A validated method to quantify folic acid in wheat flour samples using liquid chromatography: tandem mass spectrometry

Abstracts

In order to monitor the fortification level of folic acid (vitamin B9) in wheat flour samples, we developed and validated a simple and fast analytical method using the liquid chromatography with tandem triple quadrupole mass spectrometry (LC-MS/MS) technique. The compound was ionized by an electrospray source in positive mode. The tandem MS/MS instrument was set in MRM (Multiple Reaction Monitoring) mode to detect two m/z transitions used for quantitation and confirmation purposes. A total of 83 samples were prepared to validate the method according to the regulatory guidelines from the Commission Decision 2002/657/EC for food residues analysis. The method performance was evaluated in regards to linearity, accuracy, precision, sensitivity, and uncertainty. The statistical analysis of the results showed a correlation coefficient (r²) of 1.000 within the concentration range of 1.0 to 50 µg L-1. The method reproducibility of 5.6% was obtained by calculating the standard deviation sum of the intra-day and inter-day analysis at a fortification level limit of 1.50 µg g-1 of folic acid. In addition, ten blank spiked samples at the fortification level limit were extracted and analyzed with an accuracy of 85% and a precision of 3% (coefficient of variation). The method sensitivity was expressed in decision limit (CCα) of 0.06 µg g-1 and detection capability (CCβ) of 0.11 µg g-1. The total measurement uncertainty method calculated was + 0.11 µg g-1. Different brands of Brazilian wheat flour were evaluated and reported.

folic acid; validation; liquid chromatography; tandem mass spectrometry


Com o intuito de monitorar a concentração de ácido fólico (vitamina B9) em amostras de farinha de trigo, um método analítico simples e rápido foi desenvolvido e validado, utilizando a técnica de Cromatografia Líquida de Alta Eficiência com detecção por Espectrometria de Massas do tipo triplo quadrupolo (CLAE-EM/EM). O analito foi ionizado com fonte de ionização por electrospray no modo positivo e o espectrômetro de massas foi operado em modo de aquisição por Monitoramento Múltiplo de Reação (Multiple Reaction Monitoring - MRM), com o propósito de detectar duas trannsições de m/z, uma para quantificação e outra para confirmação. Um total de 83 amostras foram processadas para a validação do método de acordo como guia de procedimentos da Comissão Européia, referência 2002/657/EC, para análise de resíduos em alimentos. A performance do método foi avaliada através da linearidade, exatidão, precisão, sensibilidade e incerteza. O resultado das curvas analíticas na faixa de concentração de 1,0 a 50 µg L-1 apresentaram coeficientes de correlação (r²) de 1,000. A reprodutibilidade do método foi obtida através do cálculo da somatória dos desvios padrão dos resultados de análises "intra-dias" e "inter-dias" e apresentou o valor de 5,6%, para o nível de fortificação de 1,50 µg g-1. Além disso, dez amostras no nível de fortificação de 1,50 µg g-1 foram extraídas e analisadas com exatidão de 85% e precisão de 3% (coeficiente de variação). A sensibilidade do método foi expressa em limite de decisão (CCα) de 0,06 µg g-1 e capacidade de detecção (CCβ) de 0,11 µg g-1. A incerteza total calculada do método foi de + 0,11 µg g-1. Diferentes marcas de farinha de trigo brasileiras foram avaliadas e reportadas.


ARTICLE

A validated method to quantify folic acid in wheat flour samples using liquid chromatography - Tandem mass spectrometry

Helio A. Martins-JúniorI, * * e-mail: daniel.lebre@sciex.com ; Alexandre Y. WangI; Janete AlabourdaII; Maria A. F. PiresIII; Oscar B. VegaIII; Daniel T. LebreIV, * * e-mail: daniel.lebre@sciex.com

IApplied Biosystems do Brazil Ltda., Av. do Café, 277, 1º andar Torre B, 04311-000 São Paulo-SP, Brazil

