Data collection about failures in fixed partial dentures: 1-year monitoring

Levantamento das falhas em prótese parcial fixa: acompanhamento de um ano

ABSTRACT

Objective:

The aim of this study was to evaluate the satisfaction of patients rehabilitated with fixed partial denture and the incidence of failures/clinical complications of prostheses after one year.

Methods:

The patients answered a questionnaire about the satisfaction degree with dental treatment performed and care maintenance for prosthesis conservation. Clinical and radiographic evaluations of the prosthesis were performed. Data were tabulated for descriptive analysis of the factors assessed, with calculation of absolute frequencies and percentages, and chi-square and Fisher’s exact test..

Results:

A total of 9.67% failures were found. The most common was the prosthesis loosening (57.14%), followed by ceramic fracture (28.57%), and abutment tooth fracture (14.29%). Biological failures were observed in 30.65%. The most common failure was gingival recession (52.00%), periodontal pocket (24.00%), support periodontal involvement (16.00%), and recurrent caries (4.00%). Radiographic examination showed that 70.97% of the total number evaluated had some kind of failure. There was statistically significant association between satisfaction degree and technical failure (p=0.04).

Conclusion:

Patients were satisfied after observation period of 1 year. The main failures detected were: crown cementation failure and ceramic fracture; gingival recession, periodontal pocket; manufacture and cementation of short intraradicular post with increased diameter and crown cervical misfit.

Indexing terms:
Patient satisfaction; Fixed partial denture; Failures

RESUMO

Objetivo:

O objetivo deste estudo foi avaliar a satisfação dos pacientes reabilitados com prótese parcial fixa e a incidência de falhas/complicações clínicas das próteses instaladas após um ano.

Métodos:

Os pacientes responderam um questionário abrangendo o grau de satisfação pelo tratamento odontológico efetuado e os cuidados de manutenção realizados para a conservação da prótese. Na sequência foi realizada avaliação clínica e radiográfica das próteses. Os dados obtidos foram tabulados para análise descritiva dos fatores avaliados com cálculo de frequência absoluta e percentual, e teste de Qui-quadrado e Fischer.

Resultados:

Foram encontradas 9,67% de falhas, sendo a mais comum a descimentação (soltura da prótese) (57,14%), seguida da fratura da cerâmica (28,57%), e um caso de fratura no dente pilar (14,29%). As falhas biológicas foram observadas em 30,65%. A falha mais comum foi a recessão gengival (52,00%), seguida por bolsa periodontal (24,00%), envolvimento periodonto de suporte (16,00%), e recidiva de cárie (4,00%). No exame radiográfico 70,97% do número total avaliado apresentaram algum tipo de falha. Houve associação estatisticamente significante entre grau de satisfação e falhas mecânicas (p=0,04).

Conclusão:

Os pacientes ficaram satisfeitos após um período de acompanhamento de 1 ano. As principais falhas detectadas foram: falha de cimentação da coroa e fratura da cerâmica; Recessão gengival, bolsa periodontal; confecção e cimentação de pino intraradicular curto, pino com diâmetro aumentado e desajuste cervical da coroa.

Termos de indexação:
Satisfação do paciente; Prótese parcial fixa; falhas

INTRODUCTION

Rehabilitation performed with Fixed Partial Denture (FPD) is one of the most accepted and desired by patients [11 Reitemeier B, Hänsel K, Kastner C, Weber A, Walter MH. A prospective 10-year study of metal ceramic single crowns and fixed dental prosthesis retainers in private practice settings. J Prosthet Dent. 2013 Mar;109(3):149-55. doi: 10.1016/S0022-3913(13)60034-7
https://doi.org/10.1016/S0022-3913(13)60...
]. However, it is a long procedure and generates high expectations from patient. Once the professional knows the factors that create dissatisfaction or contribute to failures, the dentist could minimize them and thus meet all the patient´s needs and establish the most appropriate planning [22 Layton D, Walton T. Patient-evaluated dentistry: development and validation of a patient satisfaction questionnaire for fixed prosthodontic treatment. Int J Prosthodont. 2011;24(4):332-4141.].

The success of rehabilitation treatment could be evaluated by patient satisfaction and comfort and the longevity of the prosthesis. Some factors that determine treatment satisfaction are: comfort, function and aesthetics [33 Newsome PR, Wright GH. A review of patient satisfaction: 1. Concepts of satisfaction. Br Dent J. 1999;186(4 Spec No):161-5.]. These factors are strongly linked to the professional expertise, fees, professional/patient relationship and the prostheses quality.

