Impact of compression surfaces on cardiac massage during cardiopulmonary reanimation: an integrative review

Abstract

Objective

To synthesize the available evidence in the literature on the types of compression surfaces used in CPR and to analyze which characteristics of the compression surfaces impact the effectiveness of chest compression during CPR.

Method

Integrative literature review, whose selection and inclusion criteria were complete articles, in English, Portuguese or Spanish and that answered the following research question: “What are the characteristics of the compression surfaces that impact the effectiveness of chest compressions during CPR?”. It was carried out between June and July 2019.

Results

12 articles from experimental studies were included. 13 different types of mattresses were found. Regarding the boards, six different sizes and many materials were reported. Influences of the type of compression surface on the force required to perform chest compressions were found.

Conclusion

Evidence points out that larger mattresses with pressure reduction technology and larger beds have negative impacts on the quality of chest compressions.

Implication for practice

Knowledge about the influence of the type and characteristics of support surfaces on the quality of chest compressions can support professionals in the choice and incorporation of technologies in the hospital environment.

Keywords:
Cardiopulmonary Resuscitation; Cardiac Massage; Cardiac Arrest; Advanced Cardiac Vital Support; Review

Resumo

Objetivo

sintetizar as evidências disponíveis na literatura sobre os tipos de superfícies de compressão utilizadas na RCP e analisar quais características das superfícies de compressão têm impacto na eficácia da compressão torácica durante a RCP.

Método

revisão integrativa da literatura, cujos critérios de seleção e inclusão foram: artigos completos, em inglês, português ou espanhol e que respondessem a seguinte questão de pesquisa: “Quais são as características das superfícies de compressão que têm impacto na eficácia das compressões torácicas durante a RCP?”. Realizada entre os meses de junho e julho de 2019.

Resultados

inclui-se 12 artigos de estudos experimentais, cuja extração de dados revelou 13 tipos diferentes de colchões. Em relação às pranchas, seis tamanhos diferentes foram relatados, com diferentes materiais. Constatou-se influências do tipo de superfície de compressão na força necessária para realizar as compressões torácicas.

Conclusão

as evidências apontam que colchões de maiores dimensões e com tecnologia para redução de pressão e camas mais largas apresentam impactos negativos na qualidade das compressões torácicas.

Implicação para prática

o conhecimento sobre a influência do tipo e características das superfícies de apoio na qualidade das compressões torácicas podem subsidiar profissionais na escolha e incorporação de tecnologias no ambiente hospitalar.

Palavras-chave:
Reanimação Cardiopulmonar; Massagem Cardíaca; Parada Cardíaca; Suporte Vital Cardíaco Avançado; Revisão

Resumen

Objetivo

Sintetizar la evidencia disponible en la literatura sobre los tipos de superficies de compresión utilizadas en la RCP y analizar qué características de las superficies de compresión tienen un impacto en la efectividad de la compresión torácica durante la RCP.

Método

Revisión bibliográfica integradora, cuyos criterios de selección e inclusión fueron: artículos completos, en inglés, portugués o español y que respondieran a la siguiente pregunta de investigación: “¿Cuáles son las características de las superficies de compresión que inciden en la efectividad de las compresiones torácicas durante la RCP?”. Se llevó a cabo entre junio y julio de 2019.

Resultados

se incluyeron 12 artículos de estudios experimentales, cuya extracción de datos reveló 13 tipos diferentes de colchones. En cuanto a los Tabelaros, se reportaron seis tamaños diferentes, con diferentes materiales. Se encontraron influencias del tipo de superficie de compresión sobre la fuerza requerida para realizar las compresiones torácicas.

Conclusión

la evidencia señala que los colchones más grandes con tecnología de reducción de presión y las camas más grandes tienen impactos negativos en la calidad de las compresiones torácicas.

Implicación para la práctica

El conocimiento sobre la influencia del tipo y características de las superficies de apoyo en la calidad de las compresiones torácicas puede ayudar a los profesionales en la elección e incorporación de tecnologías en el ámbito hospitalario.

Palabras clave:
Reanimación Cardiopulmonar; Masaje Cardíaco; Paro Cardíaco; Soporte Vital Cardíaco Avanzado; Revisión

INTRODUCTION

Thoracic compression is one strategy that makes up the set of actions required for cardiopulmonary resuscitation (CPR). Variables such as initial cardiopulmonary arrest (CPA) rhythm, training of the professionals involved in care, available technologies, and human and material resources are known to interfere in the success of CPR11 Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: Adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259. PMid:26472993.
http://dx.doi.org/10.1161/CIR.0000000000...

2 Perkins GD, Handley AJ, Koster RW, Castrén M, Smyth MA, Olasveengen T et al. European Resuscitation Council Guidelines for Resuscitation 2015: Section2. Adult basic life support and automated external defibrillation. Resuscitation. 2015 out;95:81-99. http://dx.doi.org/10.1016/j.resuscitation.2015.07.015. PMid:26477420.
http://dx.doi.org/10.1016/j.resuscitatio...
-33 Claudia B, Sergio T, Facholi PT, Schiavo GN, Wagner da Silva SA, Agnaldo P et al. Atualização da diretriz de ressuscitação cardiopulmonar e cuidados cardiovasculares de emergência da Sociedade Brasileira de Cardiologia - 2019. Arq Bras Cardiol. 2019 set;113(3):449-663. http://dx.doi.org/10.5935/abc.20190203. PMid:31621787.
https://doi.org/10.5935/abc.20190203...
. Among these materials, compression surfaces such as bed, stretcher, mattress, and surface (backboard) where the patient is supported are highlighted in this article44 López-González A, Sánchez-López M, Garcia-Hermoso A, López-Tendero J, Rabanales-Sotos J, Martínez-Vizcaíno V. Muscular fitness as a mediator of quality cardiopulmonary resuscitation. Am J Emerg Med. 2016 set;34(9):1845-9. http://dx.doi.org/10.1016/j.ajem.2016.06.058. PMid:27344099.
http://dx.doi.org/10.1016/j.ajem.2016.06...
.

Compression surfaces need to contribute to a high-quality cardiac massage55 Sebbane M, Hayter M, Romero J, Lefebvre S, Chabrot C, Mercier G et al. Chest compressions performed by ED staff: a randomized cross-over simulation study on the floor and on a stretcher. Am J Emerg Med. 2012;30(9):1928-34. http://dx.doi.org/10.1016/j.ajem.2012.04.013. PMid:22795420.
http://dx.doi.org/10.1016/j.ajem.2012.04...
, that is to say, that which reaches a depth of between 5 cm and 6 cm, a frequency of 100 to 120 compressions per minute, and complete return of the chest33 Claudia B, Sergio T, Facholi PT, Schiavo GN, Wagner da Silva SA, Agnaldo P et al. Atualização da diretriz de ressuscitação cardiopulmonar e cuidados cardiovasculares de emergência da Sociedade Brasileira de Cardiologia - 2019. Arq Bras Cardiol. 2019 set;113(3):449-663. http://dx.doi.org/10.5935/abc.20190203. PMid:31621787.
https://doi.org/10.5935/abc.20190203...
,55 Sebbane M, Hayter M, Romero J, Lefebvre S, Chabrot C, Mercier G et al. Chest compressions performed by ED staff: a randomized cross-over simulation study on the floor and on a stretcher. Am J Emerg Med. 2012;30(9):1928-34. http://dx.doi.org/10.1016/j.ajem.2012.04.013. PMid:22795420.
http://dx.doi.org/10.1016/j.ajem.2012.04...
. However, some studies show that, in practice, the quality of chest compressions does not meet these recommendations, even in trained and often CPA-exposed teams, which contributes to low success in resuscitation66 Noordergraaf GJ, Paulussen IW, Venema A, van Berkom PF, Woerlee PH, Scheffer GJ et al. The impact of compliant surfaces on in-hospital chest compressions: effects of common mattresses and a backboard. Resuscitation. 2009;80(5):546-52. http://dx.doi.org/10.1016/j.resuscitation.2009.03.023. PMid:19409300.
http://dx.doi.org/10.1016/j.resuscitatio...
evidenced by the survival rate after intra-hospital CPA being less than 20%33 Claudia B, Sergio T, Facholi PT, Schiavo GN, Wagner da Silva SA, Agnaldo P et al. Atualização da diretriz de ressuscitação cardiopulmonar e cuidados cardiovasculares de emergência da Sociedade Brasileira de Cardiologia - 2019. Arq Bras Cardiol. 2019 set;113(3):449-663. http://dx.doi.org/10.5935/abc.20190203. PMid:31621787.
https://doi.org/10.5935/abc.20190203...
.