IIInstituto Adolfo Lutz, Departamento de Bromatologia e Química, Av. Dr. Arnaldo, 355, 01246-902 São Paulo-SP, Brazil

IIIInstituto de Pesquisas Energéticas e Nucleares, Centro de Química e Meio Ambiente (CQMA), Av. Lineu Prestes, 2258, Cidade Universitária, 05508-000 São Paulo-SP, Brazil

IVApplied Biosystems/MDS Sciex, 71 Four Valley Drive, L4K 4V8, Concord-ON, Canada

ABSTRACT

In order to monitor the fortification level of folic acid (vitamin B9) in wheat flour samples, we developed and validated a simple and fast analytical method using the liquid chromatography with tandem triple quadrupole mass spectrometry (LC-MS/MS) technique. The compound was ionized by an electrospray source in positive mode. The tandem MS/MS instrument was set in MRM (Multiple Reaction Monitoring) mode to detect two m/z transitions used for quantitation and confirmation purposes. A total of 83 samples were prepared to validate the method according to the regulatory guidelines from the Commission Decision 2002/657/EC for food residues analysis. The method performance was evaluated in regards to linearity, accuracy, precision, sensitivity, and uncertainty. The statistical analysis of the results showed a correlation coefficient (r2) of 1.000 within the concentration range of 1.0 to 50 µg L-1. The method reproducibility of 5.6% was obtained by calculating the standard deviation sum of the intra-day and inter-day analysis at a fortification level limit of 1.50 µg g-1 of folic acid. In addition, ten blank spiked samples at the fortification level limit were extracted and analyzed with an accuracy of 85% and a precision of 3% (coefficient of variation). The method sensitivity was expressed in decision limit (CCα) of 0.06 µg g-1 and detection capability (CCβ) of 0.11 µg g-1. The total measurement uncertainty method calculated was + 0.11 µg g-1. Different brands of Brazilian wheat flour were evaluated and reported.

Keywords: folic acid, validation, liquid chromatography, tandem mass spectrometry

RESUMO

Com o intuito de monitorar a concentração de ácido fólico (vitamina B9) em amostras de farinha de trigo, um método analítico simples e rápido foi desenvolvido e validado, utilizando a técnica de Cromatografia Líquida de Alta Eficiência com detecção por Espectrometria de Massas do tipo triplo quadrupolo (CLAE-EM/EM). O analito foi ionizado com fonte de ionização por electrospray no modo positivo e o espectrômetro de massas foi operado em modo de aquisição por Monitoramento Múltiplo de Reação (Multiple Reaction Monitoring - MRM), com o propósito de detectar duas trannsições de m/z, uma para quantificação e outra para confirmação. Um total de 83 amostras foram processadas para a validação do método de acordo como guia de procedimentos da Comissão Européia, referência 2002/657/EC, para análise de resíduos em alimentos. A performance do método foi avaliada através da linearidade, exatidão, precisão, sensibilidade e incerteza. O resultado das curvas analíticas na faixa de concentração de 1,0 a 50 µg L-1 apresentaram coeficientes de correlação (r2) de 1,000. A reprodutibilidade do método foi obtida através do cálculo da somatória dos desvios padrão dos resultados de análises "intra-dias" e "inter-dias" e apresentou o valor de 5,6%, para o nível de fortificação de 1,50 µg g-1. Além disso, dez amostras no nível de fortificação de 1,50 µg g-1 foram extraídas e analisadas com exatidão de 85% e precisão de 3% (coeficiente de variação). A sensibilidade do método foi expressa em limite de decisão (CCα) de 0,06 µg g-1 e capacidade de detecção (CCβ) de 0,11 µg g-1. A incerteza total calculada do método foi de + 0,11 µg g-1. Diferentes marcas de farinha de trigo brasileiras foram avaliadas e reportadas.