The quality of prostheses is related to immediate failures due to a shortage of criteria in the stages of prostheses manufacturing, such as errors in color and form selection, phonetic changes or even food impaction; as well as late failures related to biological factors such as caries, periodontal disease and endodontic complications or technical failures, such as loss of retention, cracks and subsequent fractures, loss of the coating material, metal framework fracture, spot weld, abutment tooth and marginal defects [44 Napankangas R, Raustia A. An 18-year retrospective analysis of treatment outcomes with metal-ceramic fixed partial dentures. Int J Prosthodont. 2011;24(1):314-9.

5 Svanborg P, Längström L, Lundh RM, Bjerkstig G, Ortorp A. A 5-year retrospective study of cobalt-chromium-based fixed dental prostheses. Int J Prosthodont. 2013;26(4):343-9. doi: 10.11607/ijp.3024
https://doi.org/10.11607/ijp.3024...

6 Solá-Ruiz MF, Lagos-Flores E, Román-Rodriguez JL, Highsmith Jdel R, Fons-Font A, Granell-Ruiz M. Survival rates of a lithium disilicate-based core ceramic for three-unit esthetic fixed partial dentures: a 10-year prospective study. Int J Prosthodont. 2013;26(2):175-80. doi: 10.11607/ijp.3045
https://doi.org/10.11607/ijp.3045...
-77 De Backer H, Van Maele G, De Moor N, Van den Berghe L. Long-term results of short-span versus long-span fixed dental prostheses: an up to 20-year retrospective study. Int J Prosthodont. 2008;21(1):75-85.]. These factors individually or together can influence the survival, longevity and success of the prostheses.

The main cause of prosthesis failure was lost or caries of abutment teeth and periodontal problems [44 Napankangas R, Raustia A. An 18-year retrospective analysis of treatment outcomes with metal-ceramic fixed partial dentures. Int J Prosthodont. 2011;24(1):314-9.,55 Svanborg P, Längström L, Lundh RM, Bjerkstig G, Ortorp A. A 5-year retrospective study of cobalt-chromium-based fixed dental prostheses. Int J Prosthodont. 2013;26(4):343-9. doi: 10.11607/ijp.3024
https://doi.org/10.11607/ijp.3024...
]. Caries recurrence in abutment teeth is directly related to patient hygiene and to the adjustment of prosthetic pieces by the professional. On the other hand, the periodontal disease progression may be related to deficiency in oral and general patient´s health [88 Brägger U, Hirt-Steiner S, Schnell N, Schmidlin K, Salvi GE, Pjetursson B, et al. Complication and failure rates of fixed dental prostheses in patients treated for periodontal disease. Clin Oral Implants Res. 2011;22(1):70-7. doi: 10.1111/j.1600-0501.2010.02095.x
https://doi.org/10.1111/j.1600-0501.2010...
], smoking habits and genetic factors [99 Heitz-Mayfield LJ, Huynh-Ba G. History of treated periodontitis and smoking as risks for implant therapy. Int J Oral Maxillofac Implants. 2009;24(Suppl):39-68.], besides the presence of malocclusion and bruxism [11 Reitemeier B, Hänsel K, Kastner C, Weber A, Walter MH. A prospective 10-year study of metal ceramic single crowns and fixed dental prosthesis retainers in private practice settings. J Prosthet Dent. 2013 Mar;109(3):149-55. doi: 10.1016/S0022-3913(13)60034-7
https://doi.org/10.1016/S0022-3913(13)60...
,1010 Salvi GE, Brägger U. Mechanical and technical risks in implant therapy. Int J Oral Maxillofac Implants. 2009;24(Suppl):69-85.]. The period of treatment, which is professional´s responsibility, could have an impact on supporting and protection periodontal tissues, especially in the stages of preparation, impression and prostheses contouring.