In favor of optimizing the efficacy of the compressions, recommendations for increasing CPR quality include minimizing the interruption time of the chest compressions during CPA care and positioning the victims in the supine position on a firm compression surface (floor or backboards) 11 Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: Adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259. PMid:26472993.
http://dx.doi.org/10.1161/CIR.0000000000...
. However, the American Heart Association recognizes a weak evidence base for this recommendation and, in addition to that, there is a lack of guidelines regarding the specificities of performing CPR maneuvers in hospital conditions, where the patients are bedridden11 Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: Adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259. PMid:26472993.
http://dx.doi.org/10.1161/CIR.0000000000...
.

It is noted that the variables linked to the compression surface, such as the size of the product, the material used, and densities, can exert an impact on the range of the appropriate depth in chest compression, increasing the force required to deform the thorax, which is proportional to the displacement generated by the surfaces. In an exploratory search carried out on the websites of the main suppliers of backboards available in the national and international markets, surfaces with very different characteristics were identified, varying in thickness (from 0.5 cm to 9.0 cm), length (40 cm to 100 cm), width (35 cm to 60 cm) and materials (polyethylene, acrylic, wood, medium density fiber board-MDF, and plastic).

Several studies showed a possible reduction in the efficacy of chest compressions during advanced cardiac life support due to the soft surfaces of hospital beds; this is because, when performing compressions on bedridden patients, the mattress undergoes deflection concomitantly with the patient's chest, precluding adequate chest compression77 Guimarães HP, Polastri TF, Caldeira P, Barbisan J. Suporte Básico de Vida. Manual do profissional. 6ª ed. EUA: Integracolor, LTD; 2016. 19 p.

8 Andersen LØ, Isbye DL, Rasmussen LS. Increasing compression depth during manikin CPR using a simple backboard. Acta Anaesthesiol Scand. 2007;51(6):747-50. http://dx.doi.org/10.1111/j.1399-6576.2007.01304.x. PMid:17425617.
http://dx.doi.org/10.1111/j.1399-6576.20...

9 Cloete G, Dellimore KH, Scheffer C, Smuts MS, Wallis LA. The impact of backboard size and orientation on sternum-to-spine compression depth and compression stiffness in a manikin study of CPR using two mattress types. Resuscitation. 2011;82(8):1064-70. http://dx.doi.org/10.1016/j.resuscitation.2011.04.003. PMid:21601344.
http://dx.doi.org/10.1016/j.resuscitatio...
-1010 Nishisaki A, Maltese MR, Niles DE, Sutton RM, Urbano J, Berg RA et al. Backboards are important when chest compressions are provided on a soft mattress. Resuscitation. 2012;83(8):1013-20. http://dx.doi.org/10.1016/j.resuscitation.2012.01.016. PMid:22310727.
http://dx.doi.org/10.1016/j.resuscitatio...
.

Although compression surfaces are widely used in clinical practice, there are no guidelines on their ideal characteristics, which results in variability of materials available in intra-hospital environments. In addition, the latest published updates of the American Heart Association1111 American Heart Association. Destaques das Atualizações Específicas das Diretrizes de 2017 da American Heart Association para Suporte Básico de Vida em Pediatria e para Adultos e Qualidade da ressuscitação Cardiopulmonar [Internet]. Chicago: American Heart Association; 2017 [citado 2020 ago 6]. Disponível em: https://eccguidelines.heart.org/wp-content/uploads/2017/12/2017-Focused-Updates_Highlights_PTBR.pdf
https://eccguidelines.heart.org/wp-conte...

12 American Heart Association. Destaques das Atualizações Focadas em Recomendações de 2018 da American Heart Association para RCP e ACE: Suporte Avançado de Vida Cardiovascular e Suporte Avançado de Vida em Pediatria [Internet]. Chicago: American Heart Association; 2018 [citado 2020 ago 6]. Disponível em: https://eccguidelines.heart.org/wp-content/uploads/2018/10/2018-Focused-Updates_Highlights_PTBR.pdf
https://eccguidelines.heart.org/wp-conte...
-1313 American Heart Association. Destaques das Atualizações direcionadas nas diretrizes de 2019 da American Heart Association para Ressuscitação Cardiopulmonar e Atendimento Cardiovascular de Emergência [Internet]. Chicago: American Heart Association; 2019 [citado 2020 ago 6]. Disponível em: https://eccguidelines.heart.org/wp-content/uploads/2019/11/2019-Focused-Updates_Highlights_PTBR.pdf
https://eccguidelines.heart.org/wp-conte...
guidelines for basic and advanced life support do not mention compression surfaces, the approach to this theme being restricted to the guideline published in 201511 Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: Adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259. PMid:26472993.
http://dx.doi.org/10.1161/CIR.0000000000...
, aspects that evidence the relevance of this research, justifying its conduction.

OBJECTIVE

To synthesize the evidence available in the literature on the types of compression surfaces used in CPR and to analyze which characteristics of the compression surfaces exert an impact on the efficacy of chest compression during CPR.

METHOD

Study design, period and locus

The authors made an integrative review study with the support of the PRISMA tool between June and July 2019. The review followed these methodological stages: formulation of the research question; establishment of the criteria to search for primary studies; data organization; analysis and discussion of the results; and presentation of the review1414 Mendes KDS, Silveira RCCP, Galvão CM. Integrative literature review: A research method to incorporate evidence in health care and nursing. Texto Contexto Enferm. 2008;17(4):758-64. http://dx.doi.org/10.1590/S0104-07072008000400018.
http://dx.doi.org/10.1590/S0104-07072008...
.

The search for primary studies was conducted in six online access virtual databases, as follows: Banco de Dados em Enfermagem(BDENF) – via BVS; Cumulative Index to Nursing and Allied Health Literature (CINAHL) – via EBSCO; Índice Bibliográfico Espanhol de Ciências de Saúde(INECS) – via BVS; Literatura Latino-Americana em Ciências de Saúde(LILACS) – via BVS; and Medical Literature Analysis and Retrieval System Online (Medline) – via PubMed and Scopus.

The research question “What are the characteristics of compression surfaces that exert an impact on the efficacy of chest compressions during CPR?”, which guided this search, was constructed with the aid of the PVO strategy (Problem, Variable and Outcome)1515 Fram D, Marin CM, Barbosa D. Avaliação da necessidade da revisão sistemática e a pergunta do estudo. In: Barbosa D, Taminato M, Fram D, Belasco A, editor. Enfermagem baseada em evidências. São Paulo: Atheneu; 2014. Cap. 3. p. 21-28., delimited as follows in the study: Problem: Chest Compressions; Variable: Compression surface; and Outcome: Efficacy of chest compression.

Sample; inclusion and exclusion criteria

Scientific articles selected based on the inclusion criteria were used, namely: in a full-text format available in the selected databases, in Portuguese, English, and Spanish and which were related to the research question regarding any of its elements (problem, variable or outcome). No limited time frame was established in literature searching because the absence of the apparent national and international recommendation about the compression surface to support the patient during CPR. Then, many studies as possible were included to a global understanding of the phenomenon, by synthesis. Articles whose contents were considered unsatisfactory in answering the research question were excluded, a classification based on the analysis of the relevance of their contribution in understanding the object of this review.

Study protocol

Two reviewers independently searched in literature, and the disagreements were resolved by consensus among peers or by the evaluation of a third reviewer when disagreement was maintained. The descriptors and keywords were chosen concerning the research question. Consequently, the descriptors were defined from the DeCS and Mesh Terms, using the following terms: “heart massage”(problem); “cardiopulmonary resuscitation”(outcome), in addition to using the keyword “backboards”(variable). These research terms were gathered and organized using the Boolean operator AND, adapted according to each source of information consulted. Chart 1 shows the search strategies used in this review, according to the specificities of each database.

Chart 1
Search strategies used in each database. Rio de Janeiro, RJ, Brazil, 2021.

Exploratory, selective and analytical reading was used for selecting the articles. The exploratory reading of titles and abstracts judged compliance with the inclusion criteria. In the selective reading phase, the complete text of article was read and it was evaluated whether the studies were related to the research question regarding any of its elements (problem, variable or outcome). In the last phase, the pre-selected studies were read again in full, listing the key ideas, with evaluation of the content as to the relevance of its contribution in understanding the phenomenon under study.

