, Rattray S. Aerosol deposition in infants with cystic fibrosis. Pediatr Pulmonol. 1996;21:276-81.
  • 12. Tal A, Golan H. Deposition pattern of radiolabeled salbutamol inhaled from a metered-dose inhaler by means of a spacer with mask in young children with airway obstruction. J Pediatr. 1996;128:479-84.
  • 13. Pedersen S. Delivery systems in children. In: Barnes PJ, Grunstein MM, editors. Asthma. Philadelphia: Lippincott-Raven; 1997. p. 1918-9.
  • 14. Thompson PJ. Drug delivery to small airways. Am J Respir Crit Care Med. 1998;157:199-202.
  • 15. Wildhaber J, Devadason S, Hayden MJ, Eber E, Summers QA, LeSouef PN. Aerosol delivery to wheezy infants: a comparison between a nebulizer and two small volume spacers. Ped Pulmonol. 1997;23:212-16.
  • Correspondence to

    Wilson Rocha Filho
    Avenida do Contorno, 9530/2º andar - Pulmolab
    CEP 30110-130 - Belo Horizonte, MG
    Brazil
    E-mail:
  • Publication Dates

    • Publication in this collection
      01 Dec 2004
    • Date of issue
      2004

    History

    • Accepted
      30 June 2004
    • Received
      12 Dec 2003
    Sociedade Brasileira de Pediatria Av. Carlos Gomes, 328 cj. 304, 90480-000 Porto Alegre RS Brazil, Tel.: +55 51 3328-9520 - Porto Alegre - RS - Brazil
    E-mail: jped@jped.com.br