Complications resulting from rehabilitation treatment with prostheses are factors that may occur during or after treatment [1111 Bouchard P, Renouard F, Bourgeois D, Fromentin O, Jeanneret MH, Beresniak A. Cost-effectiveness modeling of dental implant vs. bridge. Clin Oral Implants Res. 2009;20(6):583-7.]. The dentist should know such complications, in order to be able to conclude a detailed diagnosis, treatment planning and execution of procedures giving special attention to the most frequent failure factors, and thus meeting the patient’s expectations and planning the post-treatment care and maintaining [1212 Ozcan M, Niedermeier W. Clinical study on the reasons for and location of failures of metal-ceramic restorations and survival of repairs. Int J Prosthodont. 2002;15(3):299-302.].

Therefore, the aim of this study was to evaluate the satisfaction of patients rehabilitated with fixed partial denture and the incidence of failures/clinical complications of prostheses after one year in patients treated at Araçatuba Dental School - UNESP.

METHODS

The research was submitted to the analysis of Research Ethics Committee - CEP (Resolution No. 01 of 13/06/98 - cns), and after approval it was registered under the protocol 01079/2011.

Sample

The questionnaire was applied to patients enrolled at graduation clinics of Araçatuba Dental School - UNESP for treatment with fixed partial dentures. The prostheses evaluation period was one years after its installation. The study included 62 patients from both genders, over 18 years old. Patients who agreed to participate signed a consent form. The exclusion criterion was the patient's refusal to answer the questionnaire and to be examined.

The Clinical evaluation was conducted by the same examiner in all stages of the process after its calibration guided by the teacher responsible for the research. The clinical examination was carried out with the dental explorer, periodontal probe and mouth mirror to identify mechanical failures such as: fractures in metal (infrastructure), fracture of ceramic, welding, abutment tooth and perforation of the prosthesis, cement failure, adaptation retainer and contact point, and biological failures such as: gingival recession, periodontal pocket, involvement of periodontal support and recurrent caries.

After the clinical examination and applied the questionnaire, periapical radiographs were taken with the use of radiograph positioner to examine caries, endodontic lesions, periodontal problems, fractures, as well as the condition of intraradicular retainers.

Instruments used

Chart 1
This evaluation was conducted by analysis of the guided interview structured questionnaire.

Which was applied to collect the information. The questionnaire consisted of socio-demographic questions including gender, age and educational level, issues related to satisfaction with the treatment and hygiene care after cementation of the prosthesis.

Subsequently, clinical and radiological evaluations were performed to verify the prostheses and abutment teeth conditions.

Data were tabulated for descriptive analysis of the factors assessed with calculation of absolute and percentage frequency, and Chi-square and Fisher’s exact test to verify the association between variables: satisfaction degree and gender, age, education, biological and technical failures.

RESULTS

The sample consisted of 62 patients, 40 female (65%) and 22 male (35%), with mean age of 47.40 years and standard deviation of 9.90. For educational level, 18 patients had only completed primary education (29.03%), 34 the high school (54.84%), and only 10 have higher education (16:13%). When asked about their satisfaction level with the prosthetic treatment received, 17 patients manifested as unsatisfactory (27.42%), 23 as good (37.10%) and 22 considered the prosthesis as great (35.48%).

A total of 74% of the interviewer reported not presenting difficulties for prosthesis hygiene. However, (26%) reported not performing proper hygiene, due to some difficulties. In the present study, among patients who had difficulties at the time of cleaning, (56%) stated that did not know any of oral hygiene instructions. Regarding flossing, (43%) do not use. Of the 62 interviewees, (80%) reported that they had not been informed about the necessity of periodic returns.

Technical failures

Only 6 patients (9.67%) presented technical failures (Table 1).

Table 1
Technical failures.

In their fixed partial dentures. In one patient two failures were identified. Altogether, 7 technical failures were observed and the most common was the prosthesis loosening presented in 4 cases (57.14%), followed by ceramic fracture in 2 cases (28.57%), and one case of abutment tooth fracture (14.29%).

Biological failures

Biological failures were observed in 19 patients (30.65%). A total of 25 biological failures were listed, since some patients exhibited more than one occurrence. The most common failure was gingival recession in 14 cases (52%), followed by periodontal pockets in 6 cases (24%), periodontal involvement in 4 cases (16%), and recurrent caries (4%) reported in one patient (Table 2).

Table 2
Biological failures.