Data extraction from the selected primary studies was conducted with the aid of an instrument elaborated by the authors themselves. For each study included, the following information was captured: general characteristics of the study (identification, locus institution, type of publication), objective and population, manikins and surfaces, type of study and results related to the characteristics of compression surfaces and their impact on chest compressions.

Analysis of results and statistics

The information collected was later organized, in the light of the research question, in view of the knowledge synthesis. In this sense, the data from the articles were classified according to the type of compression surface (bed, mattress, backboard and stretcher) and their physical characteristics. From this classification, the data from the articles were summarized. For the synthesis of the physical characteristics, the dimensions of the surfaces were analyzed based on the application of descriptive statistics with simple and percentage frequency and measures of central tendency. Regarding the data related to the type of surface and impact on the efficacy of chest compression, at the end of the summary stage it was possible to visualize the data set, from which aggregation of evidence and its interpretation were performed for the knowledge synthesis and for understanding the phenomenon under study.

The final result of the review was first organized by presenting a chart with the main information of the corpus of articles, followed by analyzing the results that express the aggregation of evidence and knowledge synthesis. The studies were coded employing the letter E, followed by the Arabic number that indicates the presentation order.

RESULTS

The database searches resulted in 189 documents. After removing the duplicates, 187 articles remained, of which 156 were excluded in the application stage of the eligibility criteria from the reading of titles and abstracts. 19 studies were excluded from the reading of the full texts, in selective and analytical readings. Thus, 12 studies comprised the final sample of the review. The study search and selection process is graphically represented as a flowchart in Figure1.

Figure 1
Flowchart identifying the selection process of the studies selected to comprise the integrative review. Rio de Janeiro, RJ, Brazil, 2020.

All the studies included in this review were published from 2006 to 2016, in English, and were mainly produced in Korea (n=3; 25.0%) and South Africa (n=2; 16.7%). Of the total number of analyzed studies, seven were published in American journals specialized in Emergency and Resuscitation. Most of the studies were produced by professionals who were based in universities (n=7; 58.4%). As for the methodological design, all the studies were experimental and mostly used the Resusci Anne Modular System Skill Reporter Laerdal® manikin (n=5; 41.9%). As for the performer of the chest compressions, human performers predominated (n=10; 80%), with automatic compressors representing the minority (n=2; 20%).

Regarding the compression surfaces that were evaluated in the experiments, five (41.7%) studies combined bed and mattress (E5, E6, E7, E9, E10); three (25.0%) combined stretcher and mattress (E2, E3, E11); two (16.7%) combined stretcher or bed and mattress (E1 and E12); one (8.3%) used only the floor (E8) and one (8.3%) used only the mattress (E4). Regarding the use of backboards, nine (75%) studies pointed out their use (E3, E5, E6, E7, E8, E9, E10, E11, E12). The main data from the studies included in the corpus of this review were organized in Chart 2.

Chart 2
Synopsis of the articles included in the integrative review. Rio de Janeiro, RJ, Brazil, 2020.

The synthesis elaborated after aggregating the results in the selected research studies indicated that the current knowledge on the compression surfaces is organized into two synthesis units: typology and characterization of the compression surfaces and impact of the surfaces on the quality of chest compressions.

Concerning typology and characterization of the compression surfaces, 13 different types of mattresses were used in the studies. The majority was made of foam (38.4%); however, 38.4% did not describe the material the mattress was made of. Of the studies that used a bed or stretcher without a mattress as compression surface, only E1 described its dimensions. No study described the material or other composition characteristics of the beds. As for the backboards, six different sizes were described. Concerning the material, although most of the studies have not provided information (55.5%), the materials indicated were plastics (22.2%), followed by wood (11.1%) and semi-synthetic materials (11.1%). The synthesis of the compression surfaces' characteristics is shown in Table1.

Table 1
Characterization of the mattresses and backboards regarding dimensions and materials. Rio de Janeiro, RJ, Brazil, 2020.

Regarding the impact of the surfaces on the quality of chest compressions, the studies discuss influences of the type of compression surface on the force required to perform them, such force being significantly associated to the set of bed/stretcher and mattress used. Studies E2, E4, E8 and E12 considered the floor as the gold standard for high-quality chest compression depth. E2 and E4 addressed the impact of the stretcher/bed/mattress with the reduction in the quality of the compressions compared to the floor. Studies E3, E5, E6, E7, E10, E11 and E12 investigate the impact of backboards on the quality of chest compression, presenting differences in the results, sometimes showing benefits of using the board in the depth of compressions, and others demonstrating that its use does not lead to improvements in the quality of the maneuvers, and that they may even delay the start of the compressions. In relation to backboard orientation and size, it is noted that larger and longer boards produce greater lumbar support and improve the quality of chest compressions (E9).

Other characteristics of the surfaces that can influence the quality of compressions are larger mattresses and mattresses with high-technology materials for pressure reduction (E7) and wider beds (E1), which present negative impacts on the quality of chest compressions. Another characteristic that was shown to generate an impact on the quality of chest compressions is deflection of the beds and mattresses, directly related to these surfaces' characteristics (E8, E11, E12).

DISCUSSION

The results found about the characteristics and impact of the surfaces where the patient is supported (bed/stretcher, mattress and backboard) on chest compression indicate the following: the material, structure, dimensions, arrangement and knowledge on the mechanics of the mattress and backboard effects can reduce the quality of chest compressions, especially depth; lack of a uniform standard for structuring the surfaces regarding dimensions and equipment; and divergence on the indication to use backboards.

Chest compression is related to the maintenance of cardiac output and, consequently, to cerebral blood flow during CPR. CC is considered the most crucial care skill because it maintains cerebral perfusion during the CPR maneuvers. Similar to the compression frequency, some studies show that a proper compression depth range is also required. Several studies suggested that achieving a compression depth of nearly 5 cm is associated with better results when compared to more superficial compressions2424 Vadeboncoeur T, Stolz U, Panchal A, Silver A, Venuti M, Tobin J et al. Chest compression depth and survival in out-of-hospital cardiac arrest. Resuscitation. 2014;85(2):182-8. http://dx.doi.org/10.1016/j.resuscitation.2013.10.002. PMid:24125742.
http://dx.doi.org/10.1016/j.resuscitatio...
-2525 Edelson DP, Abella BS, Kramer-Johansen J, Wik L, Myklebust H, Barry AM et al. Effects of compression depth and pre-shock pauses predict defibrillation failure during cardiac arrest. Resuscitation. 2006;71(2):137-45. http://dx.doi.org/10.1016/j.resuscitation.2006.04.008. PMid:16982127.
http://dx.doi.org/10.1016/j.resuscitatio...
. According to clinical and animal studies that evaluated the quality of the CPR effect on the clinical outcome, a change from 10% to 20% (0.5-1 cm) in chest compression depth can have a clinical impact, as described in cardiac output, blood flow or successful defibrillation associated with CPR2626 Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259. PMid:26472993.
http://dx.doi.org/10.1161/CIR.0000000000...
.

CPA events in extra-hospital environments take place mainly on the floor, thus having a variable that directly favors reaching the desired depth for the chest compressions. The floor is considered the gold standard, as it is a hard surface that enables sinking the chest at the desired depth during CPR1818 Minami K, Kokubo Y, Maeda L, Hibino S. A flexible pressure sensor could correctly measure the depth of chest compression on a mattress. Am J Emerg Med. 2016 maio;34(5):899-902. http://dx.doi.org/10.1016/j.ajem.2016.02.052. PMid:26979259.
http://dx.doi.org/10.1016/j.ajem.2016.02...
,2020 Oh J, Song Y, Kang B, Kang H, Lim T, Suh Y et al. The use of dual accelerometers improves measurement of chest compression depth. Resuscitation. 2012;83(4):500-4. http://dx.doi.org/10.1016/j.resuscitation.2011.09.028. PMid:22001002.
http://dx.doi.org/10.1016/j.resuscitatio...
,2323 Oh JH, Kim CW, Kim SE, Lee DH. Does the bed frame deflection occur along with mattress deflection during in-hospital cardiopulmonary resuscitation? An experiment using mechanical devices. Hong Kong J Emerg Med. 2016;23(2):35-41. http://dx.doi.org/10.1177/102490791602300205.
http://dx.doi.org/10.1177/10249079160230...
. It is emphasized that chest compressions carried out on the floor present significantly increased quality about depth compared to a hospital bed, with a 15% reduction in the depth of the compressions carried out on such surfaces55 Sebbane M, Hayter M, Romero J, Lefebvre S, Chabrot C, Mercier G et al. Chest compressions performed by ED staff: a randomized cross-over simulation study on the floor and on a stretcher. Am J Emerg Med. 2012;30(9):1928-34. http://dx.doi.org/10.1016/j.ajem.2012.04.013. PMid:22795420.
http://dx.doi.org/10.1016/j.ajem.2012.04...
. However, in the context of health care at the in-hospital environment, where the patients are usually in beds/stretchers, performing chest compressions on the floor and being unusual can be impractical. Considering that procedures such as airway management, insertion of intravenous or intraosseous access, use of devices for monitoring and administration of drugs may be needed during in-hospital CPR, moving the patient from a bed/stretcher to the floor to perform high-quality chest compressions may not be effective and can be associated with high risks. The benefits of moving a patient from a bed/stretcher to the floor to perform CPR were not evaluated2626 Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259. PMid:26472993.
http://dx.doi.org/10.1161/CIR.0000000000...
.

