Myocardial contrast echocardiography in patients with suspected or known coronary artery disease: comparison with myocardial nuclear scintigraphy

Abstracts

OBJECTIVE: To compare myocardial contrast echocardiography (MCE) using PESDA and adenosine in bolus (ADN) with myocardial nuclear scintigraphy (NS) in patients (pts) undergoing routine investigation with a high probability of having coronary artery disease. METHODS: This study comprised 125 pts (85 men) with 58.4 ± 10.6 years, who underwent MCE and NS within 4 weeks. MCE was performed with PESDA in a continuous infusion at rest and after administration of an adenosine bolus. The LV walls was divided into 3 territories related to the coronary arteries, in a total of 375 territories. MCE was normal when an increase in contrast intensity occurred after ADN. The reduction in contrast intensity at rest or after ADN was defined as an abnormal MCE result. NS was performed according to classical protocols. When compared per patient, both examinations were considered concordant when they were normal or abnormal, independent of its location. The comparison by territory was considered concordant when perfusion defects existed or not in the same territory. The chi-square test was used to determine the significance of concordance. RESULTS: In 106/125 pts, MCE and NS were concordant (84.8% - P < 0.001). Concordance occurred in 342/375 territories (91.2% - P < 0.001). For the LAD territory, concordance was 87.2%; for the RCA, 93.6%; and for the CX, 92.8% (P < 0.001). CONCLUSION: An excellent concordance exists between MCE and NS in assessing pts for coronary artery disease; therefore, MCE may represent a good alternative for assessing myocardial perfusion.

echocardiography; myocardial perfusion; echocardiographic contrast medium; coronary artery disease; adenosine; myocardial nuclear scintigraphy


OBJETIVO: Comparar a ecocardiografia de contraste miocárdico (ECM) usando PESDA e adenosina em bolus (ADN) com a cintilografia miocárdica com radioisótopos (CM) em pacientes (pts) submetidos a investigação e alta probabilidade de doença arterial coronariana. MÉTODOS: Foram estudados 125 pts, 58,4 ± 10,6 anos, 85 homens, com ECM e CM, realizados com intervalo máximo de 4 semanas. ECM foi realizada com PESDA em infusão contínua em repouso e após ADN. As paredes do VE foram divididas em 3 territórios relacionados às artérias coronarianas, totalizando 375 territórios. ECM foi normal quando houve aumento da intensidade do contraste após ADN. Diminuição da intensidade do contraste em repouso ou após ADN foi definida como ECM anormal. CM foi realizada usando protocolos clássicos. Comparados por pacientes, foram considerados concordantes quando ambos exames eram normais ou anormais independente de localização. A comparação por território foi considerada concordante quando havia ou não déficit perfusional num mesmo território. A significância da concordância foi feita pelo teste do qui-quadrado. RESULTADOS: Em 106/125 pts ECM e CM foram concordantes (84,8% - p<0,001). Houve concordância em 342/375 territórios (91,2% - p<0,001). Para o território de DA a concordância foi de 87,2%, para CD 93,6% e para CX 92,8% (p<0,001). CONCLUSÃO: Existe uma ótima concordância entre ECM e CM nos pts em investigação para doença arterial coronariana, podendo a ECM representar uma alternativa à avaliação da perfusão miocárdica.

ecocardiografia; perfusão miocárdica; contraste ecocardiográfico; doença arterial coronariana; adenosina; cintilografia miocárdica com radioisótopos


ORIGINAL ARTICLE

Myocardial contrast echocardiography in patients with suspected or known coronary artery disease. Comparison with myocardial nuclear scintigraphy

Marcia Carrinho; Alvaro Moraes; Fernando Morcerf; Caio Medeiros; Márcia Castier; Luis José Martins Romeo

Rio de Janeiro, RJ - Brazil

ECOR Ecocardiografia and Universidade Federal Fluminense

Correspondence Correspondence to Márcia Carrinho Rua Barata Ribeiro, 432/201 Cep 22040-020 - Rio de Janeiro, RJ, Brazil E-mail: mcarrinho@cardiol.br

ABSTRACT

OBJECTIVE: To compare myocardial contrast echocardiography (MCE) using PESDA and adenosine in bolus (ADN) with myocardial nuclear scintigraphy (NS) in patients (pts) undergoing routine investigation with a high probability of having coronary artery disease.