Introduction

Vitamin B9, chemically known as either folic acid or pteroylglutamic acid, is one of the metabolically active compounds and is commonly referred to as folates. This water soluble vitamin helps the body to convert carbohydrates into glucose to produce energy. The liver converts the folates in N5-methyltetrahydrofolate, a major circulating form in blood, that is responsible for the one-carbon transfer enzymatic reactions (homocysteine to yield methionine), including the DNA synthesis. The absence of folates in the human body can cause megaloblastic anemia, and affect mental and emotional health, along with other biological effects. As well, the deficiency of this compound during pregnancy can increase the risk for neural tube birth defects including cleft palate, spina bifida, and brain damage. In general, vitamin B9 can be ingested through a diet that is rich in green and leafy vegetables, fruits, and organ meats, though many countries add folic acid in cereal and grain foods, including flours, to enrich their diet.1-3

In 1996, the U.S. Department of Health and Human Service through the Food and Drug Administration (FDA) specified that grain products required to be fortified with folic levels ranging from 0.43 mg to 1.4 mg per pound (from 0.95 to 3.10 µg g-1) of the product and all women of child-bearing age should consume 0.4 mg of folic acid daily.4 In Brazil, the Minister of Health, through ANVISA resolution RDC nº 344 from December 13th, 2002 approved the addition of folic acid in wheat and corn flours at minimum concentration level of 150 µg per 100 g (1.50 µg g-1).5

In 2004, the ANVISA released a report with the aim to identify laboratories that analyze nutrients, including folic acid, in food. From 26 laboratories located in Brazil, 61.5% (16) analyze folic acid. Among them, 10 laboratories use HPLC-UV technique, 4 use microbiological methods, 1 uses both approaches, and none of them use the LC-MS/MS technique.6

Traditionally, laboratories follow the AOAC microbiological methods, 944.12, 960.46 and 992.05, to determine total folate levels in foods, dietary supplements, and infant formulas.7 An alternative method is to use spectrophotometer or HPLC-Ultraviolet (UV) detectors, after tedious and time-consuming sample preparation, as thermal extraction, enzymatic (conjugase and amylase) incubation and purification.8,9 Other methods were developed using trienzyme sample treatment (conjugase, alfa-amylase, and protease) followed by affinity chromatography and HPLC separation coupled with UV or fluorescence (FL) detectors.10,11 The lack of specificity and sensitivity opened new opportunities for LC-MS and LC-MS/MS techniques approach.

Stokes and Webb analyzed some folate compounds in multivitamin tablets and breakfast cereal containing 200 µg per100 g. The authors applied a simple extraction procedure with solid-phase extraction clean-up, followed by LC-MS detection of the pseudo-molecular ions in negative electrospray mode.12 In 2003, Freisleblen et al. published an article comparing the quantification of folate in food using the two different detection techniques: LC-FL and stable isotopic dilution assays with LC-MS/MS. The LC-MS/MS revealed superior selective, precision and sensitivity.13 Later on, an inter laboratory comparative study was presented by Puwastien et al. using different detection strategies to measure folate levels in food, including LC-UV and LC-MS techniques.14 Spanning from early 2000 to present day, some papers reported the usage of LC/MS or LC-MS/MS (triple quadrupole) detectors to quantify folates in different matrices (food, plasma, and vitamins), although none of them are specific to analyze folic acid in wheat flour samples. 15-20

This present work shows a novel and validated method to quantify folic acid in wheat flour with a high degree of specificity and confidence using the LC-MS/MS technique.21 A simple and fast extraction procedure was developed to save time and cost. To obtain confident results the method was validated according to the regulatory guidelines from the Commission Decision 2002/657/EC, concerning the performance of analytical methods and the interpretation of results.22

Experimental

Reagents

Folic acid was purchased from Sigma-Aldrich, (St. Louis, MO, USA) with 99% purity. Ultrapure water was produced by EASYpure RF System from Barnstead (Dubuque, IA, USA). methanol HPLC grade, ammonium acetate and ammonium hydroxide were purchased from J. T. Baker, (Deventer, NE, USA).