To improve the quality of chest compressions, especially in situations where a patient suffers cardiac arrest in the bed, the guidelines dealing with resuscitation recommend that a backboard is to be inserted under the patient11 Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: Adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259. PMid:26472993.
http://dx.doi.org/10.1161/CIR.0000000000...
. However, the AHA recognizes a weak basis for this recommendation, stating insufficient evidence for or against using backboards during CPR2727 Sato Y, Weil MH, Sun S, Tang W, Xie J, Noc M et al. Adverse effects of interrupting precordial compression during cardiopulmonary resuscitation. Crit Care Med. 1997;25(5):733-6. http://dx.doi.org/10.1097/00003246-199705000-00005. PMid:9187589.
http://dx.doi.org/10.1097/00003246-19970...
. Theoretically, using a backboard would promote stability and reduce chest and mattress complacency allowing the rescuer to produce deeper chest compressions.

When the patient's chest is compressed on a firm surface, the sternum's distance is pressed toward the vertebrae directly related to the compression force applied. On a mattress, the sternum-vertebral displacement is influenced by a combination of compression force and the compression degree of the underlying surface (bed/stretcher + mattress). Mathematical data suggest that placing a backboard between the patient and the mattress improves chest compression depth, increasing mattress rigidity2828 Boe JM, Babbs CF. Mechanics of cardiopulmonary resuscitation performed with the patient on a soft bed vs. a hard surface. Acad Emerg Med. 1999;6:754-7. http://dx.doi.org/10.1111/j.1553-2712.1999.tb00449.x.
http://dx.doi.org/10.1111/j.1553-2712.19...
.

Many studies have been conducted to investigate the influence of the lumbar support provided by inserting a backboard while performing CPR66 Noordergraaf GJ, Paulussen IW, Venema A, van Berkom PF, Woerlee PH, Scheffer GJ et al. The impact of compliant surfaces on in-hospital chest compressions: effects of common mattresses and a backboard. Resuscitation. 2009;80(5):546-52. http://dx.doi.org/10.1016/j.resuscitation.2009.03.023. PMid:19409300.
http://dx.doi.org/10.1016/j.resuscitatio...
,88 Andersen LØ, Isbye DL, Rasmussen LS. Increasing compression depth during manikin CPR using a simple backboard. Acta Anaesthesiol Scand. 2007;51(6):747-50. http://dx.doi.org/10.1111/j.1399-6576.2007.01304.x. PMid:17425617.
http://dx.doi.org/10.1111/j.1399-6576.20...
,1919 Perkins GD, Smith CM, Augre C, Allan M, Rogers H, Stephenson B et al. Effects of a backboard, bed height, and operator position on compression depth during simulated resuscitation. Intensive Care Med. 2006;32(10):1632-5. http://dx.doi.org/10.1007/s00134-006-0273-8. PMid:16826385.
http://dx.doi.org/10.1007/s00134-006-027...
,2222 Oh J, Kang H, Chee Y, Lim T, Song Y, Cho Y et al. Use of backboard and deflation improve quality of chest compression when cardiopulmonary resuscitation is performed on a typical air inflated mattress configuration. J Korean Med Sci. 2013 fev;28(2):315-9. http://dx.doi.org/10.3346/jkms.2013.28.2.315. PMid:23399985.
http://dx.doi.org/10.3346/jkms.2013.28.2...
,2929 Nishisaki A, Nysaether J, Sutton R, Maltese M, Niles D, Donoghue A et al. Effect of mattress deflection on CPR quality assessment for older children and adolescents. Resuscitation. 2009;80(5):540-5. http://dx.doi.org/10.1016/j.resuscitation.2009.02.006. PMid:19342150.
http://dx.doi.org/10.1016/j.resuscitatio...

30 Perkins GD, Benny R, Giles S, Gao F, Tweed MJ. Do different mattresses affect the quality of cardiopulmonary resuscitation? Intensive Care Med. 2003;29(12):2330-5. http://dx.doi.org/10.1007/s00134-003-2014-6. PMid:14504728.
http://dx.doi.org/10.1007/s00134-003-201...

31 Tweed M, Tweed C, Perkins GD. The effect of differing support surfaces on the efficacy of chest compressions using a resuscitation manikin model. Resuscitation. 2001;51(2):179-83. http://dx.doi.org/10.1016/S0300-9572(01)00404-X. PMid:11718974.
http://dx.doi.org/10.1016/S0300-9572(01)...
-3232 Perkins GD, Kocierz L, Smith SC, McCulloch RA, Davies RP. Compression feedback devices overestimate chest compression depth when performed on a bed. Resuscitation. 2009;80(1):79-82. http://dx.doi.org/10.1016/j.resuscitation.2008.08.011. PMid:18952361.
http://dx.doi.org/10.1016/j.resuscitatio...
. Some studies suggest that chest compressions can be degraded by non-rigid support surfaces or enhanced, such as increased depth, duration, and frequency, when a board is used66 Noordergraaf GJ, Paulussen IW, Venema A, van Berkom PF, Woerlee PH, Scheffer GJ et al. The impact of compliant surfaces on in-hospital chest compressions: effects of common mattresses and a backboard. Resuscitation. 2009;80(5):546-52. http://dx.doi.org/10.1016/j.resuscitation.2009.03.023. PMid:19409300.
http://dx.doi.org/10.1016/j.resuscitatio...
,99 Cloete G, Dellimore KH, Scheffer C, Smuts MS, Wallis LA. The impact of backboard size and orientation on sternum-to-spine compression depth and compression stiffness in a manikin study of CPR using two mattress types. Resuscitation. 2011;82(8):1064-70. http://dx.doi.org/10.1016/j.resuscitation.2011.04.003. PMid:21601344.
http://dx.doi.org/10.1016/j.resuscitatio...
,2020 Oh J, Song Y, Kang B, Kang H, Lim T, Suh Y et al. The use of dual accelerometers improves measurement of chest compression depth. Resuscitation. 2012;83(4):500-4. http://dx.doi.org/10.1016/j.resuscitation.2011.09.028. PMid:22001002.
http://dx.doi.org/10.1016/j.resuscitatio...
,2323 Oh JH, Kim CW, Kim SE, Lee DH. Does the bed frame deflection occur along with mattress deflection during in-hospital cardiopulmonary resuscitation? An experiment using mechanical devices. Hong Kong J Emerg Med. 2016;23(2):35-41. http://dx.doi.org/10.1177/102490791602300205.
http://dx.doi.org/10.1177/10249079160230...
,2929 Nishisaki A, Nysaether J, Sutton R, Maltese M, Niles D, Donoghue A et al. Effect of mattress deflection on CPR quality assessment for older children and adolescents. Resuscitation. 2009;80(5):540-5. http://dx.doi.org/10.1016/j.resuscitation.2009.02.006. PMid:19342150.
http://dx.doi.org/10.1016/j.resuscitatio...
,3333 Sanri E, Karacabey S. The impact of backboard placement on chest compression quality: a mannequin study. Prehosp Disaster Med. 2019;34(2):182-7. http://dx.doi.org/10.1017/S1049023X19000153. PMid:30981288.
http://dx.doi.org/10.1017/S1049023X19000...
-3434 Cheng A, Belanger C, Wan B, Davidson J, Lin Y. Effect of emergency department mattress compressibility on chest compression depth using a standardized cardiopulmonary resuscitation board, a slider transfer board, and a flat spine board: a simulation-based study. Simul Healthc. 2017 dez;12(6):364-9. http://dx.doi.org/10.1097/SIH.0000000000000245. PMid:28697056.
http://dx.doi.org/10.1097/SIH.0000000000...
. In contrast, other papers indicate that the presence of a backboard does not significantly improve the quality of the chest compressions during CPR1717 Fischer EJ, Mayrand K, Ten Eyck RP. Effect of a backboard on compression depth during cardiac arrest in the ED: a simulation study. Am J Emerg Med. 2016;34(2):274-7. http://dx.doi.org/10.1016/j.ajem.2015.10.035. PMid:26589462.
http://dx.doi.org/10.1016/j.ajem.2015.10...
,1919 Perkins GD, Smith CM, Augre C, Allan M, Rogers H, Stephenson B et al. Effects of a backboard, bed height, and operator position on compression depth during simulated resuscitation. Intensive Care Med. 2006;32(10):1632-5. http://dx.doi.org/10.1007/s00134-006-0273-8. PMid:16826385.
http://dx.doi.org/10.1007/s00134-006-027...
,3535 Putzer G, Fiala A, Braun P, Neururer S, Biechl K, Keilig B et al. Manual versus mechanical chest compressions on surfaces of varying softness with or without backboards: a randomized, crossover Manikin study. J Emerg Med. 2016 abr;50(4):594-600.e1. http://dx.doi.org/10.1016/j.jemermed.2015.10.002. PMid:26607696.
http://dx.doi.org/10.1016/j.jemermed.201...
.