METHODS: This study comprised 125 pts (85 men) with 58.4 ± 10.6 years, who underwent MCE and NS within 4 weeks. MCE was performed with PESDA in a continuous infusion at rest and after administration of an adenosine bolus. The LV walls was divided into 3 territories related to the coronary arteries, in a total of 375 territories. MCE was normal when an increase in contrast intensity occurred after ADN. The reduction in contrast intensity at rest or after ADN was defined as an abnormal MCE result. NS was performed according to classical protocols. When compared per patient, both examinations were considered concordant when they were normal or abnormal, independent of its location. The comparison by territory was considered concordant when perfusion defects existed or not in the same territory. The chi-square test was used to determine the significance of concordance.

RESULTS: In 106/125 pts, MCE and NS were concordant (84.8% - P < 0.001). Concordance occurred in 342/375 territories (91.2% - P < 0.001). For the LAD territory, concordance was 87.2%; for the RCA, 93.6%; and for the CX, 92.8% (P < 0.001).

CONCLUSION: An excellent concordance exists between MCE and NS in assessing pts for coronary artery disease; therefore, MCE may represent a good alternative for assessing myocardial perfusion.

Key words: echocardiography, myocardial perfusion, echocardiographic contrast medium, coronary artery disease, adenosine, myocardial nuclear scintigraphy

The primary objective of perfusion studies is to assess ischemia and viability of myocardial cells. The 3 most frequently used methods are stress echocardiography, pharmacological stress echocardiography, and myocardial nuclear scintigraphy. Of these 3 methods, only scintigraphy directly assesses perfusion. Myocardial contrast echocardiography is an alternative method for scintigraphy because it provides reproducibility, real-time image assessment, in addition to being noninvasive and relatively inexpensive 1. All comparative studies between both methods have shown an acceptable concordance, validating the use of contrast echocardiography as an alternative for myocardial nuclear scintigraphy 2-5.

This study aimed at comparing myocardial contrast echocardiography using perfluorocarbon-exposed sonicated dextrose albumin (PESDA) and adenosine in bolus with myocardial nuclear scintigraphy in patients with a high probability of having coronary artery disease and undergoing routine diagnostic investigation.

Methods

This is a longitudinal, prospective study of 125 consecutive patients with suspected or known coronary artery disease, who underwent myocardial contrast echocardiography indicated by their attending physician from July 1997 to October 2001. The clinical characteristics of the patients are shown in table I.

Myocardial nuclear scintigraphy was performed in 2 large nuclear medicine centers in Rio de Janeiro at a maximum interval of 4 weeks from myocardial contrast echocardiography. The patients who had any coronary event between these 2 examinations were excluded from the study. Both examinations were assessed independently and qualitatively.

All patients provided formal consent for undergoing myocardial contrast echocardiography after oral and written clarification, according to the research protocol approved by the National Committee on Ethics and Research of the Health Ministry.

The contrast agent PESDA was prepared according to the method previously reported by Porter et al 6.

The equipment used is commercially available (HDI 3000 and HDI 5000, ATL, Bothell, WA, USA). Image preparation and contrast administration were performed according to the protocol of Morcerf et al 7, who use continuous infusion of PESDA with second harmonic imaging in the intermittent mode (1:1 with electrocardiographic QRS) at rest and after adenosine bolus injection.

The echocardiographic images were obtained with apical 4-chamber and 2-chamber views. Image acquisition began right after optimization of the equipment controls and PESDA dripping, and immediately before adenosine injection, and continued until disappearance of its effect on the myocardium. The visual gradation of the contrast in the myocardium at rest and its intensification or reduction after adenosine bolus injection was performed by 2 independent reviewers based on videotape or optical disk images 7.

Each reviewer assessed the contrast distribution and homogeneity in the left ventricular walls in all echocardiographic views at rest. After adenosine injection, the increase in, reduction in, or maintenance of contrast intensity in the left ventricular walls were assessed with the maximum effect of the drug, allowing the definition of normal and abnormal (fixed or reversible defects) patterns of myocardial perfusion 7.

Myocardial perfusion was normal when a marked intensification of the contrast occurred in all segments of the left ventricular walls after adenosine injection (fig. 1). Myocardial perfusion was abnormal when a reduction in contrast intensity occurred. If this was observed only after adenosine injection, the defect was considered reversible (fig. 2); if observed already in the evaluation at rest, with no alteration with adenosine injection, the defect was considered fixed (fig. 3).