Folic acid standard solutions

A 100 µg mL-1 stock solution of folic acid was prepared by accurately weighing 1.0 mg of the yellow standard powder, dissolving it in a beaker of 5.0 mL of ultra pure water, and using the necessary volume of ammonium hydroxide (conc.) to adjust the pH in between 7.0 to 8.0. After the powder was dissolved completely, the solution was quantitatively transferred into a 10 mL amber volumetric flask and stored at 4 ºC away from light. An intermediary solution of folic acid at 1.0 µg mL-1 in water was obtained by dilution of the stock solution. All solutions for the analytical curve were prepared from this intermediary solution at the following concentrations: 1.0, 5.0, 10, 25 and 50 ng mL-1 in water solution containing 5.0 mmol L-1 of ammonium acetate.

Sample preparation

2.0 g of wheat flour sample was weighed and transferred into a 50 mL polypropylene centrifuge tube, 40 mL of 75 mmol L-1 ammonium acetate solution (pH=7) was added, sonicated for 10 min and mixed in an orbital shaker for 60 min. After the extraction, the samples were centrifuged for 15 min at 4,000 rpm and 50 µL of the supernatant was taken and diluted with 700 µL of water solvent straight into a 2.0 mL amber vial. Volume of 20 µL of the sample solution was injected into the LC-MS/MS system. It is important to keep the sample tubes away from UV light to avoid any degradation.

Method validation

A total of 83 samples were prepared to validate the proposed methodology. The study was divided into four experiments and executed in four consecutive days to evaluate the inter-day and intra-day variations. The first three experiments included a batch of 1 blank sample, 20 blank spiked samples (63 totals) and the respective analytical curve. The blank spiked samples were analyzed in 5 different concentration levels of folic acid: 6 replicates samples at 0.75 µg g-1; 6 replicates samples at 1.50 µg g-1; 6 samples at 2.25 µg g-1; 1 sample at 3.00 µg g-1 and 1 sample at 7.5 µg g-1. In the forth experiment, 10 blank samples, 10 blank spiked samples at the concentration level of 1.5 µg g-1 and the analytical curve were analyzed. All blank spiked samples were prepared by adding the necessary volume from a standard stock solution of folic acid at a concentration level of 100 µg mL-1. After the standard addition, these samples were homogenized under circular agitation for 1 h and extracted.

LC-ESI/MS/MS

LC analysis was performed with an autosampler (20 µL injection volume), a column oven (set to 25 ºC) and a quaternary pump from the Agilent 1100 series (Agilent Technologies, Waldbronn, GE). For the separation a reverse phase column, Zorbax Eclipse XDB-C8 (150 mm × 4.6 mm, 5 µm; Agilent, California, USA) was used. LC elution conditions were as follows: mobile phase A = 5.0 mmol L-1 ammonium acetate and mobile phase B = methanol with 5.0 mmol L-1 ammonium acetate; isocratic mode: 25% mobile phase A and 75% mobile phase B; flow rate of 700 µL min-1; total run time of 3.0 min. The mass spectrometric analyses were performed on an API 4000TM (Applied Biosystems/MDS Sciex, Concord, Canada) with a TurboIonSpray® source operated in positive ion mode. The electrospray source parameters were as follows: Curtain GasTM = 10 psi, nebulizer gas = 50 psi, heater gas = 40 psi, heater temperature = 600 ºC, and electrospray voltage = + 5500V. The MRM mode was used to detect the target compounds in the triple quadrupole stage and the optimum conditions are described in Table 1. The Collision Gas (nitrogen) was maintained at 6.0 arbitrary units.

Results and Discussion

Mass Spectrometry optimization

To optimize the MRM mode, a folic acid solution at concentration of 1.0 µg mL-1 (50% methanol + 50% water + 5 mmol L-1 ammonium acetate) was infused into the electrospray ion source (positive mode) at a flow rate of 10 µL min-1. With the advances on software interface it was possible to tune the instrument automatically in a few min. Figure 1 summarizes the mass spectrometry automatic steps necessary for maximum MRM transitions sensitivity, and its optimum parameters.