Another point discussed about using backboards is related to the time needed to place the device on the patient's chest. Time is a critical factor in CPR, and it was shown that interruptions in chest compressions reduce the rate of successful defibrillation. It was shown that a delay of only 15 seconds compromises the rate of successful resuscitation during advanced cardiac life support and increases the adverse results if return of spontaneous circulation is achieved3636 Yu T, Weil MH, Tang W, Sun S, Klouche K, Povoas H et al. Adverse outcomes of interrupted precordial compression during automated defibrillation. Circulation. 2002;106(3):368-72. http://dx.doi.org/10.1161/01.CIR.0000021429.22005.2E. PMid:12119255.
http://dx.doi.org/10.1161/01.CIR.0000021...
. Actions such as placing a backboard can cause delays in starting the chest compressions or interruptions after their initiation2727 Sato Y, Weil MH, Sun S, Tang W, Xie J, Noc M et al. Adverse effects of interrupting precordial compression during cardiopulmonary resuscitation. Crit Care Med. 1997;25(5):733-6. http://dx.doi.org/10.1097/00003246-199705000-00005. PMid:9187589.
http://dx.doi.org/10.1097/00003246-19970...
,3636 Yu T, Weil MH, Tang W, Sun S, Klouche K, Povoas H et al. Adverse outcomes of interrupted precordial compression during automated defibrillation. Circulation. 2002;106(3):368-72. http://dx.doi.org/10.1161/01.CIR.0000021429.22005.2E. PMid:12119255.
http://dx.doi.org/10.1161/01.CIR.0000021...
, with adverse consequences for the success of resuscitation, and can only be accepted if the action is helpful to improve the performance of chest compression.

Carrying out good chest compressions is highly challenging and can even be more complex when we are in a scenario with a soft surface such as a mattress. Depth decreases when pressure is applied on a mattress and even using a hard surface may not be sufficient to ensure the depth indicated by the international guidelines for PCR1717 Fischer EJ, Mayrand K, Ten Eyck RP. Effect of a backboard on compression depth during cardiac arrest in the ED: a simulation study. Am J Emerg Med. 2016;34(2):274-7. http://dx.doi.org/10.1016/j.ajem.2015.10.035. PMid:26589462.
http://dx.doi.org/10.1016/j.ajem.2015.10...
. The research results suggest that the type of bed and mattress used is a relevant factor for the force needed to attain chest compression quality, such force being significantly associated to the set of bed/stretcher + mattress used55 Sebbane M, Hayter M, Romero J, Lefebvre S, Chabrot C, Mercier G et al. Chest compressions performed by ED staff: a randomized cross-over simulation study on the floor and on a stretcher. Am J Emerg Med. 2012;30(9):1928-34. http://dx.doi.org/10.1016/j.ajem.2012.04.013. PMid:22795420.
http://dx.doi.org/10.1016/j.ajem.2012.04...
-66 Noordergraaf GJ, Paulussen IW, Venema A, van Berkom PF, Woerlee PH, Scheffer GJ et al. The impact of compliant surfaces on in-hospital chest compressions: effects of common mattresses and a backboard. Resuscitation. 2009;80(5):546-52. http://dx.doi.org/10.1016/j.resuscitation.2009.03.023. PMid:19409300.
http://dx.doi.org/10.1016/j.resuscitatio...
,88 Andersen LØ, Isbye DL, Rasmussen LS. Increasing compression depth during manikin CPR using a simple backboard. Acta Anaesthesiol Scand. 2007;51(6):747-50. http://dx.doi.org/10.1111/j.1399-6576.2007.01304.x. PMid:17425617.
http://dx.doi.org/10.1111/j.1399-6576.20...
-99 Cloete G, Dellimore KH, Scheffer C, Smuts MS, Wallis LA. The impact of backboard size and orientation on sternum-to-spine compression depth and compression stiffness in a manikin study of CPR using two mattress types. Resuscitation. 2011;82(8):1064-70. http://dx.doi.org/10.1016/j.resuscitation.2011.04.003. PMid:21601344.
http://dx.doi.org/10.1016/j.resuscitatio...
,1616 Zhou XL, Sheng LP, Wang J, Li SQ, Wang HL, Ni SZ et al. Effect of bed width on the quality of compressions in simulated resuscitation: a randomized crossover manikin study. Am J Emerg Med. 2016 dez;34(12):2272-6. http://dx.doi.org/10.1016/j.ajem.2016.08.020. PMid:27592725.
http://dx.doi.org/10.1016/j.ajem.2016.08...
,1919 Perkins GD, Smith CM, Augre C, Allan M, Rogers H, Stephenson B et al. Effects of a backboard, bed height, and operator position on compression depth during simulated resuscitation. Intensive Care Med. 2006;32(10):1632-5. http://dx.doi.org/10.1007/s00134-006-0273-8. PMid:16826385.
http://dx.doi.org/10.1007/s00134-006-027...
,2121 Cloete G, Dellimore KH, Scheffer C. Comparison of experimental chest compression data to a theoretical model for the mechanics of constant peak displacement cardiopulmonary resuscitation. Acad Emerg Med. 2011 nov;18(11):1167-76. http://dx.doi.org/10.1111/j.1553-2712.2011.01213.x. PMid:22092898.
http://dx.doi.org/10.1111/j.1553-2712.20...
.

Concerning the characteristics of the mattresses, it is observed that those composed of high-technology materials negatively impact on the quality of chest compressions. Challenged to prevent pressure ulcers, the hospitals introduced different support systems for pressure relief, especially for bedridden patients, starting with simple foam mattresses. The so-called "slow foam" mattresses or "hybrid of low-pressure foam" mattresses, are increasingly used as base mattresses3737 McInnes E, Jammali-Blasi A, Bell-Syer SE, Dumville JC, Middleton V, Cullum N. Support surfaces for pressure ulcer prevention. Cochrane Database Syst Rev. 2015 set 3;(9):CD001735. http://dx.doi.org/10.1002/14651858.CD001735.pub5. PMid:26333288.
https://doi.org/10.1002/14651858.CD00173...
. These systems incorporate several layers of materials with different viscoelastic properties for weight to be equally distributed; this technology replaced the air-filled systems. Several studies showed that, when compared to the floor, support surfaces such as foam mattresses and mattresses that redistribute pressure (inflated and deflated) and mobile stretchers for ambulances, reduce chest compression depth in a resuscitation manikin3030 Perkins GD, Benny R, Giles S, Gao F, Tweed MJ. Do different mattresses affect the quality of cardiopulmonary resuscitation? Intensive Care Med. 2003;29(12):2330-5. http://dx.doi.org/10.1007/s00134-003-2014-6. PMid:14504728.
http://dx.doi.org/10.1007/s00134-003-201...
-3131 Tweed M, Tweed C, Perkins GD. The effect of differing support surfaces on the efficacy of chest compressions using a resuscitation manikin model. Resuscitation. 2001;51(2):179-83. http://dx.doi.org/10.1016/S0300-9572(01)00404-X. PMid:11718974.
http://dx.doi.org/10.1016/S0300-9572(01)...
,3838 Kim JA, Vogel D, Guimond G, Hostler D, Wang HE, Menegazzi JJA. Randomized, controlled comparasion of cardiopulmonary resuscitation performed on the floor andon a moving ambulance stretcher. Prehosp Emerg Care. 2006;10(1):68-70. http://dx.doi.org/10.1080/10903120500373108. PMid:16418093.
http://dx.doi.org/10.1080/10903120500373...
.