Biological-technical failures

Clinical evaluations did not identify the occurrence of biological-technical failures, such as occlusal and aesthetic failures, as stated by Karoussis et al.[1313 Karoussis IK, Salvi GE, Heitz-Mayfield LJ, Brägger U, Hämmerle CH, Lang NP. Long-term implant prognosis in patients with and without a history of chronic periodontitis: a 10-year prospective cohort study of the ITI Dental Implant System. Clin Oral Implants Res. 2003;14(3):329-39.] and Bragger[88 Brägger U, Hirt-Steiner S, Schnell N, Schmidlin K, Salvi GE, Pjetursson B, et al. Complication and failure rates of fixed dental prostheses in patients treated for periodontal disease. Clin Oral Implants Res. 2011;22(1):70-7. doi: 10.1111/j.1600-0501.2010.02095.x
https://doi.org/10.1111/j.1600-0501.2010...
].

Radiographic failures

Radiographic examination showed that from the total number evaluated, 44 patients (70.97%) had some kind of failure (Table 3).

Table 3
Radiographic failures.

Correlation between variables

In Table 4 was presented the correlation between variables, the chi-square test and Fisher’s exact test considered the variable “satisfaction degree” as satisfactory and unsatisfactory, the categories "Great" and "Good" was considered like Satisfactory and Unsatisfactory. Biological and technical failures were classified as present or absent. The age range was divided into 48 years or less and more than 48 years, based on the sample median.

Table 4
Correlation between satisfaction degree and gender age, educational level, mechanical and biological failures.

There was statistically significant association between satisfaction degree and technical failures (p=0.04). For the other variables, the associations were not statistically significant: gender (p=0.07); educational level (p=0.74); age range (p=0.48); and biological failures (p=0.27).

After all examinations were performed, questionnaires were evaluated individually and the overall rating of such faults was conducted as proposed by Ozcan and Niedermeier [1212 Ozcan M, Niedermeier W. Clinical study on the reasons for and location of failures of metal-ceramic restorations and survival of repairs. Int J Prosthodont. 2002;15(3):299-302.]:

Class I: The failure cause is correctable without replacing the restoration.

Class II: The failure cause is correctable without replacing the restoration, however, the support tooth structure or foundation requires repair or reconstruction.

Class III: Failure requires only the restoration replacement. Support tooth/structure is acceptable.

Class IV: Failure requires replacement of restoration, besides demanding repair or reconstruction of the tooth/foundation support structure.

Class V: Insufficiency with loss of support tooth or inability to rebuild the use of original tooth support. Replacement of fixed prosthesis remains possible through the use of other complementary support for restoration or the redesign.

Class VI: Insufficiency with loss of support tooth or impossibility to reconstruct the support tooth like the original. Replacement of conventional fixed prosthesis is not possible.

When considering technical (7 occurrences), biological (25 occurrences), biological-technical (0 occurrence) and radiographic (57 occurrences) failures, a total of 89 occurrences is reported.

The events were classified as Class I in 97.75% of cases, ie, faults could be corrected without replacement of prosthetic restoration. As an example of this occurrence cementation failures could be observed.

Radiographic findings, also classified as Class I, demonstrated the need to inform the patient about the necessity of monitoring the restorations. An example of this occurrence is the apically maladjusted posts.

A special conduct was adopted for patients with gingival recession. These patients were instructed individually for a proper brushing technique by the scholarship student.

There was one caries recurrence in the abutment tooth and this failure was classified as Class II (1.12%).

One occurrence was classified as Class IV (1.12%). This event refers to the abutment tooth fracture. In this case, the failure required the replacement of the prosthetic restoration and the abutment tooth repair, with the confection of an intraradicular retainer.

DISCUSSION

Treatment with fixed partial denture considered long and with high costs generates great expectations for the patient. Although dedication, careful planning and meticulous attention to every detail have been expended, failures and patient dissatisfaction with final results are not rare, being one of the most frustrating aspects of dental practice.

The questioning of the patients started with questions about their satisfaction regarding prostheses aesthetics and function. Most of the patients were satisfied with their prosthesis (72.58%), and only (27.42%) of the interviewees considered them unsatisfactory. Also, good patient satisfaction was found in an 18-year retrospective analysis [44 Napankangas R, Raustia A. An 18-year retrospective analysis of treatment outcomes with metal-ceramic fixed partial dentures. Int J Prosthodont. 2011;24(1):314-9.].