The two most intense transitions were chosen to build the final MRM mode. The transitions 442.2>295.2 and 442>176.2 were defined as quantifier and qualifier (confirmation), respectively.

After the MRM optimization, the electrospray ion source parameters were set to operate with maximum efficiency at mobile phase flow rate of 700 µL min-1. Folic acid is an organic acid and its dissociation constants (pKa) are 4.7, 6.8, and 8.7. At a pH level lower than 4.7 in aqueous solutions the protonated form (insoluble form below 7.0) is predominant and a pH level above 8.7 the deprotonated form (soluble) is predominant. The physicochemical information helped to chose the electrospray ionization polarity. The literature shows that it is possible to ionize folic acid in both positive and negative modes, suggesting that the amino group present in the structure is protonated in positive mode [M+H]+ and deprotonated in negative mode [M-H]-.18,19 Herein the pH of the mobile phase solutions was maintained at 7.0, adding 5.0 mmol L-1 of ammonium acetate in the mobile phase. The positive ionization mode was used to detect the [M+H]+ species.

Method optimization

For chromatographic methods, developing a separation involves demonstrating specificity, which is the ability of the method to accurately measure the analyte response in the presence of all potential sample components.23 One main advantage of the triple quadrupole mass spectrometry detector is the mechanism's ability of high compound selectivity. Therefore, the liquid chromatography (LC) separation was optimized very quickly using a C8 reverse-phase column, obtaining a sharp peak, low retention time (1.70 min), without matrix interferences. Figure 2 presents a LC-ESI/MS/MS at the optimum analyzed run conditions.


Method validation is the process of proving that an analytical method is acceptable for its intended purpose.23 The developed methodology to quantify folic acid in wheat flour sample was validated according to the Commission Decision 2002/657/EC criteria, proceeding as described in the experimental section.22 A total of 83 samples were analyzed, besides the calibration curves, to evaluate the analytical performance characteristics of specificity, accuracy, precision, repeatability, reproducibility, recovery, detection limits and uncertainty using the proposed method.

This experiment produced relative amounts of data that we summarized herein, showing some important figures of merit. The following results presented below are based on the most intense m/z transition (442.2>295.2) signal, chosen as a quantifier.

Linearity

Overall, the acceptability of linearity data is often judged by examining the correlation coefficient and y-intercept of the linear regression line for the response versus concentration plot (calibration line y = ax + b). A correlation coefficient of > 0.999 is generally considered as evidence of acceptable fit of the data to the regression line. The y-intercept should be less than a few percent of the response obtained for the analyte at the target level.23Table 2 presents the results of four analytical curves built for each experiment or batch analysis (inter-day, see experimental section). The concentration of folic acid in standard aqueous solution range from 1.0 to 50 ng mL-1, which corresponds from 0.3 to 15 µg g-1 in wheat flour samples (considering the extraction procedure volumes).

Sample extraction efficiency: correlation, recoveries and co-variations

In order to evaluate the correlation among the sample extraction efficiency, one curve was built for each day or experiment (calculated concentration vs. area signal). Table 3 shows the correlation of the extraction procedure at folic acid concentrations ranging from 0.75 µg g-1 to 7.5 µg g-1. Figure 3 shows the LC-MS/MS chromatograms of the samples extracted and analyzed at different validation level concentrations.


The accuracy was measured in terms of recovery, spiking the analyte in blank wheat flour matrices at different concentration levels (from 0.75 to 7.5 µg g-1), before the extraction process. The sample final concentration was calculated through the analytical curve built in solvent. According to the 2002/657/EC guidelines for method validation, the data corrected with the mean recovery are only acceptable when they fall within the followingranges: < 1.0 µg kg-1 from 50 to 120%; > 1.0 to 10.0 µg kg-1 from 70 to 110% and > 10.0 µg kg-1 from 80 to 110%. The last value was used as the recovery criteria. The Table 4 presents the intra-day recoveries and the coefficient of variation that did not exceed the recommendation set value of 10%, considering the concentration range validated for this analysis.