Another aspect that can affect the quality of the chest compressions carried out in hospital beds is that the beds/stretchers can oscillate when the rescuer performs chest compressions. Such oscillation can occur through deflection of the mattress or of the bed/stretcher frame itself, reducing the quality of chest compression2323 Oh JH, Kim CW, Kim SE, Lee DH. Does the bed frame deflection occur along with mattress deflection during in-hospital cardiopulmonary resuscitation? An experiment using mechanical devices. Hong Kong J Emerg Med. 2016;23(2):35-41. http://dx.doi.org/10.1177/102490791602300205.
http://dx.doi.org/10.1177/10249079160230...
,2929 Nishisaki A, Nysaether J, Sutton R, Maltese M, Niles D, Donoghue A et al. Effect of mattress deflection on CPR quality assessment for older children and adolescents. Resuscitation. 2009;80(5):540-5. http://dx.doi.org/10.1016/j.resuscitation.2009.02.006. PMid:19342150.
http://dx.doi.org/10.1016/j.resuscitatio...
,3939 Sainio M, Hellevuo H, Huhtala H, Hoppu S, Eilevstjønn J, Tenhunen J et al. Effect of mattress and bed frame deflection on real chest compression depth measured with two CPR sensors. Resuscitation. 2014;85(6):840-3. http://dx.doi.org/10.1016/j.resuscitation.2014.03.009. PMid:24657249.
http://dx.doi.org/10.1016/j.resuscitatio...
. Studies indicate that mattress deflection can be reduced, but that it cannot be removed entirely using a backboard66 Noordergraaf GJ, Paulussen IW, Venema A, van Berkom PF, Woerlee PH, Scheffer GJ et al. The impact of compliant surfaces on in-hospital chest compressions: effects of common mattresses and a backboard. Resuscitation. 2009;80(5):546-52. http://dx.doi.org/10.1016/j.resuscitation.2009.03.023. PMid:19409300.
http://dx.doi.org/10.1016/j.resuscitatio...
,88 Andersen LØ, Isbye DL, Rasmussen LS. Increasing compression depth during manikin CPR using a simple backboard. Acta Anaesthesiol Scand. 2007;51(6):747-50. http://dx.doi.org/10.1111/j.1399-6576.2007.01304.x. PMid:17425617.
http://dx.doi.org/10.1111/j.1399-6576.20...

9 Cloete G, Dellimore KH, Scheffer C, Smuts MS, Wallis LA. The impact of backboard size and orientation on sternum-to-spine compression depth and compression stiffness in a manikin study of CPR using two mattress types. Resuscitation. 2011;82(8):1064-70. http://dx.doi.org/10.1016/j.resuscitation.2011.04.003. PMid:21601344.
http://dx.doi.org/10.1016/j.resuscitatio...
-1010 Nishisaki A, Maltese MR, Niles DE, Sutton RM, Urbano J, Berg RA et al. Backboards are important when chest compressions are provided on a soft mattress. Resuscitation. 2012;83(8):1013-20. http://dx.doi.org/10.1016/j.resuscitation.2012.01.016. PMid:22310727.
http://dx.doi.org/10.1016/j.resuscitatio...
,2929 Nishisaki A, Nysaether J, Sutton R, Maltese M, Niles D, Donoghue A et al. Effect of mattress deflection on CPR quality assessment for older children and adolescents. Resuscitation. 2009;80(5):540-5. http://dx.doi.org/10.1016/j.resuscitation.2009.02.006. PMid:19342150.
http://dx.doi.org/10.1016/j.resuscitatio...
-3030 Perkins GD, Benny R, Giles S, Gao F, Tweed MJ. Do different mattresses affect the quality of cardiopulmonary resuscitation? Intensive Care Med. 2003;29(12):2330-5. http://dx.doi.org/10.1007/s00134-003-2014-6. PMid:14504728.
http://dx.doi.org/10.1007/s00134-003-201...
,4040 Lin Y, Wan B, Belanger C, Hecker K, Gilfoyle E, Davidson J et al. Reducing the impact of intensive care unit mattress compressibility during CPR: a simulation-based study. Adv Simul (Lond). 2017;2(1):22. http://dx.doi.org/10.1186/s41077-017-0057-y. PMid:29450023.
http://dx.doi.org/10.1186/s41077-017-005...
.

The impact of mattress and bed/stretcher frame deflection can be explained by a law of physics called Hooke's Law4141 Aranha N, Oliveira Jr JM, Bellio LO, Bonventi Jr W. Hooke's law and non-linear bags, a case study. Rev Bras Ensino Fis. 2016;38(4):e4305. http://dx.doi.org/10.1590/1806-9126-REBEF-2016-0102.
https://doi.org/10.1590/1806-9126-REBEF-...
. When compressing the chest, this behaves like a spring, that is, the force needed to deform it is proportional to the effective displacement. Therefore, by performing compression on the floor, which has a higher elastic constant, any force applied will be preferentially used to deform the chest. However, when the patient lies on a medium with a lower elastic constant such as mattresses, while applying a force, this will first deform the medium which presents lower resistance. Only when the force required to deform the mattress is more significant than that for deforming the chest will it start to be compressed.

Study limitations

Although all the studies that comprised this review were in-laboratory experimental studies, there are limitations for the number of papers that met the research question due to the lack of clinical studies and to the absence of a detailed description of the compression surfaces used in these studies, which limited a larger scope of the results in understanding the phenomenon under study. It is also pointed out that no guidelines or recommendations that would point to the best characteristics for the compression surfaces were identified.

Contributions to health and nursing

The study contributes to warning about the importance that the health staff is attentive to the issue of the compression surfaces on which the patient is positioned during CPR and their potential impact on high-quality chest compressions, especially in environments with critically ill patients, as these have a higher chance of evolving to CPA, which requires fast and safe assistance.

The managers need to be aware of the type and characteristics of these surfaces during the incorporation of technology in the hospital setting and the professionals who directly serve the patient when deciding on the best support surface given, optimizing the clinical outcomes. As there is yet no formal indication of the ideal surface for performing chest compressions, new clinical studies (especially in the national scope) are recommended that test the effect of the compression surfaces on the quality of CPR.

CONCLUSION

This review identified that the characteristics of the surfaces on which the patient is supported (bed/stretcher mattress and backboard) impact the quality of chest compressions, especially in their depth. These characteristics involve the material, structure, size, disposition and knowledge on the mechanics of the mattress and backboard effects.

Based on the results presented, it is observed that the bed dimensions, and mainly those of the mattress, are correlated with the quality of chest compressions, especially with their depth. Thus, large mattresses and mattresses with high-technology materials for pressure reduction and large beds exert negative impacts on the quality of chest compressions.

Another characteristic that showed to exert an impact on the quality of chest compressions is the deflection of beds and mattresses, directly related to these surfaces' characteristics. It was observed that the use of backboards could only reduce deflection. Consequently, there is a need to know more information about the damping/deflection effect of the surfaces used during cardiopulmonary resuscitation and think about strategies and technologies to reduce or eliminate this effect.

When it comes to using backboards and their impact on high-quality chest compressions, the studies analyzed present different information and, therefore, point to more robust research studies on the theme. This review suggests that new experimental studies should be conducted to understand better the impact of the surfaces where the patient is supported on the quality of chest compressions.

  • FINANCIAL SUPPORT

    Coordenação de Aperfeiçoamento de Pessoal de Nível Superior(CAPES; Master’s degree scholarship). Author: Hudson Carmo de Oliveira. Rio de Janeiro, RJ, Brazil Process: 88887.471399- 2019-00.