In each patient, the left ventricular walls were divided into 3 coronary territories or beds. The anterior wall and the anterior portion of the interventricular septum, as well as the apical 2/3 of its posterior portion, were considered part of the territory of the anterior descending artery (AD). The lateral wall was considered to belong to the territory of the circumflex artery (CX). The posterolateral and posterior walls, and the basal posterior portion of the interventricular septum were considered territories of the right coronary artery (RC).

The scintigraphic examination used in all patients was 99mTc-sestamibi SPECT associated with dipyridamole stress. The data were obtained at rest and after venous infusion of dipyridamole at the dosage of 0.56 mg/kg diluted in 20 mL of saline solution in 4 minutes. After 2 minutes, an additional dose of 0.28 mg/kg diluted in 10 mL was administered. For SPECT, 99mTc-sestamibi was injected 6 minutes after infusion of dipyridamole, with image acquisition after 1 hour. Tomographic cuts were performed in the 3 major cardiac axes (minor vertical, major vertical, and horizontal). All tests were considered effective.

When compared per patient, the examinations were considered concordant when both were normal or both had perfusion alterations independent of their locations.

The comparison between myocardial contrast echocardiography and myocardial nuclear scintigraphy per territory of coronary perfusion was considered concordant whether a perfusion defect existed or not in the same territory in both examinations, independent of the perfusion pattern.

The analysis of concordance between the 2 methods was performed per patient and per coronary perfusion territory in regard to the presence or absence of perfusion defect. The territories of coronary perfusion were analyzed both as a whole, as well as individually.

The chi-square test was used for determining the significance of the distribution of concordance. The Student t test was used for comparing the values of concordance for each coronary territory. Variability lower than 0.05 was considered significant.

Results

The overall analysis of the 125 patients studied showed concordant myocardial contrast echocardiography and myocardial nuclear scintigraphy in 106 patients (84.8%, c2 = 58.579; P<0.001). Forty-five of these examinations were normal, and 61 were abnormal (39 reversible defects and 22 fixed defects). In 19 patients (15.2%), the examinations were discordant.

In the 125 patients, 375 coronary territories were analyzed. The overall concordance in these territories was 91.2% (342 in 375 territories, c2 = 210.905; P < 0.001). The distribution of the normal and abnormal examinations is shown in table II.

The assessment of concordance for each territory (AD, RC, and CX) separately is briefly shown in table III.

Although the concordance for the alterations in the anterior descending artery territory was smaller in terms of percentage, the comparison of the 3 territories using the Student t test showed no significance (P > 0.05).

The form of adenosine administration posed neither complications nor additional risks to the examinations. The mean dose of adenosine used per examination was 3.9 ampules. The side effects related to the administration of adenosine (facial warmth sensation, chest discomfort, headache, and transient complete atrioventricular block) were reported in less than 10% of the examinations.

Discussion

For 2 decades, assessment of myocardial perfusion through scintigraphy was considered the gold-standard for diagnosing coronary artery disease and the result of the procedures of reperfusion, for studying myocardial viability, and for stratifying the risk 8.

In the past 10 years, several studies reported that contrast echocardiography produces myocardial opacification when the contrast is intravenously administered 9, and, mainly, may identify the presence of perfusion defects at rest and during pharmacological stress 10.

The study of myocardial perfusion performed with echocardiography and peripheral infusion of contrast and the study performed with scintigraphy were first compared by Kaul et al 2, who studied 30 patients with suspected or known coronary artery disease by using the FS-069 second-generation contrast and dipyridamole as the stressor agent. When the analysis was performed per patient, concordance between the 2 methods was 86%, and, when the analysis was performed in each of the 3 coronary territories, concordance was 90% (88% for the territories with abnormal perfusion and 91% with normal perfusion). These promising values contributed to continuation of the research for the development of contrast echocardiographic techniques.

Porter et al 11 compared contrast echocardiography with PESDA at rest and after stress with sestamibi SPECT using videointensity for analysis (quantitative). They found 92% sensitivity and 84% specificity for myocardial contrast echocardiography, with excellent concordance in regard to location and extension of the perfusion defects.