The method stability or reproducibility was verified thorough intra-day (sintra2) and inter-day (sinter2) variances (Table 5). The same LC-ESI/MS/MS conditions were maintained during the three days of experiments, except the samples batch that was prepared fresh daily, as described in the experimental section.

The last experiment analyzed ten blank samples and ten blank spiked samples at the validation level limit of 1.5 µg g-1. Table 6 presents the statistical results. The recovery value suggests that analyte loses just 15% of the signal, represented by matrix effect and/or extraction efficiency.

Sensitivity

Establishing the lowest analyte concentration that produces a response detectable above the noise level of the system, and corresponds to the limit of detection (LOD) and the limit of quantitation (LOQ) (which is the lowest level of analyte that can be accurately and precisely measured), were important parameters to be validated in quantitative methods. Originally, the 002/657/EC guidelines for method validation recommend the usage of decision limit (CCα) and detection capability (CCβ) to measure sensitivity.22 The method sensitivity was calculated using the following equations based on Table 3 data result and its inter-day variation:

The developed method to detect and quantify folic acid in wheat flour samples showed a CCα of 0.06 µg g-1 and CCβ of 0.11 µg g-1. These values were way below the validation method limit and the established folic acid fortification level in wheat flours of 1.5 µg g-1, which makes this methodology suitable for its purpose.

Quantitation in wheat flour samples

Before sample quantitation, the measurement uncertain (U) was calculated as 0.11 µg g-1 (11 µg per 100g) and all results shown are expressed as ± U. A total of four samples from different brands were analyzed in order to determine the folic acid concentrations. Table 7 shows the results and the variances among different brands analyzed. Figure 4 shows a chromatogram analysis from sample FT14. It was observed that the FT07 sample was way out of the minimum fortification level of 150 µg per 100g as recommended by ANVISA resolution.5 Even though, the sample FT08 showed slight higher levels of folic acid, the same does not represent any risk to human health, since the oral daily doses of 5-10 mg of folic acid appear to be well-tolerated and without toxicity in normal non-pregnant subjects.24 Overall, all samples analyzed herein are in agreement with the fortification range levels of folic acid in wheat flour suggested by FDA guidelines.4


Conclusions

We presented an alternative and effective way to analyze folic acid, which is an important dietary nutrient, in wheat flour. The use of LC-ESI/MS/MS technique allowed the determination of this compound at the required set limits (sensitivity), without matrix interferences (selective), time consuming (rapid) and tedious sample preparation, comparing with previous published works. The method was validated and applied to analyze some samples collected from Brazilian markets. Finally, this work can support and help the laboratories enrolled in food analysis for future quality control programs.

Acknowledgments

The authors would like to thank all the staff from Applied Biosystems of Brazil and the Center of Chemistry and Environmental at IPEN. We are also grateful to the researchers from Instituto Adolfo Lutz (IAL-SP). A very special thanks to Dr. Takeo Sakuma, senior researcher from MDS Sciex, who reviewed this work and contributed with his well respect knowledge in mass spectrometry and chemistry science.

Received: July 23, 2007

Web Release Date: May 29, 2008

FAPESP helped in meeting the publication costs of this article.