Despite the results have not shown statistical significance (p=0.07), men (86.36%) were more satisfied with the prosthesis than women (65%). Studies demonstrated that when questioned about satisfaction level with oral aesthetics, the female gender was also less satisfied [1414 Akarslan ZZ, Sadik B, Erten H, Karabulut E. Dental esthetic satisfaction, received and desired dental treatments for improvement of esthetics. Indian J Dent Res. 2009;20(2):195-200. doi: 10.4103/0970-9290.52902
https://doi.org/10.4103/0970-9290.52902...
,1515 Tin-Oo MM, Saddki N, Hassan N. Factors influencing patient satisfaction with dental appearance and treatments they desire to improve aesthetics. BMC Oral Health. 2011 23;11:6. doi: 10.1186/1472-6831-11-6
https://doi.org/10.1186/1472-6831-11-6...
].

While aging is seen as a gradual process, the feeling of "being old" occurs as a result of something abrupt, caused by some event that could precipitate it, such as the loss of teeth. In this study, there was no correlation between age and satisfaction with the prosthesis (p=0.48), but the older patients obtained a higher satisfaction percentage (76.67%). Newsome3 also found that age did not affect the prosthetic contentment.

The biological failure occurs when there is an inadequacy of the host, concerning the prosthesis maintenance and stability [1616 Pjetursson BE, Sailer I, Zwahlen M, Hammerle CHF. A systematic review of the survival and complication rates of all-ceramic and metal-ceramic reconstructions after an observation period of at least 3 years. Part I: single crowns. Clin Oral Implant Res. 2007;18:73-85.]. The most frequent biological failure was gingival recession, followed by periodontal involvement, cementing failures and recurrent caries, similar results were found in a retrospective study, with 59% of gingival recession [44 Napankangas R, Raustia A. An 18-year retrospective analysis of treatment outcomes with metal-ceramic fixed partial dentures. Int J Prosthodont. 2011;24(1):314-9.]. The great incidence of gingival recession can be attributed to the difficulty of hygienization (absent or very traumatic to tissues), or even to iatrogeny from the dentist when preparing the prosthesis. Biological failure did not affect the individuals' satisfaction with the prosthesis (p=0.27). A possible explanation for this occurrence is that periodontal disease does not present symptoms in the early stages, and as most of the teeth were treated endodontically, they presented no sensitivity to caries.

Periodontal disease, if not treated, causes irreversible consequences such as insertion and bone loss. According to Kourkouta et al. [1717 Kourkouta S, Hemmings KW, Laurell L. Restoration of periodontally compromised dentitions using cross-arch bridges. Principles of perio-prosthetic patient management. Br Dent J. 2007;203(4):189-95. doi: 10.1038/bdj.2007.727
https://doi.org/10.1038/bdj.2007.727...
] in patients with significant loss of periodontal support, it is preferable to opt for rehabilitation with fixed than removable prosthesis. Furthermore, the fixed restorations provide more comfort and safety, especially in cases in which there was periodontal involvement. According to literature [1313 Karoussis IK, Salvi GE, Heitz-Mayfield LJ, Brägger U, Hämmerle CH, Lang NP. Long-term implant prognosis in patients with and without a history of chronic periodontitis: a 10-year prospective cohort study of the ITI Dental Implant System. Clin Oral Implants Res. 2003;14(3):329-39.], results indicate periodontal involvement as one of the main biological failures. The prostheses hinder the natural stimulation of supporting structures, thus contributing to the accumulation of dental plaque. This plaque accumulation, gingival inflammation, insertion loss, periodontal pockets, and bone loss are possible sequelae in prostheses users [1818 Alfredo E, Garrido AD, Souza-Filho CB, Correr-Sobrinho L, Sousa-Neto MD. In vitro evaluation of the effect of core diameter for removing radicular post with ultrasound. J Oral Rehabil. 2004;31(6):590-4. doi: 10.1111/j.1365-2842.2004.01262.x
https://doi.org/10.1111/j.1365-2842.2004...
].

There was statistically significant association between satisfaction degree and technical failure (p=0.04). Satisfied patients presented lower failure rate (33.33%), contrasting with dissatisfied patients (66.67%). Even not obtaining association between radiographic examinations and patient satisfaction, the main failures encountered were short post, bone loss and inadequate endodontic treatment. The short core favors the stress concentration in certain areas, causing the root fracture. The correct length of the core inside the root is synonymous of prosthesis longevity [1818 Alfredo E, Garrido AD, Souza-Filho CB, Correr-Sobrinho L, Sousa-Neto MD. In vitro evaluation of the effect of core diameter for removing radicular post with ultrasound. J Oral Rehabil. 2004;31(6):590-4. doi: 10.1111/j.1365-2842.2004.01262.x
https://doi.org/10.1111/j.1365-2842.2004...
].