Although the techniques of the 2 examinations are analogous, significant differences exist: 1) the location of the tracer is different in each technique; the contrast medium used for echocardiography is restricted to the vascular space. When injected through the intravenous route, the gas microbubbles mix with blood, and their concentration in a myocardial region reflects the blood volume in this region 12. Myocardial nuclear scintigraphy involves the intracellular uptake of the tracer concentrated in the mitochondria 13. Contrast echocardiography studies the integrity of the microvasculature, while the nuclear technique involves the need for intact muscle cell; 2) the difference relates to the characteristic of the image. The spatial resolution of echocardiography is 1 mm, while that of scintigraphy is 1 cm. Therefore, conditions in which a reduction in the thickness of the left ventricle and abnormal motion of its walls occur (akinesia, complete left bundle-branch block) may create or emphasize the perfusion defects on myocardial nuclear scintigraphy. On the other hand, myocardial contrast echocardiography can identify perfusion defects independent of the alterations in contractility, as already reported by Moraes et al 14, even in patients with previous myocardial infarction 15 or with complete left bundle-branch block 16.

Because it is a recent technique, myocardial contrast echocardiography is still difficult to interpret, mainly because of attenuations and shadows that usually relate to the basal segments of the inferior and lateral walls, or to the greater destruction of the contrast medium by the ultrasound in the apical region. The ultrasound reflection in the microbubbles is not homogeneous, resulting in a decrease in signal intensity and amplitude, more intense in the distal field. Different types of contrast media, as well as the administration in the form of bolus injection may lead to different degrees of attenuation.

Independently of the type of echocardiographic contrast medium used and of its route of administration, researchers have emphasized the need for staff training for acquisition and analysis of the echocardiographic images, according to each protocol, aiming at minimizing the incidence of artifacts to increase the sensitivity of myocardial contrast echocardiography, since specificity is very satisfactory4.

The choice of the stressor agent is an important factor for the greater or lower degree of accuracy of perfusion echocardiography.

The blood vessels distal to an obstructive lesion are maximally dilated at rest, while the proximal areas are normal. This indicates that the myocardial region distal to the obstruction has a smaller coronary flow reserve, defined as the maximum possible increase in flow that can occur in 1 single coronary artery undergoing stress. Coronary flow reserve can be studied with physical exercise, inotropic agents, and vasodilating drugs, such as adenosine and dipyridamole. Vasodilators are the preferred agents for contrast echocardiography.

Ronderos et al 17 used dipyridamole during myocardial contrast echocardiography to study 35 patients diagnosed with perfusion defects on myocardial nuclear scintigraphy. Those authors reported that 32 of those patients also had perfusion defects on echocardiography.

As in the present study, Heinle et al 18 used adenosine for assessing coronary flow reserve. Adenosine is a potent coronary vasodilator. Its peripheral injection increases coronary flow 3 to 4 times due to a decrease in coronary resistance, resulting in an increase in blood volume 19, which causes a variation in the intensity of myocardial contrast.

Although the overall concordance between myocardial contrast echocardiography and myocardial nuclear scintigraphy was similar in our study and that of Heinle et al18 — 84.8% and 81%, respectively — significant differences are observed in the studies in regard to the analysis of each coronary territory.

Heinle et al18 reported the following concordances for the territories of the anterior descending (AD), right coronary (RC), and circumflex (CX) arteries: 81%, 76%, and 72%, respectively. In our study, the concordances were 87.2% (109/125 territories), 93.6% (117/125 territories), and 92.8% (116/125 territories), respectively (tab. III). The differences are mainly based on the protocol used for myocardial contrast echocardiography.

Heinle et al 18 used power Doppler as the image modality. It more frequently generates artifacts in the distal fields, causing "fixed defects" mainly in the lateral wall, which are not seen on scintigraphy. This can explain the lower concordance of this study for the territory of the CX. In addition, power Doppler is very sensitive to cardiac movements, which, in this study, were intensified by the use of adenosine, which is known to cause tachypnea as a reflex mechanism.

This study applied the protocol developed by Morcerf et al 7, which uses the administration of adenosine bolus. This protocol has a sensitivity of 91.7% and specificity of 98.6%, with excellent intra- and interobserver concordance (k = 0.94, and k = 0.91, respectively). The continuous infusion of the contrast medium provides longer time for the adjustment of the initial image, in an attempt to reduce, as much as possible, the artifacts inherent to the echocardiographic study with contrast medium. In addition, the effect of the use of adenosine bolus is felt within a short time, facilitating the maintenance of the same adjustments of the equipment, the same velocity of contrast infusion, and a rapid comparison with the baseline image. The characteristic that differentiates this protocol and that could justify the high concordance in the RC and CX territories (93.6% and 92.8%, respectively) is the existence of an expected normality pattern for the administration of adenosine. This pattern is the great increase in the intensity of the contrast medium, which can be visually detected in the areas without obstruction or with noncritical obstruction. Therefore, even when heterogeneity of perfusion exists at rest, due to the presence of artifacts, with an increase in contrast intensity after adenosine administration, perfusion is considered normal because coronary flow reserve is preserved.