  • 1. Kaplan, L. A; Pesce, A. J.; Kazmierczak, S. C.; Clinical Chemistry: Theory, Analysis, Correlation, 4th ed., 2003, Mosby, Inc. Elsevier Health Science, PA, USA,747.
  • 2
    http://www.umm.edu/ altmed/ConsSupplements/Print/VitaminB9FolicAcids.html, accessed November 2006.
    » link
  • 3
    http://www.scpineuniverse.com/ displayarticle.php/article1154.html, accessed November 2006.
    » link
  • 4
    HHS News: Folic Acid to Fortify U.S. Food Products to Prevent Birth Effect, http://www.fda.gov/ bbs/topics/NEWS/NEW00526.html, accessed November 2006.
  • 5
    Resolução - RDC Nº 344, de 13 Dezembro de 2002, http://e-legis.bvs.br/ leisref/public/showAct.php?id=1979, accessed November 2006.
  • 6. Silva, E. M. C; Sousa, A. T.; Mandai, M. M, Avaliação Preliminar da Capacidade Laboratorial na determinação de Ácido Fólico, Ferro Total, Glúten e Ácidos Graxos Trans em Alimentos, Reblas, ANVISA, October 2004.
  • 7. DeVries, J. W.; Rader, J. I.; Keagy, P. M.; Hudson, C. A.; Angyal, G.; Arcot, J.; Castelli, M.; Doreanu, N.; Hudson, C.; Lawrence, P.; Martin, J.; Peace, R.; Rosner, L.; Strandler, H.S.; Szpylka, J.; van den Berg, H.; Wo, C.; Wurz, C.; J. AOAC Int. 2005, 5, 88.
  • 8. Osseyi, E. S.; Wehling, R. L.; Albrecht, J. A.; J. Chromatogr. A. 1998, 235, 826.
  • 9. Ruggeri, S.; Vahteristo, L. T.; Aguzzi, A.; Finglas. P.; Carnovale, E.; J. Chromatogr. A. 1999, 237, 855.
  • 10. Konings, E. J.; J. AOAC Int. 1999, 119, 82.
  • 11. Poo-Pietro, R.; Haytowitz, D. B.; Holden, J. M.; Rogers, G.; Choumenkovitch, S. F.; Jacques, P. F.; Selhub, J.; J. Nutr. 2006, 3079, 136.
  • 12. Strokes, P.; Webb, K.; J. Chromatogr. A. 1999, 59, 864.
  • 13. Freisleben, A.; Schieberle, P.; Rychlik, M.; Anal. Biochem. 2003, 247, 315.
  • 14. Puwastien, P.; Pinprapai, N.; Judprasong, K.; Tamura, T.; J. Food Comp. Anal. 2005, 387, 18.
  • 15. Pawlosky, R. J.; Flanagan, V. P.; J. Agric. Food Chem. 2001, 1282, 49.
  • 16. Freisleben, A.; Schieberle, P.; Rychlik, M.; Anal. Bioanal. Chem. 2003, 149, 376.
  • 17. Nelson, B. C.; Dalluge, J. J.; Margolis, S. A.; J. Chromatogr. B. 2001, 141, 765.
  • 18. Kok, R. M.; Smith, D. E. C.; Dainty, J. R.; van den Akker, J. T.; Figla, P. M.; Smulders, Y. M.; Jacobs, C.; de Meer, K.; Anal. Biochem 2004, 129, 326.
  • 19. Chen, Z.; Chen, B.; Yao, S.; Anal. Chim. Acta 2006, 169, 569.
  • 20. Nelson, B. C.; Sharpless, K. E.; Sander, L.C.; J. Chromatogr. A 2006, 203, 1135.
  • 21. Wang, A. Y.; Lebre, D. T.; Martins-Junior, H. A.; Alaburda, J. K.; J. ASMS, 2005, 98S, 16 (5S).
  • 22. Commission Decision 2002/657/EC of 12 August 2002; Off. J. Eur. Commun. 2002, L221, 8.
  • 23. Green, M. J.; Anal. Chem 1996, 68, 305A.
  • 24. Butterworth Jr., C. E.; Tamura, T.; Am. J. Clin. Nutr. 1989, 50, 353.
  • *
    e-mail:
  • Publication Dates

    • Publication in this collection
      05 Aug 2008
    • Date of issue
      2008

    History

    • Accepted
      29 May 2008
    • Received
      23 July 2007
    Sociedade Brasileira de Química Instituto de Química - UNICAMP, Caixa Postal 6154, 13083-970 Campinas SP - Brazil, Tel./FAX.: +55 19 3521-3151 - São Paulo - SP - Brazil
    E-mail: office@jbcs.sbq.org.br