The longevity of fixed partial dentures depends on many factors ranging from quality until the care with which the patient preserves it. Therefore, monitoring oral hygiene of patients using fixed prosthesis is a powerful tool for the success of this type of rehabilitation. A total of 74% of the interviewees reported no difficulties of prosthesis hygiene. However, (26%) reported not performing proper hygiene because of some difficulties such as unaware of brushing techniques, and no habit of flossing. They claim that when dental floss is used, it wraps in proximal surfaces of and "causes" the bleeding of gingival tissue. Periodontal diseases that affect patients, are often related to bad prosthesis hygiene because of patient's own negligence or, in most instances, due to lack of information about how this cleaning should be performed.

Once it can not be removed, the fixed partial prosthesis requires greater technical ability of patients during cleaning, because food accumulates easily in pontics region and dental materials used retain higher amount of plaque in comparison with enamel or dentin, what favors the constant presence of this irritant factor.19 In the present study, among patients who had difficulties at the time of cleaning, (56%) stated they did not know anything about oral hygiene instructions. Regarding flossing, (43%) do not use dental floss or even devices such as “thread guides”. From a periodontal perspective, the pontics of fixed prostheses represent a serious problem in relation to hygiene and should be made following some principles for proper hygiene, through methods that eliminate biofilms [88 Brägger U, Hirt-Steiner S, Schnell N, Schmidlin K, Salvi GE, Pjetursson B, et al. Complication and failure rates of fixed dental prostheses in patients treated for periodontal disease. Clin Oral Implants Res. 2011;22(1):70-7. doi: 10.1111/j.1600-0501.2010.02095.x
https://doi.org/10.1111/j.1600-0501.2010...
].

Another tool that aids in proper oral hygiene is the periodic controls after finishing prosthetic treatment. However, this fact was not observed in patients from the current study, because from 62 interviewees, (80%) reported not having been informed about the necessity of periodic returns.

Thus, whatever the variables of prosthodontic treatment (single prostheses, or adhesive or multiple fixed prostheses), the planning of periodical controls and periodontal health, serve as a foundation for prosthesis longevity.

CONCLUSION

Based on the results, it was concluded that: after one year of prosthesis installation, patients were satisfied; the satisfaction degree was influenced mainly by technical failures; the main technical failures detected were crown cementation fault and ceramic fracture; the most relevant biological failures were gingival recession followed by periodontal pocket; the main radiographic failures were the preparation and installation of short intraradicular post with increased diameter and cervical misfit of the crown.