Based on these considerations, concordance with myocardial nuclear scintigraphy was detected in 106 (84.8%) of the 125 patients studied. The statistical study showed that this concordance was highly significant (P < 0.001). The same significance was observed when the analysis was performed by perfusion bed. Overall concordance was 91.2% (342/375 territories).

Although without statistical significance, a slight reduction in concordance among the examinations was observed in regard to the AD perfusion bed (tab. III). This may be due to the greater incidence, among the discordances in this territory, of patients with complete left bundle-branch block, which is a known factor of false-positive results in scintigraphic examinations. Of the 16 discordant examinations in the AD territories, 5 had complete left bundle-branch block.

A better understanding of the discordances is important. Initially, the results of myocardial contrast echocardiography are compared with those of myocardial nuclear scintigraphy, considered the gold standard. However, it is not possible to know which method has the correct information, because scintigraphy also has false-positive results.

In this study, discordance occurred in 19 examinations as follows: MIBI was the abnormal examination in 13 patients, and myocardial contrast echocardiography was abnormal in 6. Of the 13 patients with alteration only in MIBI, 6 had complete left bundle-branch block, 2 had previous myocardial infarction, and 1 patient had already undergone myocardial revascularization surgery. Of the 6 patients with alteration only on myocardial contrast echocardiography, 2 had previous myocardial infarction, and 3 had already undergone revascularization (2 patients with angioplasty and 1 patient with surgery). These data point to the need for correlating with another method, maybe coronary anatomy assessment (cine coronary angiography), aiming at improving the comprehension of the possibilities of myocardial contrast echocardiography in assessing myocardial perfusion.

In regard to the limitations of our study, the following issues could have favored our results: 1) the analysis of the clinical characteristics of the patients (tab. I) and of the distribution of the concordant examinations into normal and abnormal (44 normal and 61 abnormal examinations — 48.8%) identifies a group of patients with a greater probability of coronary artery disease, which is known to favor the diagnostic accuracy of the method; 2) in regard to the protocol of myocardial contrast echocardiography used, it is worth noting that its reproducibility was adequately tested in few diagnostic centers; 3) the qualitative and nonquantitative analysis could interfere with the results; 4) unlike myocardial contrast echocardiography, the scintigraphic examinations were performed in 2 different laboratories. Therefore, although the examinations were interpreted by cardiologists with certification in nuclear cardiology, who followed the same standard protocol for MIBI with dipyridamole, variation in the diagnostic criteria may have occurred.

In conclusion, an excellent concordance exists between myocardial contrast echocardiography and myocardial nuclear scintigraphy for assessing patients with coronary artery disease, and myocardial contrast echocardiography may represent a good alternative for assessing myocardial perfusion.