REFERÊNCIAS

  • 1
    Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: Adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary resuscitation and emergency cardiovascular care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259 PMid:26472993.
    » http://dx.doi.org/10.1161/CIR.0000000000000259
  • 2
    Perkins GD, Handley AJ, Koster RW, Castrén M, Smyth MA, Olasveengen T et al. European Resuscitation Council Guidelines for Resuscitation 2015: Section2. Adult basic life support and automated external defibrillation. Resuscitation. 2015 out;95:81-99. http://dx.doi.org/10.1016/j.resuscitation.2015.07.015 PMid:26477420.
    » http://dx.doi.org/10.1016/j.resuscitation.2015.07.015
  • 3
    Claudia B, Sergio T, Facholi PT, Schiavo GN, Wagner da Silva SA, Agnaldo P et al. Atualização da diretriz de ressuscitação cardiopulmonar e cuidados cardiovasculares de emergência da Sociedade Brasileira de Cardiologia - 2019. Arq Bras Cardiol. 2019 set;113(3):449-663. http://dx.doi.org/10.5935/abc.20190203. PMid:31621787.
    » https://doi.org/10.5935/abc.20190203
  • 4
    López-González A, Sánchez-López M, Garcia-Hermoso A, López-Tendero J, Rabanales-Sotos J, Martínez-Vizcaíno V. Muscular fitness as a mediator of quality cardiopulmonary resuscitation. Am J Emerg Med. 2016 set;34(9):1845-9. http://dx.doi.org/10.1016/j.ajem.2016.06.058 PMid:27344099.
    » http://dx.doi.org/10.1016/j.ajem.2016.06.058
  • 5
    Sebbane M, Hayter M, Romero J, Lefebvre S, Chabrot C, Mercier G et al. Chest compressions performed by ED staff: a randomized cross-over simulation study on the floor and on a stretcher. Am J Emerg Med. 2012;30(9):1928-34. http://dx.doi.org/10.1016/j.ajem.2012.04.013 PMid:22795420.
    » http://dx.doi.org/10.1016/j.ajem.2012.04.013
  • 6
    Noordergraaf GJ, Paulussen IW, Venema A, van Berkom PF, Woerlee PH, Scheffer GJ et al. The impact of compliant surfaces on in-hospital chest compressions: effects of common mattresses and a backboard. Resuscitation. 2009;80(5):546-52. http://dx.doi.org/10.1016/j.resuscitation.2009.03.023 PMid:19409300.
    » http://dx.doi.org/10.1016/j.resuscitation.2009.03.023
  • 7
    Guimarães HP, Polastri TF, Caldeira P, Barbisan J. Suporte Básico de Vida. Manual do profissional. 6ª ed. EUA: Integracolor, LTD; 2016. 19 p.
  • 8
    Andersen LØ, Isbye DL, Rasmussen LS. Increasing compression depth during manikin CPR using a simple backboard. Acta Anaesthesiol Scand. 2007;51(6):747-50. http://dx.doi.org/10.1111/j.1399-6576.2007.01304.x PMid:17425617.
    » http://dx.doi.org/10.1111/j.1399-6576.2007.01304.x
  • 9
    Cloete G, Dellimore KH, Scheffer C, Smuts MS, Wallis LA. The impact of backboard size and orientation on sternum-to-spine compression depth and compression stiffness in a manikin study of CPR using two mattress types. Resuscitation. 2011;82(8):1064-70. http://dx.doi.org/10.1016/j.resuscitation.2011.04.003 PMid:21601344.
    » http://dx.doi.org/10.1016/j.resuscitation.2011.04.003
  • 10
    Nishisaki A, Maltese MR, Niles DE, Sutton RM, Urbano J, Berg RA et al. Backboards are important when chest compressions are provided on a soft mattress. Resuscitation. 2012;83(8):1013-20. http://dx.doi.org/10.1016/j.resuscitation.2012.01.016 PMid:22310727.
    » http://dx.doi.org/10.1016/j.resuscitation.2012.01.016
  • 11
    American Heart Association. Destaques das Atualizações Específicas das Diretrizes de 2017 da American Heart Association para Suporte Básico de Vida em Pediatria e para Adultos e Qualidade da ressuscitação Cardiopulmonar [Internet]. Chicago: American Heart Association; 2017 [citado 2020 ago 6]. Disponível em: https://eccguidelines.heart.org/wp-content/uploads/2017/12/2017-Focused-Updates_Highlights_PTBR.pdf
    » https://eccguidelines.heart.org/wp-content/uploads/2017/12/2017-Focused-Updates_Highlights_PTBR.pdf
  • 12
    American Heart Association. Destaques das Atualizações Focadas em Recomendações de 2018 da American Heart Association para RCP e ACE: Suporte Avançado de Vida Cardiovascular e Suporte Avançado de Vida em Pediatria [Internet]. Chicago: American Heart Association; 2018 [citado 2020 ago 6]. Disponível em: https://eccguidelines.heart.org/wp-content/uploads/2018/10/2018-Focused-Updates_Highlights_PTBR.pdf
    » https://eccguidelines.heart.org/wp-content/uploads/2018/10/2018-Focused-Updates_Highlights_PTBR.pdf
  • 13
    American Heart Association. Destaques das Atualizações direcionadas nas diretrizes de 2019 da American Heart Association para Ressuscitação Cardiopulmonar e Atendimento Cardiovascular de Emergência [Internet]. Chicago: American Heart Association; 2019 [citado 2020 ago 6]. Disponível em: https://eccguidelines.heart.org/wp-content/uploads/2019/11/2019-Focused-Updates_Highlights_PTBR.pdf
    » https://eccguidelines.heart.org/wp-content/uploads/2019/11/2019-Focused-Updates_Highlights_PTBR.pdf
  • 14
    Mendes KDS, Silveira RCCP, Galvão CM. Integrative literature review: A research method to incorporate evidence in health care and nursing. Texto Contexto Enferm. 2008;17(4):758-64. http://dx.doi.org/10.1590/S0104-07072008000400018
    » http://dx.doi.org/10.1590/S0104-07072008000400018
  • 15
    Fram D, Marin CM, Barbosa D. Avaliação da necessidade da revisão sistemática e a pergunta do estudo. In: Barbosa D, Taminato M, Fram D, Belasco A, editor. Enfermagem baseada em evidências. São Paulo: Atheneu; 2014. Cap. 3. p. 21-28.
  • 16
    Zhou XL, Sheng LP, Wang J, Li SQ, Wang HL, Ni SZ et al. Effect of bed width on the quality of compressions in simulated resuscitation: a randomized crossover manikin study. Am J Emerg Med. 2016 dez;34(12):2272-6. http://dx.doi.org/10.1016/j.ajem.2016.08.020 PMid:27592725.
    » http://dx.doi.org/10.1016/j.ajem.2016.08.020
  • 17
    Fischer EJ, Mayrand K, Ten Eyck RP. Effect of a backboard on compression depth during cardiac arrest in the ED: a simulation study. Am J Emerg Med. 2016;34(2):274-7. http://dx.doi.org/10.1016/j.ajem.2015.10.035 PMid:26589462.
    » http://dx.doi.org/10.1016/j.ajem.2015.10.035
  • 18
    Minami K, Kokubo Y, Maeda L, Hibino S. A flexible pressure sensor could correctly measure the depth of chest compression on a mattress. Am J Emerg Med. 2016 maio;34(5):899-902. http://dx.doi.org/10.1016/j.ajem.2016.02.052 PMid:26979259.
    » http://dx.doi.org/10.1016/j.ajem.2016.02.052
  • 19
    Perkins GD, Smith CM, Augre C, Allan M, Rogers H, Stephenson B et al. Effects of a backboard, bed height, and operator position on compression depth during simulated resuscitation. Intensive Care Med. 2006;32(10):1632-5. http://dx.doi.org/10.1007/s00134-006-0273-8 PMid:16826385.
    » http://dx.doi.org/10.1007/s00134-006-0273-8
  • 20
    Oh J, Song Y, Kang B, Kang H, Lim T, Suh Y et al. The use of dual accelerometers improves measurement of chest compression depth. Resuscitation. 2012;83(4):500-4. http://dx.doi.org/10.1016/j.resuscitation.2011.09.028 PMid:22001002.
    » http://dx.doi.org/10.1016/j.resuscitation.2011.09.028
  • 21
    Cloete G, Dellimore KH, Scheffer C. Comparison of experimental chest compression data to a theoretical model for the mechanics of constant peak displacement cardiopulmonary resuscitation. Acad Emerg Med. 2011 nov;18(11):1167-76. http://dx.doi.org/10.1111/j.1553-2712.2011.01213.x PMid:22092898.