REFERENCES

  • 1
    Reitemeier B, Hänsel K, Kastner C, Weber A, Walter MH. A prospective 10-year study of metal ceramic single crowns and fixed dental prosthesis retainers in private practice settings. J Prosthet Dent. 2013 Mar;109(3):149-55. doi: 10.1016/S0022-3913(13)60034-7
    » https://doi.org/10.1016/S0022-3913(13)60034-7
  • 2
    Layton D, Walton T. Patient-evaluated dentistry: development and validation of a patient satisfaction questionnaire for fixed prosthodontic treatment. Int J Prosthodont. 2011;24(4):332-4141.
  • 3
    Newsome PR, Wright GH. A review of patient satisfaction: 1. Concepts of satisfaction. Br Dent J. 1999;186(4 Spec No):161-5.
  • 4
    Napankangas R, Raustia A. An 18-year retrospective analysis of treatment outcomes with metal-ceramic fixed partial dentures. Int J Prosthodont. 2011;24(1):314-9.
  • 5
    Svanborg P, Längström L, Lundh RM, Bjerkstig G, Ortorp A. A 5-year retrospective study of cobalt-chromium-based fixed dental prostheses. Int J Prosthodont. 2013;26(4):343-9. doi: 10.11607/ijp.3024
    » https://doi.org/10.11607/ijp.3024
  • 6
    Solá-Ruiz MF, Lagos-Flores E, Román-Rodriguez JL, Highsmith Jdel R, Fons-Font A, Granell-Ruiz M. Survival rates of a lithium disilicate-based core ceramic for three-unit esthetic fixed partial dentures: a 10-year prospective study. Int J Prosthodont. 2013;26(2):175-80. doi: 10.11607/ijp.3045
    » https://doi.org/10.11607/ijp.3045
  • 7
    De Backer H, Van Maele G, De Moor N, Van den Berghe L. Long-term results of short-span versus long-span fixed dental prostheses: an up to 20-year retrospective study. Int J Prosthodont. 2008;21(1):75-85.
  • 8
    Brägger U, Hirt-Steiner S, Schnell N, Schmidlin K, Salvi GE, Pjetursson B, et al. Complication and failure rates of fixed dental prostheses in patients treated for periodontal disease. Clin Oral Implants Res. 2011;22(1):70-7. doi: 10.1111/j.1600-0501.2010.02095.x
    » https://doi.org/10.1111/j.1600-0501.2010.02095.x
  • 9
    Heitz-Mayfield LJ, Huynh-Ba G. History of treated periodontitis and smoking as risks for implant therapy. Int J Oral Maxillofac Implants. 2009;24(Suppl):39-68.
  • 10
    Salvi GE, Brägger U. Mechanical and technical risks in implant therapy. Int J Oral Maxillofac Implants. 2009;24(Suppl):69-85.
  • 11
    Bouchard P, Renouard F, Bourgeois D, Fromentin O, Jeanneret MH, Beresniak A. Cost-effectiveness modeling of dental implant vs. bridge. Clin Oral Implants Res. 2009;20(6):583-7.
  • 12
    Ozcan M, Niedermeier W. Clinical study on the reasons for and location of failures of metal-ceramic restorations and survival of repairs. Int J Prosthodont. 2002;15(3):299-302.
  • 13
    Karoussis IK, Salvi GE, Heitz-Mayfield LJ, Brägger U, Hämmerle CH, Lang NP. Long-term implant prognosis in patients with and without a history of chronic periodontitis: a 10-year prospective cohort study of the ITI Dental Implant System. Clin Oral Implants Res. 2003;14(3):329-39.
  • 14
    Akarslan ZZ, Sadik B, Erten H, Karabulut E. Dental esthetic satisfaction, received and desired dental treatments for improvement of esthetics. Indian J Dent Res. 2009;20(2):195-200. doi: 10.4103/0970-9290.52902
    » https://doi.org/10.4103/0970-9290.52902
  • 15
    Tin-Oo MM, Saddki N, Hassan N. Factors influencing patient satisfaction with dental appearance and treatments they desire to improve aesthetics. BMC Oral Health. 2011 23;11:6. doi: 10.1186/1472-6831-11-6
    » https://doi.org/10.1186/1472-6831-11-6
  • 16
    Pjetursson BE, Sailer I, Zwahlen M, Hammerle CHF. A systematic review of the survival and complication rates of all-ceramic and metal-ceramic reconstructions after an observation period of at least 3 years. Part I: single crowns. Clin Oral Implant Res. 2007;18:73-85.
  • 17
    Kourkouta S, Hemmings KW, Laurell L. Restoration of periodontally compromised dentitions using cross-arch bridges. Principles of perio-prosthetic patient management. Br Dent J. 2007;203(4):189-95. doi: 10.1038/bdj.2007.727
    » https://doi.org/10.1038/bdj.2007.727
  • 18
    Alfredo E, Garrido AD, Souza-Filho CB, Correr-Sobrinho L, Sousa-Neto MD. In vitro evaluation of the effect of core diameter for removing radicular post with ultrasound. J Oral Rehabil. 2004;31(6):590-4. doi: 10.1111/j.1365-2842.2004.01262.x
    » https://doi.org/10.1111/j.1365-2842.2004.01262.x
  • 19
    Hochman N, Mitelman L, Hadani PE, Zalkind M. A clinical and radiographic evaluation of fixed partial dentures (FPDs) prepared by dental school students: a retrospective study. J Oral Rehabil. 2003;30(2):165-70.

Publication Dates

  • Publication in this collection
    Sept 2018

History

  • Received
    07 Dec 2017
  • Reviewed
    20 Apr 2018
  • Accepted
    13 June 2018
Faculdade São Leopoldo Mandic R. José Rocha Junqueira, 13, 13045-755 Campinas/SP Brasil, Tel.: (55 19) 3211-3689 - Campinas - SP - Brazil
E-mail: contato@revistargo.com.br