References

Received for publication: 9/24/2003

Accepted for publication: 07/04/2004

English version by Stela Maris Costalonga

  • 1. Porter TR, Li S, Kricsfeld D, et al. Detection of myocardial perfusion in multiple echocardiographic windows with one intravenous injection of microbubbles using transient response second harmonic imaging. J Am Col Cardiol 1997;29:791-9.
  • 2. Kaul S, Senior R, Dittrich H, et al. Detection of coronary artery disease with myocardial contrast echocardiography: comparison with 99mTc-sestamibi single-photon emission computed tomography. Circulation. 1997;96:785-92.
  • 3. Porter TR, Cwajg J, Li S, et al. Correlation of myocardial contrast enhancement determined with accelerated intermittent imaging with dual isotope SPECT imaging during exercise and dobutamine stress. J Am Soc Echocardiogr 1999;12:350.
  • 4. Marwick TH, Brunken R, Meland N, et al. Accuracy and feasibility of contrast echocardiography for detection of perfusion defects in routine practice: Comparison with wall motion and Technetium-99m sestamibi single-photon emission computed tomography. J Am Coll Cardiol 1999;32:1260-9.
  • 5. Jucquois I, Nihoyannopoulos P, D'hont AM, et al. Comparison of myocardial contrast echocardiography with NC 100100 and Tc Sestamibi Spect for detection of resting myocardial perfusion abnormalities in patients with previous myocardial infarction. Heart 2000;83:518-24.
  • 6. Porter TR, Xie F. Transient myocardial contrast following initial exposure to diagnostic ultrasound pressures with minute doses of intravenously injected microbubbles: demonstration and potential mechanisms. Circulation 1995;92:2391-5.
  • 7. Morcerf F, Moraes A, Carrinho M, et al. Estudo da reserva de fluxo coronariano com uso endovenoso de microbolhas (Ecocardiografia com contraste) e adenosina. Apresentação de protocolo para apresentação clínica em pacientes com suspeita de doença arterial coronariana. Arq Bras Cardiol2002;78:281-9.
  • 8. Ladenheim ML, Pollock BH, Rozanki A, et al. Extent and severity of myocardial hypoperfusion as predictors of prognosis in patients with suspected coronary artery disease. J Am Coll Cardiol 1986;7:464-71.
  • 9. Villanueva F, Glasheen WP, Sklenar J, et al. Successful and reproducible myocardial opacification during two-dimensional echocardiography from right heart injection of contrast. Circulation 1992;85:1557-64.
  • 10. Xie F, Porter TR. Acute myocardial ischemia and reperfusion can be visually identified non-invasively with intravenous perfluoropropane-enhanced sonicated albumin ultrasound contrast. Circulation 1994;90:1-555.
  • 11. Porter TR, Cwajg J, Li S, et al. Correlation of myocardial contrast enhancement determined with accelerated intermittent imaging with dual isotope SPECT imaging during exercise and dobutamine stress. J Am Soc Echocardiogr 1999;12:350.
  • 12. Kaul S. Clinical applications of myocardial contrast echocardiography. In: Braunwald E (Eds.) Heart Disease: A Textbook of Cardiovascular Medicine (Update 1). Philadelphia: W.B Saunders 1997;304-13.
  • 13. Dahlberg ST, Leppo JA. Physiologic properties of myocardial perfusion tracers. Cardiol. Clin 1994;12:169-85.
  • 14. Moraes A, Morcerf F, Carrinho M et al. The ability of myocardial contrast echocardiography to identify perfusion defects is independent of rest wall motion abnormalities. Circulation 1998;98(supp I):502.
  • 15. Morcerf F, Moraes A, Carrinho M et al. Intravenous infusion contrast echocardiography to assess myocardial revascularization procedures. Circulation 1998;98 (supp I):19.
  • 16. Carrinho M, Moraes A, Morcerf F, et al. Eco de Contraste Miocárdico com microbolhas e adenosina em pacientes com bloqueio completo do ramo esquerdo. Rev SOCERJ 1999;XII (suplemento A):23.
  • 17. Ronderos RE, Boskis M, Namsik C, et al. Correlation between myocardial perfusion abnormalities detected with intermittent imaging using intravenous perfluorocarbon microbubbles and radioisotope imaging during high-dose dipyridamole stress echo Clin Cardiol 2002;25:103.
  • 18. Heinle SK, Noblin J, Goree-Best P, et al. Assessment of myocardial perfusion by harmonic power Doppler imaging at rest and during adenosine stress. Comparison with 99m Tc-Sestamibi SPECT imaging. Circulation 2000;102:55-60.
  • 19. Belardinelli L, Linden J, Berne RM. The cardiac effect of adenosine. Progress in Cardiovac Dis 1989; 17: 73-97.
  • Correspondence to
    Márcia Carrinho
    Rua Barata Ribeiro, 432/201
    Cep 22040-020 - Rio de Janeiro, RJ, Brazil
    E-mail:
  • Publication Dates

    • Publication in this collection
      11 Nov 2004
    • Date of issue
      Nov 2004

    History

    • Accepted
      07 Apr 2004
    • Received
      24 Sept 2003
    Sociedade Brasileira de Cardiologia - SBC Avenida Marechal Câmara, 160, sala: 330, Centro, CEP: 20020-907, (21) 3478-2700 - Rio de Janeiro - RJ - Brazil, Fax: +55 21 3478-2770 - São Paulo - SP - Brazil
    E-mail: revista@cardiol.br