    » http://dx.doi.org/10.1111/j.1553-2712.2011.01213.x
  • 22
    Oh J, Kang H, Chee Y, Lim T, Song Y, Cho Y et al. Use of backboard and deflation improve quality of chest compression when cardiopulmonary resuscitation is performed on a typical air inflated mattress configuration. J Korean Med Sci. 2013 fev;28(2):315-9. http://dx.doi.org/10.3346/jkms.2013.28.2.315 PMid:23399985.
    » http://dx.doi.org/10.3346/jkms.2013.28.2.315
  • 23
    Oh JH, Kim CW, Kim SE, Lee DH. Does the bed frame deflection occur along with mattress deflection during in-hospital cardiopulmonary resuscitation? An experiment using mechanical devices. Hong Kong J Emerg Med. 2016;23(2):35-41. http://dx.doi.org/10.1177/102490791602300205
    » http://dx.doi.org/10.1177/102490791602300205
  • 24
    Vadeboncoeur T, Stolz U, Panchal A, Silver A, Venuti M, Tobin J et al. Chest compression depth and survival in out-of-hospital cardiac arrest. Resuscitation. 2014;85(2):182-8. http://dx.doi.org/10.1016/j.resuscitation.2013.10.002 PMid:24125742.
    » http://dx.doi.org/10.1016/j.resuscitation.2013.10.002
  • 25
    Edelson DP, Abella BS, Kramer-Johansen J, Wik L, Myklebust H, Barry AM et al. Effects of compression depth and pre-shock pauses predict defibrillation failure during cardiac arrest. Resuscitation. 2006;71(2):137-45. http://dx.doi.org/10.1016/j.resuscitation.2006.04.008 PMid:16982127.
    » http://dx.doi.org/10.1016/j.resuscitation.2006.04.008
  • 26
    Kleinman ME, Brennan EE, Goldberger ZD, Swor RA, Terry M, Bobrow BJ et al. Part 5: adult basic life support and cardiopulmonary resuscitation quality: 2015 American Heart Association Guidelines Update for Cardiopulmonary Resuscitation and Emergency Cardiovascular Care. Circulation. 2015;132(18, Suppl. 2):S414-35. http://dx.doi.org/10.1161/CIR.0000000000000259 PMid:26472993.
    » http://dx.doi.org/10.1161/CIR.0000000000000259
  • 27
    Sato Y, Weil MH, Sun S, Tang W, Xie J, Noc M et al. Adverse effects of interrupting precordial compression during cardiopulmonary resuscitation. Crit Care Med. 1997;25(5):733-6. http://dx.doi.org/10.1097/00003246-199705000-00005 PMid:9187589.
    » http://dx.doi.org/10.1097/00003246-199705000-00005
  • 28
    Boe JM, Babbs CF. Mechanics of cardiopulmonary resuscitation performed with the patient on a soft bed vs. a hard surface. Acad Emerg Med. 1999;6:754-7. http://dx.doi.org/10.1111/j.1553-2712.1999.tb00449.x
    » http://dx.doi.org/10.1111/j.1553-2712.1999.tb00449.x
  • 29
    Nishisaki A, Nysaether J, Sutton R, Maltese M, Niles D, Donoghue A et al. Effect of mattress deflection on CPR quality assessment for older children and adolescents. Resuscitation. 2009;80(5):540-5. http://dx.doi.org/10.1016/j.resuscitation.2009.02.006 PMid:19342150.
    » http://dx.doi.org/10.1016/j.resuscitation.2009.02.006
  • 30
    Perkins GD, Benny R, Giles S, Gao F, Tweed MJ. Do different mattresses affect the quality of cardiopulmonary resuscitation? Intensive Care Med. 2003;29(12):2330-5. http://dx.doi.org/10.1007/s00134-003-2014-6 PMid:14504728.
    » http://dx.doi.org/10.1007/s00134-003-2014-6
  • 31
    Tweed M, Tweed C, Perkins GD. The effect of differing support surfaces on the efficacy of chest compressions using a resuscitation manikin model. Resuscitation. 2001;51(2):179-83. http://dx.doi.org/10.1016/S0300-9572(01)00404-X PMid:11718974.
    » http://dx.doi.org/10.1016/S0300-9572(01)00404-X
  • 32
    Perkins GD, Kocierz L, Smith SC, McCulloch RA, Davies RP. Compression feedback devices overestimate chest compression depth when performed on a bed. Resuscitation. 2009;80(1):79-82. http://dx.doi.org/10.1016/j.resuscitation.2008.08.011 PMid:18952361.
    » http://dx.doi.org/10.1016/j.resuscitation.2008.08.011
  • 33
    Sanri E, Karacabey S. The impact of backboard placement on chest compression quality: a mannequin study. Prehosp Disaster Med. 2019;34(2):182-7. http://dx.doi.org/10.1017/S1049023X19000153 PMid:30981288.
    » http://dx.doi.org/10.1017/S1049023X19000153
  • 34
    Cheng A, Belanger C, Wan B, Davidson J, Lin Y. Effect of emergency department mattress compressibility on chest compression depth using a standardized cardiopulmonary resuscitation board, a slider transfer board, and a flat spine board: a simulation-based study. Simul Healthc. 2017 dez;12(6):364-9. http://dx.doi.org/10.1097/SIH.0000000000000245 PMid:28697056.
    » http://dx.doi.org/10.1097/SIH.0000000000000245
  • 35
    Putzer G, Fiala A, Braun P, Neururer S, Biechl K, Keilig B et al. Manual versus mechanical chest compressions on surfaces of varying softness with or without backboards: a randomized, crossover Manikin study. J Emerg Med. 2016 abr;50(4):594-600.e1. http://dx.doi.org/10.1016/j.jemermed.2015.10.002 PMid:26607696.
    » http://dx.doi.org/10.1016/j.jemermed.2015.10.002
  • 36
    Yu T, Weil MH, Tang W, Sun S, Klouche K, Povoas H et al. Adverse outcomes of interrupted precordial compression during automated defibrillation. Circulation. 2002;106(3):368-72. http://dx.doi.org/10.1161/01.CIR.0000021429.22005.2E PMid:12119255.
    » http://dx.doi.org/10.1161/01.CIR.0000021429.22005.2E
  • 37
    McInnes E, Jammali-Blasi A, Bell-Syer SE, Dumville JC, Middleton V, Cullum N. Support surfaces for pressure ulcer prevention. Cochrane Database Syst Rev. 2015 set 3;(9):CD001735. http://dx.doi.org/10.1002/14651858.CD001735.pub5. PMid:26333288.
    » https://doi.org/10.1002/14651858.CD001735.pub5
  • 38
    Kim JA, Vogel D, Guimond G, Hostler D, Wang HE, Menegazzi JJA. Randomized, controlled comparasion of cardiopulmonary resuscitation performed on the floor andon a moving ambulance stretcher. Prehosp Emerg Care. 2006;10(1):68-70. http://dx.doi.org/10.1080/10903120500373108 PMid:16418093.
    » http://dx.doi.org/10.1080/10903120500373108
  • 39
    Sainio M, Hellevuo H, Huhtala H, Hoppu S, Eilevstjønn J, Tenhunen J et al. Effect of mattress and bed frame deflection on real chest compression depth measured with two CPR sensors. Resuscitation. 2014;85(6):840-3. http://dx.doi.org/10.1016/j.resuscitation.2014.03.009 PMid:24657249.
    » http://dx.doi.org/10.1016/j.resuscitation.2014.03.009
  • 40
    Lin Y, Wan B, Belanger C, Hecker K, Gilfoyle E, Davidson J et al. Reducing the impact of intensive care unit mattress compressibility during CPR: a simulation-based study. Adv Simul (Lond). 2017;2(1):22. http://dx.doi.org/10.1186/s41077-017-0057-y PMid:29450023.
    » http://dx.doi.org/10.1186/s41077-017-0057-y
  • 41
    Aranha N, Oliveira Jr JM, Bellio LO, Bonventi Jr W. Hooke's law and non-linear bags, a case study. Rev Bras Ensino Fis. 2016;38(4):e4305. http://dx.doi.org/10.1590/1806-9126-REBEF-2016-0102.
    » https://doi.org/10.1590/1806-9126-REBEF-2016-0102

Edited by

ASSOCIATE EDITOR

SCIENTIFIC EDITOR

Publication Dates

  • Publication in this collection
    23 July 2021
  • Date of issue
    2021

History

  • Received
    23 Feb 2021
  • Accepted
    26 May 2021
Universidade Federal do Rio de Janeiro Rua Afonso Cavalcanti, 275, Cidade Nova, 20211-110 - Rio de Janeiro - RJ - Brasil, Tel: +55 21 3398-0952 e 3398-0941 - Rio de Janeiro - RJ - Brazil
E-mail: annaneryrevista@gmail.com